High-Field Asymmetric Waveform Ion Mobility Spectrometry

Permanent Link: http://ufdc.ufl.edu/UFE0041446/00001

Material Information

Title: High-Field Asymmetric Waveform Ion Mobility Spectrometry Evaluation of Novel Cell Geometries and Techniques
Physical Description: 1 online resource (115 p.)
Language: english
Creator: Prieto, Marilyn
Publisher: University of Florida
Place of Publication: Gainesville, Fla.
Publication Date: 2010

Subjects

Subjects / Keywords: Chemistry -- Dissertations, Academic -- UF
Genre: Chemistry thesis, Ph.D.
bibliography   ( marcgt )
theses   ( marcgt )
government publication (state, provincial, terriorial, dependent)   ( marcgt )
born-digital   ( sobekcm )
Electronic Thesis or Dissertation

Notes

Abstract: A HIGH-FIELD ASYMMETRIC WAVEFORM ION MOBILITY SPECTROMETRY: EVALUATION OF NOVEL CELL GEOMETRIES AND TECHNIQUES High-field asymmetric ion mobility spectrometry, or FAIMS, is an emerging detection technology that can operate at atmospheric pressure to separate gas-phase ions. Because FAIMS has continuous ion separation capabilities, it is amenable to use in conjunction with a mass spectrometer. This union offers two orthogonal separation methods, one separating ions according to change in their mobility through a gas as a function of field strength and the other according to the ion?s mass-to-charge ratio. FAIMS has been utilized to detect peptides, bacteria, chemical warfare agents, and even drugs. In literature, there is still much to experiment on the role between three-dimensional (3-D) curved cell geometries (spherical, hemispherical, hemi-elliptical, and ovoidal) and two-dimensional (2-D) curved geometries (cylindrical) with ion behavior. Studies encompassing this research focus on exploring the fundamental and practical issues of curved electrode geometries and novel waveform approaches. The marriage of FAIMS and mass spectrometry (MS) was operated with an atmospheric pressure ionization (API) source which was characterized with each particular geometry while applying it to explosive detection. Results from these experiments provide a greater understanding of the relationship of cell geometry to ion transmission, resolution and resolving power. Although the use of curved surfaces increases sensitivity, resolution is compromised. This effect is more pronounced with increasing curvature and/or surface area. Depending on the application needed by the analyst (more resolution or more transmission), these novel 3D curved electrodes would be valuable to employ. Subsequent to the optimization of FAIMS parameters with particular geometries, the practicability of a new waveform approach for FAIMS was examined. The high voltage asymmetric waveform typically used consists of overlapping sine waves; here an asymmetric square waveform is used. Comparison of these two waveforms shows that the implementation of a square wave on a miniature planar FAIMS would deliver better separation and resolving power, at some cost in transmission and thus sensitivity. Finally, once the performance and efficiency of geometry, waveform type, and ionizer was maximized, this novel analytical platform was applied to explosive analysis. Both 3D curved electrodes (hemispherical and spherical) displayed better resolving power and resolution than a commercial cylindrical cell. These novel FAIMS geometries were able to baseline resolve a 2 ppm mixture of five different explosives where the commercial cylindrical cell failed to resolve any. This research?s main goal is to present the first detailed investigation of novel 3-D and 2-D curved geometries while applying two unique waveforms to show its effect on ion behavior in the FAIMS apparatus. Developing and validating the APCI/FAIMS/MS interface allows for better analyte detection with increased resolution, resolving power, and transmission.
General Note: In the series University of Florida Digital Collections.
General Note: Includes vita.
Bibliography: Includes bibliographical references.
Source of Description: Description based on online resource; title from PDF title page.
Source of Description: This bibliographic record is available under the Creative Commons CC0 public domain dedication. The University of Florida Libraries, as creator of this bibliographic record, has waived all rights to it worldwide under copyright law, including all related and neighboring rights, to the extent allowed by law.
Statement of Responsibility: by Marilyn Prieto.
Thesis: Thesis (Ph.D.)--University of Florida, 2010.
Local: Adviser: Yost, Richard A.

Record Information

Source Institution: UFRGP
Rights Management: Applicable rights reserved.
Classification: lcc - LD1780 2010
System ID: UFE0041446:00001

Permanent Link: http://ufdc.ufl.edu/UFE0041446/00001

Material Information

Title: High-Field Asymmetric Waveform Ion Mobility Spectrometry Evaluation of Novel Cell Geometries and Techniques
Physical Description: 1 online resource (115 p.)
Language: english
Creator: Prieto, Marilyn
Publisher: University of Florida
Place of Publication: Gainesville, Fla.
Publication Date: 2010

Subjects

Subjects / Keywords: Chemistry -- Dissertations, Academic -- UF
Genre: Chemistry thesis, Ph.D.
bibliography   ( marcgt )
theses   ( marcgt )
government publication (state, provincial, terriorial, dependent)   ( marcgt )
born-digital   ( sobekcm )
Electronic Thesis or Dissertation

Notes

Abstract: A HIGH-FIELD ASYMMETRIC WAVEFORM ION MOBILITY SPECTROMETRY: EVALUATION OF NOVEL CELL GEOMETRIES AND TECHNIQUES High-field asymmetric ion mobility spectrometry, or FAIMS, is an emerging detection technology that can operate at atmospheric pressure to separate gas-phase ions. Because FAIMS has continuous ion separation capabilities, it is amenable to use in conjunction with a mass spectrometer. This union offers two orthogonal separation methods, one separating ions according to change in their mobility through a gas as a function of field strength and the other according to the ion?s mass-to-charge ratio. FAIMS has been utilized to detect peptides, bacteria, chemical warfare agents, and even drugs. In literature, there is still much to experiment on the role between three-dimensional (3-D) curved cell geometries (spherical, hemispherical, hemi-elliptical, and ovoidal) and two-dimensional (2-D) curved geometries (cylindrical) with ion behavior. Studies encompassing this research focus on exploring the fundamental and practical issues of curved electrode geometries and novel waveform approaches. The marriage of FAIMS and mass spectrometry (MS) was operated with an atmospheric pressure ionization (API) source which was characterized with each particular geometry while applying it to explosive detection. Results from these experiments provide a greater understanding of the relationship of cell geometry to ion transmission, resolution and resolving power. Although the use of curved surfaces increases sensitivity, resolution is compromised. This effect is more pronounced with increasing curvature and/or surface area. Depending on the application needed by the analyst (more resolution or more transmission), these novel 3D curved electrodes would be valuable to employ. Subsequent to the optimization of FAIMS parameters with particular geometries, the practicability of a new waveform approach for FAIMS was examined. The high voltage asymmetric waveform typically used consists of overlapping sine waves; here an asymmetric square waveform is used. Comparison of these two waveforms shows that the implementation of a square wave on a miniature planar FAIMS would deliver better separation and resolving power, at some cost in transmission and thus sensitivity. Finally, once the performance and efficiency of geometry, waveform type, and ionizer was maximized, this novel analytical platform was applied to explosive analysis. Both 3D curved electrodes (hemispherical and spherical) displayed better resolving power and resolution than a commercial cylindrical cell. These novel FAIMS geometries were able to baseline resolve a 2 ppm mixture of five different explosives where the commercial cylindrical cell failed to resolve any. This research?s main goal is to present the first detailed investigation of novel 3-D and 2-D curved geometries while applying two unique waveforms to show its effect on ion behavior in the FAIMS apparatus. Developing and validating the APCI/FAIMS/MS interface allows for better analyte detection with increased resolution, resolving power, and transmission.
General Note: In the series University of Florida Digital Collections.
General Note: Includes vita.
Bibliography: Includes bibliographical references.
Source of Description: Description based on online resource; title from PDF title page.
Source of Description: This bibliographic record is available under the Creative Commons CC0 public domain dedication. The University of Florida Libraries, as creator of this bibliographic record, has waived all rights to it worldwide under copyright law, including all related and neighboring rights, to the extent allowed by law.
Statement of Responsibility: by Marilyn Prieto.
Thesis: Thesis (Ph.D.)--University of Florida, 2010.
Local: Adviser: Yost, Richard A.

Record Information

Source Institution: UFRGP
Rights Management: Applicable rights reserved.
Classification: lcc - LD1780 2010
System ID: UFE0041446:00001


This item has the following downloads:


Full Text





HIGH-FIELD ASYMMETRIC WAVEFORM ION MOBILITY SPECTROMETRY:
EVALUATION OF NOVEL CELL GEOMETRIES AND TECHNIQUES





















By

MARILYN PRIETO


A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL
OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT
OF THE REQUIREMENTS FOR THE DEGREE OF
DOCTOR OF PHILOSOPHY

UNIVERSITY OF FLORIDA

2010

































2010 Marilyn Prieto



























To my parents and my beloved husband-to-be, Derick Tourne









ACKNOWLEDGMENTS

The accomplishments of this work are dedicated to those individuals who helped

me stay sane during my years at the University of Florida. I would like to extend my

deepest gratitude to my advisor, Dr. Richard Yost, for giving me the opportunity to join

his research group and for his consistent support and mentoring. His door was always

open for anything, whether it was research or personal. He taught me how to be an

independent thinker and a successful scientist.

A special thanks goes to all the members of the Yost group for their help,

suggestions and friendship. I give special thanks to Leonard Rorrer for always

answering the same questions over and over and never seeming to lose his patience. I

would like to thank Dr. Jennifer Bryant for all the late nights we spent brooding over the

designs and characterizations of the hemispherical cell. I would also like to

acknowledge Chia-Wei Tsai for helping me better understand how the square wave

performs even when the data do not make sense. I cannot forget to thank Mike

Napolitano for reminding me how important it is to stay motivated. I would also like to

acknowledge Stacey Ann Benjamin and Julianne McLaughlin who were always there to

make me laugh when all I wanted to do was cry. Very special thanks go to the

Chemistry Machine Shop members: Todd Prox, Brian Smith and Joe Shalosky for the

many hours of fruitful conversations concerning design ideas for all the cells we made.

My research would have not progressed as smoothly as it did if it was not for them.

Last, but not least, I would like to thank my family and friends for always believing

in me. They have always been my inspiration when I needed it the most. I give special

thanks to my mom for making me the woman I am today and always encouraging me to

obtain more education. I give special thanks to my dad for teaching me that there is









always something to smile about. My greatest thanks go to one person and one animal.

Derick Tourne, you make me the happiest woman in the world and I cannot wait to

spend the rest of my life by your side. Kaligula, my beloved cat, may you rest in peace

and know that I will never forget all the times you were the only one to receive me at the

door. You will always be my most prized possession.









TABLE OF CONTENTS

page

ACKNOW LEDGM ENTS ....... .. ...................................... .. ............... ............... 4

LIST O F TABLES .......... ..... .. ....................... ...... ..... .... ................... 8

LIS T O F F IG U R E S .................................................................. 9

ABSTRACT ............... .. ............................... ......... 13

CHAPTER

1 INTRODUCTION AND INSTRUMENTATION ......... .................................... 15

Background...... .................... ............... ...... ............... 15
Introduction to Ion M ability Spectrom etry............................ ........ ................ 18
Introduction to High-Field Asymmetric Waveform Ion Mobility Spectrometry.......... 20
Quadrupole Ion Trap Mass Spectrometry ....... ............ ....... ........ ......... 23
Quadrupole Ion Trap Theory ......................................... ............... ....... 24
Finnigan LCQ A Commercial QITMS .............. ............. ............ 25
Atmospheric Pressure Chemical Ionization (APCI)....................................... 27
O overview of D issertation ............. ......... ...... ......... ............... ............... 28

2 FUNDAMENTALS OF FAIMS FOR CURVED ELECTRODE GEOMETRIES ........ 36

Introduction ......... ...... .......... ......... .. ............................ 36
Description of Planar Geom etry.......................................................... .... ............... 37
D description of C ylindrical G eom etry............................................... ... .. ............... 38
Description of Hemispherical Geometry.................... .......... .. ........ ..... 40
Description and Design of a Spherical Geometry ...................... ...... ............... 42
E xpe rim e nta l............................................. ....... .......... ...... 44
In s tru m e n ta tio n ............................................................................. .............. 4 4
S am ples ............. ................................................................... ...... 45
Results and Discussion.......................................... ............... 46
Comparison of Signal Intensity..................... .... ...................... 47
Comparison of Resolving Power and Resolution ........ .......................... 49
Sensitivity of Spherical FAIMS .............. ......... ............ ..................... 52
C o n c lu s io n s ............................................................................................... 5 4

3 EVALUATION OF FAIMS WAVEFORMS APPROACHES.................................. 81

Introduction .................... ............ ............... 81
S um of S ines W aveform .............................. ........................ .............. 83
Square Waveform ........................ ......... ........... 83
Experim ental................................. ............... 85
Results and Discussion.......................................... ............... 88









Influence of Dispersion Voltage on Compensation Voltage, Peak Width and
Resolving Power .......... ..................... .......................... 88
Influence of Frequency on Compensation Voltage, Peak Width, and
Resolving Power ........................................................................... ................ 90
Influence of Frequency on Sensitivity ........................... ......... .................. 91
Influence of Duty Cycle on Sensitivity ........... ........................ ........... ....... 91
Lower Concentration Experiments ......... ......... ............................ 92
Conclusions ............. .... ...... ...... ......... .. .......... ......... 92

4 CO NCLUSIO NS ...................................... .... ................................. 106

Future W ork ............. ....... ..... ..... ......... .. ................... 108
Curved FAIMS Cells ........ ..................................... ............... 108
FAIM S W aveforms ......................... ..... ..................................... 109

LIST OF REFERENCES .................. ........... ....................... ..... .......... 110

B IO G RA PH ICA L SKETC H ............. ................. ................. .................. ............... 115









LIST OF TABLES


Table page

2-1. Comparison of dimensions between Thermo-Fisher cylindrical and
hem ispherical cell .............. .............................. .. ............ ....... ........ 75

2-2. Comparison of dimensions between all three cells. .................... .................. 75

2-3. Anticipated ions and adducts for nitramine and nitrate ester explosives .............. 76

2-4. Comparison of intensity values for RDX, HMX, NG, 2,4-DNT, PETN, TNT, and
TNB between the lonalytics cylindrical, hemispherical, and spherical FAIMS
ce ll. ............................................................................................ 7 6

2-5. Comparison of CV, intensity, and resolving power between the Thermo-Fisher
cylindrical at 16,000 V/cm and both the hemispherical and spherical FAIMS
cell at 16,072 V/cm ............ .......... .. ....... ............ ...... ......... 77

2-6. Comparison of resolution between the Thermo-Fisher cylindrical cell at 16,000
V/cm and both the hemispherical and spherical FAIMS cell at 16,072 V/cm...... 78

2-7. Comparison of CV, intensity, and resolving power between the lonalytics
cylindrical, hemispherical, and spherical FAIMS cell at 23,500 V/cm ................ 79

2-8. Comparison of resolution and peak width between the Thermo-Fisher
cylindrical, hemispherical and spherical FAIMS cell at 23,500 V/cm ................. 80

2-9. Repeatability of CV values on the spherical cell from five replicate runs for
T N T ... ... .......................................... ........ .......... ...... 8 0

2-10. Linear dynamic range and limit of detection for TNT collected by full-scan and
SIM mode. ............. ................. .............. ......... ...... ......... 80

3-1. Summary of the CV, PWHM, Rp, and signal intensity for concentrations from
10 ppm dow n to 100 ppb ........................................................................... 105









LIST OF FIGURES


Figure page

1-1. Schematic of a characteristic ion mobility spectrometer using the stacked ring
design ............ ......... ................................................................... ... .... 30

1-2. Ion motion between two plates during the application of an electric potential. A
simplified asymmetric waveform is applied to the upper plate.40 .................. 30

1-3. Hypothetical plots of the dependence of ion mobility on electric field strength
for three types of ions. ................................... ................ ..... 31

1-4. Cross sectional view of mass analyzer in the QITMS.48...................... ..... 32

1-5. Ion motion in a quadrupole ion trap.49 ........................................... 33

1-6. Mathieu stability diagram plotted in a and q space for positive ions. The shape
of the stable region is rotated about the az =0 line for negative ions. Most
commercial QITMS instruments operate along the az =0 line (RF-only
operation). Under these conditions, ions will have a stable trajectory
between qz = 0 and 0.908. The 3z =1 boundary intersects the qz axis at qz =
0 .9 0 8 .5 0 ................................................... ................................................... 3 4

1-7. Schematic of Finnigan LCQ ion trap mass spectrometer.48 ................................. 35

1-8. Schematic of the atmospheric pressure chemical ionization (APCI) source
used in the LCQ .48.... ......... ......................................... ........... 35

2-1. Schematic of hemispherical FAIMS electrodes. (1) outer electrode, (2) outer
housing, (3) inner electrode, (4) exit port, (5) base, (6) inner electrode
connection, (7) entrance port, and (8) curtain plate....................... ........... 56

2-2. Ions separated between flat plates in the FAIMS analyzer. The transmitted ion
is distributed evenly over the gap between the plates. Ions are continuously
lost through diffusion and space charge repulsion. 59................................ ... 56

2-3. Motion of a positively charged ion in a flat plate FAIMS.21 ............................... 57

2-4. Ions separated between concentric cylinders of a FAIMS analyzer. The
transmitted ion is focused towards a fixed radial distance, and distributed
around that distance through the effects of diffusion and space charge
repulsion. The effect of the focus is exaggerated for clarity in this figure.59........ 57

2-5. Three-dimensional schematic and cross section of the Thermo-Fisher
cylindrical geometry FAIMS. (a) Three-dimensional view and (b) cross
section view illustrating radial location, r, between the walls of the inner









electrode, 'a' and outer electrode 'b'. The asymmetric waveform is applied to
the inner electrode.6 ............. ....................... ......... 58

2-6. Focusing effect caused by curvature of electrodes in FAIMS. Adapted from
Barnett et al.60 ................................................................ ........ ..... ........ .......... 58

2-7. Disassembled hemispherical FAIMS cell. .............. .............................. ........ 59

2-8. Assembled hemispherical FAIMS cell. Outer casing is made of Kel-F and
curtain plate is made of stainless steel. .................. ........ ... ..... ...... ........ 59

2-9. Variation of the electric field as a function of position for the three FAIMS cells.
All three cells contain a gap of 2.50 mm. Radius of inner electrode for
cylindrical and hemispherical is 6.00 mm. DV for all three is 4000 V............... 60

2-10. Original spherical FAIMS design. Outer casing is made of Delrin, electrodes
of brass, and inner electrode of stainless steel......................... ......... .......... 61

2-11. Novel spherical FAIMS cell. Outer casing is made of PEEK, inner and outer
electrodes are made of stainless steel. ........... ................. ... ..... ...... ........ 62

2-12. Dimensions of novel spherical FAIMS cell (in inches) ....................................... 63

2-13. Disassembled spherical FAIMS electrodes. Both inner and outer electrodes
are made of stainless steel .......... ..... ................................ ........... 63

2-14. Assembled spherical FAIMS cell. Outer casing is made of Ketron PEEK.
Curtain plate is made of stainless steel. ........... ............................................ 64

2-15. Structures of several structural classes of common explosives studied in this
w ork .............. ....... ..................................... .................. 65

2-16. Comparison of signal intensity of TNT between all three FAIMS cells.
Extracted chromatogram of m/z 227 on left and corresponding mass spectra
on right. lonalytics cylindrical FAIMS (top), hemispherical FAIMS (middle),
and spherical FAIMS (bottom).............. ........... ......... ...................... 66

2-17. Comparison of signal intensity of TNB between all three FAIMS cells.
Extracted chromatogram of m/z 213 on left and corresponding mass spectra
on right. lonalytics cylindrical FAIMS (top), hemispherical FAIMS (middle),
and spherical FAIMS (bottom).............. ........... ......... ...................... 67

2-18. Comparison of signal intensity of 2,4-DNT between all three FAIMS cells.
Extracted chromatogram of m/z 182 on left and corresponding mass spectra
on right. lonalytics cylindrical FAIMS (top), hemispherical FAIMS (middle),
and spherical FAIMS (bottom).............. ........... ......... ...................... 68









2-19. CV scan of a 2 ppm mixture of explosives on the Thermo-Fisher cylindrical
cell at a DV -4000 V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB,
and (F) P ET N ................................ ........ ..... ..................... 69

2-20. CV scan of a 2 ppm mixture of explosives on hemispherical cell at a DV -
3214 V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB, and (F)
P E T N ......... ... ......... ... ............................. ............. .............. 7 0

2-21. CV scan of a 2 ppm mixture of explosives on spherical cell at a DV -3214 V.
(A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB, and (F) PETN ............ 71

2-22. CV scan of a 2 ppm mixture of explosives on cylindrical cell at a DV -4700
(23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT, (E) 2,4-
D N T and (F) 1,3-D N B .......................................... ............... .............. 72

2.23. CV scan of a 2 ppm mixture of explosives on hemispherical cell at a DV -
4700 V (23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT,
(E) 2,4-DNT, and (F) 1,3-DNB ....................................................... ............... 73

2-24. CV scan of a 2 ppm mixture of explosives on spherical cell at a DV -4700 V
(23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT, (E) 2,4-
D N T and (F) 1,3-D N B .......................................... ............... .............. 74

3-1. Schematic of (a) an asymmetric sum-of-sines waveform and (b) a rectangular
w avefo rm ............ ......... .......................................................... ..... .. ...... 94

3-2. Photograph (a), perspective and cross-sectional views (b and c) of miniature
planar FAIMS cell. Square wave generator circuit diagram is also shown (d)... 95

3-4. CV versus DV simulation data for the [M]- ion of TNT...................... ................ 97

3-5. Three CV scans of 10 ppm TNT. Peak splitting is observed as DV is increased
from (a) to (f). (a) is a DV +121 V, (b) +161 V, (c) +181 V, (d) +201 V, (e)
+221 V and (f) +242 V ...................................................... .............. 98

3-6. Scope traces of square waveforms at different frequencies for amplitudes (a)
DV 70 and (b) DV 340 where V1 is the amplitude (DV) and tl is the width of
the high voltage pulse. Waveforms are offset for ease of visualization. ............ 99

3-7. Graph displaying CV as a function of DV. For 250 kHz the duty cycle is shown
to have little effect ...... .. .. ........... ......... ....................... ............ ........ 100

3-8. Graph displaying peak width as a function of dispersion voltage. For 250 kHz
the duty cycle is shown to have no effect. ............. .................... ........... 100

3-11. Graphs displaying the effect of duty cycle on sensitivity. Frequency is fixed to
250 kHz (a) and 333 kHz (b). ................................ ............ ............... 103









3-12. Graphs displaying the effect of duty cycle on peak area. Frequency is fixed
to 250 kHz (a) and 333 kHz (b). ................ ................................ .............. 104









Abstract of Dissertation Presented to the Graduate School
of the University of Florida in Partial Fulfillment of the
Requirements for the Degree of Doctor of Philosophy

A HIGH-FIELD ASYMMETRIC WAVEFORM ION MOBILITY SPECTROMETRY:
EVALUATION OF NOVEL CELL GEOMETRIES AND TECHNIQUES

By

Marilyn Prieto

August 2010

Chair: Richard A. Yost
Major: Chemistry

High-field asymmetric ion mobility spectrometry, or FAIMS, is an emerging

detection technology that can operate at atmospheric pressure to separate gas-phase

ions. Because FAIMS has continuous ion separation capabilities, it is amenable to use

in conjunction with a mass spectrometer. This union offers two orthogonal separation

methods, one separating ions according to change in their mobility through a gas as a

function of field strength and the other according to the ion's mass-to-charge ratio.

FAIMS has been utilized to detect peptides, bacteria, chemical warfare agents, and

even drugs.

In literature, there is still much to experiment on the role between three-

dimensional (3-D) curved cell geometries (spherical, hemispherical, hemi-elliptical, and

ovoidal) and two-dimensional (2-D) curved geometries (cylindrical) with ion behavior.

Studies encompassing this research focus on exploring the fundamental and practical

issues of curved electrode geometries and novel waveform approaches. The marriage

of FAIMS and mass spectrometry (MS) was operated with an atmospheric pressure

ionization (API) source which was characterized with each particular geometry while









applying it to explosive detection. Results from these experiments provide a greater

understanding of the relationship of cell geometry to ion transmission, resolution and

resolving power. Although the use of curved surfaces increases sensitivity, resolution is

compromised. This effect is more pronounced with increasing curvature and/or surface

area. Depending on the application needed by the analyst (more resolution or more

transmission), these novel 3D curved electrodes would be valuable to employ.

Subsequent to the optimization of FAIMS parameters with particular geometries,

the practicability of a new waveform approach for FAIMS was examined. The high

voltage asymmetric waveform typically used consists of overlapping sine waves; here

an asymmetric square waveform is used. Comparison of these two waveforms shows

that the implementation of a square wave on a miniature planar FAIMS would deliver

better separation and resolving power, at some cost in transmission and thus sensitivity.

Finally, once the performance and efficiency of geometry, waveform type, and

ionizer was maximized, this novel analytical platform was applied to explosive analysis.

Both 3D curved electrodes (hemispherical and spherical) displayed better resolving

power and resolution than a commercial cylindrical cell. These novel FAIMS

geometries were able to baseline resolve a 2 ppm mixture of five different explosives

where the commercial cylindrical cell failed to resolve any.

This research's main goal is to present the first detailed investigation of novel 3-D

and 2-D curved geometries while applying two unique waveforms to show its effect on

ion behavior in the FAIMS apparatus. Developing and validating the APCI/FAIMS/MS

interface allows for better analyte detection with increased resolution, resolving power,

and transmission.









CHAPTER 1
INTRODUCTION AND INSTRUMENTATION

Background

More selective, sensitive, and efficient ion separation techniques are paramount

for next-generation detection systems. The sensitivity and selectivity of mass

spectrometry (MS) over the years, has made it an indispensable tool for modern

analytical chemistry. It should be emphasized that mass spectrometry is not a panacea

for solving all problems faced by a scientific expert; no single technique is.

Nevertheless, it would not be an overstatement to cite MS as a major contributor to the

technology that makes it possible for scientists to monitor and enforce environmental,

toxicological, and drug regulations.

Despite all the research and development that have been concentrated on ion

separation, the demand still exists for techniques with better selectivity and sensitivity

and with real-time analysis especially for explosives. With the potential for increasing

domestic terrorist attacks, the detection of hidden explosives has become a subject of

major interest. Trace detection of explosives employs a wide variety of analytical

methods, including gas chromatography-electron capture detection (GC-ECD)1, gas

chromatography-mass spectrometry (GC-MS)2-4, liquid chromatography-mass

spectrometry (LC-MS)5' 6, and thin-layer chromatography (TLC).7

The methods mentioned above all have limitations. For instance, they are

generally either susceptible to thermal degradation (GC-ECD and GC-MS) because of

the explosives' low vapor pressure and thermal instability or are too insensitive (LC)

because of broader peak shapes and lower peak heights. Furthermore, while ECD is a

sensitive means of detecting explosives, it is less selective than the mass









spectrometer.8 TLC was one of the first techniques used for explosives analysis

because of its low cost and its simultaneous analysis of samples;8 nevertheless, the

detection limits of TLC is generally in the low microgram range making it impossible to

detect trace quantities.

For the analysis of explosives, mass spectrometry is widely used when coupled

with chromatographic separation due to the advantages of high sensitivity, short

response time, as well as the additional selectivity available from MS/MS and

ion/molecule reactions.9 However, requirements for sample preparation substantially

restrict the capability of mass spectrometry for explosive detection. Different ionization

techniques were developed to overcome these limitations, including atmospheric

pressure ionization sources such as electrospray ionization (ESI)5, 6,10 and atmospheric

pressure chemical ionization (APCI).11 12

Thus far, IMS, formerly known as plasma chromatography, has been most widely

used for explosives monitoring because of its high sensitivity and amenability to

miniaturization for field portable applications.13 However, a major disadvantage of IMS

is the limited selectivity, such that some interferent ions have the same mobility as the

analyte ion, and thus can result in a false positive. FAIMS may be able to reduce this

problem because it can separate species with identical low-field mobilities.

In recent years, high-field asymmetric waveform ion mobility spectrometry (FAIMS)

has been shown to provide high sensitivity, selectivity, and specificity for many analytes

when combined with mass spectrometry.14 Interfacing these techniques allows the use

of two orthogonal separation methods, one separating ions based on their differential

ion mobility and the other, on their mass-to-charge (m/z) ratio. FAIMS is a valuable









separation technique that exhibits low parts per billion (ppb) limits of detection for

continuous vapor streams and is suitable as a gas chromatographic detector owing to

its fast response and low memory.15' 16 It also possesses capabilities of separating

isomers and providing additional information orthogonal to MS.17-19 The first FAIMS

apparatus was built by Buryakov et al.20 and consisted of two flat parallel plates for

electrodes. Since then, several different geometries have been evaluated and even

more have been theoretically proposed, with each geometry having its own specific

advantage. For example, a curved cell electrode can provide better sensitivity due to its

focusing effect on the ions.21-23

FAIMS is an ion mobility technique that operates at atmospheric pressure and

room temperature to separate gas-phase ions.15' 24,25 Separation of the ions occurs by

applying a non-constant high electric field (~10,000 V/cm), that alternates between high

and low voltages of opposite polarities. Conventional ion mobility spectrometry (IMS)

works on a somewhat similar principle but only uses low electric fields, ~200 V/cm, to

drive the ions across the drift tube. Ions in IMS are separated according to their

mobility, K, in a particular drift gas.26

FAIMS and MS are two orthogonal separation methods, which separate ions

depending on their differential ion mobility and ratio of m/z. Adding a stage of FAIMS

separation before mass spectrometry (using an ionization source capable of working

under ambient pressure) is beneficial when working with explosives because of its rapid

detection, high sensitivity, improved signal-to-noise ratio, and, when curvature is

introduced, its ion focusing capabilities.14' 27









The waveform design is an essential feature of the overall FAIMS performance,

and thus was evaluated in this research. This fundamental investigation shows a

promising technique for separating gas-phase ions at atmospheric pressure. Gaining a

more comprehensive understanding of how FAIMS geometry, and the applied FAIMS

waveforms affect operation, will allow the efficient production of next-generation

detection devices.

Introduction to Ion Mobility Spectrometry

Ion mobility spectrometry (IMS), formerly known as plasma chromatography, has

been considered a simple and useful, yet relatively low resolution technique, since its

inception into the analytical community over thirty years ago. The low resolving power

of IMS has made it unattractive to large-scale commercialization, and thus limited its

utility in mainstream analytical chemistry. IMS was developed as an analytical

technique over 30 years ago.2830 Only recently have advances in its technology and

design helped analytical chemists realize the full potential of this relatively simple and

sensitive method of analysis. During the last decade, IMS has evolved into an

inexpensive and powerful technique for the detection of many trace compounds and

different compound classes such as chemical warfare agents31, explosives32, and drugs

of abuse.33 Recently, IMS has moved into several exciting areas, including high

resolution separations, which, when used in tandem with MS,34 have allowed for

interesting investigation into gas-phase conformation of isomeric proteins and

peptides35 as well as fast separation of illicit drugs and their metabolites.33

The basic components of an ion mobility spectrometer consist of a sample inlet,

ionization source, reaction region, ion gates(s), drift region, and an ion collector or

detector. Figure 1-1 illustrates these components in a conventional, unidirectional, gas









flow system employing APCI as the ionization source and a stacked ring drift region

design. In traditional IMS, the sample is introduced into the reaction region with an inert

carrier gas. The ions formed are directed toward the drift region by an applied electric

field (1-500 V/cm) upon the drift rings. A narrow pulse of ions is injected into the drift

region, using an ion gate. Ions travel through a buffer gas at atmospheric pressure, and

collisions with the buffer gas help slow the ions down as they travel through the drift

tube. In the drift region, the ions are separated based upon their charge and size. The

analytical signal is obtained when the ions reach the detector.

The larger the collisional cross section of the ion, the more collisions it

experiences and thus the more time it requires to reach the detector. Thus, higher

mobilities will be obtained by smaller ions, as shown in equations 1-1 to 1-4. Equation

1-1 demonstrates that the velocity (vd, cm s-1) of an ion is equal to its mobility (K, cm2 V-

1 s1) times the electric field (E, V/cm). An ion's mobility can be calculated using the

length of the drift tube (d, cm) and the drift time (td, s), as shown in equation 1-2.

d
Vd =KE = (1-1)
td

K = (1-2)
td E

The ion's reduced mobility (Ko) is calculated by correcting for standard

temperature (7T in Kelvin) and pressure (P, in Torr) conditions, as shown in equation 1-

3.

273 P
K = K (1-3)
(760 T)

At low electric fields, the relationship between the ion mobility and experimental

conditions is described by equation 1-4, where q is the ion's charge (, Nis the number









density of the drift gas, / is the ion's reduced mass, kis the Boltzmann constant, Tis

the drift gas temperature in Kelvin, and fD is the collisional cross section.36'37

1
K (3q )(2)2 1 (1-4)
16N [tkT lDD

For IMS, as for any separation technique, the ability to resolve closely spaced

peaks is of considerable interest. Resolving power, RpiMs, as shown in equation 1-5,

measures the ability of such an instrument to distinguish two tightly spaced peaks in a

single broad peak by dividing the drift time of the peak (td) by its width at half height

(Wi). Although high resolution IMS has been reported in which resolving powers
2

exceeded some chromatographic systems,38 resolving powers no greater than 30 have

been reported39 for most commercial instruments. Sensitivity levels of an IMS

instrument can be enhanced by increasing the electric field strength of the drift tube or

by altering the composition of the drift gas32,34

RptI = (1-5)
p IMS W1
2

Introduction to High-Field Asymmetric Waveform Ion Mobility Spectrometry

High-field asymmetric waveform ion mobility spectrometry (FAIMS) is a relatively

new technique that separates gas-phase ions at atmospheric pressure and room

temperature.14 FAIMS works similarly to IMS, in which the motion of ions is moved

along the drift tube with an electric field at atmospheric pressure. However, FAIMS

utilizes a non-homogeneous high electric field (>10,000 V/cm) to separate the ions,

whereas IMS uses a homogeneous voltage gradient in the 500 V/cm range.

Ions in IMS are separated according to their mobility, K, in a particular drift gas,

and their mobility is independent of the applied field; ions will move through the drift









tube at a velocity given equation 1-1. At the higher electric fields employed in FAIMS, K

becomes dependent on the applied field, and is better described as the ion's high-field

mobility, or Kh.40 The dependence of an ion's mobility on field strength allows FAIMS its

separation power. Specifically, ions are separated in FAIMS by their difference in

mobility at high (Kh) and low (K) electric fields. At a constant gas number density, N, the

non-linear dependence of the ion's mobility in high electric fields is given by

Kh(E) = K[l1+a ()2 + 4 + + (1-6)

where Kis the ion mobility coefficient at low electric fields and aand /Pare coefficients

found experimentally that describe the dependence of the ion's mobility at a high

electric field in a particular drift gas.40 Equation 1-6 is an infinite polynomial series, but

at realistic field intensities the terms above the fourth order become insignificant.41'42

Equation 1-6 provides a method to describe the change in mobility of an ion in a strong

electric field, and to ascertain the mobility of that ion as a function of the strength of the

electric field.

In FAIMS, one of the electrodes is typically held at a ground potential while the

other has an asymmetric waveform applied to it. The asymmetric waveform, described

by V(t), consists of a high voltage component (Vi), labeled in this dissertation as the

dispersion voltage (DV), which lasts for a short period of time relative to a longer lasting

low voltage component (V2) of opposite polarity. The waveform is constructed so that

the voltage-time product applied to the electrode is equal to zero, as shown in equation

1-7.

VltI + V2t2 = 0 (1-7)









At high electric fields, the application of this waveform will cause an ion to

experience a net drift toward one of the plates as illustrated in figure 1-2. Ions passing

between the electrodes encounter this net displacement from their original position

because the ion's mobility during the high-voltage component is different than that from

the low-voltage mobility.40 In other words, some ions will move farther during the high-

voltage portion than during the low-voltage portion, whereas others will move a shorter

distance. These ions will continue to migrate towards one of the plates and

subsequently be lost unless a constant direct-current compensation voltage (CV) is

applied to offset the drift. The CV required to offset the drift of different ions will be

different if the Kh/Kratio of the ions are different. Thus, a mixture of compounds can be

successfully separated by scanning the CV, and allowing each ion to transmit at their

particular CV, creating a CV spectrum.

When higher electric fields are applied to ions in FAIMS, the ion can have three

possible changes in ion mobility as illustrated in figure 1-3. At high-electric fields, the

mobility of type A ions increases with increasing electric field strength, whereas the

mobility of type C ions decreases, and the mobility of type B ions initially increases

before decreasing at yet higher fields.14 Different polarity waveforms can be used with

either positive (P) or negative (N) ions. In curved geometry, the FAIMS instrument is

operated in four modes; P1, P2, N1, and N2. The P and N describe the ion polarity and

the '1' and '2' indicate the waveform polarity. The waveform with positive DV yields

spectra of types P1 and N2, whereas the reversed polarity yields P2 and N1 type

spectra.43









The ability to explore how geometry affects the performance of FAIMS is essential

to thoroughly understanding its capabilities. It has been shown that in a flat plate

design, the ions are evenly distributed within the entire analyzer region between the

plates, and are continuously being lost to diffusion and space charge repulsion.22 When

curvature is introduced into the geometry of the cell, the ions are focused towards a

fixed radial distance.22 This generally improves transmission but limits resolving power.

With a curved geometry, there is still diffusion and space charge effects, but the ions

remain distributed within the focused region. Note that ion focusing in curved

geometries only occurs in the direction of the applied electric field and thus, lateral

diffusion still takes place.

FAIMS offers many advantages as an ion separation tool,44 since it is able to

separate gas-phase ions at atmospheric pressure and temperature, provides high

sensitivity with curved cell geometry, it separates and processes ions continuously, and

it interfaces easily to a mass spectrometer.

Quadrupole Ion Trap Mass Spectrometry

The quadrupole ion trap mass spectrometer (QITMS) has been widely adopted for

chemical analysis because of its selectivity, low detection limits, and broad

applicability.45 Briefly, an ion trap is any device that is capable of trapping and storing

ions in a confined space for a period of time. The operation of a quadrupole ion trap

(QIT) was initially described by Paul and Steinwedel in a patent filed in 1953 in

Germany and awarded as a U.S. patent in 1960.46 The QITMS traps ions in a

controlled manner and is similar to a quadrupole mass filter because it uses oscillating

electric fields. However, in the QIT, the ions are trapped in a three-dimensional space,

whereas the quadrupole mass filter only confines the ions into two-dimensions.47









Quadrupole Ion Trap Theory

The QIT allows for the confinement of ions in a small volume with a three-

electrode device consisting of a ring electrode placed between two endcap electrodes

(figure 1-4). The endcap electrodes are held at ground potential while radio frequency

(RF) (and sometimes direct current (DC)) potentials are applied to the ring electrode. A

quadrupolar field is created that can trap ions. The RF potential induces oscillations of

the ions to hold the ions within the QIT and is characterized by its amplitude (0 to 8500

Vo-p) and its frequency (typically 760 kHz). A potential well difference is created within

in the trap by applying a DC voltage (U) and an AC voltage (V), at a particular

frequency. This potential difference produces a dynamic electric field for the trapping of

ions.

Ions stored in the trap follow trajectories described by the Mathieu equation.

Solutions to the second-order differential Mathieu equation are in terms of two reduced

parameters, az and qz, which are defined in equations 1-8 and 1-9. These parameters

will determine whether or not the ion will possess a stable trajectory inside the trap.

16eU
az = -2ar = (1-8)
z r m(r2 +2zO))n2

8eV
qz = -2q,= = (1-9)
m(r +2 Z2)n2

where the subscripts zand represent axial and radial motion between and

perpendicular to the endcaps, respectively; Uis the amplitude of the DC voltage applied

to the ring electrode, Vis the RF potential applied to the ring electrode, e is the charge

on an electron (1.602 x 10-19 C), m is the mass of the ion, 12 is the drive frequency (21rf,

where fis the drive frequency in s-1), and ro and zo are the trap dimensions, as defined









in figure 1-4. From this equation, on can note that q, is directly proportional to the RF

amplitude on the ring electrode, and a, is proportional to the DC voltage applied.

An ion that is trapped follows a stable trajectory that has the general appearance

of a Lissajous motion as shown in figure 1-5. An ion that is unstable will have a

trajectory that increases in magnitude towards the endcaps or the ring, and thus will be

lost through collisions with the electrodes. In order for an ion to be effectively trapped,

its trajectory must be stable in both directions. The solution parameters, az and qz, can

be mapped to generate the Mathieu stability diagram (figure 1-6). This diagram

describes the common region in (az, qz) space where both the axial and radial

components of the ions trajectory are simultaneously stable. This region of stability is

defined by the boundaries at 3P = 0, 3P = 1, r = 0, and 3r = 1. An ion that possesses an

az and qz value confined by these values will be stable in both the r and z directions,

and will be available for mass analysis.

Finnigan LCQ A Commercial QITMS

All experiments reported in this research were performed on a commercial

benchtop Thermo LCQ QITMS. This instrument is equipped with two external

atmospheric pressure ionization (API) sources: atmospheric pressure chemical

ionization (APCI), and electrospray ionization (ESI). The system is easily operated in

both positive and negative ion mode.

For APCI, the sample solution in this research is infused by a syringe pump at a

flow rate of 15 pL/min, and vaporized by the standard LCQ APCI heated nebulizer.

Once ions are formed, the fully or partially desolvated ions are sampled by a heated

capillary, as shown in figure 1-7. The ions then pass through a tube lens and skimmer









cone and into a series of octapole ion guides before making it into the ion trap. Figure

1-7 shows a schematic of the Finnigan LCQ ion trap mass spectrometer.

The exit of the heated capillary, the tube lens, and the skimmer are held under

vacuum (1 Torr). The capillary is heated to help in the desolvation process. Ions are

then gated through the skimmer cone using the tube lens. A mass-dependent potential

is applied to the tube lens so the ions are focused towards the inlet of the skimmer. The

skimmer also acts as a vacuum baffle between the skimmer region (1 Torr) and the first

octapole region (10-3 Torr). The ions are then collected by the first octapole and

transmitted to the second octapole through an interoctapole lens. The interoctapole

lens assists in focusing of the ions as well as serving as a baffle between the first

octapole region (10-3 Torr) and the analyzer region (2x10-5 Torr).

The ion optics guide ions into the QIT mass analyzer where mass analysis occurs

(figure 1-4). In order to inject ions into the trap, an offset voltage of +10V for negative

ions and -10V for positive ions is applied to the mass analyzer electrodes. The LCQ QIT

operates along the az = 0 line in the Mathieu stability diagram (figure 1-6) i.e., the DC

potential difference between the ring electrode and the endcap electrodes is zero.)

The LCQ QIT utilizes resonant ejection, or axial modulation, to eject ions from the

ion trap. Mass analysis is accomplished by ramping the RF voltage on the ring

electrode, such that ions of increasing m/z consecutively come into resonance with the

axial modulation waveform. The ions gain kinetic energy and are ejected in the z

direction. The ion detection system includes a 15-kV conversion dynode and a channel

electron multiplier.









Atmospheric Pressure Chemical Ionization (APCI)

Gas-phase ions in this research are produced with an APCI source illustrated in

figure 1-8. APCI is a gas-phase ionization technique that allows the transfer of ions to

the gas phase through ion-molecule reactions at atmospheric pressure. The ionizing

reactions of APCI are similar to those of classic CI, although CI is operated under

vacuum (~1 Torr). Low-energy electrons are emitted from a corona discharge and

initiate the ionization in the source. Analytes in solution are introduced with a syringe

pump, are nebulized by a high-speed nitrogen sheath gas flow, pass through a

temperature-controlled heater to be vaporized, and are finally ionized by the corona

discharge needle. The corona discharge initiates chemical ionization at atmospheric

pressure using nitrogen and vaporized solvent as the reagent gas. APCI provides high

ionization efficiency since the mean free path of ions at 760 Torr is very short and the

number of collisions between reagent ions and analyte molecules is consequently very

high.

In APCI, the degree of fragmentation is generally quite low, determined by the

amount of energy transferred in the ion-molecule reactions. These ion-molecule

reactions include proton transfer, charge exchange, electrophilic or nucleophilic

addition, and anion abstraction or nucleophilic displacement. For negative ions, proton

transfer takes place when a reagent ion (X-) accepts a proton from an analyte molecule

(MH), as illustrated in equation 1-10. Since H20 has the highest proton affinity of those

gases normally present at ambient conditions, it captures any free protons and, upon

collisions with other molecules of H20, forms a series of protonated water clusters.47

X- + MH XH +M- (1-10)









An alternative ionization method to proton transfer is charge exchange. This

method typically occurs when the analyte molecule does not contain any acidic or basic

sites. Charge exchange occurs in the negative mode when the reactant ion (usually 02-)

has a recombination energy higher than the ionizing energy of the sample. Additionally,

with negative ions, stable complexes may be formed by nucleophilic attack (equation 1-

11) with reagent ions of low proton affinity, such as Cl- and 02-. Finally, anion

abstraction can also occur with negative ions (equation 1-12). These reactions occur

when reagent ions react with sample molecules, resulting in fragment ions from the

sample.

M + X- -MX- (1-11)

AB + X- BX + A- (1-12)

Overview of Dissertation

This dissertation is organized into four chapters. Chapter 1 is an introduction to

the research and gives background information on ion mobility spectrometry (IMS) and

high-field asymmetric waveform ion mobility spectrometry (FAIMS). Chapter 1 also

briefly discusses the instrumentation that was utilized in this research, including ion trap

mass spectrometry and atmospheric pressure chemical ionization (APCI).

Chapter 2 explores the fundamental and practical issues of curved electrode

geometries in FAIMS. Explicit geometries explored include 3-D curved geometries

(spherical and hemispherical) with comparisons to a 2-D curved geometry (cylindrical).

Results from these experiments help provide a greater understanding of the relationship

of cell geometry to ion transmission and resolution.

In Chapter 3, a novel FAIMS waveform approach is evaluated. The advantages

and disadvantages of the application of the asymmetric square waveforms compared to









the traditional sum-of-sine waveform will be discussed. Effects of waveform frequency,

duty cycle, and voltage pulse amplitude are also discussed.

Chapter 4 summarizes the major conclusions presented in this work and gives

suggestions for future work in this area. The advantages and disadvantages of curved

geometries and different waveform approaches for FAIMS are evaluated.










Drift Gas Inlet


Wire Mesh


Gas Flow


Needle
Highest Voltage
Applied


Drift Region


Drift Rings Insulators


Figure 1-1. Schematic of a characteristic ion mobility spectrometer using the stacked
ring design.


+4000 V





OV0


-2000 V


4i *

-..... .... **........T
I- -I-Iil i T


Figure 1-2. Ion motion between two plates during the application of an electric potential.
A simplified asymmetric waveform is applied to the upper plate.40


APCI Probe
I A


T. Inlet of MS


a a a 111a a a
















1.0
S1.

I 5 Increasing Electric Field Strength--
0.95

Figure 1-3. Hypothetical plots of the dependence of ion mobility on electric field
strength for three types of ions.14


























IONS I IONS
IN TRAPPED OUT
IONS





SECOND
OCTAPOLE EXIT LENS

EXIT ENDCAP

RING
ELECTRODE


SFaCER NUT
ENTRANCE ENDCAP RINGS SR ING
ELECTRODE POST WASHER


Figure 1-4. Cross sectional view of mass analyzer in the QITMS.48






























Figure 1-5. Ion motion in a quadrupole ion trap.49











0.21

0.I 02







-0.4











O2 0.4 0.6 \1'0 1- 2 1,4 1.6
-0.2 iVr









-0.7 -i : "*'I*




Figure 1-6. Mathieu stability diagram plotted in a and q space for positive ions. The
shape of the stable region is rotated about the az =0 line for negative ions.
Most commercial QITMS instruments operate along the az =0 line (RF-only
operation). Under these conditions, ions will have a stable trajectory between
qz = 0 and 0.908. The 3z =1 boundary intersects the qz axis at qz = 0.908.50





















IONS
oUT "


ANALYZER
-- MOUNT


Figure 1-7. Schematic of Finnigan LCQ ion trap mass spectrometer.48























AP CI
FLANGE


INLET

SHEATH
GAS
INLET

APCI PROBE CORONA _


Figure 1-8. Schematic of the atmospheric pressure chemical ionization (APCI) source

used in the LCQ.48


INTEROCTAPOLE


?-,. -
TUNSLES

"/ R HOUNI









CHAPTER 2
FUNDAMENTALS OF FAIMS FOR CURVED ELECTRODE GEOMETRIES

Introduction

Buryakov et al. introduced FAIMS with a parallel plate geometry in 1993.20 Since

then, Carnahan et al. improved the sensor design by replacing the flat plates with

concentric cylinders.51 Spherical electrodes were also proposed but never reduced to

practice.52 53 It has been theorized that curved geometries help to produce significantly

better ion transmission efficiency because the electrode curvature allows unique ion

focusing capabilities.22' 23, 54 Ion losses in the analysis channel of planar electrodes

arise mainly due to ion diffusion and space-charging. Krylov proposed a method to

reduce diffusion losses by employing cylindrical electrodes in FAIMS.22 With curved

electrodes, there exists a non-uniform electric field. This electric field gradient points in

the same direction as the vector of the compensating field,22 and leads to a restoring

force that acts on the diffusing ions, leading to compression of the ion cloud.

Although the use of curved surfaces increases sensitivity, resolution is

compromised.52 Decreased resolution is due to the ability of spherical or cylindrical

surfaces to focus ions which encompass a wide range of mobilities. Since the non-

uniformity of the field allows ions with a range of different mobilities to be focused

together, they will exit the FAIMS cell at the same time, thus increasing peak width and

decreasing resolution.

The development of novel hemispherical FAIMS electrodes to balance

transmission and resolution has been pursued in this research. In a spherical or

hemispherical cavity, all ions encounter equidistant trajectories from ion inlet to outlet,

increasing focusing capabilities and transmission. The increased transmission can









allow the utilization of larger electrode radii, which will help to increase resolution.

Figure 2-1 shows the novel hemispherical cell design and table 2-1 displays all relevant

dimensions in comparison to the commercially available Thermo-Fisher Scientific

cylindrical cell. An lonalytics cylindrical cell was also employed in this research, as

noted in table 2-1. The main difference between this cell and the Thermo-Fisher cell is

the gap between the two electrodes (2.00 mm vs. 2.50 mm).

The strength of the focusing field between the two electrodes is inversely related

to the radius of the electrodes. The smaller the radius of the electrodes, the stronger

the focusing field will be; the larger the radius, the weaker it will be. A systematic

exploration of the effect of electrode radius on sensitivity will be discussed in this

research, with a comparison of the differences in sensitivity and resolution of the two

cylindrical cells.

Description of Planar Geometry

Buryakov and others first introduced a parallel plate FAIMS geometry in 1993.20

They showed that when an asymmetric waveform is applied to one of the plates while

keeping the other plate grounded, a uniform electric field is created that allows for the

transmission of only certain ions (figure 2-2). In this type of geometry, the ions are

evenly distributed among the entire gap between the two electrodes. Ion losses in the

gap of planar electrodes mainly come from lateral diffusion and space charging.

The motion of a positively charged ion in a flat plate FAIMS during the application

of a hypothetical asymmetric waveform is displayed in figure 2-3. Ions are separated in

FAIMS while they are carried by a continuous flow of gas between two closely spaced

electrodes. The waveform employed by FAIMS is characterized by significant

differences in voltage in the positive and negative polarities of the waveform. In figure









2-3 the mobility of a positive ion is higher during the positive (high voltage) portion of

this waveform than during the negative (low voltage) portion. This ion will keep

migrating towards the bottom electrode and be discharged if it remains on this path. A

low direct current (DC) potential is superimposed to offset the drift of the ion. This DC

potential is termed its compensation voltage (CV). The CV required to offset the drift of

individual components in a mixture of compounds will be different if the change in the

ions mobility between high and low field is different. Consequently, a mixture of

compounds can be successfully separated by allowing each compound to transmit at its

particular CV; scanning through these CV values creates a CV spectrum. The electric

field (Efat) between planar electrodes can be calculated using equation 2-1, where V is

the applied voltage, and (b -a) is the gap between the electrodes.

Eat = 2-1
flat (b-a)

Description of Cylindrical Geometry

Curved geometry FAIMS was introduced by Carnahan and others in 1995.51 A

schematic of this geometry is displayed in figure 2-4. Curved geometry FAIMS is

currently commercialized by Thermo-Fisher Scientific and is sold as a separation device

to be used in tandem with a mass spectrometer. A schematic of the Thermo-Fisher cell

is displayed in figure 2-5. The FAIMS process is the same as in a planar geometry a

mixture of ions may be filtered to one specific ion by selecting an appropriate

compensation voltage. The difference arises when the application of an electric field is

between curved surfaces. The field is no longer uniform and the non-uniformity of this

field is what helps to focus the ions towards a fixed radial distance within the analyzer

region. As more ions are focused towards the center, fewer ions are lost to the walls. In









essence, two-dimensional atmospheric-pressure ion focusing occurs within the

analyzing gap.

The electric field (Er) in a cylindrical geometry can be calculated at any point within

the analyzer region by using equation 2-2

E= 2-2
[r In~ (a

where V, is the applied voltage, and as shown in figure 2-5, ris the radius of the ion at

any point between the electrodes, a is the radius of the center electrode, and b is the

radius of the distance between the center of the inner electrode and inner surface of the

outer electrode.

Ion focusing in a curved geometry cell is achieved as ions near the inner electrode

are repelled to the center of the analytical gap, and ions near the outer electrode are

pulled back to the center of the analytical gap. Although ions experience a force that

directs them towards a focus point it is important to note that the force is not necessarily

strong, and the ions are distributed in space around this focus point due to the

influences of diffusion, and ion-ion repulsions (space-charging).

This effect was experimentally described by Barnett et al. and is represented in

figure 2-6. The CV that the ion needs to traverse the electrodes without being lost is

displayed by the solid line. However, the variation in the electric field allows many ions

with different CVs to be focused within the analyzer region. The electric field gradient

for an example DV where the CV is the opposite polarity of the ion is represented by the

dotted line CVcorrect in figure 2-6. The ions that are near the inner electrode (position of

highest electric field) will be repelled away towards the center of the analyzer region

where they will fall on the solid line and consequently be focused. Comparatively, ions









that are near the outer electrode will be pulled towards the center, where electric field is

higher, and also be focused. Ions that would normally collide with either the outer

electrode or inner electrode are represented by the dotted lines CV0ow and CVhigh,

respectively. Since the field is not uniform and is higher near the inner electrode, the

ion is pulled away from the outer electrode or inner electrode and successfully focused.

Lastly, the dotted line CVincorrect displays an ion with a CV that, despite focusing, is not

strong enough to deter the ions from the outer electrode.

While curved geometries in FAIMS increase transmission and thus sensitivity, they

will decrease resolution. Since curvature aids in focusing a wide range of mobilities,

peak widths will increase and resolution will decrease. The opposite is also true; if the

radii of curvature is increased, the focusing effect lessens and resolution increases at

the expense of sensitivity.

Each cell geometry has its advantages and disadvantages regarding the

optimization of resolution and ion transmission due to the enhanced focusing that

results in improved sensitivity. If the radius of curvature is decreased, focusing

improves and ion intensity increases at the expense of resolution. Clearly, there is a

compromise between resolution and transmission for both cylindrical and planar cell

geometries. Thus it is interesting to consider geometries with curvature in both

directions, i.e., spherical.

Description of Hemispherical Geometry

In a spherical cavity, all ions encounter equidistant trajectories from ion inlet to

outlet, increasing its focusing capabilities, transmission, and uniformly spaced ion

density. Nevertheless, the design and construction of a spherical cell is a daunting task;

the suspension of a central spherical electrode at the exact center of a spherical cavity









while delivering several thousand volts of RF to the central electrode is extremely

difficult.

Initial spherical designs failed to provide the precise, accurate, and rigid centering

required of the center electrode. Therefore, a second design employed a hemispherical

central electrode mounted within a hemispherical cavity, as shown in figure 2-1. This

design has shown to provide the advantages of a fully spherical cell while permitting

rigid, accurate, and reproducible positioning of the electrodes. Characterization and

performance of the hemispherical FAIMS cell in comparison to the cylindrical cell will be

reported.

The material of the inner electrode was stainless steel and was machined into a

near hemisphere and fastened with metal screws to a Kel-F base plate. Kel-F, a

polymer of chlorotrifluoroethylene, has the lowest vapor transmission rate of any plastic,

excellent thermal characteristics and maintains an operating temperature range of -204

C to 204 C. Fastening the inner electrode to the base plate ensured accurate

centering and a uniform analytical gap.

In order for the ions entering the analyzer region to 'see' a true hemispherical

electrode, the inner electrode was extended 2 mm from a true hemisphere. This way,

any path taken by the ions from inlet to exit would be identical. The asymmetric

waveform and the compensation voltage were applied to the inner electrode through the

mounting screw that was threaded in the center of the electrode (part 6 in figure 2-1).

Photos of the hemispherical FAIMS cell are shown in figures 2-7 and 2-8.

The voltage (Vhemi) and the electric field (Ehemi) in a hemispherical geometry can be

calculated at any point within the analyzer region by using equations 2-3 and 2-4









Vhemi = 2-3



\b a]
Ehemi 2-4
(- ')r

where A Vis the change in voltage, a is the inner electrode radius, b is the outer

electrode radius, ris position, and V, is the inner electrode voltage. In a planar cell, the

electric field does not vary across the gap between the electrodes, as described by

equation 2-1 and displayed by figure 2-9. The cylindrical cell, because of its curvature,

is capable of focusing ions as described by equation 2-2 and displayed in figure 2-9 by

the blue trace. In the cylindrical cell the field varies non-linearly and is higher at the

inner electrode than the outer. The hemispherical cell offers higher sensitivity than that

of the cylindrical geometry because of its curvature in both directions. The ions that are

located at various radial distances between the inner and outer electrode of the

hemispherical cell will more effectively fall into the atmospheric pressure ion trapping

region because the focusing decreases ion loss to the walls.

Figure 2-9 displays how the electric field changes as a function of position for the

three cells. As can be seen in the figure, the electric field for the hemispherical cell

(green trace) is much higher near the inner electrode than for the other two geometries.

This is important, for instance, for a cell that has the same dimensions and the same

applied voltages. One would get more efficient separations without having to do any

other modifications.

Description and Design of a Spherical Geometry

Guevremont et al. patented spherical electrodes in 2004, but the patent was never

reduced to practice.52 An initial design (figure 2-10) to suspend an inner electrode

within a spherical cavity and not disrupt the electric field was constructed without any









success. The design (figure 2-10) included a spherical inner electrode that consisted of

a 0.7 mm stainless steel rod drilled through the center of an 8.8 mm stainless steel ball.

In order to deliver the asymmetric waveform and compensation voltage, the inner

electrode needed to be suspended in this manner.

In order to minimize the disruption of the electric field within the analytical gap of

the FAIMS cell, two MACOR sleeves (4 mm in diameter) were utilized to insulate the

stainless steel rods. Minimization of any horizontal movement of the inner electrode

was attempted by adding MACOR cuffs to the MACOR sleeves. The outer electrode

was made of two adjoining brass hemispherical cavities held together tightly with thumb

screws. The entire cell was encased in Delrin.

Unfortunately, this design was not successful. The primary reason for the poor

transmission was probably imprecise centering of the electrode once the cell was

assembled. Second, the MACOR sleeves used were very large in comparison to the

size of the inner electrode. Ions entering the analyzer gap could theoretically be

attracted to the insulating MACOR sleeves and lose their initial trajectory. Also,

temperatures in excess of 150 OC were being used on the heated capillary of the MS.

Delrin has an upper temperature limit of 175 OC.

As mentioned previously, the suspension of a central spherical electrode exactly at

the center of a spherical cavity while delivering several thousand volts of RF to the

central electrode is extremely difficult. Initial designs failed to provide the precise,

accurate, and rigid centering required of the electrode. Thus, design of a novel

spherical FAIMS cell incorporating a spherical central stainless steel electrode mounted









within a spherical stainless steel cavity that has been cut in half to allow for assembly

was created.

This design has been shown to provide the ion optical advantages of a fully

spherical cell while permitting rigid, accurate, and reproducible positioning of the

electrodes. Characterization and performance of the spherical FAIMS cell shows an

improvement in sensitivity and resolution when compared to the cylindrical FAIMS cell

and will be discussed in this research.

A schematic, dimensioned cutaway drawing, and pictures of the novel spherical

FAIMS cell are shown in figures 2-11 2-14, respectively. Both inner and outer

electrodes are made of stainless steel. Outer casing is made of Ketron PEEK which is

a polymer of polyetheretherketone. The melting point of Ketron is 340 C. The

asymmetric waveform and compensation voltage is applied by a stainless steel rod that

has been drilled through the center of a 25.4 mm diameter inner electrode. A

comparison table of important dimensions between all three cells (spherical,

hemispherical, and Thermo- Fisher cylindrical) is displayed in table 2-2.

Experimental

Instrumentation

The instrumentation used in these experiments was a Thermo LCQ benchtop

quadrupole ion trap mass spectrometer (Thermo-Fisher Scientific, San Jose, CA). The

traditional sinusoidal waveform was created by either an lonalytics GPI 1000 FAIMS

system alpha prototype waveform generator or a Thermo-Fisher FAIMS waveform

generator. Both generators run with a 33% duty cycle at a frequency of 750 kHz. The

voltages for the waveform, compensation voltage, outer electrode, and curtain plate are

all supplied by these two waveform generators. The lonalytics generator required the









compensation voltage to be scanned using an external Agilent 33120-A arbitrary

waveform generator, (Loveland, CO). The Agilent generator is programmed to generate

linear voltage ramps in either positive or negative directions to scan the compensation

voltage. The Thermo-Fisher generator provided the CV ramp so no external generator

was required. Negative ions were created by atmospheric pressure chemical ionization

(APCI). Analyte samples were infused at 15 pL/min.

The LCQ does not require any modification to the interface for attachment of the

FAIMS assembly. A PEEK adapter was fitted onto the portion of the heated capillary

that extends beyond the API stack spray shield (figure 1-7). The FAIMS cell is fitted into

the PEEK adapter. The carrier gas used in the FAIMS apparatus, nitrogen (N2), was

passed through an individual gas purification cartridge equipped with molecular sieves

for removal of moisture and activated charcoal for removal of organic impurities. Flow

rate for nitrogen was controlled by an MKS M100B thermal mass flow controller

(Andover, MA) calibrated for the individual gas.

Samples

For these experiments, several structural classes of common explosives were

utilized (shown in figure 2-15), including nitroaromatic explosives such as 2,4,6-

trinitrotoluene (TNT), 1,3,5-trinitrobenzene (TNB), 2,4-dinitrotoluene (2,4-DNT), and 1,3-

dinitrobenzene (1,3-DNB), 3,4-dinitrotoluene (3,4-DNT); nitramine based explosives

such as 1,3,5-trinitroperhydro-1,3,5-triazine (RDX) and 1,3,5,7-Tetranitro-1,3,5,7-

tetraazacyclooctane (HMX); and nitrate esters pentaerythritol tetranitrate (PETN) and

1,2,3-propanetriol trinitrate (nitroglycerin, NG). Explosive standard solutions were

supplied by Dr. Yehuda Yinon of the Weizmann Institute of Science (Analytical

Laboratory, Israeli Police Headquarters).









Negative ion mode was employed since the above explosives tend to display high

electron affinities. Furthermore, less chemical noise is generated in negative ion

polarity mode thus increasing the selectivity of the analysis.48 In APCI, negative ions

are typically produced by proton abstraction to obtain an [M-H]- ion55 except for TNT,

TNB, and DNT isomers which produce mostly the [M]- ion plus other common

fragments.

All explosive solutions were diluted in a solvent mixture consisting of 65%

methanol and 35% deionized water (65:35, v/v). Utilizing a combination of methanol

and water as a solvent allows the use of lower vaporizer temperatures with the APCI ion

source to avoid the degradation of fragile explosive analytes. Additionally,

approximately 0.1% of carbon tetrachloride was added for nitramine and nitrate ester

based explosives. The addition of carbon tetrachloride allows a stable explosive adduct

ion, [M + CI]-, to be formed. Anticipated ions and adducts for nitramine and nitrate ester

explosives are shown in table 2-3. For these experiments, all solutions were diluted to a

concentration of 10 ppm unless otherwise noted.

Results and Discussion

A mixture of explosives, as well as individual analyte solutions, was tested with

each geometry to evaluate signal intensity, compensation voltage, resolution and

resolving power. The resolution between peaks in a mixture can be evaluated by

equation 2-5.

2(ACV)
Rs = (CV) (2-5)
(PW1 10%+ PW2 10%)









where ACVis the difference in CV values of the two peaks and PWi and PW2 are the full

width at 10% of peak 1 and 2, respectively. The resolving power for FAIMS (Rp FAMS) is

an instrumental parameter described by equation 2-6

CV
Rp FAIMS = PWHM (2-6)

where CVis compensation voltage and PWHMis the peak width at 50% peak height.

The CV of any particular ion is calculated by using equation 2-7.
. peak time7
CV = ptim scan range start of scan (2-7)
[total scan time

Comparison of Signal Intensity

The spatially inhomogeneous field in curved geometry FAIMS not only separates

ions but also focuses the ions that are at their correct CV values. Thus, the

transmission of ions increases with increasing curvature. Figure 2-16 shows a CV scan

(figure 2-16, left) and mass spectrum (figure 2-16, right) of TNT for all three (cylindrical,

hemispherical, and spherical) FAIMS cells. Note that these set of experiments were

done with the cylindrical FAIMS cell that has a 2.00 mm gap (lonalytics cell).

Separation parameters for all three cells include: DV of -3214 with a CV scan from 0 to

+10 V over 2 minutes (5 V/min). The CVs of TNT for all three cells are all -4.60 V.

Signal intensity for the hemispherical FAIMS shows eight-fold increase from that of the

cylindrical cell. The spherical FAIMS cell gave an intensity which is a 23-fold increase

from the cylindrical cell. The negative APCI spectrum of TNT shows the production of

[M]- at m/z 227. The mechanism that occurs with APCI with TNT is assumed to be

electron capture since TNT has three bulky electron-withdrawing nitro groups.56' 57 All

three TNT spectra also display two low-intensity fragment ions, [M-NO]- at m/z 197 and

[M-OH]- at m/z 210, which may form during the transition of the ion from atmospheric









pressure to vacuum through the heated capillary or in the tube lens skimmer region

where ions are declustered from solvent vapor.

This behavior of increasing intensity with increased FAIMS electrode surface area

was verified with TNB and 2,4-DNT, as displayed in figures 2-17 and 2-18, respectively.

For figure 2-17 the separation and scanning parameters include: DV of -3214 with a CV

scan from 0 to +10 V over 2.5 minutes (4 V/min). The CVs of TNB for all FAIMS cells

are -6.10 V. Signal intensity for the spherical FAIMS cell (7.50 x 105) shows a 13-fold

increase from the cylindrical FAIMS and the hemispherical FAIMS shows a 3-fold

increase from the cylindrical cell. The negative APCI spectrum of TNB is shown in

figure 2-17 and it displays the major ion as m/z 213, [M]-. The major fragment that is

produced in all three spectra is m/z 183, [M-NO]-. Figure 2-18 shows a compensation

voltage scan and a mass spectrum of 2,4-DNT for all three cells. Separation and

scanning parameters are the same as with TNB. The CVs of 2,4-DNT for all three cells

is ~7.50 V. Signal intensity for the hemispherical FAIMS shows a 5-fold increase from

that of the cylindrical cell. The spherical FAIMS cell gave an intensity value of 7.27 x

105 which is 17-fold increase from the cylindrical cell. The APCI spectrum of 2,4-DNT

for all three cells show the molecular ion at m/z 181 [M]- and its major fragment at m/z

135 [M-N02]-. This experiment was repeated for several other explosives including

RDX, HMX, NG, and PETN. Table 2-4 shows these results. Note that the values in

table 2-4 were taken as three repetitions of the compensation voltage scan. The

average of the centroid peak and the standard error of standard deviation of the mean

(SE) is reported in the table. Unfortunately, HMX and NG were not able to be detected

with any of the FAIMS cells while scanning the compensation voltage. This can be









attributed either to not scanning the CV slowly enough to be able to detect the analyte

peak or to the fragility of these ions causing the molecular ion to fragment before it

reached the ion trap.

An investigation of curved geometry FAIMS electrodes has shown that higher

signal intensity is achieved with the spherical and hemispherical cells. It is assumed

this is due to the two-dimensional focusing within the analytical gap of these cells. The

stronger the curvature is (when comparing between cylindrical and both the

hemispherical and spherical), the more powerful the electric field gradient will be and,

thus, the better the sensitivity. It should be noted that even though the hemispherical

and spherical cells have the same radius of curvature, they obtained different intensity

values. One would assume that a cell with the same radius of curvature should provide

identical sensitivity values. The higher sensitivity of the spherical cell could be due to its

larger surface area since that is the only real notable difference between the two cells.

Further studies with cells of different dimensions, would need to be performed in order

to verify this assumption. It should also be noted that there is no significant difference in

compensation voltage within the three CV scans of each cell, exemplifying the

reproducibility of the hemispherical and spherical FAIMS cells.

Comparison of Resolving Power and Resolution

Figure 2-19 displays one CV scan of a 2 ppm mixture of TNT, TNB, 2,4-DNT, 1,3-

DNB, and PETN on the Thermo-Fisher cylindrical cell (2.50 mm gap). The total ion

current (TIC) is shown in (A). Intensities of TIC and extracted chromatograms, RpFAIMS,

and CVs are also shown in figure. The compensation voltage for figure 2-19 was

scanned from -5 to +15 V over 2 minutes (10 V/min) with a dispersion voltage of -4000

V. This dispersion voltage corresponds to an electric field of 16,000 V/cm or 80









Townsend (Td). A Townsend is a unit of the E/N ratio, where E is the electric field and

N is the gas number density (1 Td = 10-17 V cm2). The same explosives mixture was

analyzed on the hemispherical (figure 2-20) cell. Hemispherical cell scanning

parameters include a dispersion voltage of -3214 V and a CV scan from -1 to +19 V

over 2 minutes (10 V/min). This dispersion voltage corresponds to an electric field of

16,072 V/cm and/or 81Td. Even though the absolute scanning range of the Thermo-

Fisher cylindrical cell and the hemispherical cell are not identical, the rate of scanning

(10 V/min) is the same. Resolution between peaks was calculated using equation 2-5.

Resolution and resolving power values are clearly higher for the hemispherical cell. The

higher resolution in the hemispherical cell can be attributed to its bigger inner electrode

radius and better focusing capabilities. The radius is twice the size of the cylindrical

cell, giving it a more planar-like feature, thus increasing its resolution. Also, the path

length through the cell is ~2x longer which gives 2x longer residence time, and thus 2x

more RF cycles. The bigger inner electrode radius allows for increased resolution with

increased transmission.

The same explosives mixture was attempted on the spherical cell, the results are

shown in figure 2-21. The scanning parameters for the spherical cell include a

compensation voltage scan from -1 to +19 V over 2 minutes (10 V/min) with a DV of -

3214 V. As seen above, the spherical FAIMS displays an improvement in sensitivity

(typically ~4x) for each of the extracted chromatograms and TIC compared to the

hemispherical cell. Note also that the spherical cell exhibits a ~2x decrease in resolving

power and resolution from the hemispherical cell. Again, this could be attributed to its

larger surface area. A larger surface area gives a higher potential for distortion of the









electric field which, in turn, could prevent the ion cloud to exit the cell at the appropriate

time and thus broaden its peak width. Another possible explanation for the sphericals'

lower resolution could be longer theoretical residence times within the analyzer region.

The residence time of the ion denotes how long that ion remains within the gap before

exiting towards the mass spectrometer. The calculated residence times for the

spherical and the hemispherical cell are identical (114 ms) but the ion does have a

larger surface area within the spherical cell which could increase its trajectory time.

Theoretically, if the ion has a longer trajectory and remains in the gap for a longer time

its peak width will be broader as it exits the cell. Tables 2-5 and 2-6 illustrate specific

figures of merit for all three cells to fully compare them.

In order to evaluate whether these trends were affected by dispersion voltage,

another 2 ppm mixture including TNT, TNB, 3,4-DNT, 2,4-DNT, and 1,3-DNB was

attempted with all three cells at a higher dispersion voltage. These experiments were

performed with the lonalytics cylindrical cell (2.00 mm gap). The scanning parameters

for all three cells included a dispersion voltage of -4700 V (23,500 V/cm) with a

compensation voltage scan from 0 to +20 V over 4 minutes (5 V/min). These results are

displayed in figures 2-22 (cylindrical), 2-23 (hemispherical), and 2-24 (spherical).

Resolving powers, CVs, and intensities for the chromatograms are shown on each

figure. The figures of merit for the three cells at 4700 DV are compared in tables 2-7

and 2-8. As in the lower dispersion voltage experiments, spherical FAIMS displayed an

improvement in sensitivity but a decrease in resolution when compared to the

hemispherical cell. Even though the resolution and resolving power values were lower

for the spherical cell at both dispersion voltages, they still showed an improvement from









both the two cylindrical cells. Attempting to maximize both resolution and sensitivity is

dependent upon how well the incorporation of both 'planar-like' and 'curve-like' qualities

of the cell are employed.

Sensitivity of Spherical FAIMS

Since FAIMS is able to separate analyte ions from chemical background noise and

thereby enhance selectivity, an evaluation of the analytical capabilities of the spherical

FAIMS cell was carried out in this research. Evaluation of CV repeatability, linear

dynamic range, and limit of detection for the explosive TNT was performed.

Repeatability of compensation voltage in FAIMS is a crucial parameter that

describes the consistency of the analytical approach. A repeatable compensation

voltage is similar to a reproducible retention time in chromatography and can be treated

as a verifiable index of identifying the explosive compound. The repeatability of CV

value was tested at ten different dispersion voltage values each with five repetitive

injections of a 10 ppm TNT sample while employing the same conditions over a period

of several hours. Standard deviations (SDs) and relative standard deviations (RSDs)

were calculated by analysis of variance. As shown in table 2-9, better repeatability was

obtained at higher dispersion voltages and this is assumed to be due to the increased

ion focusing and thus increased ion transmission of analyte ions. This results in a more

symmetrical and intense CV peak, which in turn, generates more precise and

reproducible CV values.

In order to fully validate the spherical FAIMS method, limit of detection (LOD) and

linear dynamic range were also characterized. LOD describes the lowest concentration

level of the analyte that may be determined to be statistically different from a blank

sample.58 In this research, the method validation values are reported in a format









consistent with the International Union of Pure and Applied Chemistry (IUPAC)

definitions. The LOD was taken as three times the standard deviation of the blank

signal, expressed in concentration. Blanks in this research are defined as infusing the

solvent solution (with no analyte) while scanning the mass spectrometer for similar

times as actual runs. The signal intensity of blanks were averaged and utilized for

statistical analysis. Solutions were flowed at 15 pL/min at concentrations of 0, 1, 10, 50,

100, 250, 500, 1000, and 10000 ng/mL (1 ppb to 10 ppb) were made and the mass

spectrometer was scanned in the full-scan and selected ion monitoring (SIM) mode.

The CV was scanned from 0 to +10 V over two minutes (5 V/min). Linear dynamic

range (LDR) was evaluated by five replicate injections of standard mixtures at eight

different levels of concentration ranging from 1 ppb to 10 ppm. The lowest point of

quantitation, specifically, the concentration yielding a signal that is ten times the

standard deviation of the blank. The highest point refers to the highest concentration

that shows a linear dependence of the intensity on concentration. Results of these

experiments are shown in table 2-10. Based on the data shown, the LOD of TNT

collected in full scan MS was 4 ng/mL and was 1 ng/mL in SIM mode. The correlation

coefficient (r2) value for the regression was 0.9951 for full scan and 0.9968 for SIM

mode, demonstrating a LDR of ~3 orders of magnitude for the FAIMS analysis of TNT.

Even though higher LODs and lower correlation coefficients were obtained with full

scan, it is still a vital mode for research analysis as more information about molecular

and fragment ions is provided.

FAIMS separation is based on a different principle than chromatography, even

though it provides chromatogram-like data. In contrast to liquid or gas chromatography,









where the analytes pass through the column over a window of retention times and are

detected, FAIMS can be used to allow only specific analytes to traverse the cell by

setting a particular compensation voltage. Thus, FAIMS/MS is more analogous to

MS/MS than to GC/MS or LC/MS.

Conclusions

In this chapter, an investigation of three-dimensional and two-dimensional curved

electrodes with respect to signal intensity, resolution, and resolving power was shown.

Although the use of curved surfaces increases sensitivity, resolution is decreased.

Decreased resolution is due to the ability of cylindrical, hemispherical, or spherical

surfaces to focus ions which encompass a wide range of mobilities. Since the non-

uniformity of the field allows ions with different mobility behaviors to be focused

together, they will exit the FAIMS cell at the same time, thus increasing peak width and

decreasing resolution. This effect is more pronounced with increasing curvature and/or

surface area as was displayed in the data. To balance transmission and resolution, the

development of novel hemispherical and spherical FAIMS electrodes were employed in

this research.

The highest sensitivity for the explosives was obtained with the spherical FAIMS

electrode. Additionally, the spherical FAIMS electrode allowed better resolving power

and resolution values than both the Thermo-Fisher cell cylindrical (2.50 mm gap) and

the lonalytics cylindrical cell (2.00 mm gap). The increased transmission allows the

utilization of larger electrode radii, which can help to increase resolution.

The sensitivity of spherical FAIMS was explored to provide information on its

quantitative capability. The best CV repeatability was obtained at higher dispersion

voltages (-2000 to -3378 V). The standard deviations for TNT ranged from 0 to 0.04









and the relative standard deviations ranged from 0 to 2.3%. Two MS scan modes, full-

scan, and SIM, were tested for LOD and LDR. Limit of detection for TNT was 4 ppb for

full scan and 1 ppb for SIM. Although full-scan MS provided higher LOD and LDR

values, it is still a vital explosive analysis mode because it provides more information

about the ion.















3P










Figure 2-1. Schematic of hemispherical FAIMS electrodes. (1) outer electrode, (2) outer
housing, (3) inner electrode, (4) exit port, (5) base, (6) inner electrode
connection, (7) entrance port, and (8) curtain plate.


Selected Ions Out
of FAIMS
C*


Figure 2-2. Ions separated between flat plates in the FAIMS analyzer. The transmitted
ion is distributed evenly over the gap between the plates. Ions are
continuously lost through diffusion and space charge repulsion. 59











V(t)



Gas Flow




Figure 2-3. Motion of a positively charged ion in a flat plate FAIMS.21
Figure 2-3 Motion o in lapla M.





Figure 2-3. Motion of a positively charged ion in a flat plate FAIMS.21


Mixture Into FAIMS

1* Z


Selected Ions Out
of FAIMS

1*=


Figure 2-4. Ions separated between concentric cylinders of a FAIMS analyzer. The
transmitted ion is focused towards a fixed radial distance, and distributed
around that distance through the effects of diffusion and space charge
repulsion. The effect of the focus is exaggerated for clarity in this figure.59













Ion inle
ion ine- Ion outlet



Inner electrode Outer electrode


b)
Gas flow a






Figure 2-5. Three-dimensional schematic and cross section of the Thermo-Fisher
cylindrical geometry FAIMS. (a) Three-dimensional view and (b) cross
section view illustrating radial location, r, between the walls of the inner
electrode, 'a' and outer electrode 'b'. The asymmetric waveform is applied to
the inner electrode.60


Ions pulled away
from the outer
electrode



9-- ~- ..- CVlow
C- 'Vcorrect
-- ----
CVhjsh


Ions repelled
away from the
inner electrode


SI I I I I I I I I I I
50 60 70 80
Dispersion Field (DF)


Figure 2-6. Focusing effect caused by curvature of electrodes in FAIMS. Adapted from
Barnett et al.60


-40-





-20-





0-







Outer
Electrode


Inner
Electrode


Curtain
Plate


Kel-F@
7 Casing


Figure 2-7. Disassembled hemispherical FAIMS cell.


Assembled
FAIMS
Cell






Figure 2-8. Assembled hemispherical FAIMS cell. Outer casing is made of Kel-F and
curtain plate is made of stainless steel.












Outer
Electrode


0.50 1.00 Analytical Gap (m 1.50 2.00
Analytical Gap (mm)


Figure 2-9. Variation of the electric field as a function of position for the three FAIMS
cells. All three cells contain a gap of 2.50 mm. Radius of inner electrode for
cylindrical and hemispherical is 6.00 mm. DV for all three is 4000 V.


Inner
Electrode


100


90


80


S70
2-
Li
.U 60

w
50


40


30


0.00




























Figure 2-10. Original spherical FAIMS design. Outer casing is made of Delrin,
electrodes of brass, and inner electrode of stainless steel.























Top














Figure 2-11. Novel spherical FAIMS cell. Outer casing is made of PEEK, inner and outer
electrodes are made of stainless steel.

















1.1400


- 2.CC00


1.s 5


.-- .C




j- .1 000


. m15


Figure 2-12. Dimensions of novel spherical FAIMS cell (in inches).


Inner
Electrode




Outer
Electrode
(bottom
half)


Outer
Electrode
(top half)


Figure 2-13. Disassembled spherical FAIMS electrodes. Both inner and outer
electrodes are made of stainless steel.


~~"~


P.5CXO ------ *"/
B~saoo ---







Assembled
FAIMS
Cell

Curtain
Plate

PEEK
Casing

Figure 2-14. Assembled spherical FAIMS cell. Outer casing is made of Ketron PEEK.
Curtain plate is made of stainless steel.





















0 0

TNT
MW=227.13


O 0



N-N N-N
o o
N

o0 0

HMX
MW=296.16


ONO
0 N N




II I
0 0
o o

TNB
MW=213.1


0
C


N


N
o~N~o11


PETN
MW=316.14


o o
I II
0 N f N

N

0 0

RDX
MW=222.12


0
=0 O=N 0
\ //
O O-N



N O
//
0
NG
MW=227.09


Figure 2-15. Structures of several structural classes of common explosives studied in
this work.



















TNT

CV +4.60 V


m/z=
226.5 -227.5


Intensity:
2.90 x 10 5


8x Increase


m/z=
226.5 -227.5
Spherical FAIMS


Intensity:
7.99 x 10 5


23x Increase


0.85 1.65 2.50 3.25 4.15 5.00 5.80 6.65 7.50 8.30 9.15 10.00
CompensationVoltage (V)


100
90
80
70,
60
50 [M-I
40-
30
20
10
^ .|l


10
9
8
7;
6
51
41
31
21
1

90
80
70
60
50
40
30
20
1
0


10]-


197.1

- 1


0*

0-

.i [M-OH]-
0*

0 197.1
0.

0*

o-
0*
0.
0*
0


0 19 .1
0 .. P -. A .'t.; .. ...... :.. ..


50 100 150


200 250 300 350 400 450 500
m/z


Figure 2-16. Comparison of signal intensity of TNT between all three FAIMS cells.

Extracted chromatogram of m/z 227 on left and corresponding mass spectra

on right. lonalytics cylindrical FAIMS (top), hemispherical FAIMS (middle),

and spherical FAIMS (bottom).


m/z=
226.5-227.5

Cylindrical FAIMS

Intensity:
3.52 x10 4














m/z= 10
90
212.5 213.5 TNB so
70
Cylindrical FAIMS CV +6.10 V 60
40 [M-NO]-
Intensity: 3
5.63 x104 20
10 182.9
L1 0
10 2131
m/z= 90
212.5 213.5 8
70
Hemispherical FAIMS 60
50
Intensity: 40
1.64 x105 3x Increase 0
10 182.9
10O
m/z= 9
212.5- 213.5 8
70
Spherical FAIMS 6
4t
Intensity: 13x Increase 3
7.50 x105 2
1(0 182.9
.l l.. .. ... .... ..... 1 .0 i 0| i ..... i .
0.00 0.66 1.35 2.00 2.70 3.35 4.00 4.70 5.35 6.05 6.70 7.35 8.05 8.70 9.40 10.0 50 100 150 200 250 300 350 400 450 500
CompensationVoltage (V) m/z

Figure 2-17. Comparison of signal intensity of TNB between all three FAIMS cells.
Extracted chromatogram of m/z 213 on left and corresponding mass spectra
on right. lonalytics cylindrical FAIMS (top), hemispherical FAIMS (middle),
and spherical FAIMS (bottom).













181.3 rM1-


2,4-DNT 100 4e'--
90
mlz= CV +7.50 V 80
181.5 182.5 70
60

Cylindrical FAIMS 0

Intensity: 3

0
4.23x104

100. 181.3
90
m/z= 5x Increase 8o
181.5 182.5 70
60
Hemispherical FAIMS 50
40
Intensity: 30
2.01 x105 20
10 135.3
10
1 0 0I 1 8 1 2
90 :
m1z= 17x Increase 80-
181.5- 182.5 70
60'
Spherical FAIMS 50
40'
Intensity: 30
7.27 x105 20
10- 135.2

0.00 0.66 1.35 2.00 2.70 3.35 4.00 4.70 5.35 6.05 6.70 7.35 8.05 8.70 9.40 10.0 50 100 150 200 250 300 350 400 450 500
CompensationVoltage(V) m/z


Figure 2-18. Comparison of signal intensity of 2,4-DNT between all three FAIMS cells.
Extracted chromatogram of m/z 182 on left and corresponding mass spectra
on right. lonalytics cylindrical FAIMS (top), hemispherical FAIMS (middle),
and spherical FAIMS (bottom).


-----
































0 1 LV O= v1- o.C_
1001 N : 0
S [NL:O
50 [PETN + CI]- 350.50s
0I


-5.00 -3.00 -1.00 1.00 3.00 5.00 7.00
Compensation Voltage (V)


-351.50


9.00 11.00 13.00 15.00


Figure 2-19. CV scan of a 2 ppm mixture of explosives on the Thermo-Fisher cylindrical
cell at a DV -4000 V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB,
and (F) PETN.











100
STIC
50
0
100O
so [TNT]-
o0
100O
Ss[TN B]-
0
M 0-
c 100
-E-
S [2,4 DNT]-
o 50
50

100
S[1,3DNB-
s [1,3 DNB]-
50
0 ] ______


1 [PETN +C-


SCRp= 7.7
S ~CV=+4.6 V


Rp=10.2
CV=+6.1 V

Rp=12.5
CV=+7.5 V

Rp=12.6
rV= +RR V


NL: 4.06E3
349.5-350.5


-1.00 0.00 1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.00 11.00 12.00 13.00 14.00 15.00 16.00 17.00 18.00
Compensation Voltage (V)

Figure 2-20. CV scan of a 2 ppm mixture of explosives on hemispherical cell at a DV -
3214 V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB, and (F)
PETN.


-


NL: 5.70E5
TIC


NL: 2.60E5
226.5-227.5

NL: 2.15E5
212.5-213.5

NL: 1.41E4
181.5-182.5

NL: 8.86E4
167.5-168.5


n


Rp= 28.3
-- V= +11_: V































Compensation Voltage (V)


Figure 2-21. CV scan of a 2 ppm mixture of explosives on spherical cell at a DV -3214
V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB, and (F) PETN.











100j


1.07E6


TIC


1001 [TN-T -
n


CV= +7.73V
Rp= 7.0: A


100 [TNB]-
0
100 [3,4-DNT]-

1001 [2,4-DNT-


U-


100 [1,3-DNB]


U- -


CV= +10.27 V
Rp =7.9


A


CV= +11.83 V
Rp =5.9

rV= +13.53 V
Rp=9.6


2.58E5
212.5 -213.5


L2^


CV= +15 13 V
Rp = 10.0


0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0


4.31E5
181.5 -182.5

3.75E4
180.5 -181.5


9.01E4
167.5 -168.5


17.0 18.0 19.0 20.0


Compensation Voltage (V)

Figure 2-22. CV scan of a 2 ppm mixture of explosives on cylindrical cell at a DV -4700
(23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT, (E) 2,4-
DNT, and (F) 1,3-DNB.


2.37E5
226.5 227.5


Um


LI-


v-









100 TC 2.39E6
1 TIC
100 TNT CV= +7.73 V 6.43E5
SRp= 11.7 226.5 -227.5
[TNB] CV= +1027 A 6.77E5
0' Rp = 14.0 212.5 -213.5
100 [3,4-DNT]- CV= +11 83 V 1.28E5
S [3,4- Rp =10.0' 181.5 -182.5
100 [2,4-DNT- c=+13.53 V 7.90E4
0 Rp = 9.3 180.5 -181.5
[1,3-DNB]- C 1513 V 167.5168.5
0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.017.0 18.0 19.0 20.0
Compensation Voltage (V)

Figure 2.23. CV scan of a 2 ppm mixture of explosives on hemispherical cell at a DV -
4700 V (23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT,
(E) 2,4-DNT, and (F) 1,3-DNB.












A


CV= +7.73V
Rp= 6.8:


[3,4-DNT] -


CV= +10.27 V
Rp=8.1


CV= +11.83 V
Rp= 5.8


[2,4-DNT] -


[1,3-DNB] -


CV= +13.53 V
Rp= 7.2


CV= +15.13 V
Rp=6.8


0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0
Compensation Voltage (V)

Figure 2-24. CV scan of a 2 ppm mixture of explosives on spherical cell at a DV -4700
V (23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT, (E)
2,4-DNT, and (F) 1,3-DNB.


TIC


[TNT] -


[TNB] -


1.13E7


-I


/x


8.35E6
226.5 .227.5


6.34E6
212.5 213.5


4.32E6
181.5 182.5


5.31 E5
180.-181.5


9.37E5
167.5 168.5


V __


I_ _


I_ _










Table 2-1. Comparison of dimensions between Thermo-Fisher cylindrical and
hemispherical cell.


Hemispherical
Cell


Thermo Fisher
Cylindrical Cell


Distance from Curtain Plate
1.32mm 2.20mm
to Outer Electrode
Curtain Plate Hole Diameter 2.00 mm 2.40 mm
Ion Inlet Diameter 1.50 mm 1.50 mm
Gap 2.00mm 2.50mm
Inner Electrode Radius 12.70 mm 6.50 mm
Outer Electrode Radius 14.73 mm 9.00 mm
Ion Outlet Diameter 1.00 mm 1.90 mm


Table 2-2. Comparison of dimensions between all three cells.


Spherical
Cell


Hemispherical
Cell


Thermo Fisher
Cylindrical Cell


Distance from Curtain Plate
1.32 mm 1.32 mm 2.20 mm
to Outer Electrode
Curtain Plate Hole Diameter 2.00 mm 2.00 mm 2.40 mm
Ion Inlet Diameter 1.50 mm 1.50mm 1.50mm
Gap 2.00 mm 2.00 mm 2.50 mm
Inner Electrode Radius 12.70 mm 12.70 mm 6.50 mm
Outer Electrode Radius 14.73 mm 14.73 mm 9.00 mm
Ion Outlet Diameter 1.00 mm 1.00mm 1.90mm









Table 2-3. Anticipated ions and adducts for nitramine and nitrate ester explosives.
APCI
Name Mode MW Characteristic Ions (m/z)
257 [M+35CI-, 259 [M+37CI]-, 479 [2M+35CI]-, 481
RDX+CI negative 257
[2M+37CI]-, 210 [M+35CI-NO2-H]-, 212 [M+37CI-NO2-H]
331 [M+35CI], 333 [M+37CI]-, 358 [M+NO03], 284 [M+35CI-
HMX+CI negative 331
SHNO3], 286 [M+37CI-HNO3]
262 [M+35CI]-, 264 [M+37CI]-, 289 [M+NO3]-, 489
NG+CI negative 227
[2M+35CI-, 491 [2M+37CI]-
PETN+CI negative 351 351 [M+35CI]-, 353 [M+37CI]-


Table 2-4. Comparison of intensity values for RDX, HMX, NG, 2,4-DNT, PETN, TNT,
and TNB between the lonalytics cylindrical, hemispherical, and spherical
FAIMS cell.
Increase Increase
Intensity of Intensity of Intensity of
Explosive over over
Cylindrical Cell Hemispherical Cell Spherical Cell
Cylindrical Cylindrical
RDX 7.55x 103 SE 38 1.96 x 104 ISE 290 >2.6x 7.25 x 104+ SE 405 ~10x
HMX ND ND 0 ND ND
NG ND ND 0 ND ND
2,4-DNT 4.23 x 104 SE 200 2.01 x 105 SE 1763 4.8x 7.27x 105 SE 3180 17x
PETN 9.77 x 103SE 26 5.28x 104 SE 750 5.5x 1.03 x 10s fSE 2577 llx
TNT 3.52 x 104 SE 264 2.90 x 10s SE 2027 8.2x 7.99 x 10s SE 1856 23x
TNB 5.63 x 104SE 218 1.64x 10 SE 2081 ~3.0x 7.50 x 105SE 1201 >13x










Table 2-5. Comparison of CV, intensity, and resolving power between the Thermo-Fisher cylindrical at 16,000 V/cm and
both the hemispherical and spherical FAIMS cell at 16,072 V/cm.
S__Cylindrical Cell Hemispherical Cell Spherical Cell


CV Value
(V)


Intensity


Resolving
Power


CV Value
(V)


Intensity


Resolving
Power


CV Value
(V)


Intensity


Resolving
Power


TNT +4.60V 1.01x 105sISE 333 2.80 +4.60V 2.60 x 10s SE 333 7.70 +4.60V 1.06 x 106 SE 5773 3.80

TNB +6.10V 1.26x 10 ISE 667 3.20 +6.10V 2.15x 10s fSE 882 10.20 +6.10V 8.33 x 105 SE 882 5.50

2,4-DNT +7.50 V 6.57x 103 SE 6 2.70 +7.50 V 1.42x 104 ISE 33 12.50 +7.50 V 2.77 x 105 SE 333 3.90

1,3-DNB +8.80 V 5.91x 104 SE 33 3.90 +8.80 V 8.88x 104 SE 88 12.60 +8.80 V 3.91 x 105sSE 333 7.40

PETN ND 0 0.00 +11.30 V 4.04x 103 SE 15 28.30 +11.30V 5.63x 103 SE 9 14.10










Table 2-6. Comparison of resolution between the Thermo-Fisher cylindrical cell at
16,000 V/cm and both the hemispherical and spherical FAIMS cell at 16,072
V/cm.


Cylindrical
Cell


Hemispherical
Cell


Spherical
Cell


Rs between TNTand TNB 0.55 0.97 0.75
Rs between TNB and 2,4-DNT 0.45 1.03 0.61
Rs between 2,4-DNTand 1,3-DNB 0.31 1.08 0.51
Rs between 1,3-DNB and PETN ND 2.17 1.56









Table 2-7. Comparison of CV, intensity, and resolving power between the lonalytics cylindrical, hemispherical, and
spherical FAIMS cell at 23,500 V/cm.
[ Cylindrical Cell I Hemispherical Cell Spherical Cell


CV Value
(V)


Intensity


Resolving
Power


CV Value
(V)


Intensity


Resolving
Power


CV Value
(V)


Intensity


Resolving
Power


TNT +7.73V 2.37x 10 SE 333 7.02 +7.73V 6.41 x 10s SE 333 11.73 +7.73 V 8.34 x 106 SE 12018 6.82

TNB +10.27 V 2.59 x 10 SE 667 7.90 +10.27 V 6.76x 10s SE 667 14.06 +10.27V 6.32x 106 SE 8819 8.11

3,4-DNT +11.33V 4.32x 104SE 88 5.92 +11.33 V 1.24x 10'SE 185 10.01 +11.33V 4.30x 106 SE 8819 5.86

2,4-DNT +13.53 V 3.74x 104 SE 58 9.66 +13.53 V 7.89x 104 SE 88 9.38 +13.53V 5.31x 10' SE 1453 7.28

1,3-DNB +15.13V 9.01 x 104 SE 58 10.09 +15.13V 2.16x 10 SE 333 11.00 +15.13V 9.34x 10 SE 1527 6.88









Table 2-8. Comparison of resolution and peak width between the Thermo-Fisher
cylindrical, hemispherical and spherical FAIMS cell at 23,500 V/cm.


Cylindrical
Cell


Hemispherical
Cell


Rs between TNTand TNB 1.65 2.33 1.27
Rs between TNB and 3,4-DNT 0.51 0.47 0.33
Rs between 3,4-DNT and 2,4-DNT 0.57 0.57 0.50
Rs between 2,4-DNT and 1,3-DNB 0.69 0.67 0.42


Table 2-9. Repeatability
TNT.


of CV values on the spherical cell from five replicate runs for


DV(V) -705 -1099 -1410 -1706 -2000 -2394 -2690 -2820 -3018 -3378

TNT CV (V) 0.3 0.4 0.7 0.9 1.5 2.4 3.3 3.8 4.3 5.0
SD 0.04 0.02 0.02 0.005 0 0.009 0 0 0.01 0
RSD% 2.3 1.4 1.2 0.2 0 0.5 0 0 0.8 0


Table 2-10. Linear dynamic range and limit of detection for TNT collected by full-scan
and SIM mode.
Full Scan SIM
LOD LOD


Linear Dynamic Corr. Coef.
Range (ng/mL) (R2)


Concn.
(ng/mL)


Linear Dynamic Corr. Coef.
Range (ng/mL) (R2)


Concn.
(ng/mL)


TNT 16-10000 0.9951 4 3 3-10000 0.9968 1


Spherical
Cell









CHAPTER 3
EVALUATION OF FAIMS WAVEFORMS APPROACHES

Introduction

High-field asymmetric ion mobility spectrometry, or FAIMS, is an emerging

detection technology which can operate at atmospheric pressure to separate gas-phase

ions as first described in detail by Buryakov, et a/.20 FAIMS is similar to conventional ion

mobility spectrometry (IMS), which utilizes relatively low electric fields to propel ions

through a drift gas chamber and separate these ions according to their drift velocity.

IMS, is an important technique for the detection of many compounds, including

narcotics, explosives, and chemical warfare agents because of its high sensitivity and

amenability to miniaturization for field-portable applications.26 In IMS, the ion drift

velocity is proportional to the field strength at low electric fields (~200 V/cm) and thus an

ion's mobility (K) is independent of the applied field. At high fields (~10,000 V/cm),

however, ion mobilities become dependent on the applied field and are better

represented by Kh (equation 1-6), a non-constant high-field mobility term. Variations in

Kh from the low-field K, and the compound-dependence of that variation affords FAIMS

its separation power. FAIMS utilizes a combination of alternating current (AC) and

direct current (DC) voltages to transmit ions of interest and filter out other ions, thus

improving signal to noise ratio. FAIMS can also help reduce false positives, since two

different compounds having the same low-field mobility can often be distinguished in a

high-field environment.61

Ions are separated in FAIMS by their difference in mobility at high (Kh) and at low

(A) electric fields. At a constant gas number density, N, the non-linear dependence of

an ion's mobility in high electric fields is given by equation 1-6 and is re-shown here









K (E) = Ko[ l+a( + (4 + .] (1-6)

where Ko is the ion mobility coefficient at zero electric field and a and / describe the

dependence of the ion's mobility at a high electric field in a particular drift gas.40 This

equation is an infinite series, but at realistic field intensities the terms above the 4th

order become insignificant.41 42 In FAIMS, one of the electrodes is typically held at a

ground potential while the other has an asymmetric waveform applied to it. The most

commonly used asymmetric waveform, described by V(t) in equation 3-2, consists of a

high-voltage component (also referred to as Vi or dispersion voltage [DV]) which lasts

for a short period of time (ti) relative to a longer lasting (t2) low-voltage component (V2)

of opposite polarity. Most FAIMS work up to date has employed a sinusoidal wave, plus

its first harmonic at twice the frequency, as shown in equation 3-2, where w is the

frequency in radians per second. The waveform is constructed so that the voltage-time

product applied to the electrode is equal to zero as displayed in equation 1-7.

V(t) = (0.61) V sin(wt) + (0.39)VI sin (2 wt ... (3-2)

At high electric fields, the application of this waveform will cause an ion to

experience a net drift toward one of the electrodes. Ions passing between the

electrodes encounter this displacement because the ion's mobility during the high-

voltage component (Kh) is different than that from the low-voltage mobility (K). In other

words, the ion will move a different distance during the high-voltage portion than during

the low-voltage portion. This ion will continue to migrate towards one of the electrodes

and subsequently be lost unless a DC compensation voltage (CV) is applied to offset

the drift. The CV values required to offset the drift of different ions will be different if the

Kh/K ratio of the ions are different. Thus, a mixture of compounds can be successfully









separated by scanning the CV, allowing each compound to transmit at its characteristic

CV, creating a CV spectrum.

When higher electric fields are applied to the FAIMS electrodes, an ion can have

three possible changes in ion mobility. The mobility of type A ions increases with

increasing electric field strength, the mobility of type C ions decreases, and the mobility

of type B ions increases initially before decreasing at yet higher fields.43' 61 Most low-

mass ions (< m/z 300) are type A ions, whereas most high-mass ions are type C ions.62

Sum of Sines Waveform

The dependence of an ion's mobility in a time-varying electric field makes the

design of the waveform in FAIMS an essential feature for its overall performance. 63 As

mentioned above, most FAIMS work to date has employed a sinusoidal wave plus its

first harmonic at twice the frequency, as described in equation 3-2. If the amplitude and

CV are being applied to the same electrode, then the addition of CV should be included

in equation 3-2. Equation 3-2 results in an asymmetric waveform of the type shown in

figure 3-1a. The shape of the waveform is a parameter that contributes to the value of

the experimentally measured (compound-dependent) CV for transmission of an ion. A

symmetrical waveform (sine or square) should result in CV = 0 V for transmission of all

types of ions. Asymmetry of the waveform is required for ion separation and is

expressed by differences in the CV.43

Square Waveform

Even though most FAIMS experiments have made use of the sum-of-sines

waveform, theoretical studies have suggested that a rectangular waveform (as shown in

figure 3-1 b) would be ideal for FAIMS analyses.6365 Analytical considerations show that

rectangular waveforms may improve ion separation efficiency, resolution and/or









sensitivity as compared to sinusoidal waveforms.62' 65-67 Unfortunately, the excessive

power load imposed by high frequency, high voltage pulses with steep rise times has

hindered the development of electronics that deliver rectangular pulses for driving

separations based on differential ion mobility.63

Intuitively, the use of an asymmetric square waveform for FAIMS would seem to

maximize the differences during the high and low field portions of the electric field.

These high to low periods of the waveform permit an ion to experience a maximum of

unequal voltages maximizing the CV. In previous studies, there have been concerns

that the time it takes an ion to respond to the idealized asymmetric square waveform

and reach "steady state," or terminal, drift velocity might be sufficiently long enough to

introduce error due to the transient electric field.42 Lin et al. showed that, to the first

order, this can be neglected if the time for reaching terminal velocity is small relative to

the total drift time. Since the estimated time necessary to reach this velocity in a

transient electric field is in the picosecond range and the drift time is in the millisecond

range, this factor can be ignored.

Because FAIMS has continuous ion separation capabilities, it is attractive to use in

conjunction with a mass spectrometer. This union was utilized to offer orthogonal

detection methods, one separating ions according to their mobilities through a gas and

the other separating ions according to their mass-to-charge ratios. In this chapter we

describe how the independent control (frequency, amplitude, and duty cycle) of a

square waveform can help to more completely characterize FAIMS in terms of

transmission, resolution, and separation in comparison to a sum-of-sines waveform.

Previous experiments have shown the duty cycle to have a strong influence on ion









separation with planar electrodes.63 Here, the duty cycle is defined as the ratio of the

width of the high-voltage pulse to waveform period. A miniature FAIMS cell powered by

a digitally driven circuit producing a square waveform will be described in comparison to

an analog sum-of-sines waveform. A full comparison of separation and sensitivity is

performed over the overlapping range provided by the waveform generators. This

chapter also includes SIMION 868 modeling to help validate experimental results.

Experimental

The traditional sinusoidal waveform was created by an lonalytics GPI 1000 FAIMS

system alpha prototype waveform generator which only runs with a 2:1 ratio (33% duty

cycle) at a frequency of 750 kHz. This waveform was applied to one of the plates and a

reference lead to the other plate. Each individual component waveform was

capacitance tuned for the electrodes and added together to produce a waveform similar

to the one shown in figure 3-la. The full range of operation for the lonalytics alpha

prototype is (+/-) 300 V to (+/-) 4000 V; the range utilized for this research was from (+/-)

300 V to (+/-) 800 V, since higher voltages could arc within the short distances of the

cell. Data were processed using the mass spectrometer instrument software (described

below).

The square waveform generator used in these studies was constructed by Implant

Sciences Corporation; a schematic is shown in figure 3-2. In this computer-generated

waveform, three of the four variables ( V1 V2, ti, and t2) are specified; the fourth is

deduced from the balanced equation 1-7. The waveform generator is the subject of a

patent application.

The frequency, duty cycle, and amplitude on the square waveform are all

independently controlled. The square generator can provide positive DV values from









+50 V to +750 V. For this reason all experiments with the lonalytics alpha prototype

were only run with positive amplitude values. Three different frequencies (250, 333,

500 kHz) and six different duty cycles (25%, 30%, 33.3%, 35%, 40%, and 45%, defined

as ti/(ti + t2)) were evaluated on this waveform generator unless otherwise noted. All

experiments employed a CV separation over a 10 V range at a scan rate of 5 V/min.

Data were processed using instrument software interface (DMS Interface Version 0.25)

created by Implant Sciences.

The planar geometry FAIMS cell, as shown in figure 3-2a c, was designed by

the University of Florida Chemistry Department Machine Shop and consists of two

parallel stainless steel plates (5 mm wide, 15 mm long, 1 mm thick) encased and

recessed in a PEEK support (8mm wide, 18 mm long, 3 mm thick) that provides

mechanical stability and electrical insulation. Electrical connections to the individual

plates were made via posts spot welded to the plates through holes in the PEEK

support. The gap within the FAIMS cell (0.38 mm) was maintained by fused silica

capillaries as shown in the figure 3-2c The top and bottom plates were then secured to

each other through the PEEK with 4 screws to ensure mechanical stability and

alignment.

The FAIMS cell was interfaced directly to the heated capillary inlet of an LCQ

(Thermo Scientific) quadrupole ion trap mass spectrometer (QITMS) with the aid of a

home-built end-cap connector piece. All experiments were performed utilizing this

commercial bench-top MS. The end-cap piece was made of polyaryletheretherketone

(PEEK) and is displayed in figure 3-2 (top left). End-cap piece and FAIMS cell were

wrapped in Teflon to ensure no signal loss occurred between these connections. No









curtain plate or curtain gas was used and no attempt to perform additional ion

desolvation was made. Conventional atmospheric pressure chemical ionization (APCI)

was employed to produce gas phase ions. Nitrogen was used as the nebulizing

(sheath) and auxiliary gases. APCI and heated capillary temperatures were held at 75

C. Analyte samples were infused with a 500 pL Hamilton syringe at 15 pL/min.

Solutions were injected directly via the syringe pump of the LCQ. Data acquisitions

were performed using full scan mode (50-500 m/z) with a maximum injection time of

100 ms for automatic gain control (AGC). Data were processed using the instrument

software interface (Xcalibur version 1.4 SR1).

In this chapter, a common explosive, 2,4,6-trinitrotoluene (TNT) was utilized. This

explosive was provided by Dr. Jehuda Yinon of the Weizmann Institute of Science, and

was obtained from the Analytical Laboratory of the Israeli Police Headquarters.

Negative ion mode was employed with APCI. The explosive solution was diluted in a

solvent mixture consisting of 65% methanol and 35% de-ionized water. Utilizing a

combination of methanol and water as a solvent allows the use of lower vaporizer

temperatures than water alone; the low vaporizer temperature help to avoid the

degradation of fragile explosive analytes. In this research, all solutions were diluted to

a concentration of 10 ppm unless otherwise noted.

SIMION 8, the ion trajectory modeling program68, was utilized in an effort to

optimize the performance of the miniature FAIMS cell and validate experimental results.

The influences of pneumatic and electrostatic forces at atmospheric pressure are

shown. The variables used in this investigation include gas flow, DV, CV, analytical gap

(spacing between the plates), frequency of the asymmetric waveform, its shape









sinusoidall versus square), and its duty cycle. The potential array of the FAIMS

analyzer consisted of parallel plates (5 mm wide, 15 mm long, and separated by a gap,

G = 0.5 mm). Statistical Diffusion Simulation68 (SDS) user program was used to model

the ion motion at atmospheric pressure. By modeling both diffusional and mobility

terms of ions in a neutral gas, the effects of high pressure collisions are taken into

account by the SDS user program. All simulations were performed using TNT, [M]-,

ions (m/z = 227) for which the low field and high field mobilities are known.62' 69-71

Results and Discussion

Influence of Dispersion Voltage on Compensation Voltage, Peak Width and
Resolving Power

The characterization and performance of the square waveform on planar FAIMS

ion separation was investigated. In FAIMS, ions are separated based on the change in

their mobility at high vs. low electric fields. Thus, theoretically, the higher the field, the

better separation that would be obtained. Generally, the magnitude of the CV increases

as the DV increases for low mass ions (m/z usually below 300). This behavior is

verified with both the sum-of-sines and square waveform generators, as displayed in

figure 3-3a. The plot shows CV as a function of DV for both waveforms with the sine

wave at 750 kHz and the square wave at three different frequencies (250, 333, and 500

kHz). The duty cycle for the sine waveform was 33% (not adjustable); a duty cycle of

25% was selected for the square waveform. It is evident that, for both square and sine

waves, CV increases with DV, with square waveform giving higher CV magnitudes

within the overlapped DV range (300-350 V). The frequency of the square waveform

has no significant effect on the CV. Further insight into the behavior of the [M]- ion in

the planar geometry FAIMS cell can be gained by simulation of ion trajectories with









SIMION. Figure 3-4 compares the plots for CV vs. DV for both waveforms, over a wider

range of DV that can be achieved with our current electronics. Comparing these plots

with the experimental data in figure 3-3a shows the same trend of higher CV values for

square wave than for sine wave at the same DV (although only a 1.5x increase, less

than the 2.5x increase observed experimentally). Note that the simulated CV values are

generally smaller than those observed experimentally, perhaps because of the

increased CV values recently observed in the presence of solvent vapor in the FAIMS

cell.72

In a planar geometry FAIMS cell, ions are presumably distributed evenly across

the gap between the two electrodes. There is typically better resolution with this

geometry with an uniform electric field than with the focusing effect observed with

curved electrodes.73 Peak shapes, as well as transmission, are affected by lateral

diffusion and space charging (also known as ion-ion repulsions). Figure 3-3b shows

that experimental peak widths (full width at half maximum, FWHM) generally increase

as DV increases for both waveforms, although the peak width for the square wave

increases only slightly over the available DV range. The dramatic increase in peak

width for the square waveforms around DV 150 V is due to splitting of the main TNT

peak into two peaks. Peak splitting is shown in figure 3-5. The figure displays three CV

scans of the extracted chromatograms of m/z 227. The CV scan is from -2 to 8 V over

two minutes (5 V/min). The frequency and duty cycle applied were 250 kHz and 25%,

respectively. The figure shows a progressive splitting of the CV peak of mass 227 as

the dispersion voltage is increased (3-5a 3-5f). It was suspected that this ion, with

identical mass (m/z 227), is not TNT and therefore has a different compensation









voltage. As the electric field increases within the gap of the FAIMS cell the difference in

their mobility becomes more apparent. Figure 3-5 is a great example of how FAIMS

can be utilized as an orthogonal detection method when applied to MS.

Resolving power (RpFAiMS) indicates the ability of the FAIMS cell to resolve a peak

at a particular compensation voltage.74 RpFAIMsiS defined as the CV divided by the peak

width at FWHM and I, as shown in equation 2-6. Figure 3-3c displays that the RpFAIMS

increases with DV for both waveforms, although at any given DV, higher RpFAIMsvalues

are obtained for the square waveform. Even though broader peaks are observed at

higher dispersion voltages, as shown in figure 3-3b, the even greater increase in CV

(figure 3-3a) results in enhanced RpFAIMS.75

Influence of Frequency on Compensation Voltage, Peak Width, and Resolving
Power

The rectangular asymmetric waveforms were monitored at different amplitudes

and frequencies with a 100 MHz Tektronix TDS 1012B oscilloscope, as shown in figure

3-6. The requested values of the width, t1, and the amplitude, V1, of the high voltage

pulse are indicated on the figure. It should be noted that the traces are offset vertically

from the 250 kHz trace for ease of visualization. All the traces displayed in figure 3-6

are at 25% duty cycle. Here, the duty cycle, d, is defined as

d = t- (3-3)
(tl+t2)

Figure 3-3 shows that the frequency of the waveforms has little effect on CV,

PWHM, or RpFAIMS. This is as expected, since the frequency of the waveform (and thus

the rate of ion oscillations perpendicular to the gas flow) should not affect the CV where

the ion is transmitted.









While the duty cycle of the sine waveform is not variable, it is quite simple to vary

the duty cycle of the square waveform. At lower duty cycles (where the waveform

deviates more from symmetrical) RpFAiMS increases, as displayed in figure 3-9. Note

that at DV values between 00 and 200 V, a 25% duty cycle yields an RpFAIMS ~5x

higher than a 45% duty cycle.

Influence of Frequency on Sensitivity

Higher frequencies are generally thought to improve ion separation at the expense

of transmission.65 The loss of transmission can be better tolerated by increasing the

frequency, which however is limited by the electronics and creates more distortion in the

rectangular shape of the waveform. Figure 3-10 displays the effect of frequency on

sensitivity (3-10a) and peak area (3-10b) as a function of dispersion voltage. It is

known that a reduction of the ion oscillation amplitude is due to the increased

frequencies. This increased frequency helps to intensify the transmission of the

specified ion and this is verified in figure 3-10. It should be noted that a 25% duty cycle

was selected for these graphs to be able to show 500 kHz frequency data. Higher

cycles displayed the same behavior.

Influence of Duty Cycle on Sensitivity

The duty cycle is one FAIMS parameter that has not been explored in the past

since the duty cycle of the sine waveform (used in all commercial FAIMS systems) is

fixed to 33%. In this work, the effect of duty cycle of the square waveform on sensitivity

has been explored by holding the frequency at a fixed value of 250 kHz and 333 kHz.

The results shown in figure 3-11 indicate that the signal intensity increases as the duty

cycle decreases. Indeed, decreasing duty cycle from 45% to 25% increases the signal









by a factor of 2 to 3. Lower duty cycles give greater sensitivity at all frequencies

examined.

Lower Concentration Experiments

To evaluate whether the performance of square wave planar FAIMS is a function

of concentration, samples varying over a 100x concentration range were evaluated; the

results (CV, PWHM, Rp, and signal intensity) are displayed in table 3-1. These data

verify that the reproducibility of CV peak positions and Rp are unaffected by

concentration within experimental error. Note as well that signal intensity varies linearly

with concentration over two decades.

Conclusions

This paper demonstrated for the first time the practical implementation and

characterization of square wave FAIMS on a miniature planar system. The independent

control of frequency, amplitude, and duty cycle of a computer-generated square

waveform can provide in a more complete characterization of FAIMS in terms of

transmission, resolution, and separation compared to a sum of sines waveform. Square

wave FAIMS delivers better separation and resolving power, as predicted with ion

trajectory simulation. The increase of frequency was shown to increase sensitivity but

had little effect on CV, peak width, and Rp. The lower the duty cycle the higher our peak

area and signal intensity. Duty cycle decreases from 45% to 25% also showed a 5x

increase in Rp. Comparison of results with the sum of sines waveform shows that the

square waveform can provide better separation and resolving power, albeit at some

cost in transmission and thus sensitivity. Further studies should be performed to

compare square and sine waveforms with additional explosives and mixture of

explosives. It would also be interesting to compare the two waveforms with different cell









dimensions and with different cell geometries. The major limitation of the square

waveform is the limited voltage and power that can be obtained with practical waveform

generators.









(a)


(b)


Period


Period


Time----


Figure 3-1. Schematic of (a) an asymmetric sum-of-sines waveform and (b) a
rectangular waveform.












(a) shown (d)(d)

= M










SainlessSte Male




PEEK


Figure 3-2. Photograph (a), perspective and cross-sectional views (b and c) of
miniature planar FAIMS cell. Square wave generator circuit diagram is also
shown (d).



















40m

350

300

250

1m


(a) I 15
050

00





130

120

110

(b) ,


(c)


S ,1 ZD 3D 4 5spe D EOD 7Vdt D S

DspersimVdtage(V)


0 100 200 300 40 50 60 700 900
Iispersin Vltage (V)


-A-S e 333&, 25%o Ey Oyde
150 ------ --/ --





0 100 300 0 500 600 70 800 900


Figure 3-3. Graphs displaying behavior of CV (a), peak width (b), and resolving power

(c) as a function of DV.


- SreT73-k 333'/oCty Qde
- aeM-, 25 /oCDy q/dE
Saae rSM-k, 2/o q/dE
-cs- sea5D+, 25/oCayO/dE_


---2f
























Ak


-E


4.00

3.50

3.00

2.50

2.00

1.50

1.00

0.50

0.00

-0.50


3-Sine: 750
kHz, 33%
DC

D-Square:
750 kHz,
33% DC


Dispersion Voltage (V)


Figure 3-4. CV versus DV simulation data for the [M]- ion of TNT.


S200 400 600 800 1000 1200 1400 1600
200 400 600 800 1000 1200 1400 1600

















loo (a)
80
60
40
20



60
40
20
0
S(c)

60
40

20
1O 0
1 :(d)


40

20

10D
BO (e)
60
40
20





40
20

00 0 5 10 1 .5 2,0 25 3, 35 4.0 4.5 50 55 60
Trme (mmn)


Figure 3-5. Three CV scans of 10 ppm TNT. Peak splitting is observed as DV is

increased from (a) to (f). (a) is a DV +121 V, (b) +161 V, (c) +181 V, (d) +201

V, (e) +221 V, and (f) +242 V.


004_10ppmTNT_Square_250kHz_25%_DV121


5/16/200 12:44:13 PM


I0ppmTNT. Square: 250kHz: 250; CV 2(oB









250 kHz, 70 DV
O 333 kHz, 70 DV
"-- 500 kHz, 70 DV

Kr


_I


kHz, 340 DV
kHz, 340 DV
kHz, 340 DV


Figure 3-6. Scope traces of square waveforms at different frequencies for amplitudes
(a) DV 70 and (b) DV 340 where V1 is the amplitude (DV) and tl is the width
of the high voltage pulse. Waveforms are offset for ease of visualization.


(a)


V1





V2


(b)















2 bU
--- 25% Duty Cycle
30% Duty Cycle
2 -- 33 3% Duty Cycle
2 00 35% Duty Cycle
40% Duty Cycle
-A- 45% Duty Cycle
S550

0

1 00-


E
8 0 50 -



0 00 .
S50 1V00 150 200 250 300 350 4 0


-0 50
Dispersion Voltage (V)


Figure 3-7. Graph displaying CV as a function of DV. For 250 kHz the duty cycle is
shown to have little effect.












1.20 -- -25% Duty Cyle
-+--30% Duty Cycle
1.10M -A- 33.3% DutyCycle
-0- 35% Duty Cycle
1.00 --- 40% Duty Cycle
-A-45% Duty Cycle

0.90-


S0.80


0.70


0.60


0.50
50 100 150 200 250 300 350

Dispersion Voltage (V)




Figure 3-8. Graph displaying peak width as a function of dispersion voltage. For 250
kHz the duty cycle is shown to have no effect.


100
















350


300


250


0200
0

21 50
0


1 00


050


000


100 150 200

Dispersion Voltage (V)


250 300 350


Figure 3-9. RpFAIM as a function of DV at six different duty cycles.


101



















SSne 750kHz 33%
Sq250kHz_25%
350E+06 Sq333kHz25%

Sq500kHz_25%
300E+06

(a)
250E+06


S 200E,06


150E+06


100E+06


500E+05 r
O oo. ,u -

000E+00 "MOCO 0" C C
o 100 200 300 400 500 600 700 800 900
Dispersion Voltage (V)





350E+07
(b) --Sine750kHz33%
SSq250kHz_25%
300 E+07 X Sq333kHz_25% -
& -Sq500kHz 25%


150E+07



100E+07



150E+07



100E+07








0 100 200 300 400 500 600 700 800 900
Dispersion Voltage (V)




Figure 3-10. Graphs displaying effect of square wave frequency on sensitivity (a) and

peak area (b).





























102



















-- 25%
-A-33%
A35%
2 50E06\ 45









150E+06
}\


1 00E+06




5 00E+05




o oOE+ooO O -------------_ 0 ^ ^ i =.-^ ^ ^ ^ B E ----
S50 100 150 200 250 300 350 400
Dispersion Voltage (V)





4 00E+06
--25%
33%
350E+06 35%


3 00E+06 45Y



2 50E+06






150E+06






5 00E+05





Dlpersion Voltage (V)






fixed to 250 kHz (a) and 333 kHz (b).


103


















2 50E+07
--25%
&33%
R 35%
2 OOE07 45%





150E,07



S(a)

100E+07





5 OOE+06 \





SOOO00EO ---0
0 50 100 150 200 250 300 350 400
Dipersion Voltage (V)






300 E+07
--25%



250E+07 -0-45%-




200E+07









100E+07




500E+06




0 OOE+00 --
0 50 100 150 200 250 300 350 400
Dipersion Voltage (V)




Figure 3-12. Graphs displaying the effect of duty cycle on peak area. Frequency is

fixed to 250 kHz (a) and 333 kHz (b).


104









Table 3-1. Summary of the CV, PWHM, Rp, and signal intensity for concentrations from
10 ppm down to 100 ppb.
CV PWHM Signal
Concentration RCV PWHM Sgn
(V) (V) Intensity
10ppm 0.16 0.6 0.26 3.3E+06
1ppm 0.16 0.6 0.25 3.3E+05
500ppb 0.19 0.6 0.30 1.5E+05
250ppb 0.17 0.6 0.29 6.5E+04
100ppb 0.12 0.6 0.19 3.2E+04


105









CHAPTER 4
CONCLUSIONS

This dissertation presents high-field asymmetric waveform ion mobility

spectrometry, or FAIMS, combined with mass spectrometry (FAIMS/MS) as a viable

technique for the separation of gas-phase explosive ions at atmospheric pressure.

FAIMS can behave as an ion filter that is capable of transmitting only selected

compounds in a mixture to pass through to the mass spectrometer. Gaining a better

understanding of how this technique separates explosive ions can lead to improved

explosive detection. The application of FAIMS/MS to explosive analysis can improve

signal-to-noise, while reducing analysis time. The main purpose of this dissertation was

to investigate the applicability of novel cell geometries and waveform approaches for

FAIMS as a separation tool to be utilized in combination with an atmospheric pressure

chemical ionization/mass spectrometry (APCI/MS) system.

The three different geometries of FAIMS cells studied and were cylindrical,

hemispherical, and spherical. FAIMS experiments were conducted on the molecular

ions of several explosives to determine how their signal intensity, resolution and

resolving power were affected in relation to the geometry. The highest sensitivity was

obtained by the spherical FAIMS electrode with every explosive that was attempted.

Additionally, the spherical FAIMS electrode allowed better resolving power and

resolution values than either cylindrical cell. The increased transmission allows the

utilization of larger electrode radii, which will help to increase resolution. Hemispherical

FAIMS was also shown to improve sensitivity, resolution, and resolving power when

compared to the cylindrical cell. It was found that the hemispherical electrodes allow for

better resolution when compared to the spherical electrodes. The need of these novel


106









electrode geometries would be dependent on what the analyst needs, whether it be

more resolution with increased but adequate sensitivity (hemispherical), or increased

sensitivity with adequate resolution (spherical). The analytical performance of spherical

FAIMS/MS was evaluated for TNT to provide information on its quantitative capability.

The best CV repeatability was obtained at higher dispersion voltages (-2000 to -3378

V). Two MS scan modes, full-scan and SIM, were tested for LOD and LDR. The

method provided limits of detection for TNT down to 4 ppb for fulllscan and 1 ppb for

SIM. Although full-scan MS provided higher LOD and LDR values, it is still a vital

explosive analysis mode because it provides more information about the ion.

The implementation of square wave FAIMS on a miniature planar system was also

demonstrated. The ability to independently control frequency, amplitude, and duty cycle

of a square waveform can aid in a more complete characterization of FAIMS in terms of

transmission, resolution, and separation in comparison to a sum of sines waveform was

shown. Square wave FAIMS was displayed to provide better separation and resolving

power within the available voltage range and this behavior was verified at higher

voltages with the SIMION modeling program. The increase of frequency was shown to

increase sensitivity as well as peak area, but had no effect on compensation voltage or

resolving power. Experiments showed that the lower the duty cycle the higher the

signal intensity. Duty cycle decreases also showed a 30-35% improvement in

compensation voltage and resolving power. Comparison of experiments with the sum

of sines waveform shows that a square wave technology would be ideal in terms of

separation and resolving power.


107









Future Work


Curved FAIMS Cells

There are many opportunities for future work in this project. The variation of the

analytical gap and dispersion voltage would affect the ion transmission and resolution in

curved geometry FAIMS cells. Gap size is related to transmission efficiency and

resolution. For example, ions with higher mobility values will come closer to the

electrodes because of their larger ion amplitude oscillation in the high portion of the

asymmetric waveform. These ions will have a better chance of being lost to the

electrodes, which will reduce the transmission efficiency through the analyzer region.

Since the non-uniformity of the electric field in a curved electrode is aiding the focusing

of the ions, it will be advantageous to explore if gap size is a dominant factor in

transmission loss or is an insignificant factor when compared to the strong focusing

properties in the cell. In other words, it would be beneficial to evaluate whether or not

reducing the gap size in a curved geometry will have a negative effect on transmission

as it does in a planar cell.

Higher dispersion voltages are needed in order to improve transmission. Further

studies can evaluate explosive ion transmission and separation at higher dispersion

voltages on curved geometries. Efforts should also continue to explore other

geometries in order to characterize all advantages/disadvantages of these novel

electrodes. Further studies are also required to improve the sensitivity to be able to use

FAIMS as a quantitative method. Finally, computer simulations of various electrode

radius sizes and geometries should also be performed in order to compare transmission

and resolution.


108









FAIMS Waveforms

It would be advantageous to improve our understanding of how ions behave under

the influence of different parameters of FAIMS, such as different waveforms, in order to

apply the optimum condition when more sensitivity or resolution is required for the

application of interest. Further studies should be performed to compare square and

sine waveforms with different explosives and mixture of explosives as a function of cell

dimensions with different cell geometries. Additional experiments using FAIMS

waveforms could include the evaluation of other waveform types, higher frequencies,

and duty cycles. Finally, computer simulations of various waveform types should be

performed to explore the viability of these waveforms.


109










LIST OF REFERENCES


1. Moore, D. S., Review of Scientific Instruments 2004, 75 (8), 2499-2512.

2. Yinon, J., Journal of Chromatography A 1996, 742 (1-2), 205-209.

3. Sigman, M. E.; Clark, C. D.; Fidler, R.; Geiger, C. L.; Clausen, C. A., Rapid
Commun. Mass Spectrom. 2006, 20 (19), 2851-2857.

4. Sigman, M. E.; Ma, C. Y., Journal of Forensic Sciences 2001, 46 (1), 6-11.

5. Pan, X. P.; Zhang, B. H.; Tian, K.; Jones, L. E.; Liu, J.; Anderson, T. A.; Wang, J.
S.; Cobb, G. P., Rapid Commun. Mass Spectrom. 2006, 20 (14), 2222-2226.

6. Mathis, J. A.; McCord, B. R., Rapid Commun. Mass Spectrom. 2005, 19 (2), 99-
104.

7. Yinon, J.; Zitrin, S., The Analysis of Explosives. Pergamon Press: New York, NY,
1981; Vol. 4.

8. Beveridge, A., Forensic Investigations of Explosives. Taylor & Francis Ltd.:
Bristol, PA, 1998.

9. Na, N.; Zhang, C.; Zhao, M. X.; Zhang, S. C.; Yang, C. D.; Fang, X.; Zhang, X.
R., Journal of Mass Spectrometry 2007, 42 (8), 1079-1085.

10. Gapeev, A.; Yinon, J., Journal of Forensic Sciences 2004, 49 (2), 227-237.

11. Fu, X. F.; Zhang, Y.; Shi, S. H.; Gao, F.; Wen, D. W.; Li, W.; Liao, Y. P.; Liu, H.
W., Rapid Commun. Mass Spectrom. 2006, 20 (19), 2906-2914.

12. Tachon, R.; Pichon, V.; Le Borgne, M. B.; Minet, J. J., Journal of
Chromatography A 2007, 1154 (1-2), 174-181.

13. Buryakov, I. A., Journal of Chromatography B-Analytical Technologies in the
Biomedical and Life Sciences 2004, 800 (1-2), 75-82.

14. Purves, R. W.; Guevremont, R.; Day, S.; Pipich, C. W.; Matyjaszczyk, M., Review
of Scientific Instruments 1998, 69, 4094-4105.

15. Eiceman, G. A.; Tadjikov, B.; Krylov, E.; Nazarov, E. G.; Miller, R. A.; Westbrook,
J.; Funk, P., Journal of Chromatography A 2001, 917 (1-2), 205-217.

16. Eiceman, G. A.; Nazarov, E. G.; Miller, R. A.; Krylov, E. V.; Zapata, A. M.,
Analyst (Cambridge, United Kingdom) 2002, 127 (4), 466-471.


110









17. Barnett, D. A.; Purves, R. W.; Ells, B.; Guevremont, R., Journal of Mass
Spectrometry 2000, 35, 976-980.

18. Robinson, E. W.; Sellon, R. E.; Williams, E. R., Interational Journal of Mass
Spectrometry 2007, 259 (1-3), 87-95.

19. Tang, K. Q.; Li, F. M.; Shvartsburg, A. A.; Strittmatter, E. F.; Smith, R. D.,
Analytical Chemistry 2005, 77 (19), 6381-6388.

20. Buryakov, I. A.; Krylov, E. V.; Nazarov, E. G.; Rasulev, E. K., International
Journal of Mass Spectrometry and Ion Processes 1993, 128, 143-148.

21. Guevremont, R.; Purves, R. W., Review of Scientific Instruments 1999, 70, 1370-
1383.

22. Krylov, E. V., Technical Physics 1999, 44, 113-116.

23. Guevremont, R.; Purves, R. W.; Barnett, D. A.; Ding, L. Y., International Journal
of Mass Spectrometry 1999, 193 (1), 45-56.

24. Buryakov, I. A.; Kolomiets, Y. N.; Luppu, B. V., Journal of Analytical Chemistry
2001, 56 (4), 336-340.

25. Buryakov, I. A., Talanta 2003, 61 (3), 369-375.

26. Eiceman, G. A.; Karpas, Z., Ion Mobility Spectrometry. 2nd ed; CRC Press: Boca
Raton, FL, 2005.

27. Barnett, D. A.; Ells, B.; Guevremont, R.; Purves, R. W., Journal of the American
Society for Mass Spectrometry 1999, 10, 1279-1284.

28. Karasek, F. W., Research-Development 1970, 21 (3), 34-37.

29. Cohen, M. J.; Karasek, F. W., Journal of Chromatographic Science 1970, 8 (6),
330-337.

30. Karasek, F. W.; Kilpatri.Wd; Cohen, M. J., Analytical Chemistry 1971, 43 (11),
1441-1447.

31. Zimmermann, S.; Barth, S.; Baether, W. K. M.; Ringer, J., Analytical Chemistry
2008, 80 (17), 6671-6676.

32. Ewing, R. G.; Atkinson, D. A.; Eiceman, G. A.; Ewing, G. J., Talanta 2001, 54 (3),
515-529.


111









33. Lu, Y.; O'Donnell, R. M.; Harrington, P. B., Forensic Science International 2009,
189 (1-3), 54-59.

34. Kemper, P. R.; Dupuis, N. F.; Bowers, M. T., International Journal of Mass
Spectrometry 2009, 287 (1-3), 46-57.

35. Tao, L.; Dahl, D. B.; Perez, L. M.; Russell, D. H., JAm Soc Mass Spectrom 2009,
20 (9), 1593-1602.

36. Creaser, C. S.; Griffiths, J. R.; Bramwell, C. J.; Noreen, S.; Hill, C. A.; Thomas,
C. L. P., Analyst, 2004, 129 (11), 984-994.

37. Revercomb, H. E.; Mason, E. A., Analytical Chemistry 1975, 47 (7), 970-983.

38. Dugourd, P.; Hudgins, R. R.; Clemmer, D. E.; Jarrold, M. F., Review of Scientific
Instruments 1997, 68, 1122-1130.

39. Kanu, A. B.; Haigh, P. E.; Hill, H. H., Analytica Chimica Acta 2005, 553 (1-2),
148-159.

40. Guevremont, R., Journal of Chromatography A 2004, 1058 (1-2), 3-19.

41. Viehland, L. A.; Guevremont, R.; Purves, R. W.; Barnett, D. A., International
Journal of Mass Spectrometry 2000, 197, 123-130.

42. Shvartsburg, A. A.; Tang, K. Q.; Smith, R. D., Journal of the American Society for
Mass Spectrometry 2004, 15 (10), 1487-1498.

43. Purves, R. W.; Guevremont, R., Analytical Chemistry 1999, 71 (13), 2346-2357.

44. Barnett, D. A.; Ells, B.; Guevremont, R., Journal of the American Society for
Mass Spectrometry 2000, 11 (12), 1125-1133.

45. Ouyang, Z.; Noll, R. J.; Cooks, R. G., Analytical Chemistry 2009, 81 (7), 2421-
2425.

46. Paul, W.; Steinwedel, H. U.S. Patent; Apparatus for Separating Charged
Particles of Different Specific Charges. 1960; Vol. 2939952.

47. Watson, J. T., Types of Mass Spectrometers. In Introduction to Mass
Spectrometry, 3rd ed.; Lippincott-Raven Publishers: Philadelphia, PA, 1997; p 80.

48. Finnigan LCQ Hardware Manual Classic Series, 97000-97053 Revision B ed.;
San Jose, CA, 2003; pp 2-17 2-24.


112









49. Wuerker, R. F.; Shelton, H.; Langmuir, R. V., Journal of Applied Physics 1959, 30
(3), 342-349.

50. March, R. E., Journal of Mass Spectrometry 1997, 32 (4), 351-369.

51. Carnahan, B. L.; Tarassov, A. S. Ion mobility spectrometer, 1995.

52. Guevremont, R.; Purves, R.; Barnett, D. Spherical side-to-side FAIMS. 359644 &
226965, February 7, 2003.

53. Guevremont, R.; Purves, R. W. FAIMS Apparatus and Method with Ion Diverting
Device. 6825461, November 30, 2004.

54. Guevremont, R.; Ding, L. Y.; Ells, B.; Barnett, D. A.; Purves, R. W., Journal of the
American Society for Mass Spectrometry 2001, 12 (12), 1320-1330.

55. Hoffman, E. D.; Stroobant, V., Mass Spectrometry Principles and Applications.
2nd ed.; John Wiley and Sons: West Sussex, England, 2002.

56. Spangler, G. E.; Lawless, P. A., Analytical Chemistry 1978, 50 (7), 884-892.

57. Spangler, G. E.; Lawless, P. A., Analytical Chemistry 1978, 50 (2), 290-294.

58. Long, G. L.; Winefordner, J. D., Analytical Chemistry 1983, 55 (7), A712-&.

59. Guevremont, R., FAIMS. www.faims.com, September 23, 2009.

60. Barnett, D. A.; Belford, M.; Dunyach, J. J.; Purves, R. W., Journal of the
American Society for Mass Spectrometry 2007, 18, 1653-1663.

61. Purves, R. W.; Guevremont, R.; Day, S.; Pipich, C. W.; Matyjaszczyk, M. S.,
Review of Scientific Instruments 1998, 69 (12), 4094-4105.

62. Krylov, E.; Nazarov, E. G., International Journal of Mass Spectrometry 2009,
285, 149-156.

63. Papanastasiou, D.; Wollnik, H.; Rico, G.; Tadjimukhamedov, F.; Mueller, W.;
Eiceman, G. A., Journal of Physical Chemistry A 2008, 112 (16), 3638-3645.

64. Krylov, E. V., Instruments and Experimental Techniques 1997, 40 (5), 628-631.

65. Shvartsburg, A. A.; Tang, K.; Smith, R. D., Journal of the American Society for
Mass Spectrometry 2005, 16 (1), 2-12.

66. Krylov, E. V.; Nazarov, E. G.; Miller, R. A., International Journal of Mass
Spectrometry 2007, 266, 76-85.


113









67. Shvartsburg, A. A.; Smith, R. D., Journal of the American Society for Mass
Spectrometry 2008, 19 (9), 1286-1295.

68. Manura, D. J.; Dahl, D., SIMION Version 8.0 User Manual. 2007.

69. Eiceman, G. A.; Krylov, E.; Krylova, N.; Nazarov, E. G.; Miller, R. A., Analytical
Chemistry 2004, 76, 4937-4944.

70. Matz, L. M.; Tornatore, P. S.; Hill, H. H., Talanta 2001, 54 (1), 171-179.

71. Tam, M.; Hill, H. H., Analytical Chemistry 2004, 76 (10), 2741-2747.

72. Rorrer III, L. C.; Yost, R. A., International Journal of Mass Spectrometry 2010,
(In Press).

73. Krylov, E. V., International Journal of Mass Spectrometry 2003, 225, 39-51.

74. Rokushika, S.; Hatano, H.; Baim, M. A.; Hill, H. H., Analytical Chemistry 1985, 57
(9), 1902-1907.

75. Shvartsburg, A. A.; Smith, R. D., Journal of the American Society for Mass
Spectrometry 2007, 18, 1672-1681.


114









BIOGRAPHICAL SKETCH

Marilyn Prieto was born in 1977, in Rio Piedras, Puerto Rico. She attended

Florida International University where she received a Bachelor of Science degree in

chemistry in 2005 along with a forensic science certification. Her undergraduate

research involved the development of a universal GC/MS/MS method for the detection

of ignitable liquid residues (ILR) under the advisement of Dr. Jose R. Almirall. After

graduating, she traveled to Midland, Michigan for an internship in Dow Corning

Corporation where her research dealt with making emulsions that consisted of smaller

particle sizes from novel anionic surfactants.

After completing the internship, she travelled to Gainesville, Florida in 2006 to

commence her graduate career in analytical chemistry under the supervision of Dr.

Richard A. Yost. Her graduate interests have been mostly focused on the basic

development and characterization of a portable explosives detection instrument which

utilizes and maximizes the technology of High-Field Asymmetric Waveform Ion Mobility

Spectrometry/Mass Spectrometry (FAIMS/MS). She completed her graduate career

and received her PhD degree in August 2010.


115





PAGE 1

1 HIGH-FIELD ASYMMETRIC WAVEFORM ION MOBILITY SPECTROMETRY: EVALUATION OF NOVEL CELL GEOMETRIES AND TECHNIQUES By MARILYN PRIETO A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLORIDA IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY UNIVERSITY OF FLORIDA 2010

PAGE 2

2 2010 Marilyn Prieto

PAGE 3

3 To my parents and my beloved husbandto -be, Derick Tourne

PAGE 4

4 ACKNOWLEDGMENTS The accomplishments of this work are dedicated to those individuals who helped me stay sane during my years at the University of Florida. I would like to extend my deepest gratitude to my advisor, Dr. Richard Yost, for giving me the opportunity to join his research group and for his consistent support and mentoring. His door was always open for anything, whether it was research or personal. He taught me how to be an independent thinker and a successful scientist. A special thanks goes to all the members of the Yost group for their help, suggestions and friendship. I give special thanks to Leonard Rorrer for always answering the same questions over and over and never seeming to lose his patience. I would like to thank Dr. Jennifer Bryant for all the late nights we spent brooding over the designs and characterizations of the hemispherical cell. I would also like to acknowledge Ch ia -Wei Tsai for helping me better understand how the square wave performs even when the data do not make sense. I cannot forget to thank Mike Napolitano for reminding me how important it is to stay motivated. I would also like to acknowledge Stacey Ann Benjamin and Julianne McLaughlin who were always there to make me laugh when all I wanted to do was cry. Very special thanks go to the Chemistry Machine Shop members: Todd Prox, Brian Smith and Joe Shalosky for the many hours of fruitful conversations concerning design ideas for all the cells we made. My research would have not progressed as smoothly as it did if it was not for them. Last, but not least, I would like to thank my family and friends for always believing in me. They have always been my inspiration when I needed it the most. I give special thanks to my mom for making me the woman I am today and always encouraging me to obtain more education. I give special th anks to my dad for teaching me that there is

PAGE 5

5 always something to smile about. My greatest thanks go to one person and one animal. Derick Tourne, you make me the happiest woman in the world and I cannot wait to spend the rest of my life by your side. Kaligula, my beloved cat, may you rest in peace and know that I will never forget all the times you were the only one to receive me at the door. You will always be my most prized possession.

PAGE 6

6 TABLE OF CONTENTS page ACKNOWLEDGMENTS .................................................................................................. 4 LIST OF TABLES ............................................................................................................ 8 LIST OF FIGURES .......................................................................................................... 9 ABSTRACT ................................................................................................................... 13 CHAPTER 1 INTRODUCTION AND INSTRUMENTATION ........................................................ 15 Background ............................................................................................................. 15 Introduction to Ion Mobility Spectrometry ................................................................ 18 Introduction to High-Field Asymmetric Waveform Ion Mobility Spectrometry .......... 20 Quadrupole Ion Trap Mass Spectrometry ............................................................... 23 Quadrupole Ion Trap Theory ............................................................................ 24 Finnigan LCQ A Commercial QITMS ............................................................ 25 Atmospheric Pressure Chemical Ionization (APCI) ................................................. 27 Overview of Dissertation ......................................................................................... 28 2 FUNDAMENTALS OF FAIMS FOR CURVED ELECTRODE GEOMETRIES ........ 36 Introduction ............................................................................................................. 36 Description of Planar Geometry .............................................................................. 37 Description of Cylindrical Geometry ........................................................................ 38 Description of Hemispherical Geometry .................................................................. 40 Description and Design of a Spherical Geometry ................................................... 42 Experimental ........................................................................................................... 44 Instrumentation ................................................................................................ 44 Samples ........................................................................................................... 45 Results and Discussion ........................................................................................... 46 Comparison of Signal Intensity ......................................................................... 47 Comparison of Resolving Power and Resolution ............................................. 49 Sensitivity of Spherical FAIMS ......................................................................... 52 Conclusions ............................................................................................................ 54 3 EVALUATION OF FAIMS WAVEFORMS APPROACHES ..................................... 81 Introduction ............................................................................................................. 81 Sum of Sines Waveform ......................................................................................... 83 Square Waveform ................................................................................................... 83 Experimental ........................................................................................................... 85 Results and Discussion ........................................................................................... 88

PAGE 7

7 Influence of Dispersion Voltage on Compensation Voltage, Peak Width and Resolving Power ........................................................................................... 88 Influence of Frequency on Compensation Voltage, Peak Width, and Resolving Power ........................................................................................... 90 Influence of Frequency on Sensitivity ............................................................... 91 Influence of Duty Cycle on Sensitivity .............................................................. 91 Lower Concentration Experiments ................................................................... 92 Conclusions ............................................................................................................ 92 4 CONCLUSIO NS ................................................................................................... 106 Future Work .......................................................................................................... 108 Curved FAIMS Cells ....................................................................................... 108 FAIMS Waveforms ......................................................................................... 109 LIST OF REFERENCES ............................................................................................. 110 BIOGRAPHICAL SKETCH .......................................................................................... 115

PAGE 8

8 LIST OF TABLES Table page 2-1. Comparison of dimensions between Thermo-Fisher cylindrical and hemispherical cell. .............................................................................................. 75 2-2. Comparison of dimensions between all three cells. .............................................. 75 2-3. Anticipated ions and adducts for nitramine and nitrate ester explosives. .............. 76 2-4. Comparison of intensity values for RDX, HMX, NG, 2,4-DNT, PETN, TNT, and TNB between the Ionalytics cylindrical, hemispherical, and spherical FAIMS cell. ..................................................................................................................... 76 2-5. Comparison of CV, intensity, and resolving power between the Thermo-Fisher cylindrical at 16,000 V/cm and both the hemispherical and spherical FAIMS cell at 16,072 V/cm. ............................................................................................ 77 2-6. Comparison of resolution between the Thermo-Fisher cylindrical cell at 16,000 V/cm and both the hemispherical and spherical FAIMS cell at 16,072 V/cm. ..... 78 2-7. Comparison of CV, intensity, and resolving power between the Ionalytics cylindrical, hemispherical, and spherical FAIMS cell at 23,500 V/cm. ................ 79 2-8. Comparison of resolution and peak width between the Thermo-Fisher cylindrical, hemispherical and spherical FAIMS cell at 23,500 V/cm. ................. 80 2-9. Repeatability of CV values on the spherical cell from five replicate runs for TNT. ................................................................................................................... 80 2-10. Linear dynamic range and limit of detection for TNT collected by full-scan and SIM mode. .......................................................................................................... 80 3-1. Summary of the CV, PWHM, R p and signal intensity for concentrations from 10 ppm down to 100 ppb. ................................................................................. 105

PAGE 9

9 LIST OF FIGURES Figure page 1-1. Schematic of a characteristic ion mobility spectrometer using the stacked ring design. ................................................................................................................ 30 1-2. Ion motion between two plates during the application of an electric potential. A simplified asymmetric waveform is applied to the upper plate. 40 ........................ 30 1-3. Hypothetical plots of the dependence of ion mobility on electric field strength for three types of ions. 14 ..................................................................................... 31 1-4. Cross sectional view of mass analyzer in the QITMS. 48 ........................................ 32 1-5. Ion motion in a quadrupole ion trap. 49 ................................................................... 33 1-6. Mathieu stability diagram plotted in a and q space for positive ions. The shape of the stable region is rotated about the a z =0 line for negative ions. Most commercial QITMS instruments operate along the a z =0 line (RF-only operation). Under these conditions, ions will have a stable trajectory between q z = 0 and 0.908. The z =1 boundary intersects the q z axis at q z = 0.908. 50 ............................................................................................................... 34 1-7. Schematic of Finnigan LCQ ion trap mass spectrometer. 48 .................................. 35 1-8. Schematic of the atmospheric pressure chemical ionization (APCI) source used in the LCQ. 48 .............................................................................................. 35 2-1. Schematic of hemispherical FAIMS electrodes. (1) outer electrode, (2) outer housing, (3) inner electrode, (4) exit port, (5) base, (6) inner electrode connection, (7) entrance port, and (8) curtain plate. ........................................... 56 2-2. Ions separated between flat plates in the FAIMS analyzer. The transmitted ion is distributed evenly over the gap between the plates. Ions are continuously lost through diffusion and space charge repulsion. 59 ......................................... 56 2-3. Motion of a positively charged ion in a flat plate FAIMS. 21 .................................... 57 2-4. Ions separated between concentric cylinders of a FAIMS analyzer. The transmitted ion is focused towards a fixed radial distance, and distributed around that distance through the effects of diffusion and space charge repulsion. The effect of the focus is exaggerated for clarity in this figure. 59 ........ 57 2-5. Three-dimensional schematic and cross section of the Thermo-Fisher cylindrical geometry FAIMS. (a) Three-dimensional view and (b) cross section view illustrating radial location, r, between the walls of the inner

PAGE 10

10 the inner electrode. 60 .......................................................................................... 58 2-6. Focusing effect caused by curvature of electrodes in FAIMS. Adapted from Barnett et al. 60 .................................................................................................... 58 2-7. Disassembled hemispherical FAIMS cell. ............................................................. 59 2-8. Assembled hemispherical FAIMS cell. Outer casing is made of Kel-F and curtain plate is made of stainless steel. .............................................................. 59 2-9. Variation of the electric field as a function of position for the three FAIMS cells. All three cells contain a gap of 2.50 mm. Radius of inner electrode for cylindrical and hemispherical is 6.00 mm. DV for all three is 4000 V. ................ 60 2-10. Original spherical FAIMS design. Outer casing is made of Delrin, electrodes of brass, and inner electrode of stainless steel. .................................................. 61 2-11. Novel spherical FAIMS cell. Outer casing is made of PEEK, inner and outer electrodes are made of stainless steel. .............................................................. 62 2-12. Dimensions of novel spherical FAIMS cell (in inches). ........................................ 63 2-13. Disassembled spherical FAIMS electrodes. Both inner and outer electrodes are made of stainless steel. ................................................................................ 63 2-14. Assembled spherical FAIMS cell. Outer casing is made of Ketron PEEK. Curtain plate is made of stainless steel. ............................................................. 64 2-15. Structures of several structural classes of common explosives studied in this work. ................................................................................................................... 65 2-16. Comparison of signal intensity of TNT between all three FAIMS cells. Extracted chromatogram of m/z 227 on left and corresponding mass spectra on right. Ionalytics cylindrical FAIMS (top), hemispherical FAIMS (middle), and spherical FAIMS (bottom). ........................................................................... 66 2-17. Comparison of signal intensity of TNB between all three FAIMS cells. Extracted chromatogram of m/z 213 on left and corresponding mass spectra on right. Ionalytics cylindrical FAIMS (top), hemispherical FAIMS (middle), and spherical FAIMS (bottom). ........................................................................... 67 2-18. Comparison of signal intensity of 2,4-DNT between all three FAIMS cells. Extracted chromatogram of m/z 182 on left and corresponding mass spectra on right. Ionalytics cylindrical FAIMS (top), hemispherical FAIMS (middle), and spherical FAIMS (bottom). ........................................................................... 68

PAGE 11

11 2-19. CV scan of a 2 ppm mixture of explosives on the Thermo-Fisher cylindrical cell at a DV -4000 V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB, and (F) PETN. .................................................................................................... 69 2-20. CV scan of a 2 ppm mixture of explosives on hemispherical cell at a DV 3214 V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB, and (F) PETN. ................................................................................................................. 70 2-21. CV scan of a 2 ppm mixture of explosives on spherical cell at a DV -3214 V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB, and (F) PETN. .............. 71 2-22. CV scan of a 2 ppm mixture of explosives on cylindrical cell at a DV -4700 (23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT, (E) 2,4DNT, and (F) 1,3-DNB. ....................................................................................... 72 2.23. CV scan of a 2 ppm mixture of explosives on hemispherical cell at a DV 4700 V (23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT, (E) 2,4-DNT, and (F) 1,3-DNB. ........................................................................... 73 2-24. CV scan of a 2 ppm mixture of explosives on spherical cell at a DV -4700 V (23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT, (E) 2,4DNT, and (F) 1,3-DNB. ....................................................................................... 74 3-1. Schematic of (a) an asymmetric sumof -sines waveform and (b) a rectangular waveform. ........................................................................................................... 94 3-2. Photograph (a) perspective and cross-sectional views (b and c) of miniature planar FAIMS cell. Square wave generator circuit diagram is also shown (d) .. 95 3-4. CV versus DV simulation data for the [M] ion of TNT. .......................................... 97 3-5. Three CV scans of 10 ppm TNT. Peak splitting is observed as DV is increased from (a) to (f). (a) is a DV +121 V, (b) +161 V, (c) +181 V, (d) +201 V, (e) +221 V, and (f) +242 V. ...................................................................................... 98 3-6. Scope traces of square waveforms at different frequencies for amplitudes (a) DV 70 and (b) DV 340 where V 1 is the amplitude (DV) and t 1 is the width of the high voltage pulse. Waveforms are offset for ease of visualization. ............ 99 3-7. Graph displaying CV as a function of DV. For 250 kHz the duty cycle is shown to have little effect. ............................................................................................ 100 3-8. Graph displaying peak width as a function of dispersion voltage. For 250 kHz the duty cycle is shown to have no effect. ........................................................ 100 3-11. Graphs displaying the effect of duty cycle on sensitivity. Frequency is fixed to 250 kHz (a) and 333 kHz (b) ........................................................................... 103

PAGE 12

12 3-12. Graphs displaying the effect of duty cycle on peak area. Frequency is fixed to 250 kHz (a) and 333 kHz (b) ....................................................................... 104

PAGE 13

13 Abstract of Dissertation Presented to the Graduate School of the University of Florida in Partial Fulfillment of the Requirements for the Degree of Doctor of Philosophy A HIGH-FIELD ASYMMETRIC WAVEFORM ION MOBILITY SPECTROMETRY: EVALUATION OF NOVEL CELL GEOMETRIES AND TECHNIQUES By Marilyn Prieto August 2010 Chair: Richard A. Yost Major: Chemistry High-field asymmetric ion mobility spectrometry, or FAIMS, is an emerging detection technology that can operate at atmospheric pressure to separate gas-phase ions. Because FAIMS has continuous ion separation capabilities, it is amenable to use in conjunction with a mass spectrometer. This union offers two orthogonal separation methods, one separating ions according to change in their mobility through a gas as a function of field strength and the ot to -charge ratio. FAIMS has been utilized to detect peptides, bacteria, chemical warfare agents, and even drugs. In literature, there is still much to experiment on the role between threedimensional (3-D) curved cell geometries (spherical, hemispherical, hemi-elliptical, and ovoidal) and two-dimensional (2-D) curved geometries (cylindrical) with ion behavior. Studies encompassing this research focus on exploring the fundamental and practical issues of curved electrode geometries and novel waveform approaches. The marriage of FAIMS and mass spectrometry (MS) was operated with an atmospheric pressur e ionization (API) source which was characterized with each particular geometry while

PAGE 14

14 applying it to explosive detection. Results from these experiments provide a greater understanding of the relationship of cell geometry to ion transmission resolution and resolving power. Although the use of curved surfaces increases sensitivity, resolution is compromised. This effect is more pronounced with increasing curvature and/or surface area. Depending on the application needed by the analyst (more resolution or more transmission), these novel 3D curved electrodes would be valuable to employ. Subsequent to the optimization of FAIMS parameters with particular geometries, the practicability of a new waveform approach for FAIMS was examined. The high voltage asymmetric waveform typically used consists of overlapping sine waves; here an asymmetric square waveform is used. Comparison of these two waveforms shows that the implementation of a square wave on a miniature planar FAIMS would deliver better separation and resolving power, at some cost in transmission and thus sensitivity Finally, once the performance and efficiency of geometry, waveform type, and ionizer was maximized, this novel analytical platform was applied to explosive analysis. Both 3D curved electrodes (hemispherical and spherical) displayed better resolving power and resolution than a commercial cylindrical cell. These novel FAIMS geometries were able to baseline resolve a 2 ppm mixture of five different explosives where the commercial cylindrical cell failed to resolve any. D and 2-D curved geometries while applying two unique waveforms to show its effect on ion behavior in the FAIMS apparatus. Developing and validating the APCI/FAIMS/MS interface allows for better analyte detection with increased resolution, resolving power, and transmission.

PAGE 15

15 CHAPTER 1 INTRODUCTION AND INSTRUMENTATION Background More selective, sensitive, and efficient ion separation techniques are paramount for next-generation detection systems. The sensitivity and selectivity of mass spectrometry (MS) over the years, has made it an indispensible tool for modern analytical chemistry. It should be emphasized that mass spectrometry is not a panacea for solving all problems faced by a scientific expert; no single technique is Nevertheless, it would not be an overstatement to cite MS as a major contributor to the technology that makes it possible for scientists to monitor and enforce environmental, toxicological, and drug regulations. Despite all the research and development that ha ve been concentrated on ion separation, the demand still exists for techniques with better selectivity and sensitivity and with real-time analysis especially for explosives With the potential for increasing domestic terrorist attacks, the detection of hidden explosives has become a subject of major interest. Trace detection of explosives employs a wide variety of analytical methods, including gas chromatography-electron capture detection (GC-ECD) 1 gas chromatography-mass spectrometry (GC-MS) 2-4 liquid chromatography-mass spectrometry (LC-MS) 5, 6 and thin-layer chromatography (TLC). 7 The methods mentioned above all have limitations. For instance, they are generally either susceptible to thermal degradation (GC-ECD and GC-MS) because of por pressure and thermal instability or are too insensitive (LC) because of broader peak shapes and lower peak heights. Furthermore, while ECD is a sensitive means of detecting explosives, it is less selective than the mass

PAGE 16

16 spectrometer. 8 TLC was one of the first techniques used for explosives analysis because of its low cost and its simultaneous analysis of samples; 8 nevertheless, the detection limits of TLC is generally in the low microgram range making it impossible to detect trace quantities. For the analysis of explosives, mass spectrometry is widely used when coupled with chromatographic separation due to the advantages of high sensitivity, short response time, as well as the additional selectivity available from MS/MS and ion/molecule reactions. 9 However, requirements for sample preparation substantially restrict the capability of mass spectrometry for explosive detection. Different ionization techniques were developed to overcome these limitations, including atmospheric pressure ionization sources such as electrospray ionization (ESI) 5, 6, 10 and atmospheric pressure chemical ionization (APCI). 11, 12 Thus far, IMS, formerly known as plasma chromatography, has been most widely used for explosives monitoring because of its high sensitivity and amenability to miniaturization for field portable applications. 13 However, a major disadvantage of IMS is the limited selectivity, such that some interferent ions have the same mobility as the analyte ion, and thus can result in a false positive. FAIMS may be able to reduce this problem because it can separate species with identical low-field mobilities. In recent years, high-field asymmetric waveform ion mobility spectrometry (FAIMS) has been shown to provide high sensitivity, selectivity, and specificity for many analytes when combined with mass spectrometry. 14 Interfacing these techniques allows the use of two orthogonal separation methods, one separating ions based on their differential ion mobility and the other, on their massto -charge ( m/ z ) ratio FAIMS is a valuable

PAGE 17

17 separation technique that exhibits low parts per billion (ppb) limits of detection for continuous vapor streams and is suitable as a gas chromatographic detector owing to its fast response and low memory. 15, 16 It also possesses capabilities of separating isomers and providing additional information orthogonal to MS. 1719 The first FAIMS apparatus was built by Buryakov et al. 20 and consisted of two flat parallel plates for electrodes. Since then, several different geometries have been evaluated and even more have been theoretically proposed, with each geometry having its own specific advantage. For example, a curved cell electrode can provide better sensitivity due to its focusing effect on the ions. 2123 FAIMS is an ion mobility technique that operates at atmospheric pressure and room temperature to separate gas-phase ions. 15, 24, 25 Separation of the io ns occurs by applying a non-constant high electric field (~10,000 V/cm), that alternates between high and low voltages of opposite polarities. Conventional ion mobility spectrometry (IMS) works on a somewhat similar principle but only uses low electric fields, ~200 V/cm, to drive the ions across the drift tube. Ions in IMS are separated according to their mobility, K in a particular drift gas. 26 FAIM S and MS are two orthogonal separation methods, which separate ions depending on their differential ion mobility and ratio of m/z Adding a stage of FAIMS separation before mass spectrometry (using an ionization source capable of working under ambient pressure) is beneficial when working with explosives because of its rapid detection, high sensitivity, improved signalto -noise ratio, and, when curvature is introduced, its ion focusing capabilities. 14, 27

PAGE 18

18 The waveform design is an essential feature of the overall FAIMS performance, and thus was evaluated in this research. This fundamental investigation shows a promising technique for separating gas-phase ions at atmospheric pressure. Gaining a more comprehensive understanding of how FAIMS geometry, and the applied FAIMS waveforms affect operation, will allow the efficient production of next-generation detection devices. Introduction to Ion Mobility Spectrometry Ion mobility spectrometry (IMS), formerly known as plasma chromatography, has been considered a simple and useful, yet relatively low resolution technique, since its inception into the analytical community over thirty years ago. The low resolving power of IMS has made it unattractive to large-scale commercialization, and thus limited its utility in mainstream analytical chemistry. IMS was developed as an analytical technique over 30 years ago. 2830 Only recently have advances in its technology and design helped analytical chemists realize the full potential of this relatively simple and sensitive method of analysis. During the last decade, IMS has evolved into an inexpensive and powerful technique for the detection of many trace compounds and different compound classes such as chemical warfare agents 31 explosives 32 and drugs of abuse. 33 Recently, IMS has moved into several exciting areas, including high resolution separations, which, when used in tandem with MS, 34 have allowed for interesting investigation into gas-phase conformation of isomeric proteins and peptides 35 as well as fast separation of illicit drugs and their metabolites. 33 The basic components of an ion mobility spectrometer consist of a sample inlet, ionization source, reaction region, ion gates(s), drift region, and an ion collector or detector. Figure 1-1 illustrates these components in a conventional, unidirectional, gas

PAGE 19

19 flow system employing APCI as the ionization source and a stacked ring drift region design. In traditional IMS, the sample is introduced into the reaction region with an inert carrier gas. The ions formed are directed toward the drift region by an applied electric field (1-500 V/cm) upon the drift rings. A narrow pulse of ions is injected into the drift region, using an ion gate. Ions travel through a buffer gas at atmospheric pressure, and collisions with the buffer gas help slow the ions down as they travel through the drift tube. In the drift region, the ions are separated based upon their charge and size. The analytical signal is obtained when the ions reach the detector. The larger the collisional cross section of the ion, the more collisions it experiences and thus the more time it requires to reach the detector. Thus, higher mobilities will be obtained by smaller ions, as shown in equations 1-1 to 1-4. Equation 1-1 demonstrates that the velocity ( cm s -1 ) of an ion is equal to its mobility (K, cm 2 V 1 s -1 ) times the electric field (E, V/cm). length of the drift tube (d, cm) and the drift time (t d, s), as shown in equation 12. = = (1 1) = (1 2) K 0) is calculated by correcting for standard temperature (T, in Kelvin) and pressure (P, in Torr) conditions, as shown in equation 13. 0 = 273 ( 760 ) (1 3) At low electric fields, the relationship between the ion mobility and experimental conditions is described by equation 1-4, where q s charge ( N is the number

PAGE 20

20 density of the drift gas, k is the Boltzmann constant, T is the drift gas temperature in Kelvin, and D is the collisional cross section. 36, 37 = 3 16 2 1 2 1 (1 4) For IMS, as for any separation technique, the ability to resolve closely spaced peaks is of considerable interest. Resolving power, R p IMS, as shown in equation 1-5, measures the ability of such an instrument to distinguish two tightly spaced peaks in a single broad peak by dividing the drift time of the peak (t d) by its width at half height ( 1 2 ). Although high resolution IMS has been reported in which resolving powers exceeded some chromatographic systems, 38 resolving powers no greater than 30 have been reported 39 for most commercial instruments. Sensitivity levels of an IMS instrument can be enhanced by increasing the electric field strength of the drift tube or by altering the composition of the drift gas 32, 34 = 1 2 (1 5) Introduction to High-Field Asymmetric Waveform Ion Mobility Spectrometry High-field asymmetric waveform ion mobility spectrometry (FAIMS) is a relatively new technique that separates gas-phase ions at atmospheric pressure and room temperature. 14 FAIMS works similarly to IMS, in which the motion of ions is moved along the drift tube with an electric field at atmospheric pressure. However, FAIMS utilizes a non-homogeneous high electric field (>10,000 V/cm) to separate the ions, whereas IMS uses a homogeneous voltage gradient in the 500 V/cm range. Ions in IMS are separated according to their mobility, K, in a particular drift gas, and their mobility is independent of the applied field; ions will move through the drift

PAGE 21

21 tube at a velocity given equation 1-1. At the higher electric fields employed in FAIMS, K becomes dependent on the applied field, -field mobility, or K h. 40 separation power. Specifically, ions are separated in FAIMS by their difference in mobility at high (K h) and low (K) electric fields. At a constant gas number density, N, the non= 1+ 2 + 4 + (1 6) where K is the ion mobility coefficient at low electric fields and and are coefficients found experimentally that describe the dep electric field in a particular drift gas. 40 Equation 1-6 is an infinite polynomial series, but at realistic field intensities the terms above the fourth order become insignificant. 41, 42 Equation 1-6 provides a method to describe the change in mobility of an ion in a strong electric field, and to ascertain the mobility of that ion as a function of the strength of the electric field. In FAIMS, one of the electrodes is typically held at a ground potential while the other has an asymmetric waveform applied to it. The asymmetric waveform, described by V(t), consists of a high voltage component (V 1), labeled in this dissertation as the dispersion voltage (DV), which lasts for a short period of time relative to a longer lasting low voltage component (V 2) of opposite polarity. The waveform is constructed so that the voltage-time product applied to the electrode is equal to zero, as shown in equation 17. 1 1 + 2 2 =0 (1 7)

PAGE 22

22 At high electric fields, the application of this waveform will cause an ion to experience a net drift toward one of the plates as illustrated in figure 1-2. Ions passing between the electrodes encounter this net displacement from their original position -voltage component is different than that from the low-voltage mobility. 40 In other words, some ions will move farther during the highvoltage portion than during the low-voltage portion, whereas others will move a shorter distance. These ions will continue to migrate towards one of the plates and subsequently be lost unless a constant direct-current compensation voltage (CV) is applied to offset the drift. The CV required to offset the drift of different ions will be different if the K h /K ratio of the ions are different. Thus, a mixture of compounds can be successfully separated by scanning the CV, and allowing each ion to transmit at their particular CV, creating a CV spectrum. When higher electric fields are applied to ions in FAIMS, the ion can have three possible changes in ion mobility as illustrated in figure 1-3. At high-electric fields, the mobility of type A ions increases with increasing electric field strength, whereas the mobility of type C ions decreases, and the mobility of type B ions initially increases before decreasing at yet higher fields. 14 Different polarity waveforms can be used with either positive (P) or negative (N) ions. In curved geometry, the FAIMS instrument is operated in four modes; P1, P2, N1, and N2. The P and N describe the ion polarity and the waveform polarity. The waveform with positive DV yields spectra of types P1 and N2, whereas the reversed polarity yields P2 and N1 type spectra. 43

PAGE 23

23 The ability to explore how geometry affects the performance of FAIMS is essential to thoroughly understanding its capabilities. It has been shown that in a flat plate design, the ions are evenly distributed within the entire analyzer region between the plates, and are continuously being lost to diffusion and space charge repulsion. 22 When curvature is introduced into the geometry of the cell, the ions are focused towards a fixed radial distance. 22 This generally improves transmission but limits resolving power. With a curved geometry, there is still diffusion and space charge effects, but the ions remain distributed within the focused region. Note that ion focusing in curved geometries only occurs in the direction of the applied electric field and thus, lateral diffusion still takes place. FAIMS offers many advantages as an ion separation tool, 44 since it is able to separate gas-phase ions at atmospheric pressure and temperatu re provides high sensitivity with curved cell geometry it separates and process es ions continuously and it interfaces easily to a mass spectrometer. Quadrupole Ion Trap Mass Spectrometry The quadrupole ion trap mass spectrometer (QITMS) has been widely adopted for chemical analysis because of its selectivity, low detection limits, and broad applicability. 45 Briefly, an ion trap is any device that is capable of trapping and storing ions in a confined space for a period of time. The operation of a quadrupole ion trap (QIT) was initially described by Paul and Steinwedel in a patent filed in 1953 in Germany and awarded as a U.S. patent in 1960. 46 The QITMS traps ions in a controlled manner and is similar to a quadrupole mass filter because it uses oscillating electric fields. However, in the QIT, the ions are trapped in a three-dimensional space, whereas the quadrupole mass filter only confines the ions into two-dimensions. 47

PAGE 24

24 Quadrupole Ion Trap Theory The QIT allows for the confinement of ions in a small volume with a threeelectrode device consisting of a ring electrode placed between two endcap electrodes (figure 1-4). The endcap electrodes are held at ground potential while radio frequency (RF) ( and sometimes direct current (DC)) potentials are applied to the ring electrode. A quadrupolar field is created that can trap ions. The RF potential induces oscillations of the ions to hold the ions within the QIT and is characterized by its amplitude (0 t o 8500 V 0-p ) and its frequency (typically 760 kHz). A potential well difference is created within in the trap by applying a DC voltage ( U ) and an AC voltage ( V ), at a particular frequency. This potential difference produces a dynamic electric field for the trapping of ions. Ions stored in the trap follow trajectories described by the Mathieu equation Solutions to the second-order differential Mathieu equation are in terms of two reduced parameters, a z and q z which are defined in equations 1-8 and 1-9. These parameters will determine whether or not the ion will possess a stable trajectory inside the trap. = 2 = 16 0 2 + 2 0 2 2 (1 8) = 2 = 8 0 2 + 2 0 2 2 (1 9) where the subscripts z and r represent axial and radial motion between and perpendicular to the endcaps, respectively; U is the amplitude of the DC voltage applied to the ring electrode, V is the RF potential applied to the ring electrode, e is the charge on an electron (1.602 x 10 19 C), m is the mass of the ion, is the drive frequency (2 f, where f is the drive frequency in s -1 ), and r 0 and z 0 are the trap dimensions, as defined

PAGE 25

25 in figure 1-4. From this equation, on can note that q r is directly proportional to the RF amplitude on the ring electrode, and a r is proportional to the DC voltage applied. An ion that is trapped follows a stable trajectory that has the general appearance of a Lissajous motion as shown in figure 15. An ion that is unstable will have a trajectory that increases in magnitude towards the endcaps or the ring, and thus will be lost through collisions with the electrodes. In order for an ion to be effectively trapped, its trajectory must be stable in both directions. The solution parameters, a z and q z can be mapped to generate the Mathieu stability diagram (figure 1-6). This diagram describes the common region in (a z, q z) space where both the axial and radial components of the ions trajectory are simultaneously stable. This region of stability is defined by the boundaries at z = 0, z = 1, r = 0, and r = 1. An ion that possesses an a z and q z value confined by these values will be stable in both the r and z directions, and will be available for mass analysis. Finnigan LCQ A Commercial QITMS All experiments reported in this research were performed on a commercial benchtop Thermo LCQ QITMS. This instrument is equipped with two external atmospheric pressure ionization (API) sources: atmospheric pressure chemical ionization (APCI), and electrospray ionization (ESI). The system is easily operated in both positive and negative ion mode. For APCI, the sample solution in this research is infused by a syringe pump at a Once ions are formed, the fully or partially desolvated ions are sampled by a heated capillary, as shown in figure 1-7 The ions then pass through a tube lens and skimmer

PAGE 26

26 cone and into a series of octapole ion guides before making it into the ion trap. Figure 1-7 shows a schematic of the Finnigan LCQ ion trap mass spectrometer. The exit of the heated capillary, the tube lens, and the skimmer are held under vacuum (1 Torr). The capillary is heated to help in the desolvation process. Ions are then gated through the skimmer cone using the tube lens. A mass-dependent potential is applied to the tube lens so the ions are focused towards the inlet of the skimmer. The skimmer also acts as a vacuum baffle between the skimmer region (1 Torr) and the first octapole region (10 -3 Torr). The ions are then collected by the first octapole and transmitted to the second octapole through an interoctapole lens. The interoctapole lens assists in focusing of the ions as well as serving as a baffle between the first octapole region (10 -3 Torr) and the analyzer region (2x10 -5 Torr). The ion optics guide ions into the QIT mass analyzer where mass analysis occurs (figure 1-4). In order to inject ions into the trap, an offset voltage of +10V for negative ions and -10V for positive ions is applied to the mass analyzer electrodes. The LCQ QIT operates along the a z = 0 line in the Mathieu stability diagram (figure 16) i.e., the DC potential difference between the ring electrode and the endcap electrodes is zero.) The LCQ QIT utilizes resonant ejection, or axial modulation, to eject ions from the ion trap. Mass analysis is accomplished by ramping the RF voltage on the ring electrode, such that ions of increasing m/z consecutively come into resonance with the axial modulation waveform. The ions gain kinetic energy and are ejected in the z direction. The ion detection system includes a 15-kV conversion dynode and a channel electron multiplier.

PAGE 27

27 Atmospheric Pressure Chemical Ionization (APCI) Gasphase ions in this research are produced with an APCI source illustrated in figure 18. APCI is a gas-phase ionization technique that allows the transfer of ions to the gas phase through ion-molecule reactions at atmospheric pressure. The ionizing reactions of APCI are similar to those of classic CI, although CI is operated under vacuum (~1 Torr). Low-energy electrons are emitted from a corona discharge and initiate the ionization in the source. Analytes in solution are introduced with a syringe pump, are nebulized by a high-speed nitrogen sheath gas flow, pass through a temperature-controlled heater to be vaporized, and are finally ionized by the corona discharge needle. The corona discharge initiates chemical ionization at atmospheric pressure using nitrogen and vaporized solvent as the reagent gas. APCI provides high ionization efficiency since the mean free path of ions at 760 Torr is very short and the number of collisions between reagent ions and an alyte molecules is consequently very high In APCI, the degree of fragmentation is generally quite low, determined by the amount of energy transferred in the ion-molecule reactions. These ion-molecule reactions include proton transfer, charge exchange, electrophilic or nucleophilic addition, and anion abstraction or nucleophilic displacement. For negative ions, proton transfer takes place when a reagent ion (X -) accepts a proton from an analyte molecule (MH), as illustrated in equation 1-10. Since H 2 O has the highest proton affinity of those gases normally present at ambient conditions, it captures any free protons and, upon collisions with other molecules of H 2 O, forms a series of protonated water clusters. 47 + + (1 -10)

PAGE 28

28 An alternative ionization method to proton transfer is charge exchange. This method typically occurs when the analyte molecule does not co nt ain any acidic or basic sites. Charge exchange occurs in the negative mode when the reactant ion (usually O 2 ) has a recombination energy higher than the ionizing energy of the sample. Additionally, with negative ions, stable complexes may be formed by nucleophilic attack (equation 111) with reagent ions of low proton affinity, such as Cl and O 2 Finally, anion abstraction can also occur with negative ions (equation 1-12). These reactions occur when reagent ions react with sample molecules, resulting in fragment ions from the sample. + (1 -11) + + (1 12) Overview of Dissertation This dissertation is organized into four chapters. Chapter 1 is an introduction to the research and gives background information on ion mobility spectrometry (IMS) and high-field asymmetric waveform ion mobility spectrometry (FAIMS). Chapter 1 also briefly discusses the instrumentation that was utilized in this research, including ion trap mass spectrometry and atmospheric pressure chemical ionization (APCI). Chapter 2 explores the fundamental and practical issues of curved electrode geometries in FAIMS. Explicit geometries explored include 3-D curved geometries (spherical and hemispherical) with comparisons to a 2-D curved geometry (cylindrical). Results from these experiments help provide a greater understanding of the relationship of cell geometry to ion transmission and resolution. In Chapter 3, a novel FAIMS waveform approach is evaluated. The advantages and disadvantages of the application of the asymmetric square waveforms compared to

PAGE 29

29 the traditional s um of -sine waveform will be discussed. Effects of waveform frequency, duty cycle, and voltage pulse amplitude are also discussed. Chapter 4 summarizes the major conclusions presented in this work and gives suggestions for future work in this area. The advantages and disadvantages of curved geometries and different waveform approaches for FAIMS are evaluated.

PAGE 30

30 Figure 1-1. Schematic of a characteristic ion mobility spectrometer using the stacked ring design. Figure 1-2. Ion motion between two plates during the application of an electric potential. A simplified asymmetric waveform is applied to the upper plate. 40

PAGE 31

31 Figure 1-3. Hypothetical plots of the dependence of ion mobility on electric field strength for three types of ions. 14

PAGE 32

32 Figure 1-4. Cross sectional view of mass analyzer in the QITMS. 48

PAGE 33

33 Figure 1-5. Ion motion in a quadrupole ion trap. 49

PAGE 34

34 Figure 1-6. Mathieu stability diagram plotted in a and q space for positive ions. The shape of the stable region is rotated about the a z =0 line for negative ions. Most commercial QITMS instruments operate along the a z =0 line (RF-only operation) Under these conditions, ions will have a stable trajectory betwe en q z = 0 and 0.908. The z =1 boundary intersects the q z axis at q z = 0.908. 50

PAGE 35

35 Figure 1-7. Schematic of Finnigan LCQ ion trap mass spectrometer. 48 Figure 1-8. Schematic of the atmospheric pressure chemical ionization (APCI) source used in the LCQ. 48

PAGE 36

36 CHAPTER 2 FUNDAMENTALS OF FAIMS FOR CURVED ELECTRODE GEOMETRIES Introduction Buryakov et al introduced FAIMS with a parallel plate geometry in 1993. 20 Since then, Carnahan et al. improved the sensor design by replacing the flat plates with concentric cylinders. 51 Spherical electrodes were also proposed but never reduced to practice. 52, 53 It has been theorized that curved geometries help to produce significantly better ion transmission efficiency because the electrode curvature allows unique ion focusing capabilities. 22, 23, 54 Ion losses in the analysis channel of planar electrodes arise mainly due to ion diffusion and space-charging. Krylov proposed a method to reduce diffusion losses by employing cylindrical electrodes in FAIMS. 22 With curved electrodes, there exists a non-uniform electric field. This electric field gradient points in the same direction as the vector of the compensating field, 22 and leads to a restoring force that acts on the diffusing ions, leading to compression of the ion cloud. Although the use of curved surfaces increases sensitivity, resolution is compromised. 52 Decreased resolution is due to the ability of spherical or cylindrical surfaces to focus ions which encompass a wide range of mobilities. Since the nonuniformity of the field allows ions with a range of different mobilities to be focused together, they will exit the FAIMS cell at the same time, thus increasing peak width and decreasing resolution. The development of novel hemispherical FAIMS electrodes to balance transmission and resolution has been pursued in this research. In a spherical or hemispherical cavity, all ions encounter equidistant trajectories from ion inlet to outlet, increasing focusing capabilities and transmission. The increased transmission can

PAGE 37

37 allow the utilization of larger electrode radii, which will help to increase resolution. Figure 2-1 shows the novel hemispherical cell design and table 2-1 displays all relevant dimensions in comparison to the commercially available Thermo-Fisher Scientific cylindrical cell. An Ionalytics cylindrical cell was also employed in this research, as noted in table 2-1. The main difference between this cell and the Thermo-Fisher cell is the gap between the two electrodes (2.00 mm vs. 2.50 mm). The strength of the focusing field between the two electrodes is inversely re lated to the radius of the electrodes. The smaller the radius of the electrodes, the stronger the focusing field will be; the larger the radius, the weaker it will be. A systematic exploration of the effect of electrode radius on sensitivity will be discussed in this research, with a comparison of t he differences in sensitivity and resolution of the two cylindrical cells. Description of Planar Geometry Buryakov and others first introduced a parallel plate FAIMS geometry in 1993. 20 They showed that when an asymmetric waveform is applied to one of the plates while keeping the other plate ground ed a uniform electric field is created that allows for the transmission of only certain ions (figure 22) In this type of geometry, the ions are evenly distributed among the entire gap between the two electrodes. Ion losses in the gap of planar electrodes mainly come from lateral diffusion and space charging. The motion of a positively charged ion in a flat plate FAIMS during the application of a hypothetical asymmetric waveform is displayed in figure 2-3. Ions are separated in FAIMS while they are carried by a continuous flow of gas between two closely spaced electrodes The waveform employed by FAIMS is characterized by significant differences in voltage in the positive and negative polarities of the waveform. In figure

PAGE 38

38 2-3 the mobility of a positive ion is higher during the positive (high voltage) portion of this waveform than during the negative (low voltage) portion. This ion will kee p migrating towards the bottom electrode and be discharged if it remains on this path. A low direct current ( DC ) potential is superimposed to offset the drift of the ion. This DC potential is termed its compensation voltage ( CV ). The CV required to offset the drift of individual components in a mixture of compounds will be different if the change in the ions mobility between high and low field is different. Consequently, a mixture of compounds can be successfully separated by allowing each compound to transmit at its particular CV; scanning through these CV values creat es a CV spectrum. The electric field (E flat) between planar electrodes can be calculated using equation 2-1, where Va is the applied voltage, and (b a) is the gap between the electrodes. = 2-1 Description of Cylindrical Geometry Curved geometry FAIMS was introduced by Carnahan and others in 199 5. 51 A schematic of this geometry is displayed in figure 2-4. Curved geometry FAIMS is currently commercialized by Thermo-Fisher Scientific and is sold as a separation device to be used in tandem with a mass spectrometer. A schematic of the Thermo-Fisher cell is displayed in figure 25. The FAIMS process is the same as in a planar geometry a mixture of ions may be filtered to one specific ion by selecting an appropriate compensation voltage. The difference arises when the application of an electric field is between curved surfaces. The field is no longer uniform and the non-uniformity of this field is what helps to focus the ions towards a fixed radial distance within the analyzer region. As more ions are focused towards the center, fewer ions are lost to the walls I n

PAGE 39

39 essence, two-dimensional atmospheric-pressure ion focusing occurs within the analyzing gap. The electric field (Er ) in a cylindrical geometry can be calculated at any point within the analyzer region by using equation 2-2 = 2-2 where V a is the applied voltage, and as shown in figure 25, r is the radius of the ion at any point between the electrodes, a is the radius of the center electrode, and b is the radius of the distance between the center of the inner electrode and inner surface of the outer electrode. Ion focusing in a curved geometry cell is achieved as ions near the inner electrode are repelled to the center of the analytical gap, and ions near the outer electrode are pulled back to the center of the analytical gap. Although ions experience a force that directs them towards a focus point it is important to note that the force is not necessarily strong, and the ions are distributed in space around this focus point due to the influences of diffusion, and ion-ion repulsions (space-charging). This effect was experimentally described by Barnett et al. and is represented in figure 2-6. The CV that the ion needs to traverse the electrodes without being lost is displayed by the solid line. However, the variation in the electric field allows many ions with different CVs to be focused within the analyzer region. The electric field gradient for an example DV where the CV is the opposite polarity of the ion is represented by the dotted line CV correct in figure 2-6. The ions that are near the inner electrode (position of highest electric field) will be repelled away towards the center of the analyzer region where they will fall on the solid line and consequently be focused. Comparatively, ions

PAGE 40

40 that are near the outer electrode will be pulled towards the center, where electric field is higher, and also be focused. Ions that would normally collide with either the outer electrode or inner electrode are represented by the dotted lines CV low and CV high respectively. Since the field is not uniform and is higher near the inner electrode, the ion is pulled away from the outer electrode or inner electrode and successfully focused. Lastly, the dotted line CV incorrect displays an ion with a CV that, despite focusing, is not strong enough to deter the ions from the outer electrode. While curved geometries in FAIMS increase transmission and thus sensitivity, they will decrease resolution. Since curvature aids in focusing a wide range of mobilities, peak widths will increase and resolution will decrease. The opposite is also true; if the radii of curvature is increased, the focusing effect lessens and resolution increases at the expense of sensitivity. Each cell geometry has its advantages and disadvantages regarding the optimization of resolution and ion transmission due to the enhanced focusing that results in improved sensitivity. If the radi us of curvature is decreased, focusing improves and ion intensity increases at the expense of resolution. Clearly, there is a compromise between resolution and transmission for both cylindrical and planar cell geometries. Thus it is interesting to consider geometries with curvature in both directions, i.e., spherical. Description of Hemispherical Geometry In a spherical cavity, all ions encounter equidistant trajectories from ion inlet to outlet, increasing its focusing capabilities, transmission, and uniformly spaced ion density. Nevertheless, the design and construction of a spherical cell is a daunting task; the suspension of a central spherical electrode at the exact center of a spherical cavity

PAGE 41

41 while delivering several thousand volts of RF to the central electrode is extremely difficult. Initial spherical designs failed to provide the precise, accurate, and rigid centering required of the center electrode. Therefore, a second design employ ed a hemispherical central electrode mounted within a hemispherical cavity, as shown in figure 2-1 This design has shown to provide the advantages of a fully spherical cell while permitting rigid, accurate, and reproducible positioning of the electrodes. Characterization and performance of the hemispherical FAIMS cell in comparison to the cylindrical cell will be reported. The material of the inner electrode was stainless steel and was machined into a near hemisphere and fastened with metal screws to a Kel-F base plate. Kel-F, a pol ymer of chlorotrifluoroethyle ne, has the lowest vapor transmission rate of any plastic, excellent thermal characteristics and maintains an operating temperature range of -204 C to 204 C. Fastening the inner electrode to the base plate ensured accurate centering and a uniform analytical gap. hemispherical electrode, the inner electrode was extended 2 mm from a true hemisphere. This way, any path taken by the ions from inlet to exit would be identical. The asymmetric waveform and the compensation voltage were applied to the inner electrode through the mounting screw that was threaded in the center of the electrode (part 6 in figure 2-1). Photos of the hemispherical FAIMS cell are shown in figures 2-7 and 28. The voltage (V hemi) and the electric field (E hemi) in a hemispherical geometry can be calculated at any point within the analyzer region by using equations 2-3 and 2-4

PAGE 42

42 = 1 1 2-3 = 1 1 2 24 where is the change in voltage, a is the inner electrode radius, b is the outer electrode radius, r is position, and V a is the inner electrode voltage. In a planar cell, the electric field does not vary across the gap between the electrodes, as described by equation 2-1 and displayed by figure 2-9. The cylindrical cell, because of its curvature, is capable of focusing ions as described by equation 2-2 and displayed in figure 2-9 by the blue trace. In the cylindrical cell the field varies non-linearly and is higher at the inner electrode than the outer. The hemispherical cell offers higher sensitivity than that of the cylindrical geometry because of its curvature in both directions. The ions that are located at various radial distances between the inner and outer electrode of the hemispherical cell will more effectively fall into the atmospheric pressure ion trapping region because the focusing decreases ion loss to the walls. Figure 2-9 displays how the electric field changes as a function of position for the three cells. As can be seen in the figure, the electric field for the hemispherical cell (green trace) is much higher near the inner electrode than for the other two geometries. This is important, for instance, for a cell that has the same dimensions and the same applied voltages. One would get more efficient separations without having to do any other modifications. Description and Design of a Spherical Geometry Guevremont et al. patented spherical electrodes in 2004, but the patent was never reduced to practice. 52 An initial design (figure 2-10) to suspend an inner electrode within a spherical cavity and not disrupt the electric field was constructed without any

PAGE 43

43 success. The design (figure 2-10) included a spherical inner electrode that consisted of a 0.7 mm stainless steel rod drilled through the center of an 8.8 mm stainless steel ball. In order to deliver the asymmetric waveform and compensation voltage, the inner electrode needed to be suspended in this manner. In order to minimize the disruption of the electric field within the analytical gap of the FAIMS cell, two MACOR sleeves (4 mm in diameter) were utilized to insulate the stainless steel rods. Minimization of any horizontal movement of the inner electrode was attempted by adding MACOR cuffs to the MACOR sleeves. The outer electrode was made of two adjoining brass hemispherical cavities held together tightly with thumb screws. The entire cell was encased in Delrin. Unfortunately, this design was not successful. The primary reason for the poor transmission was probably imprecise centering of the electrode once the cell was assembled. Second, the MACOR sleeves used were very large in comparison to the size of the inner electrode. Ions entering the analyzer gap could theoretically be attracted to the insulating MACOR sleeves and lose their initial trajectory. Also temperatures in excess of 150 C were being used on the heated capillary of the MS. Delrin has an upper temperature limit of 175 C As mentioned previously, the suspension of a central spherical electrode exactly at the center of a spherical cavity while delivering several thousand volts of RF to the central electrode is extremely difficult. Initial designs failed to provide the precise, accurate, and rigid centering required of the electrode. Thus, design of a novel spherical FAIMS cell incorporating a spherical central stainless steel electrode mounted

PAGE 44

44 within a spherical stainless steel cavity that has been cut in half to allow for assembly was created. This design has been shown to provide the ion optical advantages of a fully spherical cell while permitting rigid, accurate, and reproducible positioning of the electrodes. Characterization and performance of the spherical FAIMS cell shows an improvement in sensitivity and resolution when compared to the cylindrical FAIMS cell and will be discussed in this research. A schematic, dimensioned cutaway drawing, and pictures of the novel spherical FAIMS cell are shown in figures 211 2-14, respectively. Both inner and outer electrodes are made of stainless steel. Outer casing is made of Ketron PEEK which is a polymer of polyetheretherketone. The melting point of Ketron is 340 C. The asymmetric waveform and compensation voltage is applied by a stainless steel rod that has been drilled through the center of a 25.4 mm diameter inner electrode. A comparison table of important dimensions between all three cells (spherical, hemispherical, and ThermoFisher cylindrical) is displayed in table 22. Experimental Instrumentation The instrumentation used in these experiments was a Thermo LCQ benchtop quadrupole ion trap mass spectrometer (Thermo-Fisher Scientific, San Jose, CA). The traditional sinusoidal waveform was created by either an Ionalytics GPI 1000 FAIMS system alpha prototype waveform generator or a Thermo-Fisher FAIMS waveform generator. Both generators run with a 33% duty cycle at a frequency of 750 kHz. The voltages for the waveform, compensation voltage, outer electrode, and curtain plate are all supplied by these two waveform generators The Ionalytics generator required t he

PAGE 45

45 compensation voltage to be scanned using an external Agilent 33120-A arbitrary waveform generator, (Loveland, CO). The Agilent generator is programmed to generate linear voltage ramps in either positive or negative directions to scan the compensation voltage. The Thermo-Fisher generator provided the CV ramp so no external generator was required. Negative ions were created by atmospheric pressure chemical ionization (APCI). Analyte samples were infused at 15 L/min. The LCQ does not require any modification to the interface for attachment of the FAIMS assembly. A PEEK adapter was fitted onto the portion of the heated capillary that extends beyond the API stack spray shield (figure 17) The FAIMS cell is fitted into the PEEK adapter. The carrier gas used in the FAIMS apparatus, nitrogen (N 2 ), was passed through an individual gas purification cartridge equipped with molecular sieves for removal of moisture and activated charcoal for removal of organic impurities. Flow rate for nitrogen was controlled by an MKS M100B thermal mass flow controller (Andover, MA) calibrated for the individual gas. Samples For these experiments several structural classes of common explosives were utilized (shown in figure 215), including nitroaromatic explosives such as 2,4,6trinitrotoluene (TNT), 1,3,5-trinitrobenzene (TNB), 2,4di nitrotoluene (2,4-DNT), and 1,3dinitrobenzene (1,3-DNB), 3,4-dinitrotoluene (3,4-DNT); nitramine based explosives such as 1,3,5-trinitroperhydro-1,3,5-triazine ( RDX ) and 1,3,5,7-Tetranitro-1,3,5,7tetraazacyclooctane (HMX); and nitrate esters pentaerythritol tetranitrate (PETN) and 1,2,3-propanetriol trinitrate (n itroglycerin NG). Explosive standard solutions were supplied by Dr. Yehuda Yinon of the Weizmann Institute of Science (Analytical Laboratory, Israeli Police Headquarters).

PAGE 46

46 Negative ion mode was employed since the above explosives tend to display high electron affinities. Furthermore, less chemical noise is generated in negative ion polarity mode thus increasing the selectivity of the analysis. 48 In APCI, negative ions are typically produced by proton abstraction to obtain an [M H] ion 55 except for TNT, TNB, and DNT isomers which produce mostly the [M] ion plus other common fragments. All explosive solutions were diluted in a solvent mixture consistin g of 65 % methanol and 35% deionized water (65:35, v/v) Utilizing a combination of methanol and water as a solvent allows the use of lower vaporizer temperatures with the APCI ion source to avoid the degradation of fragile explosive analytes. Additionally, approximately 0.1% of carbon tetrachloride was added for nitramine and nitrate ester based explosives. The addition of carbon tetrachloride allows a stable explosive adduct ion, [M + Cl] to be formed. Anticipated ions and adducts for nitramine and nitrate ester explosives are shown in table 2-3 For these experiments, all solutions were diluted to a concentration of 10 ppm unless otherwise noted. Results and Discussion A mixture of explosives, as well as individual analyte solutions, w as tested with each geometry to evaluate signal intensity, compensation voltage resolution and resolving power. The resolution between peaks in a mixture can be evaluated by equation 25. = 2( ) ( 1 10 % + 2 10 %) (2 5)

PAGE 47

47 where is the difference in CV values of the two peaks and PW 1 and PW 2 are the full width at 10% of peak 1 and 2, respectively. The resolving power for FAIMS (R p FAIMS) is an instrumental parameter described by equation 2-6 = (2 6) where CV is compensation voltage and PWHM is the peak width at 50% peak height. The CV of any particular ion is calculated by using equation 27. = (2 7) Comparison of Signal Intensity The spatially inhomogeneous field in curved geometry FAIMS not only separates ions but also focuses the ions that are at their correct CV values. Thus, the transmission of ions increases with increasing curvature. Figure 2-16 shows a CV scan (figure 2-16, left) and mass spectrum (figure 2-16, right) of TNT for all three (cylindrical, hemispherical, and spherical) FAIMS cells. Note that these set of experiments were done with the cylindrical FAIMS cell that has a 2.00 mm gap (Ionalytics cell) Separation parameters for all three cells include: DV of -3214 with a CV scan from 0 to +10 V over 2 minutes (5 V/min). The CVs of TNT for all three cells are all ~4. 60 V. Signal intensity for the hemispherical FAIMS shows eight-fold increase from that of the cylindrical cell. The spherical FAIMS cell gave an intensity which is a 23-fold increase from the cylindrical cell. The negative APCI spectrum of TNT shows the production of [M] at m/z 227. The mechanism that occurs with APCI with TNT is assumed to be electron capture since TNT has three bulky electron-withdrawing nitro groups. 56, 57 All three TNT spectra als o display two low-intensity fragment ions, [M-NO] at m/z 197 and [M -OH] at m/z 210, which may form during the transition of the ion from atmospheric

PAGE 48

48 pressure to vacuum through the heated capillary or in the tube lens skimmer region where ions are declustered from solvent vapor. This behavior of increasing intensity with increased FAIMS electrode surface area was verified with TNB and 2,4DNT as displayed in figures 2-17 and 2-18, respectively. For figure 2-17 the separation and scanning parameters include: DV of -3214 with a CV scan from 0 to +10 V over 2.5 minutes (4 V/min). The CVs of TNB for all FAIMS cells are ~6.10 V. Signal intensity for the spherical FAIMS cell (7.50 x 10 5 ) shows a 13-fold increase from the cylindrical FAIMS and the hemispherical FAIMS shows a 3-fold increase from the cylindrical cell. The negative APCI spectrum of TNB is shown in figure 2-17 and it displays the major ion as m/z 213, [M] The major fragment that is produced in all three spectra is m/z 183, [M-NO] Figure 2-18 shows a compensation voltage scan and a mass spectrum of 2,4DNT for all three cells. Separation and scanning parameters are the same as with TNB. The CVs of 2,4-DNT for all three cells is ~7.50 V. Signal intensity for the hemispherical FAIMS shows a 5-fold increase from that of the cylindrical cell. The spherical FAIMS cell gave an intensity value of 7. 27 x 10 5 which is 17-fold increase from the cylindrical cell. The APCI spectrum of 2,4-DNT for all three cells show the molecular ion at m/ z 181 [M] and its major fragment at m/ z 135 [MNO 2 ] This experiment was repeated for several other explosives including RDX, HMX, NG, and PETN. Table 2-4 shows these results. Note that the values in table 2-4 were taken as three repetitions of the compensation voltage scan. The average of the centroid peak and the standard error of standard deviation of the mean ( S E ) is reported in the table. Unfortunately, HMX and NG were not able to be detected with any of the FAIMS cells while scanning the compensation voltage. This can be

PAGE 49

49 attributed either to not scanning the CV slow ly enough to be able to detect the analyte peak or to the fragility of these ions causing the molecular ion to fragment before it reached the ion trap. An investigation of curved geometry FAIMS electrodes has shown that higher signal intensity is achieved with the spherical and hemispherical cells It is assumed this is due to the two-dimensional focusing within the analytical gap of these cells. The stronger the curvature is (when comparing between cylindrical and both the hemispherical and spherical), the more powerful the electric field gradient will be and, thus, the better the sensitivity. It should be noted that even though the hemispherical and spherical cells have the same radius of curvature, they obtained different intensity values. One would assume that a cell with the same radius of curvature should provide identical sensitivity values. The higher sensitivity of the spherical cell could be due to its larger surface area since that is the only real notable difference between the two cells. Further studies with cells of different dimensions, would need to be performed in order to verify this assumption. It should also be noted that there is no significant difference in compensation voltage within the three CV scans of each cell, exemplifying the reproducibility of the hemispherical and spherical FAIMS cells. Comparison of Resolving Power and Resolution Figure 2-19 displays one CV scan of a 2 ppm mixture of TNT, TNB, 2,4-DNT, 1,3DNB, and PETN on the Thermo-Fisher cylindrical cell (2.50 mm gap) The total ion current (TIC) is shown in (A). Intensities of TIC and extracted chromatograms, R p FAIMS, and CVs are also shown in figure. The compensation voltage for figure 219 was scanned from -5 to +15 V over 2 minutes (10 V/min) with a dispersion voltage of -4000 V This dispersion voltage corresponds to an electric field of 16,000 V/cm or 80

PAGE 50

50 Townsend ( Td ) A Townsend is a unit of the E/N ratio, where E is the electric field and N is the gas number density (1 Td = 10 17 V cm 2 ). The same explosives mixture was analyzed on the hemispherical (figure 2-20) cell. Hemispherical cell scanning parameters include a dispersion voltage of -3214 V and a CV scan from -1 to +19 V over 2 minutes (10 V/min). This dispersion voltage corresponds to an electric field of 16,072 V/cm and/or 81 Td. Even though the absolute scanning range of the ThermoFisher cylindrical cell and the hemispherical cell are not identical, the rate of scanning (10 V/min) is the same. Resolution between peaks was calculated using equation 2-5 Resolution and resolving power values are clearly higher for the hemispherical cell. The higher resolution in the hemispherical cell can be attributed to its bigger inner electrode radius and better focusing capabilities. The radius is twice the size of the cylindrical cell, giving it a more planar-like feature, thus increasing its resolution. Also, the path length through the cell is ~2x longer which gives 2x longer residence time, and thus 2x more RF cycles. The bigger inner electrode radius allows for increased resolution wi th increased transmission. The same explosives mixture was attempted on the spherical cell, the results are shown in figure 2-21. The scanning parameters for the spherical cell include a compensation voltage scan from -1 to +19 V over 2 minutes (10 V/mi n) with a DV of 3214 V. As seen above, the spherical FAIMS displays an improvement in sensitivity (typically ~4x) for each of the extracted chromatograms and TIC compared to the hemispherical cell. Note also that the spherical cell exhibits a ~2x decrease in resolving power and resolution from the hemispherical cell Again, this could be attributed to its larger surface area. A larger surface area gives a higher potential for distortion of the

PAGE 51

51 electric field which, in turn, could prevent the ion cloud to exit the cell at the appropriate time and thus broaden its peak width lower resolution could be longer theoretical residence times within the analyzer region. The residence time of the ion denotes how long that ion remains within the gap before exiting towards the mass spectrometer. The calculated residence times for the spherical and the hemispherical cell are identical (114 ms) but the ion does have a larger surface area within the spherical cell which could increase its trajectory time. Theoretically, if the ion has a longer trajectory and remains in the gap for a longer time its peak width will be broader as it exits the cell. Tables 2-5 and 2-6 illustrate specific figures of merit for all three cells to fully compare them. In order to evaluate whether these trends were affected by dispersion voltage another 2 ppm mixture including TNT, TNB, 3,4-DNT, 2,4-DNT, and 1,3-DNB was attempted with all three cells at a higher dispersion voltage. These experiments were performed with the Ionalytics cylindrical cell (2.00 mm gap). The scanning parameters for all three cells included a dispersion voltage of -4700 V (23,500 V/cm) with a compensation voltage scan from 0 to +20 V over 4 minutes (5 V/min). These results are displayed in figures 2-22 (cylindrical), 2-23 (hemispherical), and 2-24 (spherical). Resolving powers, CVs, and intensities for the chromatograms are shown on each figure. The figures of merit for the three cells at 4700 DV are compared i n tables 27 and 2-8. As in the lower dispersion voltage experiments, spherical FAIMS displayed an improvement in sensitivity but a decrease in resolution when compared to the hemispherical cell Even though the resolution and resolving power values were lower for the spherical cell at both dispersion voltages, they still showed an improvement from

PAGE 52

52 both the two cylindrical cells. Attempting to maximize both resolution and sensitivity is dependent upon how well the incorporation of both of the cell are employed. Sensitivity of Spherical FAIMS Since FAIMS is able to separate analyte ions from chemical background noise and thereby enhance selectivity a n evaluation of the analytical capabilities of the spherical FAIMS cell was carried out in this research. Evaluation of CV repeatability, linear dynamic range, and limit of detection for the explosive TNT was performed. Repeatability of compensation voltage in FAIMS is a crucial parameter that describes the consistency of the analytical approach. A repeatable compensation voltage is similar to a reproducible retention time in chromatography and can be treated as a verifiable index of identifying the explosive compound. The repeatability of CV value was tested at te n different dispersion voltage values each with five repetitive injections of a 10 ppm TNT sample while employing the same conditions over a period of several hours. Standard deviations (SDs) and relative standard deviations (RSDs) were calculated by analysis of variance. As shown in table 2-9, better repeatability was obtained at higher dispersion voltages and this is assumed to be due to the increased ion focusing and thus increased ion transmission of analyte ions. This results in a more symmetrical and intense CV peak, which in turn, generates more precise and reproducible CV values. In order to fully validate the spherical FAIMS method, limit of detection (LOD) and linear dynamic range were also characterized. LOD describes the lowest concentratio n level of the analyte that may be determined to be statistically different from a blank sample. 58 In this research, the method validation values are reported in a format

PAGE 53

53 consistent with the International Union of Pure and Applied Chemistry (IUPAC) definitions. The LOD was taken as three times the standard deviation of the blank signal, expressed in concentration. Blanks in this research are defined as infusing the solvent solution (with no analyte) while scanning the mass spectrometer for similar times as actual runs. The signal intensity of blanks were averaged and utilized for statistical analysis. Solutions were flowed at 15 L/min at concentrations of 0, 1, 10, 50, 100, 250, 500, 1000, and 10000 ng/mL (1 ppb to 10 ppb) were made and the mass spectrometer was scanned in the full-scan and selected ion monitoring (SIM) mode. The CV was scanned from 0 to +10 V over two minutes (5 V/min). Linear dynamic range (LDR) was evaluated by five replicate injections of standard mixtures at eight different levels of concentration ranging from 1 ppb to 10 ppm. The lowest point of quantitation, specifically, the concentration yielding a signal that is ten times the standard deviation of the blank. The highest point refers to the highest concentration that shows a linear dependence of the intensity on concentration. Results of these experiments are shown in table 2-10. Based on the data shown, the LOD of TNT collected in full scan MS was 4 ng /mL and was 1 ng/mL in SIM mode. The correlation coefficient (r 2 ) value for the regression was 0.9951 for full scan and 0.9968 for SIM mode, demonstrating a LDR of ~3 orders of magnitude for the FAIMS analysis of TNT Even though higher LODs and lower correlation coefficients were obtained with full scan, it is still a vital mode for research analysis as more information about molecular and fragment ions is provided. FAIMS separation is based on a different principle than chromatography, even though it provides chromatogram-like data. In contrast to liquid or gas chromatography,

PAGE 54

54 where the analytes pass through the column over a window of retention times and are detected, FAIMS can be used to allow only specific analytes to traverse the cell by setting a particular compensation voltage. Thus, FAIMS/MS is more analogous to MS/MS than to GC/MS or LC/MS. Conclusions In this chapter, an investigation of three-dimensional and two-dimensional curved electrodes with respect to signal intensity, resolution, and resolving power was shown. Although the use of curved surfaces increases sensitivity, resolution is decreased. Decreased resolution is due to the ability of cylindrical, hemispherical, or spherical surfaces to focus ions which encompass a wide range of mobilities. Since the nonuniformity of the field allows ions with different mobility behaviors to be focused together, they will exit the FAIMS cell at the same time, thus increasing peak width and decreasing resolution. This effect is more pronounced with increasing curvature and/or surface area as was displayed in the data. To balance transmission and resolution, the development of novel hemispherical and spherical FAIMS electrodes were employed in this research. The highest sensitivity for the explosives was obtained with the spherical FAIMS electrode Additionally, the spherical FAIMS electrode allowed better resolving power and resolution values than both the Thermo-Fisher cell cylindrical (2.50 mm gap) and the Ionalytics cylindrical cell (2.00 mm gap) The increased transmission allows the utilization of larger electrode radii, which can help to increase resolution. The sensitivity of spherical FAIMS was explored to provide information on its quantitative capability. The best CV repeatability was obtained at higher dispersion voltages (-2000 to -3378 V). The standard deviations for TNT ranged from 0 to 0.04

PAGE 55

55 and the relative standard deviations ranged from 0 to 2.3%. Two MS scan modes, fullscan, and SIM, were tested for LOD and LDR. Limit of detection for TNT was 4 ppb for full scan and 1 ppb for SIM. Although full-scan MS provided higher LOD and LDR values, it is still a vital explosive analysis mode because it provides more information about the ion.

PAGE 56

56 Figure 2-1. Schematic of hemispherical FAIMS electrodes. (1) outer electrode, (2) outer housing, (3) inner electrode, (4) exit port, (5) base, (6) inner electrode connection, (7) entrance port, and (8) curtain plate. Figure 2-2. Ions separated between flat plates in the FAIMS analyzer. The transmitted ion is distributed evenly over the gap between the plates. I on s are continuously lost through diffusion and space charge repulsion. 59

PAGE 57

57 Figure 2-3. Motion of a positively charged ion in a flat plate FAIMS. 21 Figure 2-4. Ions separated between concentric cylinders of a FAIMS analyzer. The transmitted ion is focused towards a fixed radial distance, and distributed around that distance through the effects of diffusion and space charge repulsion. The effect of the focus is exaggerated for clarity in this figure. 59

PAGE 58

58 Figure 2-5. Three-dimensional schematic and cross section of the Thermo-Fisher cylindrical geometry FAIMS. (a) Three-dimensional view and (b) cross section view illustrating radial location, r, between the walls of the inner the inner electrode. 60 Figure 2-6. Focusing effect caused by curvature of electrodes in FAIMS. Adapted from Barnett et al. 60

PAGE 59

59 Figure 2-7. Disassembled hemispherical FAIMS cell. Figure 2-8. Assembled hemispherical FAIMS cell. Outer casing is made of Kel-F and curtain plate is made of stainless steel.

PAGE 60

60 Figure 2-9. Variation of the electric field as a function of position for the three FAIMS cells. All three cells contain a gap of 2.50 mm. Radius of inner electrode for cylindrical and hemispherical is 6.00 mm. DV for all three is 4000 V. Inner Electrode Outer Electrode

PAGE 61

61 Figure 2-10. Original spherical FAIMS design. Outer casing is made of Delrin, electrodes of brass, and inner electrode of stainless steel.

PAGE 62

62 Figure 2-11. Novel spherical FAIMS cell. Outer casing is made of PEEK, inner and outer electrodes are made of stainless steel. Top

PAGE 63

63 Figure 2-12. Dimensions of novel spherical FAIMS cell (in inches). Figure 2-13. Disassembled spherical FAIMS electrodes. Both inner and outer electrodes are made of stainless steel.

PAGE 64

64 Figure 2-14. Assembled spherical FAIMS cell. Outer casing is made of Ketron PEEK. Curtain plate is made of stainless steel.

PAGE 65

65 Figure 2-15. Structures of several structural classes of common explosives studied in this work.

PAGE 66

66 Figure 2-16. Comparison of signal intensity of TNT be tween all three FAIMS cells. Extracted chromatogram of m/z 227 on left and corresponding mass spectra on right. Ionalytics cylindrical FAIMS (top), hemispherical FAIMS (middle), and spherical FAIMS (bottom).

PAGE 67

67 Figure 2-17. Comparison of signal intensity of TNB between all three FAIMS cells. Extracted chromatogram of m/z 213 on left and corresponding mass spectra on right. Ionalytics cylindrical FAIMS (top), hemispherical FAIMS (middle), and spherical FAIMS (bottom).

PAGE 68

68 Figure 2-18. Comparison of signal intensity of 2,4-DNT between all three FAIMS cells. Extracted chromatogram of m/z 182 on left and corresponding mass spectra on right. Ionalytics cylindrical FAIMS (top), hemispherical FAIMS (middle), and spherical FAIMS (bottom).

PAGE 69

69 Figure 2-19. CV scan of a 2 ppm mixture of explosives on the Thermo-Fisher cylindrical cell at a DV -4000 V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB, and (F) PETN.

PAGE 70

70 Figure 2-20. CV scan of a 2 ppm mixture of explosives on hemispherical cell at a DV 3214 V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB, and (F) PETN.

PAGE 71

71 Figure 2-21. CV scan of a 2 ppm mixture of explosives on spherical cell at a DV -3214 V. (A) TIC, (B) TNT, (C) TNB, (D) 2,4-DNT, (E) 1,3-DNB, and (F) PETN.

PAGE 72

72 Figure 2-22. CV scan of a 2 ppm mixture of explosives on cylindrical cell at a DV -4700 (23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT, (E) 2,4DNT, and (F) 1,3-DNB.

PAGE 73

73 Figure 2.23. CV scan of a 2 ppm mixture of explosives on hemispherical cell at a DV 4700 V (23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT, (E) 2,4-DNT, and (F) 1,3-DNB.

PAGE 74

74 Figure 2-24. CV scan of a 2 ppm mixture of explosives on spherical cell at a D V -4700 V (23,500 V/cm). (A) total ion current, (B) TNT, (C) TNB, (D) 3,4-DNT, (E) 2,4-DNT, and (F) 1,3-DNB.

PAGE 75

75 Table 2-1. Comparison of dimensions between Thermo-Fisher cylindrical and hemispherical cell. Table 2-2. Comparison of dimensions between all three cells.

PAGE 76

76 Table 2-3. Anticipated ions and adducts for nitramine and nitrate ester explosives. Table 2-4. Comparison of intensity values for RDX, HMX, NG, 2,4-DNT, PETN, TNT, and TNB between the Ionalytics cylindrical, hemispherical, and spherical FAIMS cell.

PAGE 77

77 Table 2-5. Comparison of CV, intensity, and resolving power between the Thermo-Fisher cylindrical at 16,000 V/cm and both the hemispherical and spherical FAIMS cell at 16,072 V/cm.

PAGE 78

78 Table 2-6. Comparison of resolution between the Thermo-Fisher cylindrical cell at 16,000 V/cm and both the hemispherical and spherical FAIMS cell at 16,072 V/ cm.

PAGE 79

79 Table 2-7. Comparison of CV, intensity, and resolving power between the Ionalytics cylindrical, hemispherical, and spherical FAIMS cell at 23,500 V/cm.

PAGE 80

80 Table 2-8. Comparison of resolution and peak width between the ThermoFi sher cylindrical, hemispherical and spherical FAIMS cell at 23,500 V/cm. Table 2-9. Repeatability of CV values on the spherical cell from five replicate runs for TNT. Table 2-10. Linear dynamic range and limit of detection for TNT collected by full-scan and SIM mode.

PAGE 81

81 CHAPTER 3 EVALUATION OF FAIMS WAVEFORMS APPROACHES Introduction High-field asymmetric ion mobility spectrometry, or FAIMS, is an emerging detection technology which can operate at atmospheric pressure to separate gas-phase ions as first described in detail by Buryakov, et al 20 FAIMS is similar to conventional ion mobility spectrometry (IMS), which utilizes relatively low electric fields to propel ions through a drift gas chamber and separate these ions according to their drift velocity. IMS, is an important technique for the detection of many compounds, including narcotics, explosives, and chemical warfare agents because of its high sensitivity and amenability to miniaturization for field-portable applications. 26 In IMS, the ion drift velocity is proportional to the field strength at low electric fields (~200 V/cm) and thus an mobility ( K ) is independent of the applied field. At high fields (~10,000 V/cm), however, ion mobilities become dependent on the applied field and are better represented by K h (equation 1-6), a non-constant high-field mobility term. Variations in K h from the low-field K, and the compound-dependence of that variation affords FAIMS its separation power. FAIMS utilizes a combination of alternating current (AC) and direct current (DC) voltages to transmit ions of interest and filter out other ions, thus improving signal to noise ratio. FAIMS can also help reduce false positives, since two different compounds having the same low-field mobility can often be distinguished in a high-field environment. 61 Ions are separated in FAIMS by their difference in mobility at high (K h) and at low (K) electric fields. At a constant gas number density, N, the non-linear dependence of an equation 1-6 and is re-shown here

PAGE 82

82 ( )= 0 1+ 2 + 4 + (1-6) where K 0 is the ion mobility coefficient at zero electric field and and describe the 40 This equation is an infinite series, but at realistic field intensities the terms above the 4 th order become insignificant. 41, 42 In FAIMS, one of the electrodes is typically held at a ground potential while the other has an asymmetric waveform applied to it. The most commonly used asymmetric waveform, described by V(t) in equation 3-2, consists of a high-voltage component (also referred to as V 1 or dispersion voltage [DV]) which lasts for a short period of time (t 1) relative to a longer lasting (t 2) low-voltage component (V 2) of opposite polarity. Most FAIMS work up to date has employed a sinusoidal wave, plus its first harmonic at twice the frequency, as sh own in equation 3-2, where is the frequency in radians per second. The waveform is constructed so that the voltage-time product applied to the electrode is equal to zero as displayed in equation 1-7. = 0. 61 1 + 0. 39 1 sin 2 2 (3 2) At high electric fields, the application of this waveform will cause an ion to experience a net drift toward one of the electrodes. Ions passing between the electrodes encounter this displacement because the voltage component (K h) is different than that from the low-voltage mobility (K). In other words, the ion will move a different distance during the high-voltage portion than during the low-voltage portion. This ion will continue to migrate towards one of the electrodes and subsequently be lost unless a DC compensation voltage (CV) is applied to offset the drift. The CV values required to offset the drift of different ions will be different if the K h /K ratio of the ions are different. Thus, a mixture of compounds can be successfully

PAGE 83

83 separated by scanning the CV, allowing each compound to transmit at its characteristic CV, creating a CV spectrum. When higher electric fields are applied to the FAIMS electrodes an ion can have three possible changes in ion mobility. The mobility of type A ions increases with increasing electric field strength, the mobility of type C ions decreases, and the mobility of type B ions increases initially before decreasing at yet higher fields. 43, 61 Most lowmass ions (< m/z 300) are type A ions, whereas most high-mass ions are type C ions. 62 Sum of Sines Waveform The dependence of an -varying electric field makes the design of the waveform in FAIMS an essential feature for its overall performance. 63 As mentioned above, most FAIMS work to date has employed a sinusoidal wave plus its first harmonic at twice the frequency, as described in equation 3-2. If the amplitude and CV are being applied to the same electrode, then the addition of CV should be included in equation 3-2. Equation 3-2 results in an asymmetric waveform of the type shown in figure 3-1a. The shape of the waveform is a parameter that contributes to the value of the experimentally measured (compound-dependent) CV for transmission of an ion. A symmetrical waveform (sine or square) should result in CV = 0 V for transmission of all types of ions. Asymmetry of the waveform is required for ion separation and is expressed by differences in the CV 43 Square Waveform Even though most FAIMS experiments have made use of the sumof -sines waveform, theoretical studies have suggested that a rectangular waveform (as shown in figure 3-1b) would be ideal for FAIMS analyses. 6365 Analytical considerations show that rectangular waveforms may improve ion separation efficiency, resolution and/or

PAGE 84

84 sensitivity as compared to sinusoidal waveforms. 62, 6567 Unfortunately, the excessive power load imposed by high frequency, high voltage pulses with steep rise times has hindered the development of electronics that deliver rectangular pulses for driving separations based on differential ion mobility. 63 Intuitively, the use of an asymmetric square waveform for FAIMS would seem to maximize the differences during the high and low field portions of the electric field. These high to low periods of the waveform permit an ion to experience a maximum of unequal voltages maximizing the CV. In previous studies, there have been concerns that the time it takes an ion to respond to the idealized asymmetric square waveform introduce error due to the transient electric field. 42 Lin et al. showed that, to the first order, this can be neglected if the time for reaching terminal velocity is small relative to the total drift time. Since the estimated time necessary to reach this velocity in a transient electric field is in the picosecond range and the drift time is in the millisecond range, this factor can be ignored. Because FAIMS has continuous ion separation capabilities, it is attractive to use in conjunction with a mass spectrometer. This union was utilized to offer orthogonal detection methods, one separating ions according to their mobilities through a gas and the other separating ions according to their massto -charge ratios. In this chapter we describe how the independent control (frequency, amplitude, and duty cycle) of a square waveform can help to more completely characterize FAIMS in terms of transmission, resolution, and separation in comparison to a sumof -sines waveform. Previous experiments have shown the duty cycle to have a strong influence on ion

PAGE 85

85 separation with planar electrodes. 63 Here, the duty cycle is defined as the ratio of the width of the high-voltage pulse to waveform period. A miniature FAIMS cell powered by a digital ly driven circuit producing a square waveform will be described in comparison to an analog sumof -sines waveform. A full comparison of separation and sensitivity is performed over the overlapping range provided by the waveform generators. This chapter also includes SIMION 8 68 modeling to help validate experimental results. Experimental The traditional sinusoidal waveform was created by an Ionalytics GPI 1000 FAIMS system alpha prototype waveform generator which only runs with a 2:1 ratio (33% duty cycle) at a frequency of 750 kHz. This waveform was applied to one of the plates and a reference lead to the other plate. Each individual component waveform was capacitance tuned for the electrodes and added together to produce a waveform similar to the one shown in figure 3-1a. The full range of operation for the Ionalytics alpha prototype is (+/-) 300 V to (+/-) 4000 V; the range utilized for this research was from (+/) 300 V to (+/-) 800 V, since higher voltages could arc within the short distances of the cell. Data were processed using the mass spectrometer instrument software (described below). The square waveform generator used in these studies was constructed by Implant Sciences Corporation; a schematic is shown in figure 3-2. In this computer-generated waveform, three of the four variables (V 1, V 2, t 1, and t 2) are specified; the fourth is deduced from the balanced equation 1-7. The waveform generator is the subject of a patent application. The frequency, duty cycle, and amplitude on the square waveform are all independently controlled. The square generator can provide positive DV values from

PAGE 86

86 +50 V to +750 V. For this reason all experiments with the Ionalytics alpha prototype were only run with positive amplitude values. Three different frequencies (250, 333, 500 kHz) and six different duty cycles (25%, 30%, 33.3%, 35%, 40%, and 45%, defined as t 1/(t 1 + t 2) ) were evaluated on this waveform generator unless otherwise noted. All experiments employed a CV separation over a 10 V range at a scan rate of 5 V/min. Data were processed using instrument software interface (DMS Interface Version 0.25) created by Implant Sciences. The planar geometry FAIMS cell, as shown in figure 32a c, was designed by the University of Florida Chemistry Department Machine Shop and consists of two parallel stainless steel plates (5 mm wide, 15 mm long, 1 mm thick) encased and recessed in a PEEK support (8mm wide, 18 mm long, 3 mm thick) that provides mechanical stability and electrical insulation. Electrical connections to the individual plates were made via posts spot welded to the plates through holes in the PEEK support. The gap within the FAIMS cell (0.38 mm) was maintained by fused silica capillaries as shown in the figure 3-2c The top and bottom plates were then secured to each other through the PEEK with 4 screws to ensure mechanical stability and alignment. The FAIMS cell was interfaced directly to the heated capillary inlet of an LCQ (Thermo Scientific) quadrupole ion trap mass spectrometer ( QI TMS) with the aid of a home-built end-cap connector piece. All experiments were performed utilizing this commercial bench-top MS. The end-cap piece was made of polyaryletheretherketone (PEEK) and is displayed in figure 3-2 (top left). End-cap piece and FAIMS cell were wrapped in Teflon to ensure no signal loss occurred between these connections. No

PAGE 87

87 curtain plate or curtain gas was used and no attempt to perform additional ion desolvation was made. Conventional atmospheric pressure chemical ionization (APCI) was employed to produce gas phase ions. Nitrogen was used as the nebulizing (sheath) and auxiliary gases. APCI and heated capillary temperatures were held at 75 C. Analyte samples were infused with a 500 L Hamilton syringe at 15 L/min. Solutions were injected directly via the syringe pump of the LCQ. Data acquisitions were performed using full scan mode (50-500 m/z ) with a maximum injection time of 100 ms for automatic gain control (AGC). Data were processed using the instrument software interface (Xcalibur version 1.4 SR1). In this chapter, a common explosive, 2,4,6-trinitrotoluene (TNT) was utilized. This explosive was provided by Dr. Jehuda Yinon of the Weizmann Institute of Science, and was obtained from the Analytical Laboratory of the Israeli Police Headquarters. Negative ion mode was employed with APCI. The explosive solution was diluted in a solvent mixture consisting of 65% methanol and 35% de-ionized water. Utilizing a combination of methanol and water as a solvent allows the use of lower vaporizer temperatures than water alone; the low vaporizer temperature help to avoid the degradation of fragile explosive analytes. In this research, all solutions were diluted to a concentration of 10 ppm unless otherwise noted. SIMION 8, the ion trajectory modeling program 68 was utilized in an effort to optimize the performance of the miniature FAIMS cell and validate experimental results. The influences of pneumatic and electrostatic forces at atmospheric pressure are shown. The variables used in this investigation include gas flow, DV CV analytical gap (spacing between the plates), frequency of the asymmetric waveform, its shape

PAGE 88

88 (sinusoidal versus square), and its duty cycle. The potential array of the FAIMS analyzer consisted of parallel plates (5 mm wide, 15 mm long, and separated by a gap, G = 0.5 mm). Statistical Diffusion Simulation 68 (SDS) user program was used to model the ion motion at atmospheric pressure. By modeling both diffusional and mobility terms of ions in a neutral gas, the effects of high pressure collisions are taken into account by the SDS user program. All simulations were performed using TNT, [M] ions (m/z = 227) for which the low field and high field mobilities are known. 62, 6971 Results and Discussion Influence of Dispersion Voltage on Compensation Voltage, Peak Width and Resolving Power The characterization and performance of the square waveform on planar FAIMS ion separation was investigated. In FAIMS, ions are separated based on the change in their mobilit y at high vs. low electric fields. Thus, theoretically, the higher the field, the better separation that would be obtained. Generally, the magnitude of the CV increases as the DV increases for low mass ions ( m/z usually below 300). This behavior is verified with both the sumof -sines and square waveform generators, as displayed in figure 3-3a. The plot shows CV as a function of DV for both waveforms with the sine wave at 750 kHz and the square wave at three different frequencies (250, 333, and 500 kHz). The duty cycle for the sine waveform was 33% (not adjustable); a duty cycle of 25% was selected for the square waveform. It is evident that, for both square and sine waves, CV increases with DV, with square waveform giving higher CV magnitudes within the overlapped DV range (300-350 V). The frequency of the square waveform has no significant effect on the CV. Further insight into the behavior of the [M]ion in the planar geometry FAIMS cell can be gained by simulation of ion trajectories with

PAGE 89

89 SIMION. Figure 3-4 compares the plots for CV vs. DV for both waveforms, over a wider range of DV that can be achieved with our current electronics. Comparing these plots with the experimental data in figure 3-3a shows the same trend of higher CV values for square wave than for sine wave at the same DV (although only a 1.5x increase, less than the 2.5x increase observed experimentally). Note that the simulated CV values are generally smaller than those observed experimentally, perhaps because of the increased CV values recently observed in the presence of solvent vapor in the FAIMS cell. 72 In a planar geometry FAIMS cell, ions are presumably distributed evenly across the gap between the two electrodes. There is typically better resolution with this geometry with an uniform electric field than with the focusing effect observed with curved electrodes. 73 Peak shapes, as well as transmission, are affected by lateral diffusion and space charging (also known as ion-ion repulsions). Figure 3-3b shows that experimental peak widths (full width at half maximum, FWHM) generally increase as DV increases for both waveforms, although the peak width for the square wave increases only slightly over the available DV range. The dramatic increase in peak width for the square waveforms around DV 150 V is due to splitting of the main TNT peak into two peaks. Peak splitting is shown in figure 3-5. The figure displays three CV scans of the extracted chromatograms of m/z 227. The CV scan is from -2 to 8 V over two minutes (5 V/min). The frequency and duty cycle applied were 250 kHz and 25%, respectively. The figure shows a progressive splitting of the CV peak of mass 227 as the dispersion voltage is increased (3-5a 3-5f). It was suspected that this ion, with identical mass ( m/z 227 ), is not TNT and therefore has a different compensation

PAGE 90

90 voltage A s the electric field increases within the gap of the FAIMS cell the difference in their mobility becomes more apparent. Figure 3-5 is a great example of how FAIMS can be utilized as an orthogonal detection method when applied to MS. Resolving power (R p FAIMS) indicates the ability of the FAIMS cell to resolve a peak at a particular compensation voltage. 74 R p FAIMS is defined as the CV divided by the peak width at FWHM and I, as shown in equation 2-6. Figure 3-3c displays that the R p FAIMS increases with DV for both waveforms, although at any given DV, higher R p FAIMS values are obtained for the square waveform. Even though broader peaks are observed at higher dispersion voltages, as shown in figure 3-3b, the even greater increase in CV (figure 3-3a) results in enhanced R p FAIMS. 75 Influence of Frequency on Compensation Voltage, Peak Width, and Resolving Power The rectangular asymmetric waveforms were monitored at different amplitudes and frequencies with a 100 MHz Tektronix TDS 1012B oscilloscope, as shown in figure 3-6 The requested values of the width, t 1 and the amplitude, V 1 of the high voltage pulse are indicated on the figure. It should be noted that the traces are offset vertically from the 250 kHz trace for ease of visualization. All the traces displayed in figure 3-6 are at 25% duty cycle. Here, the duty cycle, d, is defined as = 1 1 + 2 (3 3) Figure 3-3 shows that the frequency of the waveforms has little effect on CV, PWHM, or R p FAIMS. This is as expected, since the frequency of the waveform (and thus the rate of ion oscillations perpendicular to the gas flow) should not affect the CV where the ion is transmitted.

PAGE 91

91 While the duty cycle of the sine waveform is not variable, it is quite simple to vary the duty cycle of the square waveform. At lower duty cycles (where the waveform deviates more from symmetrical ) R p FAIMS increases, as displayed in figure 3-9. Note that at DV values between100 and 200 V, a 25% duty cycle yields an R p FAIMS ~5x higher than a 45% duty cycle. Influence of Frequency on Sensitivity Higher frequencies are generally thought to improve ion separation at the expense of transmission. 65 The loss of transmission can be better tolerated by increasing the frequency, which however is limited by the electronics and creates more distortion in the rectangular shape of the waveform. Figure 310 displays the effect of frequency on sensitivity (310a) and peak area (310b) as a function of dispersion voltage It is known that a reduction of the ion oscillation amplitude is due to the increased frequencies. This increased frequency helps to intensify the transmission of the specified ion and this is verified in figure 310 It should be noted that a 25% duty cycle was selected for these graphs to be able to show 500 kHz frequency data. Higher cycles displayed the same behavior. Influence of Duty Cycle on Sensitivity The duty cycle is one FAIMS parameter that has not been explored in the past since the duty cycle of the sine waveform (used in all commercial FAIMS systems) is fixed to 33%. In this work, the effect of duty cycle of the square waveform on sensitivity has been explored by holding the frequency at a fixed value of 250 kHz and 333 kHz. The results shown in figure 311 indicate that the signal intensity increases as the duty cycle decreases. Indeed, decreasing duty cycle from 45% to 25% increases the signal

PAGE 92

92 by a factor of 2 to 3. Lower duty cycles give greater sensitivity at all frequencies examined Lower Concentration Experiments To evaluate whether the performance of square wave planar FAIMS is a function of concentration, samples varying over a 100x concentration range were evaluated; the results (CV, PWHM, R p and signal intensity) are displayed in table 3-1. These data verify that the reproducibility of CV peak positions and R p are unaffected by concentration within experimental error. Note as well that signal intensity varies linearly with concentration over two decades. Conclusions This paper demonstrated for the first time the practical implementation and characterization of square wave FAIMS on a miniature planar system. The independent control of frequency, amplitude, and duty cycle of a computer-generated square waveform can provide in a more complete characterization of FAIMS in terms of transmission, resolution, and separation compared to a sum of sines waveform. Square wave FAIMS delivers better separation and resolving power, as predicted with ion trajectory simulation. The increase of frequency was shown to increase sensitivity but had little effect on CV, peak width, and R p The lower the duty cycle the higher our peak area and signal intensity. Duty cycle decreases from 45% to 25% also showed a 5x increase in R p Comparison of results with the sum of sines waveform shows that the square waveform can provide better separation and resolving power, albeit at some cost in transmission and thus sensitivity. Further studies should be performed to compare square and sine waveforms with additional explosives and mixture of explosives. It would also be interesting to compare the two waveforms with different cell

PAGE 93

93 dimensions and with different cell geometries. The major limitation of the square waveform is the limited voltage and power that can be obtained with practical waveform generators.

PAGE 94

94 Figure 3-1. Schematic of (a) an asymmetric sumof -sines waveform and (b) a rectangular waveform. Period

PAGE 95

95 Figure 3-2. Photograph (a) perspective and cross-sectional views (b and c) of miniature planar FAIMS cell. Square wave generator circuit diagram is also shown (d) (a) (b) (c) (d)

PAGE 96

96 Figure 3-3. Graphs displaying behavior of CV (a) peak width (b) and resolving power (c) as a function of DV ( a ) ( b ) ( c )

PAGE 97

97 Figure 3-4. CV versus DV simulation data for the [M] ion of TNT. 0.50 0.00 0.50 1.00 1.50 2.00 2.50 3.00 3.50 4.00 0 200 400 600 800 1000 1200 1400 1600 Compensation Voltage (V) Dispersion Voltage (V) SIMION CV/DV Curve for TNT Sine: 750 kHz, 33% DC Square: 750 kHz, 33% DC

PAGE 98

98 Figure 3-5. Three CV scans of 10 ppm TNT. Peak splitting is observed as DV is increased from (a) to (f). (a) is a DV +121 V, (b) +161 V, (c) +181 V, (d) +201 V, (e) +221 V, and (f) +242 V.

PAGE 99

99 Figure 3-6. Scope traces of square waveforms at different frequencies for amplitudes (a) DV 70 and (b) DV 340 where V 1 is the amplitude (DV) and t 1 is the width of the high voltage pulse. Waveforms are offset for ease of visualization. t 1 (a) t 1 (b) V 1 V 1 V 2 V 2 250 kHz, 70 DV 333 kHz, 70 DV 500 kHz, 70 DV 250 kHz, 340 DV 333 kHz, 340 DV 500 kHz, 340 DV

PAGE 100

100 Figure 3-7. Graph displaying CV as a function of DV For 250 kHz the duty cycle is shown to have little effect. Figure 3-8. Graph displaying peak width as a function of dispersion voltage. For 250 kHz the duty cycle is shown to have no effect.

PAGE 101

101 Figure 3-9. R p FAIMS as a function of DV at six different duty cycles.

PAGE 102

102 Figure 3-10. Graphs displaying effect of square wave frequency on sensitivity (a) and peak area (b) (a) (b)

PAGE 103

103 Figure 311. Graphs displaying the effect of duty cycle on sensitivity. Frequency is fixed to 250 kHz (a) and 333 kHz (b) (a) (b)

PAGE 104

104 Figure 3-12. Graphs displaying the effect of duty cycle on peak area. Frequency is fixed to 250 kHz (a) and 333 kHz (b) (a) (b)

PAGE 105

105 Table 3-1. Summary of the CV, PWHM, R p and signal intensity for concentrations from 10 ppm down to 100 ppb. Concentration CV PWHM R p Signal (V) (V) Intensity 10ppm 0.16 0.6 0.26 3.3E+06 1ppm 0.16 0.6 0.25 3.3E+05 500 ppb 0.19 0.6 0.30 1.5E+05 250ppb 0.17 0.6 0.29 6.5 E+04 100ppb 0.12 0.6 0.19 3.2 E+04

PAGE 106

106 CHAPTER 4 CONCLUSIONS This dissertation presents high-field asymmetric waveform ion mobility spectrometry, or FAIMS, combined with mass spectrometry (FAIMS/MS) as a viable technique for the separation of gas-phase explosive ions at atmospheric pressure. FAIMS can behave as an ion filter that is capable of transmitting only selected compounds in a mixture to pass through to the mass spectrometer Gaining a better understanding of how this technique separates explosive ions can lead to improved explosive detection. The application of FAIMS/MS to explosive analysis can improve signalto -noise, while reducing analysis time. The main purpose of this dissertation was to investigate the applicability of novel cell geometries and waveform approaches for FAIMS as a separation tool to be utilized in combination with an atmospheric pressure chemical ionization/mass spectrometry (APCI/MS) system. The three different geometries of FAIMS cells studied and were cylindrical, hemispherical, and spherical. FAIMS experiments were conducted on the molecular ions of several explosives to determine how their signal intensity, resolution and resolving power were affected in relation to the geometry. The highest sensitivity was obtained by the spherical FAIMS electrode with every explosive that was attempted. Additionally, the spherical FAIMS electrode allowed better resolving power and resolution values than either cylindrical cell. The increased transmission allows the utilization of larger electrode radii, which will help to increase resolution. Hemispherical FAIMS was also shown to improve sensitivity, resolution, and resolving power when compared to the cylindrical cell. It was found that the hemispherical electrodes allow for better resolution when compared to the spherical electrodes. The need of these novel

PAGE 107

107 electrode geometries would be dependent on what the analyst needs, whether it be more resolution with increased but adequate sensitivity (hemispherical), or increased sensitivity with adequate resolution (spherical). The analytical performance of spherical FAIMS/MS was evaluated for TNT to provide information on its quantitative capability. The best CV repeatability was obtained at higher dispersion voltages (-2000 to -3378 V). Two MS scan modes, full-scan and SIM, were tested for LOD and LDR. The method prov id ed limits of detection for TNT down to 4 ppb for full1scan and 1 ppb for SIM. Although full-scan MS provided higher LOD and LDR values, it is still a vital explosive analysis mode because it provides more information about the ion. The implementation of square wave FAIMS on a miniature planar system was also demonstrated. The ability to independent ly control frequenc y, amplitude, and duty cycle of a square waveform can aid in a more complete characterization of FAIMS in terms of transmission, resolution, and separation in comparison to a sum of sines waveform was shown. Square wave FAIMS was displayed to provide better separation and resolving power within the available voltage range and this behavior was verified at higher voltages with the SIMION modeling program. The increase of frequency was shown to increase sensitivity as well as peak area, but had no effect on compensation voltage or resolving power. Experiments showed that the lower the duty cycle the higher the signal intensity. Duty cycle decreases also showed a 30-35% improvement in compensation voltage and resolving power. Comparison of experiments with the sum of sines waveform shows that a square wave technology would be ideal in terms of separation and resolving power

PAGE 108

108 Future Work Curved FAIMS Cells There are many opportunities for future work in this project. The variation of the analytical gap and dispersion voltage would affect the ion transmission and resolution in curved geometry FAIMS cells. Gap size is related to transmission efficiency and resolution. For example, ions with higher mobility values will come closer to the electrodes because of their larger ion amplitude oscillation in the high portion of the asymmetric waveform. These ions will have a better chance of being lost to the electrodes, which will reduce the transmission efficiency through the analyzer region. Since the non-uniformity of the electric field in a curved electrode is aiding the focusing of the ions, it will be advantageous to explore if gap size is a dominant factor in transmission loss or is an insignificant factor when compared to the strong focusing properties in the cell. In other words, it would be beneficial to evaluate whether or not reducing the gap size in a curved geometry will have a negative effect on transmission as it does in a planar cell. Higher dispersion voltages are needed in order to improve transmission. Further studies can evaluate explosive ion transmission and separation at higher dispersion voltages on curved geometries. Efforts should also continue to explore other geometries in order to characterize all advantages/disadvantages of these novel electrodes. Further studies are also required to improve the sensitivity to be able to use FAIMS as a quantitative method. Finally, computer simulations of various electrode radius sizes and geometries should also be performed in order to compare transmission and resolution.

PAGE 109

109 FAIMS Waveforms It would be advantageous to improve our understanding of how ions behave under the influence of different parameters of FAIMS, such as different waveforms, in order to apply the optimum condition when more sensitivity or resolution is required for the application of interest. Further studies should be performed to compare square and sine waveforms with different explosives and mixture of explosives as a function of cell dimensions with different cell geometries. Additional experiments using FAIMS waveforms could include the evaluation of other waveform types, higher frequencies, and duty cycles. Finally, computer simulations of various waveform types should be performed to explore the viability of these waveforms.

PAGE 110

110 LIST OF REFERENCES 1. Moore, D. S., Review of Scientific Instruments 2004, 75 (8), 2499-2512. 2. Yinon, J., Journal of Chromatography A 1996, 742 (1 -2), 205-209. 3. Sigman, M. E.; Clark, C. D.; Fidler, R.; Geiger, C. L.; Clausen, C. A., Rapid Commun. Mass Spectrom. 2006, 20 (19), 2851-2857. 4. Sigman, M. E.; Ma, C. Y., Journal of Forensic Sciences 2001, 46 (1), 611. 5. Pan, X. P.; Zhang, B. H.; Tian, K.; Jones, L. E.; Liu, J.; Anderson, T. A.; Wang, J. S.; Cobb, G. P., Rapid Commun. Mass Spectrom. 2006, 20 (14), 2222-2226. 6. Mathis, J. A.; McCord, B. R., Rapid Commun. Mass Spectrom. 2005, 19 (2), 99104. 7. Yinon, J.; Zitrin, S., The Analysis of Explosives Pergamon Press: New York, NY, 1981; Vol. 4. 8. Beveridge, A., Forensic Investigations of Explosives Taylor & Francis Ltd.: Bristol, PA, 1998. 9. Na, N.; Zhang, C.; Zhao, M. X.; Zhang, S. C.; Yang, C. D.; Fang, X.; Zhang, X. R., Journal of Mass Spectrometry 2007, 42 (8), 1079-1085. 10. Gapeev, A.; Yinon, J., Journal of Forensic Sciences 2004, 49 (2), 227237. 11. Fu, X. F.; Zhang, Y.; Shi, S. H.; Gao, F.; Wen, D. W.; Li, W.; Liao, Y. P.; Liu, H. W., Rapid Commun. Mass Spectrom. 2006, 20 (19), 2906-2914. 12. Tachon, R.; Pichon, V.; Le Borgne, M. B.; Minet, J. J., Journal of Chromatography A 2007, 1154 (1 -2), 174-181. 13. Buryakov, I. A., Journal of Chromatography B-Analytical Technologies in the Biomedical and Life Sciences 2004, 800 (1 -2), 75-82. 14. Purves, R. W.; Guevremont, R.; Day, S.; Pipich, C. W.; Matyjaszczyk, M., Review of Scientific Instruments 1998, 69, 4094-4105. 15. Eiceman, G. A.; Tadjikov, B.; Krylov, E.; Nazarov, E. G.; Miller, R. A.; Westbrook, J.; Funk, P., Journal of Chromatography A 2001, 917 (1 -2), 205-217. 16. Eiceman, G. A.; Nazarov, E. G.; Miller, R. A.; Krylov, E. V.; Zapata, A. M., Analyst (Cambridge, United Kingdom) 2002, 127 (4), 466-471.

PAGE 111

111 17. Barnett, D. A.; Purves, R. W.; Ells, B.; Guevremont, R., Journal of Mass Spectrometry 2000, 35 976-980. 18. Robinson, E. W.; Sellon, R. E.; Williams, E. R., International Journal of Mass Spectrometry 2007, 259 (1 -3), 8795. 19. Tang, K. Q.; Li, F. M.; Shvartsburg, A. A.; Strittmatter, E. F.; Smith, R. D., Analytical Chemistry 2005, 77 (19), 6381-6388. 20. Buryakov, I. A.; Krylov, E. V.; Nazarov, E. G.; Rasulev, E. K., International Journal of Mass Spectrometry and Ion Processes 1993, 128 143-148. 21. Guevremont, R.; Purves, R. W., Review of Scientific Instruments 1999, 70 13701383. 22. Krylov, E. V., Technical Physics 1999, 44 113116. 23. Guevremont, R.; Purves, R. W.; Barnett, D. A.; Ding, L. Y., International Journal of Mass Spectrometry 1999, 193 (1), 45-56. 24. Buryakov, I. A.; Kolomiets, Y. N.; Luppu, B. V., Journal of Analytical Chemistry 2001, 56 (4), 336-340. 25. Buryakov, I. A., Talanta 2003, 61 (3), 369-375. 26. Eiceman, G. A.; Karpas, Z., Ion Mobility Spectrometry 2nd ed; CRC Press: Boca Raton, FL, 2005. 27. Barnett, D. A.; Ells, B.; Guevremont, R.; Purves, R. W., Journal of the Americ an Society for Mass Spectrometry 1999, 10, 1279-1284. 28. Karasek, F. W., Research-Development 1970, 21 (3), 3437 29. Cohen, M. J.; Karasek, F. W., Journal of Chromatographic Science 1970, 8 (6), 330-337. 30. Karasek, F. W.; Kilpatri.Wd; Cohen, M. J., Analytical Chemistry 1971, 43 (11), 1441-1447. 31. Zimmermann, S.; Barth, S.; Baether, W. K. M.; Ringer, J., Analytical Chemistry 2008, 80 (17), 6671-6676. 32. Ewing, R. G.; Atkinson, D. A.; Eiceman, G. A.; Ewing, G. J., Talanta 2001, 54 (3), 515-529.

PAGE 112

112 33. Lu, Y.; O'Donnell, R. M.; Harrington, P. B., Forensic Science International 2009, 189 (1 -3), 54-59. 34. Kemper, P. R.; Dupuis, N. F.; Bowers, M. T., International Journal of Mass Spectrometry 2009, 287 (1 -3), 4657. 35. Tao, L.; Dahl, D. B.; Perez, L. M.; Russell, D. H., J Am Soc Mass Spectrom 2009, 20 (9), 1593-1602. 36. Creaser, C. S.; Griffiths, J. R.; Bramwell, C. J.; Noreen, S.; Hill, C. A.; Thomas, C. L. P., An alyst, 2004, 129 (11), 984-994. 37. Revercomb, H. E.; Mason, E. A., Analytical Chemistry 1975, 47 (7), 970-983. 38. Dugourd, P.; Hudgins, R. R.; Clemmer, D. E.; Jarrold, M. F., Review of Scientific Instruments 1997, 68 1122-1130. 39. Kanu, A. B.; Haigh, P. E.; Hill, H. H., Analytica Chimica Acta 2005, 553 (1 -2), 148-159. 40. Guevremont, R., Journal of Chromatography A 2004, 1058 (1 -2), 319. 41. Viehland, L. A.; Guevremont, R.; Purves, R. W.; Barnett, D. A., International Journal of Mass Spectrometry 2000, 197 123130. 42. Shvartsburg, A. A.; Tang, K. Q.; Smith, R. D., Journal of the American Society for Mass Spectrometry 2004, 15 (10), 1487-1498. 43. Purves, R. W.; Guevremont, R., Analytical Chemistry 1999, 71 (13), 2346-2357. 44. Barnett, D. A.; Ells, B.; Guevremont, R., Journal of the American Society for Mass Spectrometry 2000, 11 (12), 1125-1133. 45. Ouyang, Z.; Noll, R. J.; Cooks, R. G., Analytical Chemistry 2009, 81 (7), 24212425. 46. Paul, W.; Steinwedel, H. U.S. Patent; Apparatus for Separating Charged Particles of Different Specific Charges. 1960; Vol. 2939952. 47. Watson, J. T., Types of Mass Spectrometers. In Introduction to Mass Spectrometry 3r d ed.; Lippincott-Raven Publishers: Philadelphia, PA, 1997; p 80. 48. Finnigan LCQ Hardware Manual Classic Series, 97000-97053 Revision B ed.; San Jose, CA, 2003; pp 2-17 2-24.

PAGE 113

113 49. Wuerker, R. F.; Shelton, H.; Langmuir, R. V., Journal of Applied Physics 1959, 30 (3), 342-349. 50. March, R. E., Journal of Mass Spectrometry 1997, 32 (4), 351-369. 51. Carnahan, B. L.; Tarassov, A. S. Ion mobility spectrometer, 1995. 52. Guevremont, R.; Purves, R.; Barnett, D. Spherical sideto -side FAIMS. 359644 & 226965, February 7, 2003. 53. Guevremont, R.; Purves, R. W. FAIMS Apparatus and Method with Ion Diverting Device. 6825461, November 30, 2004. 54. Guevremont, R.; Ding, L. Y.; Ells, B.; Barnett, D. A.; Purves, R. W., Journal of the American Society for Mass Spectrometry 2001, 12 (12), 1320-1330. 55. Hoffman, E. D.; Stroobant, V., Mass Spectrometry Principles and Applications 2nd ed.; John Wiley and Sons: West Sussex, England, 200 2. 56. Spangler, G. E.; Lawless, P. A., Analytical Chemistry 1978, 50 (7), 884-892. 57. Spangler, G. E.; Lawless, P. A., Analytical Chemistry 1978, 50 (2), 290-294. 58. Long, G. L.; Winefordner, J. D., Analytical Chemistry 1983, 55 (7), A712&. 59. Guevremont, R., FAIMS. www.faims.com September 23, 2009. 60. Barnett, D. A.; Belford, M.; Dunyach, J. J.; Purves, R. W., Journal of the American Society for Mass Spectrometry 2007, 18, 1653-1663. 61. Purves, R. W.; Guevremont, R.; Day, S.; Pipich, C. W.; Matyjaszczyk, M. S., Review of Scientific Instruments 1998, 69 (12), 4094-4105. 62. Krylov, E.; Nazarov, E. G., International Journal of Mass Spectrometry 2009, 285, 149-156. 63. Papanastasiou, D.; Wollnik, H.; Rico, G.; Tadjimukhamedov, F.; Mueller, W.; Eiceman, G. A., Journal of Physical Chemistry A 2008, 112 (16), 3638-3645. 64. Krylov, E. V., Instruments and Experimental Techniques 1997, 40 (5), 628-631. 65. Shvartsburg, A. A.; Tang, K.; Smith, R. D., Journal of the American Society for Mass Spectrometry 2005, 16 (1), 212. 66. Krylov, E. V.; Nazarov, E. G.; Miller, R. A., International Journal of Mass Spectrometry 2007, 266 7685.

PAGE 114

114 67. Shvartsburg, A. A.; Smith, R. D., Journal of the American Society for Mass Spectrometry 2008, 19 (9), 1286-1295. 68. Manura, D. J.; Dahl, D., SIMION Version 8.0 User Manual 2007. 69. Eiceman, G. A.; Krylov, E.; Krylova, N.; Nazarov, E. G.; Miller, R. A., Analytical Chemistry 2004, 76 4937-4944. 70. Matz, L. M.; Tornatore, P. S.; Hill, H. H., Talanta 2001, 54 (1), 171-179. 71. Tam, M.; Hill, H. H., Analytical Chemistry 2004, 76 (10), 2741-2747. 72. Rorrer III, L. C.; Yost, R. A., International Journal of Mass Spectrometry 2010, (In Press). 73. Krylov, E. V., International Journal of Mass Spectrometry 2003, 225 3951. 74. Rokushika, S.; Hatano, H.; Baim, M. A.; Hill, H. H., Analytical Chemistry 1985, 57 (9), 1902-1907. 75. Shvartsburg, A. A.; Smith, R. D., Journal of the American Society for Mass Spectrometry 2007, 18 16721681.

PAGE 115

115 BIOGRAPHICAL SKETCH Marilyn Prieto was born in 1977, in Rio Piedras, Puerto Rico. She attended Florida International University where she received a Bachelor of Science degree in chemistry in 2005 along with a forensic science certification. Her undergraduate research involved the development of a universal GC/MS/MS method for the detection of ignitable liquid residues (ILR) under the advisement of Dr. Jose R. Almirall. After graduating, she traveled to Midland, Michigan for an internship in Dow Corning Corporation where her research dealt with making emulsions that consisted of smaller particle sizes from novel anionic surfactants. After completing the internship, she travelled to Gainesville, Florida in 2006 to commence her graduate career in analytical chemistry under the supervision of Dr. Richard A. Yost. Her graduate interests have been mostly focused on the basic development and characterization of a portable explosives detection instrument which utilizes and maximizes the technology of High-Field Asymmetric Waveform Ion Mobility Spectrometry/Mass Spectrometry (FAIMS/MS). She completed her graduate career and received her PhD degree in August 2010.