Convection-Enhanced Delivery of Macromolecules into Nervous Tissue

Permanent Link: http://ufdc.ufl.edu/UFE0022443/00001

Material Information

Title: Convection-Enhanced Delivery of Macromolecules into Nervous Tissue Computational Modeling, MR Imaging, and Experiments
Physical Description: 1 online resource (155 p.)
Language: english
Creator: Chen, Xiaoming
Publisher: University of Florida
Place of Publication: Gainesville, Fla.
Publication Date: 2008

Subjects

Subjects / Keywords: albumin, biotransport, biphasic, ced, contrast, csf, delivery, edema, indentation, infusion, mri, poroelastic, relaxivity, sciatic, spinal
Mechanical and Aerospace Engineering -- Dissertations, Academic -- UF
Genre: Mechanical Engineering thesis, Ph.D.
bibliography   ( marcgt )
theses   ( marcgt )
government publication (state, provincial, terriorial, dependent)   ( marcgt )
born-digital   ( sobekcm )
Electronic Thesis or Dissertation

Notes

Abstract: Convection-enhanced delivery (CED) is a local drug delivery technique which directly infuses drug into the parenchyma of nervous tissue via a cannula. CED provides a promising drug delivery method for treating diseases of the nervous system such as brain tumors, spinal cord injury, and Parkinson?s disease since it circumvents the blood brain barrier, provides more accurate targeting, and reduces the potential of systemic toxicity. CED enhances drug transport into nervous tissue by introducing convectional transport in addition to diffusional transport. This is particularly important for therapy where a large drug distribution volume on the order of centimeters is necessary. Drugs delivered to undesired regions may cause side effects or damage healthy tissues, especially for tumor therapy where many therapeutic agents are toxic. Therapeutic doses should also be in an appropriate range since high doses may be toxic while low doses may result in ineffectual treatment. Concentration distribution of therapeutic agents after infusion is significantly related to CED protocol such as cannula shape and size, infusion site and rate, and infusate concentration. Thus, the CED protocol needs to be well designed. To improve CED protocol design, previous studies have used mathematical models to investigate the mechanics of infusion. These studies provided limited prediction ability. MR imaging methods have also been used to monitor drug distribution during CED. These imaging studies provided limited quantification capacity. Additional prediction tools and imaging techniques need to be developed for potential application in clinical CED protocol optimization and planning. The purpose of this dissertation is to provide fundamental methodologies, in computational modeling and MR imaging techniques, for CED to improve the CED technique and application. First, a biphasic finite element (FE) modeling of CED into nervous tissue was developed. The FE-based model solved for both the fluid flow and macromolecular transport. The model also accounted for deformation-dependent hydraulic permeability, which is difficult to account for using analytical methods. Infusion from a constant pressure source was modeled. In addition, parametric analysis was conducted to determine the sensitivity of macromolecular distribution to changes in tissue properties and infusion parameters. The developed FE model provides a platform upon which anisotropic transport and realistic anatomical boundaries may be modeled in the future. Second, experimental micro-indentation to measure biphasic properties, i.e., Young's modulus and hydraulic permeability, was investigated. Characterization of these tissue properties is necessary for further development of computational transport models. Previous property measurements using micro-indentation are limited to specific indentation configurations. This dissertation used a biphasic FE model for micro-indentation. Micro-indentation using a spherical impermeable indenter on a thin hydrogel contact lens was conducted. Curve-fitting method based on the biphasic FE simulation and micro-indentation was developed. Systemic analysis to understand how biphasic properties and indenter velocity affect the indentation response and fitting result was also investigated. Such an analysis will aid in the design of a more efficient micro-indentation strategy. Third, a non-invasive MR imaging method to quantify the concentration distribution during CED was developed. Previous studies have used MR imaging to quantify the distribution volume of MR-visible tracers in tissues. These studies did not quantify local concentration profiles, and used a simple linear relationship between signal intensity and tracer concentration. In this dissertation, a basic relationship between signal intensity and tracer concentration was quantified. An MR imaging scheme and data analysis method was designed. The method maximally utilized the intrinsic material properties (e.g., T1 and R1) to determine the concentration, and reduces the effect of instrumental factors (e.g., magnetic field inhomogeneity) on concentration calculation. Developing such a non-invasive quantification method is important since spatial concentration profiles can provide more information on the dose at different tissue regions. Comparison between computational transport model predictions and measured concentration profiles may also provide important transport parameters of drugs in nervous tissue in the future. Finally, experiments of CED into spinal cord via the peripheral nerve infusion were conducted. A CED protocol for delivery of compound agents into the rat spinal cord was developed. A real-time in vivo MRI methodology to monitor the agent transport during CED was also developed. Experimental parameters and distribution characteristics of drug transport from the sciatic nerve to the spinal cord was investigated for macromolecular tracers. The developed MR imaging technique and CED protocol provided the infrastructure for further CED studies on rat models. Results of these animal experiments also provided fundamental knowledge of transport pathways from the sciatic nerve to the spinal cord.
General Note: In the series University of Florida Digital Collections.
General Note: Includes vita.
Bibliography: Includes bibliographical references.
Source of Description: Description based on online resource; title from PDF title page.
Source of Description: This bibliographic record is available under the Creative Commons CC0 public domain dedication. The University of Florida Libraries, as creator of this bibliographic record, has waived all rights to it worldwide under copyright law, including all related and neighboring rights, to the extent allowed by law.
Statement of Responsibility: by Xiaoming Chen.
Thesis: Thesis (Ph.D.)--University of Florida, 2008.
Local: Adviser: Sarntinoranont, Malisa.
Electronic Access: RESTRICTED TO UF STUDENTS, STAFF, FACULTY, AND ON-CAMPUS USE UNTIL 2009-02-28

Record Information

Source Institution: UFRGP
Rights Management: Applicable rights reserved.
Classification: lcc - LD1780 2008
System ID: UFE0022443:00001

Permanent Link: http://ufdc.ufl.edu/UFE0022443/00001

Material Information

Title: Convection-Enhanced Delivery of Macromolecules into Nervous Tissue Computational Modeling, MR Imaging, and Experiments
Physical Description: 1 online resource (155 p.)
Language: english
Creator: Chen, Xiaoming
Publisher: University of Florida
Place of Publication: Gainesville, Fla.
Publication Date: 2008

Subjects

Subjects / Keywords: albumin, biotransport, biphasic, ced, contrast, csf, delivery, edema, indentation, infusion, mri, poroelastic, relaxivity, sciatic, spinal
Mechanical and Aerospace Engineering -- Dissertations, Academic -- UF
Genre: Mechanical Engineering thesis, Ph.D.
bibliography   ( marcgt )
theses   ( marcgt )
government publication (state, provincial, terriorial, dependent)   ( marcgt )
born-digital   ( sobekcm )
Electronic Thesis or Dissertation

Notes

Abstract: Convection-enhanced delivery (CED) is a local drug delivery technique which directly infuses drug into the parenchyma of nervous tissue via a cannula. CED provides a promising drug delivery method for treating diseases of the nervous system such as brain tumors, spinal cord injury, and Parkinson?s disease since it circumvents the blood brain barrier, provides more accurate targeting, and reduces the potential of systemic toxicity. CED enhances drug transport into nervous tissue by introducing convectional transport in addition to diffusional transport. This is particularly important for therapy where a large drug distribution volume on the order of centimeters is necessary. Drugs delivered to undesired regions may cause side effects or damage healthy tissues, especially for tumor therapy where many therapeutic agents are toxic. Therapeutic doses should also be in an appropriate range since high doses may be toxic while low doses may result in ineffectual treatment. Concentration distribution of therapeutic agents after infusion is significantly related to CED protocol such as cannula shape and size, infusion site and rate, and infusate concentration. Thus, the CED protocol needs to be well designed. To improve CED protocol design, previous studies have used mathematical models to investigate the mechanics of infusion. These studies provided limited prediction ability. MR imaging methods have also been used to monitor drug distribution during CED. These imaging studies provided limited quantification capacity. Additional prediction tools and imaging techniques need to be developed for potential application in clinical CED protocol optimization and planning. The purpose of this dissertation is to provide fundamental methodologies, in computational modeling and MR imaging techniques, for CED to improve the CED technique and application. First, a biphasic finite element (FE) modeling of CED into nervous tissue was developed. The FE-based model solved for both the fluid flow and macromolecular transport. The model also accounted for deformation-dependent hydraulic permeability, which is difficult to account for using analytical methods. Infusion from a constant pressure source was modeled. In addition, parametric analysis was conducted to determine the sensitivity of macromolecular distribution to changes in tissue properties and infusion parameters. The developed FE model provides a platform upon which anisotropic transport and realistic anatomical boundaries may be modeled in the future. Second, experimental micro-indentation to measure biphasic properties, i.e., Young's modulus and hydraulic permeability, was investigated. Characterization of these tissue properties is necessary for further development of computational transport models. Previous property measurements using micro-indentation are limited to specific indentation configurations. This dissertation used a biphasic FE model for micro-indentation. Micro-indentation using a spherical impermeable indenter on a thin hydrogel contact lens was conducted. Curve-fitting method based on the biphasic FE simulation and micro-indentation was developed. Systemic analysis to understand how biphasic properties and indenter velocity affect the indentation response and fitting result was also investigated. Such an analysis will aid in the design of a more efficient micro-indentation strategy. Third, a non-invasive MR imaging method to quantify the concentration distribution during CED was developed. Previous studies have used MR imaging to quantify the distribution volume of MR-visible tracers in tissues. These studies did not quantify local concentration profiles, and used a simple linear relationship between signal intensity and tracer concentration. In this dissertation, a basic relationship between signal intensity and tracer concentration was quantified. An MR imaging scheme and data analysis method was designed. The method maximally utilized the intrinsic material properties (e.g., T1 and R1) to determine the concentration, and reduces the effect of instrumental factors (e.g., magnetic field inhomogeneity) on concentration calculation. Developing such a non-invasive quantification method is important since spatial concentration profiles can provide more information on the dose at different tissue regions. Comparison between computational transport model predictions and measured concentration profiles may also provide important transport parameters of drugs in nervous tissue in the future. Finally, experiments of CED into spinal cord via the peripheral nerve infusion were conducted. A CED protocol for delivery of compound agents into the rat spinal cord was developed. A real-time in vivo MRI methodology to monitor the agent transport during CED was also developed. Experimental parameters and distribution characteristics of drug transport from the sciatic nerve to the spinal cord was investigated for macromolecular tracers. The developed MR imaging technique and CED protocol provided the infrastructure for further CED studies on rat models. Results of these animal experiments also provided fundamental knowledge of transport pathways from the sciatic nerve to the spinal cord.
General Note: In the series University of Florida Digital Collections.
General Note: Includes vita.
Bibliography: Includes bibliographical references.
Source of Description: Description based on online resource; title from PDF title page.
Source of Description: This bibliographic record is available under the Creative Commons CC0 public domain dedication. The University of Florida Libraries, as creator of this bibliographic record, has waived all rights to it worldwide under copyright law, including all related and neighboring rights, to the extent allowed by law.
Statement of Responsibility: by Xiaoming Chen.
Thesis: Thesis (Ph.D.)--University of Florida, 2008.
Local: Adviser: Sarntinoranont, Malisa.
Electronic Access: RESTRICTED TO UF STUDENTS, STAFF, FACULTY, AND ON-CAMPUS USE UNTIL 2009-02-28

Record Information

Source Institution: UFRGP
Rights Management: Applicable rights reserved.
Classification: lcc - LD1780 2008
System ID: UFE0022443:00001


This item has the following downloads:


Full Text

PAGE 1

CONVECTION-ENHANCED DELIVERY OF MACROMOLECULES INTO NERVOUS TISSUE: COMPUTATIONAL MODELING, MR IMAGING, AND EXPERIMENTS By XIAOMING CHEN A DISSERTATION PRESENTED TO THE GRADUATE SCHOOL OF THE UNIVERSITY OF FLOR IDA IN PARTIAL FULFILLMENT OF THE REQUIREMENTS FOR THE DEGREE OF DOCTOR OF PHILOSOPHY UNIVERSITY OF FLORIDA 2008 1

PAGE 2

2008 Xiaoming Chen 2

PAGE 3

To my parents, my wife Han, and my newborn son! 3

PAGE 4

ACKNOWLEDGMENTS First of all, I would like to thank my advisor, Dr. Malisa Sarntinoranont, for everything she has done throughout my doctoral study at the University of Florida. This dissertation would not be possible without her. Dr. Sarn tinoranont introduced me to th is research area and provided financial support. Her insight and experience in this area guided me through my research and she gave numerous valuable suggestions to this diss ertation. I am also thankf ul to Dr. Thomas H. Mareci for his advice throughout the magnetic re sonance imaging experiments and for allowing me to use research resources in his lab. The MR I experiments in this dissertation would not be possible without him. I would also like to th ank my other excellent committee members, Dr. Roger Tran-Son-Tay and Dr. Jacob N. Chung, for providing valuable advice to this dissertation. My doctoral research was a happy cooperation with many people. Dr. Sarntinoranont has been involved with the whole progress. I have collaborated with Dr. Ma reci and Garrett W. Astary on the MRI experiments. I have also colla borated with Dr. Greg W. Sawyer and Alison C. Dunn on the micro-indentation experiments. Friends in my lab, Dr. Marecis lab, my depa rtment, and the University of Florida are much appreciated for sharing my joys and sadnesses. Special thanks go to Garrett W. Astary with whom the imaging night is fun with hamburge r, pizza, rats and red eyes. Thanks also go to Greg Pishko, Jianbing Zhao, Sung Jin Lee, H ector Sepulveda, Mansi Parekh, Jung Hwan Kim, Ana Saaibi, Jessica Cobb, Jessica Meloy, and Alison C. Dunn with whom I have enjoyed a happy and multi-cultural life at UF. I am also fortunate to enjoy the friendship of Ming Zhao, Jing Xu, Yiwen Wang, Yuchu Tong, Jianlong Xu, Junqiang Wang, Yawei Li, Fei Liu, and their families. Staffs in the RF lab in the McKnight Brain In stitute of UF are also appreciated for their help in my MRI experiments. 4

PAGE 5

Last, but not least, I woul d like to thank my parents and my wife Han for their understanding and love during th e past few years. Their support and encouragement are my source of strength. Thanks also go to my newborn son who brings so much happiness to my life. This research was supported in part by the grants NIH R21-NS052670. I also have received financial support from the Departme nt of Mechanical and Aerospace Engineering, College of Engineering, University of Florida. 5

PAGE 6

TABLE OF CONTENTS page ACKNOWLEDGMENTS...............................................................................................................4 LIST OF TABLES................................................................................................................. ..........9 LIST OF FIGURES.......................................................................................................................10 LIST OF NOMENCLATURE.......................................................................................................13 ABSTRACT...................................................................................................................................16 CHAPTER 1 INTRODUCTION..................................................................................................................19 1.1 Background of Convection-Enhanced Delivery (CED)...................................................19 1.2 Literature Review and Motivation....................................................................................21 1.2.1 Mathematical Modeling of CED............................................................................23 1.2.2 Characterization of Soft Biomaterials and Tissues................................................26 1.2.3 In Vivo Magnetic Resonance Imaging (MRI) Monitoring of Drug Transport during CED..................................................................................................................29 1.2.4 In Vivo CED into Rat Spinal Cord via Sciatic Nerve Infusion and Real-time MRI..............................................................................................................................31 1.3 Contributions....................................................................................................................32 2 BIPHASIC FINITE ELEMENT MODEL OF SOLUTE TRANSPORT FOR DIRECT INFUSION INTO NE RVOUS TISSUE.................................................................................42 2.1 Introduction............................................................................................................... ........42 2.2 Theory and Methods.........................................................................................................45 2.2.1 Mechanics Model...................................................................................................45 2.2.2 Solute Transport Model..........................................................................................48 2.2.3 Numerical Implementation.....................................................................................49 2.2.4 Tissue Infusion Model............................................................................................50 2.3 Results...............................................................................................................................52 2.3.1 Validation of the Finite Element Model.................................................................52 2.3.2 Biphasic Sensitivity Analysis.................................................................................53 2.3.3 Solute Transport Sensitivity Analysis....................................................................55 2.4 Discussion.........................................................................................................................56 2.5 Conclusions.......................................................................................................................60 3 CHARACTERIZATION OF BIOMATERIAL/TISSUE USING MICROINDENTATION.....................................................................................................................70 3.1 Introduction............................................................................................................... ........70 6

PAGE 7

3.2 Materials and Methods.....................................................................................................74 3.2.1 Material and Experimental Testing........................................................................74 3.2.2 Mechanics Model...................................................................................................75 3.2.3 Finite Element Modeling........................................................................................77 3.2.4 Determination of Biphasic Properties....................................................................78 3.3 Results...............................................................................................................................79 3.3.1 Experimental Micro-indentation Results................................................................79 3.3.2 Sensitivity of the Predicted F-D Response.............................................................80 3.3.3 Biphasic Parameter Estimates................................................................................80 3.4 Discussion.........................................................................................................................82 3.5 Conclusions.......................................................................................................................87 4 QUANTITATIVE ASSESSMENT OF MACROMOLECULAR CONCENTRATION DURING CED BY USING CONTRAST-ENHANCED MRI: A HYDROGEL PHANTOM STUDY..............................................................................................................94 4.1 Introduction............................................................................................................... ........94 4.2 Theory...............................................................................................................................97 4.2.1 Relationship between Concentration and Signal Intensity.....................................97 4.2.2 Signal in the Presence of Noise..............................................................................98 4.2.3 Porous Media Model of Direct Infusion.................................................................99 4.3 Materials and Methods.....................................................................................................99 4.3.1 Materials.................................................................................................................99 4.3.2 MR Imaging..........................................................................................................100 4.3.3 MRI Data Processing............................................................................................100 4.3.4 Quantitative Optical Measurement of Concentration...........................................101 4.3.5 Simulation of Infusion into Porous Media...........................................................101 4.4 Results.............................................................................................................................103 4.4.1 Measurement of T10, T20, R1, and R2 for the Hydrogel.........................................103 4.4.2 Infusion Concentration Analysis..........................................................................103 4.4.3 MRI Results..........................................................................................................103 4.5 Discussion.......................................................................................................................105 5 CONVECTION-ENHANCED DELIVERY INTO THE RAT SPINAL CORD VIA THE SCIATIC NERVE INFUSION AND REAL-TIME MR IMAGING..........................116 5.1 Introduction............................................................................................................... ......116 5.2 Materials and Methods...................................................................................................120 5.2.1 Infusion System Setup..........................................................................................120 5.2.2 Animal Preparation...............................................................................................120 5.2.3 MR Imaging..........................................................................................................121 5.2.4 MRI Data Processing............................................................................................122 5.3 Results.............................................................................................................................122 5.4 Discussions.....................................................................................................................124 6 SUMMARY AND FUTURE WORK..................................................................................138 7

PAGE 8

APPENDIX A ANALYTICAL SOLUTION OF INFUSION INTO A SPHERICAL POROUS MEDIA..142 LIST OF REFERENCES.............................................................................................................143 BIOGRAPHICAL SKETCH.......................................................................................................155 8

PAGE 9

LIST OF TABLES Table page 2-1 Material parameter ranges of gray ma tter brain tissue used in simulation studies............61 3-1 Estimates of biphasic propertie s at different Poisson ratios..............................................88 5-1 Experimental parameters for CED into rat sciatic nerves................................................129 9

PAGE 10

LIST OF FIGURES Figure page 1-1 Schematic of the bl ood-brain-barrier (BBB).....................................................................37 1-2 Drug release from a polymeric implant.............................................................................37 1-3 A) Convection-enhanced delivery into a monkey brain. B) Infusate backflow into adjacent white matter.........................................................................................................3 8 1-4 Changes of drug concentration in blood plasma following a single dose administration of a therapeutic agent. () safe dose, (---) unsafe dose, (-) ideal dose [3]...............................................................................................................................38 1-5 Electron micrograph of a small region of the cerebral cortex (t he surface layer of gray matter of the cerebrum) of a rat.................................................................................39 1-6 Micro-indentation test..................................................................................................... ...39 1-7 Nano-indentation method proposed by Oliver and Pharr [39]...........................................40 1-8 Creep test of biphasic materi als proposed by Mow et al. [43]...........................................40 1-9 Distribution of 14C-albumin after CED directly into the dorsal column (white matter) of the rat spinal cord......................................................................................................... .41 1-10 Coronal MR image of the bilateral lower extremities of a primate infused with 40 L Gd-albumin in each sciatic nerve.......................................................................................41 2-1 Finite element (FE) mesh and boundary c onditions used to model pressure-controlled infusion into tissue........................................................................................................... ..62 2-2 Validation analysis comparing transient FE and analytical solutions for infusion into tissue..................................................................................................................................63 2-3 Effect of Youngs modulus on pressu re-controlled infu sion into tissue............................64 2-4 Effect of hydraulic permeability on pr essure-controlled infusion into tissue....................65 2-5 Effect of nonlinear hydrau lic permeability on pressure-controlled infusion into tissue....66 2-6 Effect of infusion pressure on infusion into tissue.............................................................67 2-7 Effect of infusion pressure on infusion rate.......................................................................67 2-8 Effect of varying tissue stiffness and hydraulic permeability on albumin distribution during pressure-controlled infusion...................................................................................68 10

PAGE 11

2-9 A) Albumin distribut ion and B) Peclet number. C) E ffect of retardation coefficient, ,on albumin transport at t = 3min....................................................................................69 3-1 Schematic representation of the microindentation experiment on contact lenses............89 3-2 Cantilever force response during micro-indentation (unsubmerged)................................89 3-3 Axisymmetric FE mesh of the contac t lens and imposed boundary conditions on the top and bottom surfaces.....................................................................................................90 3-4 Fitting between computational and expe rimental curves at different indenter velocities (1.2 m/s and 20 m/s)......................................................................................90 3-5 Changes in the AUC (area under curve) co rrespond to changes in F-D behavior A). Sensitivity of F-D curves to B) indenter velocity and C) permeability.............................91 3-6 Mean-square-error (MSE) map generated to compare experimental and simulated FD response for varying E and k ..........................................................................................91 3-7 Mean-square-error (MSE) maps generated to compare simulated and experimental FD responses at indenter velocities of 1.2 m/s and 20 m/s.............................................92 3-8 Predicted pore pressure distribution for different indenter velo cities using optimal parameters..................................................................................................................... .....93 4-1 Schematic of the infusion system setup...........................................................................111 4-2 Optical imaging method used to measure the concentration of Evans blue labeled Gd-Albumin in the hydrogel slice....................................................................................111 4-3 Plot of the theoretical relationship between concentration and signal intensity used to determine optimal infu sate concentration........................................................................112 4-4 A) Evolution of T1-weighted MR signal during infu sion of Gd-Albumin into 1% agarose hydrogel at a rate of 0.29 L/min. B) The signal profile in radial direction was obtained by sampling in the small recta ngular regions of interest (ROIs). C) Contours of MR signal inte nsity at time = 91 min..........................................................113 4-5 Magnetic resonance-derived concentrat ion of Evans blue labeled Gd-Albumin compared with profiles measured directly using quantitative optical imaging in a 1 mm thick hydrogel slice...................................................................................................114 4-6 Magnetic resonance-derived concentration of Gd-Albumin compared with predicted porous media transport profiles at time = 30 and 91 min................................................115 5-1 Schematic of the anatomy of the spinal column region...................................................130 11

PAGE 12

5-2 External anatomy of the sciatic nerve, the spinal cord and other connected spinal nerves...............................................................................................................................130 5-3 A) Schematic of the sciatic nerve infusion. B) The isolated rat sciatic nerve. C) The dual RF coil system with an anesthet ized rat used to collect MRI data..........................131 5-4 A) A typical transverse MR image of the rat spinal cord. B) Th e distribution of GdDTPA within the sciatic nerve (arrows) at the end of direct nerve infusion in rat #1.....132 5-5 Dynamic contrast-enhanced MRI scans during infusion of Gd-DTPA into the right sciatic nerve of rat #2.......................................................................................................133 5-6 Dynamic contrast-enhanced MRI scans during infusion of Gd-DTPA-albumin into the right sciatic nerve of rat #3........................................................................................134 5-7 Evolving Gd-DTPA-albumin distribution pattern during CED into the right sciatic nerve of rat #4................................................................................................................ ..135 5-8 Evolving Gd-DTPA-albumin distribution pattern during CED into the right sciatic nerve in rat #5 (Top) and rat #6 (Bottom)........................................................................136 5-9 Graph of infused Gd-DTPA-albumin vol ume versus estimated tissue distribution volume in the spinal cavity (rats #3 to 6).........................................................................137 6-1 Framework for CED pred iction and optimization...........................................................141 12

PAGE 13

LIST OF NOMENCLATURE BBB Blood brain barrier CED Convection-enhanced delivery CNS Central neural system CSF Cerebrospinal fluid DCE-MRI Dynamic contrast-enhanced MRI ECS Extracellular space FE Finite element FOV Field of view Gd Gadolinium Gd-DTPA Gd-diethylenetriamine penta-acetic acid GW Gray matter MR Magnetic resonance MRI Magnetic resonance imaging MW Molecular weight RF Radio frequency ROI Region of interest SE Spin echo SNR Signal noise ratio TE Time of echo TR Time for recovery WM White matter a Radius of infusion cavity c Solute concentration in tissue 13

PAGE 14

effD Effective diffusivity of solute in tissue e Volume dilatation (= ) () Tr E Youngs modulus AH Aggregate modulus (=2 ) k Hydraulic permeability M Material constant for deformation-dependent hydraulic permeability p Pore (interstitial) fluid pressure eP Peclet number (/eff L vD ) 12, R R Relaxivities of a contrast agent 0S Proton density ()Sc Signal intensity with an agent concentration c 0t Ramping time of load 1T Longitudinal relaxation time of water 2T Transverse relaxation time of water u Displacement vector v Volume-averaged bulk velocity cv Solute velocity f v Intrinsic fluid velocity s v Solid velocity Retardation coefficient Infinitesimal deformation tensor f Volume fraction of fluid phase; Porosity 14

PAGE 15

s Volume fraction of solid phase Lam elastic constant of solid phase Shear modulus of solid phase Poisson ratio E Contact stress tensor f Cauchy stress tensor of fluid phase s Cauchy stress tensor of solid phase 15

PAGE 16

Abstract of Dissertation Pres ented to the Graduate School of the University of Florida in Partial Fulfillment of the Requirements for the Degree of Doctor of Philosophy CONVECTION-ENHANCED DELIVERY OF MACROMOLECULES INTO NERVOUS TISSUE: COMPUTATIONAL MODELING, MR IMAGING, AND EXPERIMENTS By Xiaoming Chen August 2008 Chair: Malisa Sarntinoranont Major: Mechanical Engineering Convection-enhanced delivery (C ED) is a local drug delivery technique which directly infuses drug into the parenchyma of nervous ti ssue via a cannula. CED provides a promising drug delivery method for treating diseases of the nervous system such as brain tumors, spinal cord injury, and Parkinsons di sease since it circumvents the blood brain barrier, provides more accurate targeting, and reduces the potential of sy stemic toxicity. CED enhances drug transport into nervous tissue by introducing convectional tr ansport in addition to diffusional transport. This is particularly important for therapy wher e a large drug distribution volume on the order of centimeters is necessary. Drugs delivered to undesired regions may cause side effects or damage healthy tissues, especially for tumor therapy where many therapeu tic agents are toxic. Therapeutic doses should also be in an appropriate ra nge since high doses may be toxi c while low doses may result in ineffectual treatment. Concentration distributi on of therapeutic agents after infusion is significantly related to CED protocol such as can nula shape and size, infusion site and rate, and infusate concentration. Thus, the CED protocol needs to be well designed. To improve CED protocol design, previous studies have used mathematical models to investigate the mechanics of infusion. These studies provided limited prediction ability. MR imaging methods have also been 16

PAGE 17

used to monitor drug distribution during CED. These imaging studies provided limited quantification capacity. Additional prediction tools and imaging techniques need to be developed for potential application in clinical CED protocol optimization and planning. The purpose of this dissertation is to provide fundamental methodologies, in computational modeling and MR imaging techniqu es, for CED to improve the CED technique and application. First, a biphasic finite element (FE) mode ling of CED into nervous tissue was developed. The FE-based model solved for both the fluid flow and macromolecular transport. The model also accounted for deformation-dependent hydrau lic permeability, which is difficult to account for using analytical methods. Infusion from a constant pressure source was modeled. In addition, parametric analysis was conducted to determine th e sensitivity of macrom olecular distribution to changes in tissue properties and infusion para meters. The developed FE model provides a platform upon which anisotropic transport and realistic anatomi cal boundaries may be modeled in the future. Second, experimental micro-indentation to measure biphasic properties, i.e., Youngs modulus and hydraulic permeability, was investigate d. Characterization of these tissue properties is necessary for further development of comput ational transport models. Previous property measurements using micro-indentation are limited to specific indentati on configurations. This dissertation used a biphasic FE model for microindentation. Micro-indent ation using a spherical impermeable indenter on a thin hydrogel contact lens was conducted. Curve-fitting method based on the biphasic FE simulation and micro-indentation was developed. Systemic analysis to understand how biphasic properties a nd indenter velocity affect the indentation response and fitting result was also investigated. Such an analys is will aid in the design of a more efficient micro-indentation strategy. 17

PAGE 18

Third, a non-invasive MR imaging method to quantify the concentration distribution during CED was developed. Previous studies have used MR imaging to quantify the distribution volume of MR-visible tracers in tissues. Thes e studies did not quantify local concentration profiles, and used a simple linear relationship between signal intensity an d tracer concentration. In this dissertation, a basic relationship between signal intensity and tr acer concentration was quantified. An MR imaging scheme and data analysis method was designed. The method maximally utilized the intrin sic material properties (e.g., T1 and R1) to determine the concentration, and reduces the e ffect of instrumental factors (e.g., magnetic field inhomogeneity) on concentration calculation. Developing such a non-invasive quantification method is important since spatial concentration profiles can provide more information on the dose at different tissue regions. Comparison between co mputational transport model predictions and measured concentration profiles may also provide important transport parameters of drugs in nervous tissue in the future. Finally, experiments of CED into spinal co rd via the peripheral nerve infusion were conducted. A CED protocol for delivery of com pound agents into the rat spinal cord was developed. A real-time in vivo MRI methodology to monitor the agent transport during CED was also developed. Experimental para meters and distribution character istics of drug transport from the sciatic nerve to the spinal cord was inves tigated for macromolecular tracers. The developed MR imaging technique and CED protocol provided the infrastructure for further CED studies on rat models. Results of these animal experime nts also provided fundamental knowledge of transport pathways from the sciatic nerve to the spinal cord. 18

PAGE 19

CHAPTER 1 1. INTRODUCTION In this chapter, the background of convectionenhanced delivery (CED), that is, direct infusion, of drug into nervous tissue are presented. Previous studies of CED are reviewed and the motivation of this dissertation is discussed. 1.1 Background of Convection-Enhanced Delivery (CED) A growing number of potentially therap eutic compounds, including antibodies, gene vectors, growth factors, and immunotoxins, exist for treatment of cancer, spinal cord injury, Parkinsons disease, and other diseases of th e nervous system. However, biotransport issues continue to be a hurdle to effective therapy. This especially holds in the central nervous system (CNS) which is protected from blood-borne pa thogens by the blood-brain-barrier (BBB) as shown in Fig. 1-1. Only a small class of drugs, e.g., small molecules with high lipid solubility and a low molecular weight (MW) of le ss than 400-500 Da, can cross the BBB [1]. Unfortunately, many new drugs with therapeutic pot ential are proteins w ith high macromolecules and cannot bypass the BBB. To circumvent the BBB, local drug delivery me thods have been developed that rely on transport through the extracellular space (ECS) (F ig. 1-5). One of these local drug delivery methods is drug-loaded polymeric implants in which biocompatible polymers are used as a carrier for drugs [2, 3]. By implanting drug-loaded polymers into the target region of diseased tissue, drugs are released into the surrounding tiss ue (Fig. 1-2). The main release mechanism is diffusion whereby the drugs migrates into the su rrounding tissue as driven by the concentration gradient. However, the efficiency of this local delivery method is restricted by the poor diffusion of drugs, especially for high molecular we ight compounds, through the brain or tumor 19

PAGE 20

interstitium1. For example, diffusion of Immunoglobulin2 G (IgG) in tumor tissue requires three days to transport 1 mm from the point of origin [4]. The penetration distance is also determined by the balance between the drug source and si nk rates (e.g., binding and metabolism). Such balances may lead to a steady-state penetrati on distance of only a few millimeters [5]. In addition, the diffusional requirement of a concen tration gradient may lead to a very high concentration level within the vicinity of the source, which may be toxic to the surrounding tissue. To avoid limitations encountered by the pol ymeric implant method, a group of researchers from the NIH introduced a convection-enhanced delivery (CED) technique [5-8]. CED directly infuses drugs into the parenchyma3 of nervous tissue via a can nula (Fig. 1-3A). Convectional transport is introduced by applying a pressure gradient between the in jection site and the surrounding tissue. In the early work by Bobo et al. [6], a macromolecular 111In-labeled transferin molecule (MW = 80 kDa) was infused in to the corona radiata of living cat brains. A distribution diameter over centimeters was obser ved after infusing for several hours and the distribution of 111In-labeled transferin was found to be homogeneous. This technique was also found to be clinically safe: cats remained neurologically normal after infusion; edema4 resolved by one week; and only mild gliosis5 in the infusion area was found. Similar results, i.e., large distribution region and clinical sa fety of CED, were further conf irmed in a study by Lonser et al. 1 The small space between tissue cells, i.e., the extracellular space excluding blood vessels. 2 A specific protein substance that is produced by plasma cells and used by the immune system to identify and neutralize foreign objects, such as bacteria and viruses (also know as antibody). 3 The tissue that makes up the speciali zed parts of an organ, rather than the blood vessels and connective or supporting tissue. 4 The presence of abnormally larg e amounts of fluid in the inter cellular tissue spaces of the body. 5 The production of a dense fibrous network of neuroglia. Neuroglia is a glial cell which surround neurons, providing mechanical and p hysical support and electrical insulation between neurons. 20

PAGE 21

[9], where Gd-bound albumin (MW = 72 kDa) was infused into th e pontine region of the living primate brainstem. Application of the CED techni que to drug delivery has been conducted to investigate the safety and feasibility of therapy in animals (e.g., rats and monkeys)[10-14]. For example, Hamilton et al. [10] examined the potential of using CED to deliver glial-derived neurotrophic factor (GDNF) into the rat brain. GDNF (MW = ~33-45 kDa) has been found to support the survival of neurons and is proposed to be useful in the treatment of neurodegenerative disorders such as Parkinsons desease [ 15, 16]. GDNF which is heparin-bi nding was infused or coinfused with heparin into the rat stratium. A much larg er distribution volume of GDNF was obtained for CED when coinfusing with heparin since the binding of heparin on GDNF may prevent the binding of GDNF to growth factors in the extr acellular matrix which limits its transport. Limited clinical studies using CED have also been reported on humans. Pilot studies using this technique have been conducted to delive r tumor-targeting immunot oxin conjugates (e.g., TFCRM107 and IL4-Pseudomons exotoxin) [17-19], ch emotherapeutic agents (e.g., paclitaxel) [20, 21], and antiglioblastoma gene therapy (e.g., HSV1-tk) [22] to cancer patients. More recently, Patel et al. [23] investigated the safety and feasibility of CED of Cotara for the treatment of malignant glioma. Cotara (MW = 150-170 kDa) is a radio-immunotherapeutic agent, which targets a universal intracellular antigen (i.e., hi stone H1) in the necrotic tumor core and thus delivers a cytotoxic dose of 131I radiation to the su rrounding living tumor cells. An initial study involving 51 patients showed evidence of therapeutic efficacy in several patients, both in terms of extended survival and improved MRI and neurological signs. 1.2 Literature Review and Motivation CED provides a promising drug delivery met hod. As a local drug delivery method, it has advantages of bypassing the BBB, reducing the pote ntial of systemic toxicity, and providing 21

PAGE 22

more accurate targeting compared with convent ional administration such as intravascular injection. CED enhances drug transport into ne rvous tissue by introducing convectional transport in addition to diffusional transport. This is part icularly important for therapy where a large drug distribution volume on the order of centimeters is necessary. Nevertheless, challenges exist for CED to deliv er drugs only to the place where they are needed and at the level they ar e required. Drug delivered to an undesired region may cause side effects or damage healthy tissues, especially for tumor therapy where many therapeutic agents are toxic. The level of drug should be in an appropriate concentration range since a high concentration level may be toxic, and a low conc entration level may not be effective (Fig. 1-4). These challenges arise from (1) the complex micr ostructure of nervous tissue, e.g., oriented fibrous, porous structure in white matter; (2) drug-nervous tissue interactions, e.g., drug absorption and metabolism; and (3) the infu sion mechanics of CED, e.g., needle-tissue interaction. All of these factors affect the infusate flow a nd drug distribution in nervous tissue during CED. Variations of these factors between different pati ents make the prediction more difficult and the CED clinical planning more ch allenging. In spite of these challenges, drugtargeting using the CED technique could be improved with the aid of engineering tools such as computational prediction models real-time monitoring technique s, and improved design of the CED protocol. To address these issues, previous theore tical modeling and experiments have been conducted. These efforts mainly focused on three aspects. (1) Theoretical modeling of infusion using continuum mechan ics and pharmacokinetic6 methods with the pu rpose to predict drug distribution. Developing such a tool is important since it may be us eful for the clinical planning 6 The body's reaction to drugs, including their absorption, metabolism, and elimination. 22

PAGE 23

and optimization of CED protocols, such as the se lection of infusion rate and infusion site. (2) Non-invasive monitoring of drug distribution du ring CED. By developing such methods, the concentration levels and the regions to which dr ugs are delivered can be quantified in real time during the CED procedure. (3) Methods to avoid backflow along the needle tract during infusion. High infusion rates or infusion pressures may cause backflow, resulting in the drug being delivered to undesired regions of tissue (Fig. 1-3B). Instead of on specific clinical application, the focus of this di ssertation is to develop basic engineering tools to improve the CED technique. In the follo wing sections, previous CED studies that address these respects are reviewed. The motivation for th is dissertation is discussed. 1.2.1 Mathematical Modeling of CED Previous studies. To understand and predict drug transport during CED, mechanics models for direct infusion have been develope d [5, 24-28]. Nervous tissue is generally modeled as a porous media, which includes a solid matrix (solid phase) consis ting of cells and microvascular structure, and connect ed pores composed of extracell ular space and saturated with interstitial fluid (fluid phase) (Fig. 1-5). For simplicity, it is al so called a biphasic material, and its related properties are called biphasic properties which include mechanical properties of the solid matrix (Youngs modulus and Poisson ratio) and transport properties of the fluid phase (hydraulic permeability). Hydraulic permeability is defined as the water conductivity in a porous media driven by a pressure gradient. For isotropic gray matter, Morrison et al. [5] have developed a mathematical model based on a rigid porous media assumption. Their model accounted for intracellular uptake and metabolism of the infused species. Analytical so lutions allow for prediction of time-dependent drug concentration profiles duri ng and after CED from extracellular volume fraction, volumetric flow parameters, and metabolic and tissue transp ort parameters. A rather uniform distribution 23

PAGE 24

over a large volume of brain was predicted which wa s consistent with their experimental results for infusion into the rat striatum. To understand how drug transport depends on in fusion pressure, flow rate, and material properties of brain tissues, biphasic, poroelastic, and poroviscoela stic models which couple the interaction between tissue deformation and fluid fl ow have been used for direct infusion [24-27, 29]. Most of these studies considered an idealized scenario, i.e., infusion into an infinite porous media with a spherical infusion cavity at the center. Basser [24] and Barry and Aldis [25] analytically solved for the pore pressure, tissue displacement, and fluid velocity. A major difference between these two models is Basser considered a constant hydraulic permeability while Barry and Aldis considered a deformati on-dependent hydraulic permeability. These studies found a significant increase in dila tation (interstitial volume) and fl uid fraction near the site of injection. However, solute transport was not in cluded in their studies. Using a poroviscoelastic approach to account for the viscoe lastic behavior of the solid ma trix during infusion into gels, Netti et al. [29] analytically determined the evolution of fluid flow, matrix dilatation, and stress distribution about the point of in fusion, and numerically solved for the distribution of a solute for tissues with a constant hydraulic permeability. More recently, to investigate the effect of a tumor on interstitial transport during CED, Smith et al. [26] solved for the interstitial fluid pressure and ve locity when infusing into brain tumor tissue. Stokes flow was included as a sour ce term in their poroelastic model to account for the transvascular fluid exchange. In sensitivity analysis, interstitial tr ansport was found to be sensitive to the vascular conductivity Interstitial flow was found to increase in the vicinity of a tumor for infusion into the tumor and impeded by the outward flow at the tumors periphery for infusion proximal to the tumor. 24

PAGE 25

In addition, Morrison et al. [28] developed a non-spherical infusion model based on Stokes flow, Darcys law, and elastic brain tissue deforma tion to investigate the po tential fluid backflow along the needle tract which may occur at high infu sion flow rates. Using their model, a finite backflow distance was determined, which implies that backflow is minimized by the use of small diameter cannula. Backflow is also more likely to be reduced when infusing into tissue with a higher hydraulic permeability and using a more viscous infusate. Motivation Distribution of therapeutic agents after infusion is significantly related to the CED protocol, e.g., infusion site, infusate co ncentration, and infusion rate, and plays an important role in determining the efficacy of treatment. Computational and analytical models of infusion may be used to predict drug distribution and optimize CED protocols. Previous studies have analytically solved for tissue deformation, pore pressure and fluid velocity during infusion based on simple infusion scenarios, i.e., is otropic porous media with constant hydraulic permeability. However, most of these studies did not couple the effect of local tissue swelling on solute transport, and no studies have conducted a parametric analysis of this effect. Previous experimental infusion studies have identified that the macromolecular distribution is governed by tissue properties and anatomi cal boundaries. For example, highly anisotropic transport of albumin along fiber alignment direct ions has been observed in white matter during infusion into the rat brain [6, 30, 31]. Little albumin is transported across the boundary from white matter into gray matter following infusion into the rat spinal cord (Fig. 1-9) [6, 32]. Macromolecular transport into soft tissues is al so sensitive to tissue deformation. For example, hydraulic permeability may alter significantly due to deformation-induced changes in tissue microstructure such as pore size [33, 34]. In tu mor tissue studies, Zhang et al. [35] determined the hydraulic permeability to va ry over orders of magnitude depending on the perfusion 25

PAGE 26

conditions. When considering defo rmation-dependent tissue propert ies, anisotropic transport, and complex boundary conditions, an alytical solutions become di fficult to solve. Previous models based on idealized infusion scenarios provide fundamental an alysis for infusion mechanics, but have limited abil ity to predict the drug distribu tion for more realistic infusion scenarios where the above factors are involved. In such cases, a computational or finite element (FE) approach, which is also able to account for more realis tic anatomical boundaries, may be useful. In addition, in spite of trans port sensitivity to tissue deformation, brain tissue properties are not well characterized. Parametric analysis of infusion models to determine the sensitivity of macromolecular transport to varying tissue propert ies and infusion conditions will help towards a better understanding of the mechanics of infu sion and the design of infusion protocols. 1.2.2 Characterization of Soft Biomaterials and Tissues Previous studies. Tissue deformations, such as expa nsion of the infusion cavity and changes of pore size, are mainly determined by tissue Youngs modulus for a given load such as infusion pressure. For example, a larger expans ion of the infusion cavity is expected when infusing into a porous tissue with a lower Youngs modulus. In addition, interstitial fluid flow is sensitive to the tissue hydraulic permeability. A higher fluid velocity is expected for a tissue with expanded pores and higher hydraulic permeability. As a result, these two ti ssue properties play a significant role in determining the solute transport in porous tissues. Measurement of these tissue properties is no t only important to provide quantitative characteristic of the tissue but also necessary for accurate predicti on of the concentration distribution using computational models. However, using traditional characterization techniques, such as tensile or compression testing, and dynamic mechanical analysis, may be challenging because nervous tissues are very soft and heterogeneous. Preparing samples of nervous tissue for 26

PAGE 27

these traditional tests is difficult. Instead, microindentation exists as an alternative testing technique for measuring the mechanical properties of biomaterials or ti ssues [36-43]. This is because (1) micro-indentation is an indentation on the micron scale; (2) it is a non-destructive test, has small sample capacity and is able to hon e in on localized regions of interest. In practice, an indenter tip is pressed on the surface of th e sample. The applied force and the penetration depth of the tip into the sample are used to create force-displa cement (F-D) or force-time (F-t) curves (Fig. 1-6). However, efficient methods to extract soft biomaterial or tissue properties from these experimental data need to be developed. Oliver and Pharr [39] proposed a formula to calculate the material stiffness from nanoindentation response curves, i. e., loading-unloading F-D curve (Fig. 1-7). The material is expected to undergo elastic and pl astic deformation during the load ing process. In contrast, only elastic deformation happens in the unloading process since plastic deformation is not recoverable. The material stiffness was derive d using the elastic cont act mechanics during the initial period of unloading (Fig. 1-7). Their deve loped formula is good for hard materials such as metals. However, their formula dose not accoun t for fluid-solid interactions in porous biomaterials and tissues, and cannot measure the transport property hydraulic permeability. More recently, Elkin et al. [44] used atomic force microscope (AFM ) nano-indentation to measure the apparent elastic modulus of brain ti ssue at different loca tions. A formula based on the Hertz theory of contact betw een elastic solids was used to calculate the apparent elastic modulus. Heterogeneity of elastic modulus was observed at different lo cations of the brain tissue. A major difference between Hertzs method and Oliver-Pharrs method is the former used the loading curve and assumed only elastic defo rmation. Their method also does not account for fluid-solid interactions in biphasic biomaterials and tissues. 27

PAGE 28

Biphasic analysis has been used to address the complex indentation response during indentation tests (indent ation depth ~0.1-0.5 mm) of fluid-satu rated materials such as cartilage and hydrogels [42, 43, 45]. Mow and coworkers [43] provided an analytical solution for the transient displacement, U(t), of the porous cyli ndical indenter during a creep test. Biphasic properties, i.e., Youngs modulus, Poisson rati o, and hydraulic permeability, can be derived from U(t) vs. log(t) curve (Fig. 1-8). That is Poisson ratio can be uniquely determined by the slope of the experimental curve; Youngs modulus can be determined by equilibrium displacement U( ) value; and the hydraulic permeability ca n be determined by the shift, S, from master solutions where hydraulic permeability is zero. However, their method is limited to cylindical porous indenter and creep tests. In add ition, idealized boundary conditions, such as infinit sample size and no friction between the porous indenter and the samp le, were used in their analytical solution. Hale et al. [42] used a numerical method and fit finite element simulations to indentation test data (F-D response) to back out th e Youngs modulus and hydraulic permeability. Single curve matching between the experimental F-D re sponse and simulation data was used. However, this method may lead to non-unique values if mu ltiple combinations of biphasic values have approximately the same F-D response. Systemic an alysis has not yet been reported in their study to identify how varying biphasic prop erties and indenter velocities a ffected the fitted properties. In addition, Miller et al. [46] used a hyperviscoelastic model to fit indentation test data to back out material parameters of brain ti ssue. Material parameters were based on a hyperviscoelastic model. Such properties may be us ed to describe the mechanical behavior of brain tissues, but they cannot back out the bipha sic properties of hydraulic permeability, which is critical in the development of transport models. 28

PAGE 29

Motivation. Characterization of nervous tissues is challenging since they are porous, very soft, and heterogeneous. With the ability to account for these challenging conditions, microindentation provides a promising method for ne rvous tissue characteri zation. Previous data analysis methods [39, 43] provide a simple and efficient way to extract material properties from micro-indentation data. However, these methods are limited to specific micro-indentation conditions. The formula proposed by Oliver et al. [39] is limited to hard materials (one phase: solid) and does not account for fluid-solid intera ctions in porous media. The method proposed by Mow et al. [43] is limited to the cylindrical porous indenter and idealized boundary conditions. A finite element simulation of the micro-indentation experiment will provide a more flexible method to extract material properties for porous media since the FE method can account for flexible boundary conditions and geometries of th e indenter and samples. Simple FE analysis by Hale et al.[42] showed that this method can be used to back out the biphasic properties. However, systemic analysis is needed to understand how varying biphasic properties and indenter velocities affect the indentation response and fitted result. Such an analysis may also aid in the design of a more effici ent micro-indentation strategy and provide a more reliable data analysis method. 1.2.3 In Vivo Magnetic Resonance Imaging (MRI ) Monitoring of Drug Transport during CED Previous studies. Due to its noninvasive nature, MR imaging methods have been developed to monitor the distribu tion of agents during CED [9, 4749]. Contrast agents which are visible by MRI were used as tracers for drug transport. Widely used contrast agents are Gd-based compounds such as Gd-DTPA, which are either bound to or co-infused with therapeutic agents into nervous tissue. For instance, Lonser et al [9] used serial MR imaging to monitor the distribution of Gd-bound albumin during CED into the pontine region of primate brainstem. 29

PAGE 30

Infusate volume of distribution, homogeneity, and anatomical dist ribution were visualized and quantified based on enhanced signal intensity. Krau ze et al. [47] used real-time MR imaging of Gd-labeled liposomes to quantify distribution vol umes after infusion into the primate brain. A linear relationship between signa l intensity and tracer concentration was assumed in these CED studies, and only distribution volum es rather than spatial concentration profiles were quantified. Recently, Kim et al. [50] used a signal-concentr ation reference table to determine the spatial concentration profile of a drug re leased from an implant into th e vitreous of rabbit eyes. The table was obtained by imaging calibration vials with varying concentrations of Gd-DTPA in gels which mimic vitreous. Motivation. Real-time monitoring of agents transported within nervous tissues will provide information on the drug-affected region, and improve the understanding of drug pharmacokinetics and aid clinical protocol desi gn. Previous studies ha ve used a contrastenhanced MR imaging technique to monitor the drug distribution dur ing CED [9, 49, 51]. However, these previous CED studies assume d signal intensity was proportional to tissue concentration of Gd-labeled tracers. Few studies have examined the relationship between signal intensity and tracer concentrati on starting from the basic MR imaging theory, and investigated the imaging strategy by which the concentration could be derived from the signal intensity. Methods using signal-concentrati on reference table [50] may be difficult to implement in vivo where calibration samples may not capture the he terogeneity of nervous tissues. Accuracy of reference table may also decrease if the magnetic field is inhomogeneous. Further examination of the mechanism of signal intensity enhancement by contrast agents may improve and simplify the determination of spatial concentr ation distribution during CED. 30

PAGE 31

1.2.4 In Vivo CED into Rat Spinal Cord via Sc iatic Nerve Infusion and Real-time MRI Previous studies. Previous CED studies [7, 32] ha ve investigated the distribution characteristics after direct CED into the spinal cord of rats, primates and swines. Lonser et al. [7] infused various volumes ( 50 l) of labeled or unlabeled albu min directly into the spinal cord dorsal columns or lateral columns of swines and primates. Wood et al. [32] infused various volumes (~ 4 l) of 14C-labled albumin into the dorsal columns of rat spinal cords. Macromolecular infusate in the white matter were found to spread into the interstitial space parallel to fibers of passage, reflecting low transport resistan ce along the axonal fibers. A large volume of the spinal cord can be filled with infusate, and the distribution is homogeneous, allowing clinically significant distribution volume to be obtaine d in targets (Fig. 1-9). In subsequent studies of CED into peripheral nerves [52, 53], similar infusate distribution characteristics were observed in the peripheral nerves. Infusate was well contained within the peripheral nerve and filled a long segment of nerve (Fig. 1-10). In two primates, Ratiliff et al. [53] also conducted an experiment of CED into the spinal cord through remote sciatic nerve infusion. Gd-labeled albumin (~ 60-85 l) was infuse d. Infusate filling in the spinal cord gray matter was observed in the MR images acquired at the end of infusion. Motivation. Direct CED into the spinal cord tissues likely involves invasive surgery including a laminectomy to remove protective ve rtebral bones. Due to th e invasive surgery and sensitivity of the target area, direct CED into the spinal cord tissue is still not a developed clinical technique. Instead, drug delivery into the spinal cord through remote peripheral nerve CED is of potential interest si nce it may provide a remote deliv ery site to the spinal cord, requires less invasive surgery and reduces the risk of spinal cord injury during surgery. It may also allow targeting of specific neurons at nerve root entry. A similar idea has been used in previous studies which use retr ograde axonal transport for gene delivery to the spinal cord 31

PAGE 32

through sciatic nerve and muscle injections [54, 55]. However, a specific vector is required as a delivery vehicle in retrograde a xonal transport. Inefficient deliv ery due to a limited number of gene-expressing cells in the sp inal cord may be a major concern. Previous CED studies [7, 32, 52, 53] have shown that macromolecular transpor t in white matter and peripheral nerves were preferentially parallel to neural tracts and a long dist ribution distance can be achieved. By taking advantage of this low resistance transport along axonal fibers, remote infusion from peripheral nerve to a specific region of spin al cord is possible. In a limited number of primates (n=2), Ratiliff et al. [53] evaluated this possibility and infusate spread in the spinal cord was evident. Nevertheless, the characteristics of this remote CE D have yet to be studied, such as the pathways that the infusate follows and wh ere else the infusate goes. To capture the transport characteristics of this remote CED technique, real-time in vivo MR imaging is useful to monito r the transport dynamics. Previous studies [9, 51-53] have used contrast-enhanced MRI to visualize the final di stribution region and only a few have conducted a real-time imaging to monitor the agent transpor t during CED. For example, Krauze et al. [51] used consecutive MR scans to monitor the Gd-l abeled lipsomes distribution during CED into non-human primate brains. However, real-time in vivo MR imaging of the CED transport in the rat peripheral nerve and spinal co rd using is challenging due to the small dimensions of these features (e.g., ~ 4 mm diameter of the rat spinal cord), and motion-induced artifacts (e.g., rat breathing) in MR images. Partic ular MR device designs and MR imaging parameters may be evaluated such as RF coil configuration. Such a real-time in vivo MR I methodology to monitor rat spinal cord infusion may also provide a useful tool for future CED studies in small animals. 1.3 Contributions This dissertation provides fundamental engi neering tools and met hodologies for CED with the intension to improve CED techniques and its application. Specifically, research studies (1) 32

PAGE 33

developed a 3-D finite element model for CED; (2 ) considered the effect of tissue properties and infusion parameters on solute transport during CE D by using simulation anal ysis; (3) designed a methodology for biomaterial characterization us ing micro-indentation; (4) developed a methodology for real-time MR imaging and quantif ication of concentra tion distribution during CED; and (5) investigated the feasibility of in vivo CED into rat spinal cord via the sciatic nerve infusion and developed a methodology for real-t ime monitoring of agen t transport during rat spinal cord CED using in vivo MRI. Research results based on the above scope provided an infrastructure for future CED studies, which include computational models fo r prediction of CED, ch aracterization methods for nervous tissues, and MR imaging methods to monitor CED. As a whole, fundamental methods were developed which may be used to predict the drug transpor t during infusion, aid in CED protocol design, and monito r the drug transport into ne rvous tissue during CED. The specific contributions of this dissertation are as follows: Finite element modeling of CED into nervous ti ssue and parametric analysis of CED using the FE model. A biphasic FE model, which integrates tissu e swelling and macromol ecular transport and accounts for deformation-dependent hydraulic permeability, was developed and implemented using a commercial FE software package ADI NA and custom subroutines. Biphasic mixture theory was used to model the macromolecular transport in nervous tissue. A deformationdependent hydraulic permeability, which decreases exponentially with the pore size, was used to account for the effect of tissue swelling on the tran sport property. Constant pressure infusion into an infinite isotropic media with a spherical infusion cavity was modeled. This computational model was validated by comparing with analy tical solutions obtain ed by Basser [24]. Sensitivities of tissue swelling and concentration distribution to tissue parameters such as 33

PAGE 34

Youngs modulus and hydraulic permeability, and experimental parameters such as infusion pressure were also investigate d. The parametric analysis of di rect infusion was conducted using the developed FE model for infusion of albumin tracer into infinite gray matter. A range of infusion pressure, Youngs modulus, hydraulic permeability, and deformation dependence of hydraulic permeability was simulated. Parameter va lues were obtained from previous published experimental data. The effects of these parame ters on albumin concentration distribution were investigated. Previous studies used analytical methods, solved for fluid flow, and were based on idealized scenarios. These studies are good for fundamental understanding of infusion mechanics, but have limited abil ity to predict drug distribution in a realistic infusion scenario with complex tissue behavior. In contrast, this dissertation used a FE approach, which solves for both fluid flow and solute transport, and cons iders a more realistic deformation-dependent hydraulic permeability condition. The developed FE model may also provide a platform to account for anisotropic transport an d realistic anatomical geometry in future studies. In addition, parametric analysis is important since it provi des information on how the drug distribution is sensitive to changes in tissue properties and infusion parameters. A methodology to determine biphasic properties of biomaterials or tissues using microindentation. A methodology was proposed to determine the biphasic properties of porous media. The methodology used FE simulation and micro-indent ation test. A biphasic FE model for microindentation on porous media was developed. Micro-indentation of hydrogel-based contact lenses using an impermeable spherical indenter was modeled to validate the methodology. Systemic analysis of the model was also conducted to understand how varying bi phasic properties and indenter velocities affect the inde ntation response and fitted result. 34

PAGE 35

Use of micro-indentation and biphasic FE simulation provides a more flexible method to back out tissue properties since analytical techniques are restri cted to a specific indentation configuration or can only back out tissue elasti c modulus. Sensitivity analysis may improve the understanding of micro-indentati on response on biphasic materials, and aid in the design of a more efficient micro-indentation strategy a nd a more reliable data analysis method. A non-invasive MRI methodology for real-time and quantitative monitoring of agent concentration during CED. A quantitative MRI method was proposed a nd validated to quantify the spatial concentration profile and dist ribution volume during CED. A dynamic contrast-enhanced MR imaging technique was used. To validate th e methodology, CED of Gd-DTPA-labeled albumin into an agarose hydrogel was conducted. The ag arose hydrogel phantom was used as a tissue substitute. Gd-DTPA-labeled albumin (MW = 70 kD a) is a contrast agent which enhances the MR signal. It was used to mimic the tran sport of a macromolecular drug. A theoretical relationship between agent concen tration and signal intensity of MR image was derived from fundamental MRI theory. An imaging strategy an d data processing method was developed. As a validation test, MR-deriv ed concentration profiles were found comparable to both results measured directly using quantitative optical imag ing and results from a computational transport model in porous media. Developing a non-invasive quantitative MRI method is important since spatial concentration profiles can provide more info rmation, e.g., dose at different regions, than distribution volume. Comparison between computational drug transport models and concentration profiles may also provide impor tant pharmacokinetic parameters such as diffusivity and binding rate in future studies. To the best of our knowledge, MR studies that quantify concentration profiles for CED have not been previously reported. 35

PAGE 36

In vivo CED into the rat spinal cord via the sciatic nerve infusion a nd real-time MRI. An experimental protocol of CED into rat sciatic nerves was developed. A real-time in vivo MRI method to monitor the agent transpor t during CED was presented. Experimental CED parameters and distribution characteristics were examined. Dynamic contrast-enhanced MR images were acquired using an 11.1-T magnet system. Infusions of Gd-DTPA and Gd-DTPAalbumin tracers were conducted, and MR images showed infusate to track along the peripheral nerves and enter CSF and spinal nerve root para llel to the spinal cord. Depending on the nerve placement of the infusion cannula, the majority of infusate was found to divert to either the CSF or spinal nerve root; however, uptake in spinal cord tissues (gray matter or white matter) appeared to be limited. Additional factors such as the distance from the peripheral nerve infusion site to spinal cord entry, peri pheral nerve branching, total infusi on volume, fluid resistance of the spinal cord system, and infusate leakage around the cannula tip also influence the transport of infusate into spinal cord tissues. To the best of our knowledge, characteriza tion of agent transport dynamics from the peripheral nerve to the spinal cord during CED has not been reported before. This experiment provides an in vivo MRI methodology for rat spinal cord CED. Results of this experiment also provide knowledge of pathways of drug transport from peripheral nerves to the spinal cord. In addition, the developed experiment al methodology may provide an in frastructure for future CED studies on small animals. 36

PAGE 37

Brain Blood vessel Neuron Blood-Brain Barrier Endothelial cell Tight junction Figure 1-1. Schematic of the bl ood-brain-barrier (BBB). The BBB is created by the tight junction of endothelial cells lining the interior su rface of blood vessels in the brain, forming a barrier between the circulating blood and the brain parenchyma. The BBB prevents the penetration of macromolecules from the blood to the brain parenchyma. (Adapted from [56] and [57]) Figure 1-2. Drug release from a polymeric implan t. Diffusion is a main release mechanism, whereby the drug migrates from its initial position to the surrounding tissue driven by the concentration gradient [2]. 37

PAGE 38

Figure 1-3. A) Convection-enhanced delivery into a monkey brain. The infusion cannula is identified by the dark line in the brai n. The white region is the region where the infusate distributed [9]. B) Infusate back flow into the adjacent white matter (arrow) when using a high infusion rate a nd a large diameter cannula [31]. Figure 1-4. Changes of drug concentrati on in blood plasma following a single dose administration of a therapeutic agent. () safe dose, (---) unsafe dose, (-) ideal dose [3]. B A 38

PAGE 39

m Figure 1-5. Electron micrograph of a small region of the cerebral co rtex (the surface layer of gray matter of the cerebrum) of a rat. Th e black areas between cells indicate the extracellular space (ECS), which may have been reduced in size as a consequence of the histological processing [58]. Figure 1-6. Micro-indentati on test. F(t) is the force applied on the tip and D is the penetration depth of the tip into the sample. 39

PAGE 40

Figure 1-7. Nano-indentation met hod proposed by Oliver and Pharr [39]. Material stiffness can be derived from the slope, S, in the unloading curve. Figure 1-8. Creep test of biphasi c materials proposed by Mow et al [43]. Biphasic properties of the porous material can be derived from the transient displacement, U(t), of the porous cylindrical indenter. 40

PAGE 41

Figure 1-9. Distribution of 14C-albumin after CED directly into the dorsal column (white matter) of the rat spinal cord (L: coronal view; R: transverse view. Adapted from [32]). Figure 1-10. Coronal MR image of the bilateral lower extremities of a primate infused with 40 L Gd-albumin in each sciatic nerve. Arro ws indicate the areas of infusion (Adapted from [52]). 41

PAGE 42

CHAPTER 2 2. BIPHASIC FINITE ELEMENT MODEL OF SOLUTE TRANSPORT FOR DIRECT INFUSION INTO NERVOUS TISSUE This chapter presents a computational model for direct infusion in to nervous tissue by using a finite element method. Biphasic mixture theory was used to model the deformationtransport behavior of soft porous nervous tissue during direct infusion. Parametric analysis was conducted to investigate the sensit ivity of drug transport to the ti ssue properties and experiment protocol, such as infusion pressure, for an idealized geometry and constant pressure infusion. Content in this chapter is based on a paper published in Annals of Biomedical Engineering [59]. 2.1 Introduction Distribution of therapeutic agents after in fusion is significantly related to the CED protocol, e.g., infusion site, infusate concentra tion, and infusion rate, and is critical in determining the efficacy of treatme nt. Computational and analyti cal models of infusion may be used to predict drug distribution and optimize CED protocols. For isotropic gray matter, simple mathematical models have been developed that allow for prediction of time-dependent concentration profiles from extr acellular volume fraction, volu metric flow parameters, and metabolic and tissue transport para meters [5]. In addition, macr omolecular distributions may be significantly influenced by local tissue swelling and edema induced by local infusion. Biphasic models that account for coupled fluid flow and solid deform ation have been previously developed for soft tissues [60-62]. These studies assume the interstitial fluid space to be a fluid phase embedded in a linear elastic or viscoelastic solid matrix consisting of extracellular matrix, vasculature, and cells. Previous studies have used biphasic, po roelastic, and poroviscoelastic models to account for local tissue deformati on and fluid flow for infusion [24, 25, 27-29]. Analytical solutions for interstitial fluid pressu re and the velocity field have been obtained by Basser [24] and Barry and Aldis [25] assuming a s pherical infusion source in infinite media. 42

PAGE 43

These studies found a significant in crease in dilatation (interstitial volume) and fluid fraction near the site of injection. Us ing a poroviscoelastic approach to account for the viscoelastic behavior of the solid matrix during infusion into ge ls, Netti et al. [29] analytically determined the evolution of fluid flow, matr ix dilatation, and stress distribu tion about the point of infusion, and numerically solved for the distribution of a solute for tissues w ith constant hydraulic permeability. In addition to these infusion st udies, a non-spherical infusion model based on Stokes flow, Darcys law, and elastic brain tissue deformation wa s developed by Morrison et al. [28] to investigate fluid backflow along the n eedle tract which may occur at high infusion flow rates. However, most of these studies did not coup le the effect of local tissue swelling on solute transport, and no studies have conducted a para metric analysis of this effect. Previous infusion studies have identified th at the macromolecular di stribution is governed by tissue properties and anatomical boundaries [6 28, 30, 31]. For example, highly anisotropic transport of albumin along fiber alignment direct ions have been observed in white matter during infusion into the rat brain [6, 30, 31]. Little albumin is tran sported across the boundary from white matter into gray matter following infusion into the rat spinal cord [32]. Macromolecular transport into soft tissues is also sensitiv e to tissue deformation. For example, hydraulic permeability may alter significantly due to deformation-induced changes in tissue microstructure, e.g., pore size [33, 34]. In tumor tissue studies, Zhang et al.[35] determined the hydraulic permeability to vary over orders of magnitude depending on the perfusion conditions. When considering deformation-de pendent tissue properties, anisot ropic transport, and complex boundary conditions, analytical solutions become difficult to solve. In such cases, a computational or finite element (FE) approach wh ich is also able to account for more realistic anatomical boundaries can be use d. In addition, in spite of tr ansport sensitivity to tissue 43

PAGE 44

deformation, brain tissue properties are not well characterized. Pa rametric analysis to determine the sensitivity of macromolecular transport to varying tissue properties and infusion conditions will help towards a better understanding of the m echanics of infusion and the design of infusion protocols. Previous computational solute transport models that incor porate deformation have been developed for soft tissues and hydrogels for various loading conditions [63-67]. Most of these studies did not consider an infusion scenario. Ferguson et al. [65] developed an axisymmetric, poroelastic FE model of cyclic loading of the in tervertebral disk that was sequentially coupled with solute transport. Studies by Sengers et al. [66] and Mauck et al. [64] used FE and finite difference approaches, respectively, to determin e the effect of dynamic compressive loading on solute transport within gels for tissue engineering applications. In this study, we (i) built upon previous analytical models and used a FE methodology to couple local tissue swelling (including defo rmation-dependent hydraulic permeability) and interstitial solute transport within tissues and (ii) used the developed model to investigate the effect of pressure-induced swelling on macrom olecular distributions for a simple infusion scenario by parametric analysis. To validate th e FE model, numerical solutions for constant pressure infusion were compared with analytic al solutions of flow and deformation [24]. Development of such FE based models will allow for consideration of more realistic anatomical and infusion device geometries in future studies The developed FE biphasic modeling approach was used to model interstitial fluid flow and tr acer transport following low pressure infusion into gray matter. The infusion point was modele d as a pressurized, sphe rical cavity surrounded by isotropic, biphasic media with a linear elas tic solid phase. Such isotropic transport considerations work well within gray matter re gions which exhibit isot ropic distribution of 44

PAGE 45

agents [6]. Tissue porosity and hydraulic permeability were assumed to depend on the extent of deformation. Various relationships for deforma tion-dependent permeability have been proposed [33, 34, 68, 69]. We used an expone ntial form which has been measured for agarose gels and biological soft tissues such as cartilage [33, 34, 69], though such deformation-dependent hydraulic permeability has not been quantified for nervous tissue. The developed biphasic FE model was used to determine the sensitivity of macromolecular tracer infusion to pressureinduced swelling, i.e., deformation-dependent hydraulic permeability, tissue stiffness, and infusion pressure. The range of infusion scenario s for which the biphasic FE approach is most useful was determined. Ultimately with furthe r development, such a computational modeling approach may be used in CED protocol desi gn and optimization for treatment of specific structures of the brain and spinal cord. 2.2 Theory and Methods 2.2.1 Mechanics Model A brief description of the biphasic theory us ed in the model is presented. For a more expanded description, the reader is referred to [62, 70]. Nervous tissue was treated as a mixture, which includes a solid phase (label: s ) and a fluid phase (label: f ). Both solid and fluid phases were assumed to be incompressible with th e solid matrix fully saturated with fluid. The fluid phase included the infusate which was assu med to have the same fluid properties as the interstitial fluid. Low solute concentrations were assumed, and the influence of the solute on fluid flow and tissue deformation was considered negligible. Also, osmotic effects were not considered. The constitutive equations for solid and fluid phases are s sEp I ( 2Ee I ) (2-1) ff p I (2-2) 45

PAGE 46

where s and f are the Cauchy stress tensors of the solid and fluid phases; is the contact stress from deformation of the solid matrix [71]; E s and f are the solid and fluid volume fractions ( ); is the infinitesimal strain tensor of the solid matrix (1f s 1 2[]T uu where is the displacement vector); e is the dilatation ; and are the Lam elastic constants of the solid matrix; u=Tr() e p is the pore (interstit ial) fluid pressure; and I is the identity tensor. Fluid flow is described by Darcys law as s kp vv (2-3) where s sfvvvfis the volume-averaged bulk velocity; s vand f v are the velocity vectors of solid and fluid phases; and k is the hydraulic permeability. Hydraulic permeability has been found to be deformation-dependent due to localized changes in porosity for soft tissues such as cartilage and hydrogels [33, 34, 69, 72]. For small deformation, Lai et al. [34] proposed an exponential relationship 0exp() kkMe (2-4) where M is a material constant and k0 is the baseline hydraulic permeability at zero strain (no deformation). The spatially varying porosity, f due to solid deformation is related to the initial porosity, 0 f and the Jacobian, 0/ JdVdV by For small deformation and the porosity is calculated by 01(1)/ffJ1 J e 01 f fe e (2-5) The conservation of mass for tissue is given by f q v (2-6) 46

PAGE 47

where f q is the source term for the fluid phase. We assumed no fluid source term for the fluid phase. Absorption of fluid by capillaries was a ssumed to be negligible, and there are no lymphatics in nervous tissue [73]. In addition, although there exists sl ow cerebrospinal fluid (CSF) circulation within the brain, which arises out of the continuous bulk flow of CSF from the choroids plexus formation sites to the arachnoid villi absorptions sites [74], this bulk flow was considered negligible compared with induced flow due to infusion. Taking divergence on both sides of Eq. (2-3) and appl ying Eq. (2-6) results in () f e kpq t (2-7) where fp q t in the FE formulation (see below). Neglecting inertia and body force terms, the balance of momentum for th e solid-fluid mixture requires (2-8) ()()sf Ep I 0The nature of the coupled solid-f luid interaction can be further illustrated by rewriting Eq. (2-8) using and taking divergence on both sides = e u 22(2) ep (2-9) Assuming initial conditions, p(x,t ) = e (x,t )=0 at t =0, results in A p He ( 2AH ) (2-10) and Eq. (2-7) can be written as ()Ae kHe t ( 1Ak k H ) (2-11) which is similar in form to equations of heat conduction or diffusion. The FE formulation assumed =constant, even though no fluid source term was assumed. Hydraulic permeability of the tissue, was related to input hydraulic permeability, k using Eq. (2-11). k 47

PAGE 48

2.2.2 Solute Transport Model Mass conservation for the solute in tissue is given by ()cc c cMc M cq t v (2-12) where c is the solute concentration in mole per unit volume of the whole mixture; c M is the molecular weight of the solute; is the velocity of solute; and is the source term for the solute. We consider solute transport that is confined to the fluid and solute phases only (extracellular). Transport behavior is described by Ficks law cvcq ()c effcc vvD (2-13) where is the effective diffusion tensor of the solute in the porous media. effD ,1fc vv is the density-averaged velocity of the fluid and solute mixture, is the apparent density of constituent and, fc We assumed the solute concentration was too low to influence the density of the mixture a nd the velocity of the fluid. Thus, can be approximated by v f v. In addition, the solid matrix-solute interact ion will hinder the convection transport, which is corrected by including a retardation coefficient, [75]. Thus, f vv. Substituting Eq. (213) into Eq. (2-12) results in the relation (c f eff cc cc tM vD) q (2-14) Volumetric extravasation, absorption, and degradation of the tracer solute were assumed negligible ( qc=0) during direct infusion. In addition, and are also affected by the porosity [76] which changes with tissue deformati on. In this study, unl ess otherwise mentioned, effD1 and was assumed to be independent of tissue deformation. Changes in diffusional effD 48

PAGE 49

transport may be small compared to the overall dist ribution if interstitial transport is dominated by convection. 2.2.3 Numerical Implementation The computational model was developed usi ng the FE software package ADINA (version 8.2.2, ADINA R&D Inc., Watertown, MA) along with user-defined subroutines and a custom C++ program. Three modules were used to solve for solid deformation, fluid flow, and solute transport equations (denoted by ADINA-S, AD INA-T, and ADINA-F, respectively). The coupled solid deformation and fluid flow equations [Eqs. (2-7) a nd (2-8) expresse d in terms of u and p] were solved using ADINA-S and ADINAT modules simultaneously. The NewtonRaphson iteration method was used to solve FE-dis cretized equations, and an Euler-backward integration scheme was used for the transient solutions. u and p solutions were obtained at each time point. Note that solu tions were obtained assuming = constant and re lated to the case = 0 using the hydraulic permeability relation defined by Eq. (2-11). The quasi-static biphasic solution was incorporated in the solute transport problem [Eq. (214)] using the ADINA-F module. Th e biphasic-solute transport solution interface was achieved using a custom C++ program which: (1) calcul ated the nodal dilatati on, porosity, and fluid velocity at each time step; and (2) created and compiled th e model for solute transport computation using the fluid veloc ity field at that time step. Nodal deformation velocity was calculated by dividing the displacement differenc e between two neighboring time steps with the time step, vf was calculated using the relations ()tts ttt vuu / f f ss vvv and Eq. (25). Since fluid velocity was output at the integration point, nodal fluid velocity was then approximated by averaging the fl uid velocities at the surround ing integration points. 49

PAGE 50

2.2.4 Tissue Infusion Model We considered solute infusion into gray matter which was idealized as a homogeneous, isotropic, biphasic media with no fluid sour ce or sink regions and negligible endogenous interstitial fluid flow. A symmetric, spherical geometry was modeled and the infusion site was a spherical cavity with radius, a. The initial radius of the infusion cavity corresponded to the external diameter of a 28 gauge cannula, ao=0.18 mm. The outer radius was 20 ao=3.6 mm. Previous infusion analyses show that pore pres sure, displacement and fluid velocity change negligibly at radial positions more than 20 ao away [24, 25, 29]. A FE mesh was created using 4node tetrahedral elements (~42,000 elements) with finer meshing in th e region close to the infusion site (Fig. 2-1). Zero initial pore pressure, strain a nd fluid flow were assumed. Pressure within the infusion cavity was assu med uniform, and a ramp-hold pressure was applied at the spherical boundary. Solid, fluid, and solute transport boundary conditions were applied separately. Previous studies by Kenyon et al.[71] and Hou et al.[77] used a zero contact stress ( ) applied to the solid phase at the interf ace between fluid and porous media. Since the stress calculated in the solid module of ADINA -S was the total stress for the bulk material (E 0 s f), the infusion pressure was applied at the porous media-fluid interface at r = a, i.e. infusion cavity surface, which m oves during infusion. Also, a c onstant solute concentration boundary condition was applied on this surface. Ze ro pore pressure and free displacement were applied along the outer tissue boundary. Symme tric boundary conditions were applied to symmetry faces (zero displacement, flow flux, and mass flux normal to the surface). Infusion parameters were varied, and sensitivity to changes in infusion pressure,0 p over the range 1 to 10 kPa (7.5 to 75 mmHg), was determined. Infusi on pressure is likely on the lower end of this range based on experiments of Prabhu et al. [78] who observed a range of infusion pressures in 50

PAGE 51

the rat caudate of 1.6 to 4.2 kPa (12 to 32 mmHg) for infusion rates varying between 0.17 and 1.5 L/min (25-gauge needle). The lower pressure value is also in the vicinity of the consolidated tissue pressure (~2.4 kPa) m easured after an hour of infusion at 0.5 L/min into the white matter of the corona radiata of cats [6 ]. The time to reach constant pressure, to, was considered of short duration. The influence of material parameters on pr essure-induced tissue swelling and solute transport was considered. In addition, biphasic and solute transport solutions were compared with rigid model solutions. Table 21 lists the range of parameters used in this study. The value of Youngs modulus of the solid matrix was set to range from 1 to 10 kPa [79, 80]. This range corresponds well with modulus values estimated fo r small strains tested under low strain rate conditions by Miller and Chinzei, E ~1 kPa [81]. The range of Poi sson ratio has been previously estimated by Mostachfi et al. [82] to range betwee n 0.3 to 0.4, based on literature values and the compliant behavior of brain tissues. Very few experimental studies have attemp ted to measure the hydraulic permeability of nervous tissue. The baseline hydraulic permeab ility for the gray matter was chosen between 1.0e-13 to 1.0e-12 m4N-1s-1. This range was established from the spread of dye through the brain following cold-induced edema by Reulen et al. [ 83] and the estimated ranges of previous poroelastic brain models [24, 28, 84]. A deformation-dependent hydraulic permeability was also considered and we used the exponential relation by Lai et al. [34]. The value for the material constant M was varied between 0 and 5 based on a prev ious range establishe d for cartilage and hydrogels [34]. Porosity was varied between 0.2 to 0.3. The lower range of porosity corresponds to measures by radiotracer methods and iontophoretic measurements of tetramethylammonium (TMA+) in non-infused tissues [85]. The lower porosity values also match the 51

PAGE 52

volume ratio, Vinfusion/Vdistribution, of CED striatum distribution studies of 14C-albumin by Chen et al. [31]. The upper porosity range is characteristic of values reported elsewhere for edematous states, which occur after prolonged infusion or local damage to tissu e [83, 86]. Diffusivity of the solute in gray matter was set to correspond to the macromolecular tracer albumin (MW = ~66 kDa). The apparent diffusion coefficient of fluo rescently-labeled bovine serum albumin has been measured by Tao and Nicholson [87] in rat cortic al slices using an integrative optical imaging system, Deff =1.6e-11 m2/s. The FE biphasic solution was validated by comp aring with previous analytical solutions by Basser [24] for infusion into an infinite biphasi c media with constant hydraulic permeability. Solutions for pore pressure and fluid velocity following a step change in pressure infusion ( ) were compared (Rather than instantaneously applied pressure, simulations ramp infusion pressure rapidly with a ramp time 0.02 s). Al so, an analytical solution of displacement was solved (see Appendix A) and comp ared to the FE solution. 00 t2.3 Results 2.3.1 Validation of the Finite Element Model FE transient solutions were compared with an alytical solutions for a simplified infusion case: constant pressure infusion ( p= po) and constant hydraulic permeability ( k = k0). Numerical solutions for radial fluid velocity, pore pressure displacement, and dilatation match closely with analytical solutions with an error less than 2% in the region of interest for simulations up to 3 min (Fig. 2-2). Following the initiation of infusi on, regions of elevated pore pressure increased around the point of infusion with time as fl uid entered the tissue Corresponding infusion velocity was not constant with time and decrease d as the pressure gradient became less steep. The solid tissue was predicted to deform in the positive radial direction with largest displacement 52

PAGE 53

at the fluid/porous media boundary. Displacements increased with time as the pressure-driven flow penetrated the tissue. Dilatation which is a measure of pressure-induced tissue swelling also increased in the radial direction with time as the pressure-driven fluid flow penetrated the tissue. Longer infusions are expected to matc h rigid media solutions as the solutions reach steady-state (increased pore pressure and lower fluid velocity profiles). 2.3.2 Biphasic Sensit ivity Analysis The developed model was used to determine the sensitivity of nervous tissue transport to changes in material parameters for the case of pressure-controlled infusion. The more general case of a deformation-dependent hydraulic permeab ility was considered. Biphasic FE solutions for interstitial pore pressure, radial fluid ve locity, radial displacement, and dilatation are presented for a range of tissue stiffness, hydraulic permeability, and nonlinear parameter, M in Figs. 2-3 to 2-5 at a single time point, t =30 s. Decreasing tissue stiffness was predicted to incr ease the ease at which fluid entered tissue. Overall, the soft tissue analysis predicted flows that were considerably higher and pressures that were considerably lower than for rigid porous medi a (which is the special case of infinitely large E ). Decreasing E over an order of magnitude (10 to 1 kP a) resulted in increased radial fluid velocity (Fig. 2-3A). With decreasing E, pore pressure decreased a nd pressure gradients increased which enhanced the fluid flow (Fig. 23B). Also, radial displacement increased with decreasing E resulting in a larger ex pansion of the infusion cavity (Fig. 2-3C). Radial compression was offset by a predicted tension in th e circumferential direction. The net effect was a local increase in tissue volume or dilatation. Dilatation was considered a measure of the pressure-induced changes in tissue volume, i.e., swelling. Decreasing E increased dilatation and porosity [Fig. 2-3D; porosity is related to di latation by Eq. (2-5)]. Significant changes in dilatation and radial flui d velocity with varying E was confined to the regions near the infusion 53

PAGE 54

cavity, r/ao < 5. These behaviors are consistent with increased tissue ex pansion and hydraulic conductivity within the soft tissue. It should be noted that the infusion cavity was determined to expand considerably for low E Maximum predicted radial displacement at t = 30s was 120 m for E =1 kPa. Increasing hydraulic permeability also increased the ease at which fluid entered tissue. Increasing the baseline value, k0, over an order of magnitude (1.0e-13 to 1.0e-12 m4N-1s-1) resulted in increases in predicte d radial fluid velocity, pore pressure, radial displacement, and dilatation in the vicinity of the infusion site, r/ao <5 (Fig. 2-4). The effect of nonlinear, deformation-dependent hydraulic permeability was determined by varying the parameter M in Eq. (2-4) (Fig. 2-5). For E =10 kPa, varying M (0 to 5) had little eff ect on the biphasic solutions for flow and deformation compared with the constant permeability case ( M =0). More significant increases in pore pressure and fl uid velocity profiles were fou nd with smaller Youngs modulus. This was considered reasonable given the larger dilatation that was predicted as E decreased (Fig. 2-3D). Sensitivity to changes in infusion pressure was also determined. Increasing infusion pressure enhanced fluid flow by establishing a la rger pressure gradient (Fig. 2-6). Radial displacement also increased, resulting in greater expansion of the infusion cavity. In addition, the predicted relation between infu sion pressure and infusion rate was nonlinear. Small increases in pressure resulted in larger increases in infusi on rate (Fig. 2-7). Infusion rate decreased with time as pressure gradients became less steep as flow penetrated tissue. The difference between transient biphasic and rigid media solutions depends on the mechanical parameters: E v, and k The time scale for fluid redistribution can be characterized by 24/( ) A R kH where R is a characteristic radius, e.g ., range of influence [29]. If is 54

PAGE 55

small with respect to the total infusion tim e, transport simulations may be simplified by assuming the rigid media solution for interstitial fl uid velocity in the so lute transport equation [Eq. (2-14)]. Assuming E =10 kPa, v =0.35, and k =1.0e-13 m4N-1s-1 and considering a range of influence of r/ ao < 5, is approximately 106 min. 2.3.3 Solute Transport Sensitivity Analysis The effect of varying mechanical parameters was determined on albumin transport at early infusion times (Fig. 2-8). For the baseline parameters (E =10 kPa, v =0.35, k0=1.0e-13 m4N-1s-1, M =2, and p0= 1 kPa, t0 = 1 s), infusions of 30 s and 3 min correspond to infusion volumes of ~0.02 and ~0.07 l, respectively. It should be noted that infusion volume varies with changes in E k0, and p0. Macromolecular solute distribution was convection-driv en and distribution profiles were dropped off relatively steeply at the advanci ng front. Distribution was sensitive to changes in k0 and E Decreasing E by one order of magnitude resulted in increased distribution volume by increasing pressure-induced swelling effect s, i.e., hydraulic permeability and pressure gradient. The change of di stribution radius, which is defined as the radius at c =0.5 c0, is nonlinear with changes of E For example, a drop of E from 10 to 5 kPa (50%) and from 10 to 1 kPa (90%) resulted in a ~5% and ~27% in crease in distribution radius at t =3 min, respectively. Solute transport was slightly more sensitiv e to the order of magnitude changes in k0, with increased distribution volume predicted as k0 increased since fluid and solute entered the tissue more easily. The effect of deformati on-dependent hydraulic permeability on albumin distribution appears to be minimal for M up to 5 when comparing with constant permeability simulations. Biphasic solute transport was also compared w ith solute transport in a rigid media (Fig. 29A). Biphasic solute models predicted larger distribution volumes. This difference in 55

PAGE 56

distribution volume decreased with longer in fusion times. For the baseline simulation parameters, biphasic transport predicted an increased distributi on volume of ~58% (t =30 s) and ~26% ( t =3 min) greater than that predicted using a rigid media model. In both cases, solute transport is convection-dominated in the vicinity of infusion site (r/a0 < 5). The calculated Peclet number, (assuming L = 1 mm), was as high as ~ 60 for the biphasic model and ~35 for the rigid media model (Fig. 2-9B). In additi on, the effect of retardation coefficient on macromolecular transport was investigated. Decreasi ng the retardation coefficient from 1.0 to 0.5 reduced the macromolecular distribution radius by ~60% (Fig. 2-9C). Si nce the solute no longer moves at the same velocity as the extracellula r fluid, the retardation coefficient plays an important role in reducing the convection transport. Pe/refLvD f2.4 Discussion Many new and promising therapeutic agents de veloped for the treatment of neurological disorders are macromolecules, e.g., protein ligands and viral vectors. Transport of these large macromolecules, when infused into tissue by CED, is sensitive to bulk/interstitial fluid flow and, potentially, to local tissue swelling. To inve stigate the effect of local tissue swelling on macromolecular transport, a bipha sic FE model of pressure-contr olled infusion was developed. Interstitial fluid flow solutions were validated using the simplified case of constant pressure infusion and constant hydraulic permeability. Comparison with analytical solutions show matching results for pore pressure, fluid velocity displacement, and dilatation solutions [24]. The validated model showed that the net effect of infusion was a local increase in tissue dilatation, i.e., localized tissue swelling. Also biphasic analysis predicted flows that were considerably higher, pres sures that were considerably lower, and solute distribution volume that 56

PAGE 57

were larger than for rigid porous media. Cons idering an elastic solid phase, the predicted transient response was due to fluid redistribution within the tissue. In parametric analysis, enhan ced interstitial flow was predicted with increasing hydraulic permeability or decreasing tissue stiffness. Sensitivity of model results can be related to the diffusivity parameter, kHA, in Eq. (2-11). HA is proportional to E Therefore, when keeping kHA a constant, increasing k is equivalent to decreasing E when considering changes in the pore pressure and its gradient solutions. However, when increasing k the net change in fluid flow is enhanced according to Darcys law. Thus, fluid flow and solute transport is more sensitive to changes of k than E For the baseline properties simulated, biphasic solutions were not greatly affected by the deformation dependence of permeability over the range simulated ( M < 5). This is due to the small tissue strains predicted in ou r model and the fact that significant changes in porosity are confined to within th e vicinity of infusion site (r/a0 < 5). The current model is constrained by a small strain assumption. Previo us study by Spilker et al. [88] suggests that strain below 25% is reasonably mode led using linear biphasic theory for cartilage. Larger strains predicted using lower modulus values or higher infusion pressures show corresponding increases in predicted deformation and fluid velocity. Fu ture models which incorporate a nonlinear largestrain formulation will allow for further investigat ion into the effects of softer tissue properties and higher infusion pressures. Sensitivity of macromolecular solute transport to pressure-induced swelling effects was determined for a range of nervous tissue properties. Transient beha vior was investigated at early time (on the order of minutes) after initiation of infusion. Solute transport was predicted to be enhanced with increased convective transport. Thus, factors which increase interstitial fluid velocity, i.e., increasing k0 or decreasing E, also increased the solute distribution. Swelling57

PAGE 58

induced changes in interstitial fluid velocity, dilata tion, and solute transport were localized in the immediate vicinity of the cannula ( r/ao <5). Distribution results de viated most from rigid porous media analysis at short time. Overall, the deve loped biphasic FE model was most useful in the analysis of low pressure (~1 kPa), short infusi on (e.g. < 1 hour), and small infusion volumes ( < 1 L) analysis. The model can also be used for larger infusion volumes by simulating higher infusion pressures (e.g., ~7.4 L after 1 hour was predicted when using p0=30 mmHg). Thus, the biphasic models would be us eful in predicting microinjec tion of compounds at multiple infusion sites in the brain [89]. Biphasic analysis may also be necessary for longer or larger infusions if precise control of the final distribu tion volume is required such as in the delivery of potentially toxic compounds, or for small-vol ume tissue infusion where boundary contours may play an important role [28, 90]. In addition, the biphasic model can account for the increased injection volume that is infused due to enha nced transport during th e transient phase. This implies that extending the transient phase (e.g., by cyclic loading) may further enhance distribution of CED. However, if the transient phase is short (such as, for large Youngs modulus and permeability) and the infusion time is long, our analysis shows that total distribution can be approximately predicted by a rigid media model. Sensitivity analysis may provide an estimate of upper and lower distribution volumes for in vivo infusion with regard to th e uncertainties of properties ( E v k ). Direct in vivo characterization is challenging, and v and k have not been directly reported for nervous tissue. Also, for nervous tissue, further examina tion of the deformation-dependent hydraulic permeability may be needed depending on the tissue microstructure. Given these material uncertainties and the sensitivity of infusion to th ese properties, infusion rate should be measured 58

PAGE 59

with care during pressure-controll ed infusion. Infusion controls should be implemented to limit final infusion volumes. For the small infusion volumes that we simulated, direct comparison with experimental infusions volumes cited in previous literature is difficult. Additional tissue distribution studies are required for direct comparison. However, previous studies by Ch en et al. [31] and Neeves et al.[30] show approximately spheri cal distribution pattern s similar to those predicted in the gray matter at low infusion rates (0.1-0.5 L/min) for larger total infusion volumes (4-5 L). In addition, the nonlinear relationshi p that was determined between infusion pressure and infusion rate is consistent with experime ntal measures in other soft tissu es, i.e., tumors [91] and hydrogel [27]. In future studies, the developed FE a pproach may be used to create three-dimensional transport models which take into account more complex considerations, e.g., adjacent CSF or white matter [90], anisotropic transport in white matter [92], and cannula tip shape and placement. Local vasculature and endogenous flows can also have a role in local infusion. In the current model, we assumed the extent of vasc ular leakage due to local trauma is negligible, and we also neglected absorption and leakage of fluid by capillaries. Future models will account for microvasculature and endogenous flows by including a volumetric source/sink term in tissue governed by Starlings law. With further de velopment and quantifica tion of solute-tissue interactions, tissue binding and metabolism may also be considered. In conclusion, the developed biphasic FE model was used to describe transient behavior of local tissue swelling and solute transport during pr essure-controlled CED infusion. Simu lation results show that pressureinduced tissue swelling enhanced macromolecula r transport for short times or small volume infusions. 59

PAGE 60

2.5 Conclusions In this chapter, a biphasic FE model was developed using biphasic mixture theory. The developed biphasic FE model describes the transient behavior of br ain tissue swelling and solute transport during pressure-controlled CED infusion A parametric analysis was also conducted with varying tissue properties, i.e., Youngs modulus, hydraulic permeability, a material parameter for non-linear hydraulic permeability, a nd infusion pressure. It is concluded that Significant tissue swelling is confined to the near regions of infusion cavity (r/a0 < 5). Pressure-induced tissue swelling influences macromolecular transport. Macromolecules penetrate further in tissues that are softer or more permeable. The biphasic approach is necessary to predic t macromolecular distribu tions for short times or small volume infusions (e.g. < 10 min or <1 L). 60

PAGE 61

Table 2-1. Material parameter ranges of gray matter brain tissue used in simulation studies Parameter Range References Youngs modulus, E 1-10 kPa [79-81] Poisson ratio, v 0.3-0.4 [82] Baseline hydraulic permeability, k0 10-13-10-12 m4N-1s-1[24, 28, 83, 84] Nonlinear parameter, M 0-5 [34] Porosity, f 0.2-0.3 [31, 83, 85, 86] Diffusivity (albumin), Deff 1.610-11 m2/s [87] 61

PAGE 62

Figure 2-1. Finite element (FE) mesh and boundary conditions used to model pressure-controlled infusion into tissue. The infusion cav ity boundary conditions are applied at r = a ( ao=0.18 mm). The other radius of the tissue boundary is sufficiently distant that pore pressure is assumed negligible ( r = 20 ao ). 62

PAGE 63

A B C D Figure 2-2. Validation analysis co mparing transient FE and analyti cal solutions for infusion into tissue. A) Volume-averaged radial fluid velocity, vr=fvf r; B) pore pressure, p; C) radial displacement, ur; and D) dilatation, e, with distance from the infusion cavity boundary. Model simulation parameters: E = 10 kPa, v = 0.35, k0= 1.0e-13 m4N-1s-1, and p0= 1 kPa with instantaneous loading. 63

PAGE 64

A B C D Figure 2-3. Effect of Youngs m odulus on pressure-controlled infu sion into tissue. A) Radial fluid velocity, vr=fvf r; B) pore pressure, p; C) radial displacement, ur; and D) porosity, f, at t =30 sec. Model simulation parameters: k0=1.0e-13 m4N-1s-1, M =2, v =0.35, p0= 1 kPa, and t0=1 s. Rigid media solutions for radial fluid velocity and pore pressure are designate d by solid lines (A and B). 00.2f 64

PAGE 65

A B C D Figure 2-4. Effect of hydrauli c permeability on pressure-controlled infusion into tissue. A) Radial fluid velocity, vr=fvf r; B) pore pressure, p ; C) radial displacement, ur; and D) porosity, f, at t =30 sec. Model simulation parameters: E =10 kPa, v =0.35, M =2, p0= 1 kPa, and t0=1 s. 00.2f 65

PAGE 66

A B C D Figure 2-5. Effect of nonlinear hydraulic permeability on pressure-c ontrolled infusion into tissue. Hydraulic permeability was dependent on ti ssue swelling according to Eq. (2-4) where M is a material constant. M =0 corresponds to the cas e of constant hydraulic permeability. A) Radial fluid velocity, vr=fvf r; B) pore pressure, p; C) radial displacement, ur; and D) porosity, f, at t =30 sec. Model simulation parameters: E =10 kPa, v =0.35, k0=1.0e-13 m4N-1s-1, p0=1 kPa, and t0=1 s. 00.2f 66

PAGE 67

A B C Figure 2-6. Effect of infusion pressure on infusion into tissue. A) Radial fluid velocity, vr=fvf r; B) normalized pore pressure, p/po; and C) radial displacement, ur, at t =30 sec. Model simulation parameters: E =10 kPa, v =0.35, k0=1.0e-13 m4N-1s-1, M =2, and t0=1 s. 00.2f Figure 2-7. Effect of infusi on pressure on infusion rate. Model simulation parameters: E =10 kPa, v =0.35, k0=1.0e-13 m4N-1s-1, M =2, and t0=1 s. 00.2f 67

PAGE 68

t = 30 s t =3 min A B C Figure 2-8. Effect of varying tissue stiffness and hydraulic permeability on albumin distribution during pressure-controlled in fusion. A) Youngs modulus, E B) baseline hydraulic permeability, ko, and C) deformation dependence, M, were compared at two different time points (column 1) t =30 s and (column 2) t =3 min. Baseline simulation parameters: E =10 kPa, v =0.35, k0=1.0e-13 m4N-1s-1, M =2, p0=1 kPa, and t0=1 s. 00.2f 68

PAGE 69

A B C Figure 2-9. A) Albu min distribution and B) Peclet number, Pe /refLvDf (length scale, L =1 mm), following pressure-controlled infusion into rigid and soft biphasic media at t = 30 s and 3 min (Concentration was normalized by dividing by C0). C) Effect of retardation coefficient, ,on albumin transport at t = 3min. Baseline model simulation parameters: E =10 kPa, v =0.35, k0=1.0e-13 m4N-1s-1, M =2, p0= 1 kPa, and t0=1 s. 00.2f 69

PAGE 70

CHAPTER 3 3. CHARACTERIZATION OF BIOMATERIA L/TISSUE USING MICRO-INDENTATION In Chapter 2, a computational model was presented to predict the concentration distribution during direct infusion. The model used a biphasic model to describe the tissue swelling and flow transport. Important tissue parameters include Youngs modulus and hydraulic permeability, which determine the tissu e deformation and interstitial flow. The accuracy of these two parameters for a tissue will determine the accuracy of prediction by computational models. In this chapter, a met hod is proposed to measure these parameters by micro-indentation tests combined with com putational simulations. By comparing microindentation data with simulati on results, tissue parameters can be backed out. As a methodology investigation, hydrogel-based contac t lenses were used as test ma terials. A constant velocity micro-indentation with an impermeable spherical i ndenter was used to indent curved thin contact lenses. By testing on thin soft contact lenses, th e capacity of the proposed method for small-sized samples was evaluated. Previous published data for the contact lens was also compared to validate the proposed method. Content of this chapter is based on a published paper in Journal of Biomechanical Engineering Transactions of the ASME [93]. This study was collaborated with Alison C. Dunn (Mechanical and Aerospace Engineering, UF) an d Dr. W.G. Sawyer (Mechanical and Aerospace Engineering, UF). Alison collected and helped analyze the experimental micro-indentation data. Dr. Sawyer provided techni cal advice and use of the NanoTribometer system. 3.1 Introduction Soft contact lenses are composed of polymeric hydrogels. The porous structure of these hydrogels allows for transport of oxygen, water, and other nutrients thr ough the lens to the cornea of the eye. Numerous contact lens studies have focused on oxygen permeation through 70

PAGE 71

the lens [94, 95]. However, the flow of water through the lens, governed by the hydraulic permeability property, may also contribute to important lens functions including on-eye lens movement, comfort, and wettability [94, 96-98]. In addition to fluid considerations, there is a direct correlation between the ocul ar comfort of contact lenses a nd the bulk stiffness of contact lenses. For example, stiffer contact lenses may not fit the cornea well and exhibit edge fluting, causing a foreign-body sensation. Also, highly stiff contact lens es increase the occurrence of complications, e.g., SEALs (superior epit helial arcuate lesions) [96, 97]. Hydraulic permeability and Youngs modulus of the hydrogel matrix can be determined by micro-indentation of the lens under certain loading conditions. In such cases, matrix deformation and fluid transport through the hydrogel can be related to bulk viscoelastic be havior through biphasic theory [62]. Therefore biphasic characterization may be us eful in the development and te sting of comfortable lenses. Stiffness properties for various hydrogels ha ve been reported by a number of researchers using traditional techniques, including tensile testing, compression testing, and dynamic mechanical analysis [99-104]. For the hydrogel used in this study, Etafilcon A, Enns [103] used tensile tests to measure the Youngs modulus, obt aining a bulk value of approximately 255 kPa. Hydraulic permeability has been measured di rectly by permeation experiments. Yasuda et al. [105] used low-pressure ultrafiltration cells to measure hydraulic permeability of hydrogels with different equilibrium water content (EWC), which is the weight percentage of water in the equilibrium hydrogel. Measured values were 2.8910-17 m4/N-s for 21% EWC pHEMA-EG (poly-hydroxyethyl methacrylat e-ethylene glycol) and 1.2510-15 m4/N-s for 64% EWC pGMA (poly-glycerol methacrylate). Te sting pHEMA membranes, Refojo [106] measured the hydraulic permeability of 38.7% and 53.8% EWC to be 8.410-18 and 1.0510-17 m4/N-s, respectively using a pressure-driven permeameter. More recently, Monticelli et al. [82] used a similar system 71

PAGE 72

to test 38% EWC HEMA-based hydrogel membranes (polymacon) and measured hydraulic permeability was 4.010-18 m4/N-s. Besides such studies that measure stiffness and permeability properties separately, Chiarelli et al. [107] used a poroelastic anal ysis of the stress relaxation response of thin strips to estimate poroelastic properties of the hydr ogel polyvinyl alcoholpolyacrylic acid. For 19 mm thick samples, Youngs modulus was estimated to be 750 kPa and hydraulic permeability was estimated to be 1.210-17 m4/N-s. Also, biphasic analysis by LeRoux and Setton [108] have extracted properties from tensile tests on canine meniscus. In their study, custom biphasic FEM codes were used to predict the stress-rela xation behavior, and permeabilities were obtained by fitting computa tional predictions with experimental data. In addition to traditional testing methods, which generally involve complex sample preparation, indentation exists as an alternative, non-destructiv e testing technique for measuring the mechanical properties of materials [36-43]. Micro-indentation refers to indentation on the micron scale, and nano-indentation refers sub-micron sc ale testing. In practice, an indenter tip is pushed into the surface of the sample. The applied force and the penetration depth of the tip into the sample are used to create a force-displacement (F-D ) curve. Indentation is increasingly being used in the mechanical assessment of soft hydrated materials, e.g., biological tissues, because of its non-destructive nature, small sample capacity, a nd ability to hone in on localized regions of interest [36, 37, 42]. For its application to soft hydrated mate rials, various data analysis techniques have been used to determine mechan ical properties. Using the analysis technique developed by Oliver and Pharr [39] fo r elastic/plastic materials, Ebenstein et al. [36] determined reduced modulus values of porcine aorta by nano-indentation (indentation depth 1-2 m). Also, biphasic analysis has been used to address the complex indentation resp onse during indentation tests (indentation depth ~0.1-0.5 mm) on car tilage and hydrogels [41-43, 45, 88]. Mow and 72

PAGE 73

coworkers [43, 45] solved for the displacement of a cylindrical porous inde nter during creep then used the analytical solution to extract Youngs modulus, Poiss on ratio and permeability from indentation tests (~0.2 mm indentation depth) of cartilage. FEM models are able to account for complex boundary conditions and allow for the consideration of spherical and impermeable indenters, for which analytic solutions are not mathematically tractable. A linear biphasic FEM model has been developed by Spilker et al. [88] to predict the indentation stress-relaxation response of articular cartilage taking into account friction at the indenter-tissue interface and tissue thickness. Using the same linear biphasic model, Hale et al. [42] have predicted the indentation F-D response for canin e articular cartilage. They m odeled an impermeable spherical tip moving with a constant indentation sp eed (~0.3 mm indentation depth and ~25% compression), and these predic ted F-D profiles were used to determine Youngs modulus, Poisson ratio, and permeability. In this study, a linear biphasic model is used to predict the rate-dependent behavior of a curved contact lens duri ng micro-indentation (~20 m indentation depth). A FEM model was developed that accounts for the thin geometry, the spherical indenter shape, and the time-varying contact interface between the impermeable indenter and hydrogel surface. Specifically, the FEM model was used to predict F-D behavior during co nstant velocity indentation. The effect of varying experimental parameters, including inde nter velocity, hydrogel permeability, and matrix stiffness, on predicted F-D curves were parameterized. The results of this sensitivity analysis were used as a guide for experime ntal design, and model results were optimally fit with test data obtained at multiple indentation rates. The results of this study provide data for hydraulic permeability and Youngs modulus of the polymer ma trix of the contact le ns material, Etafilcon A. 73

PAGE 74

3.2 Materials and Methods 3.2.1 Material and Experimental Testing The contact lenses (ACUVUETM, Vistakon, Jacksonville, FL) used in this study were composed of Etafilcon-A (copolymer of 2-hydroxye thyl methacrylate and methacrylic acid), which has an equilibrium water content (EWC) of 58% (by weight). Lenses were approximately 100 m thick at the apex and had a 7.68 mm base radius of curvature along the apex. Samples were fixed onto a rigid, impermeable, and c onformal polymer foundation (which is rigid and impermeable in comparison to the hydrogel). A NanoTribometerTM (CSM Instruments, Peseux, Switzerland) was used to perform in dentation tests. The test appa ratus consisted of a cantilever arm whose base was coarsely controlled by a stepper motor and finely controlled by a piezoelectric cell (maximum vertical movement ~ 100 m, and vertical resolution, ~20 nm), Fig. 3-1. The stainless steel spherical in denter (radius = 1 mm) was bonded to the end of the cantilever, which had a measured cantilever stiffness of 0.7194 mN/m. The cantilever displacements were measured at a fixed loca tion along the cantilever using built in optical sensors. The entire apparatus was located on a vi bration isolated granite table. For analysis, force vs. time (F-t), and force vs. penetration depth (F-D) into the hydrogel were collected and plotted. The experiments were conducted by moving the ba se of the cantilever using a controlled velocity through the initial gap and into the le ns until a penetration de pth of approximately 20 m was reached (strain ~ 20%). Two indenter velo cities exhibiting distinct F-D behavior were used: a high velocity ~ 20 m/s (19.7-21.8 m/s, n = 5) and a low velocity ~1.2 m/s, (1.1-1.3 m/s, n = 4). A new lens from commercial packag ing was used for each test. For experiments that were performed in air the assembly and inde ntation process occurred in less than 2 minutes 74

PAGE 75

from the time the lens was removed from the packing; this was done to minimize lens dehydration. For these tests, thr ee distinct stages were identified from the F-t curve: (1) indenter movement through air, (2) indenter movement th rough the fluid film on the surface of the contact lens (determined to be several micrometers thic k), and (3) indenter cont act with the hydrogel. The indenter contact point was determined by first fitting a line to the for ce response for stage 2, then determining the point from which there was a departure from this line (due to contact with a stiffer material, i.e. the hydrogel), Fig. 3-2. For low velocity tests, the contact point was distinct, marked by a discrete change in slope. However, at high indenter velocities determination of the contact point was difficult because of limited samp ling of data through stage 2 and difficulty in separating out meniscus effects (the adhesion of fluid to the indenter due to surface tension). For high velocity tests, the sample was indented us ing a submerged configuration. These submerged tests reduced meniscus effects and increased th e number of data point s sampled in the fluid before contact, facilitating the determination of the contact point. (Submerged tests at low velocity were found to have increased noise. Therefore, unsubmerged tests were used at low velocity). For these tests, a polymer dam was created around the contact lens set-up. The dam was filled with deionized water, and the contact lens was held in place by overlaying a steel plate with a circle hole cut-out 3.2.2 Mechanics Model Hydrogels are composed of a polymer netw ork saturated with water. The water in hydrogels exists in two forms, i.e. bound water and free water [109]. It is reported that Etafilcon A has approximately 54% free water out of a tota l 58% water content [110]. The free water is responsible for the fluid transport propertie s of hydrogels. Polymer chains and bound water create the polymer network that contributes to the solid prop erties of the hydrogel. The mechanical behavior of porous hydrogels can be described by the biphasic model by Mow et al 75

PAGE 76

[62], which has been extensively used in the anal ysis of articular cartilage [43, 45, 62]. Biphasic theory is based on theory of mixtures in which eac h spatial point in the mi xture is assumed to be occupied simultaneously by a material point of a fluid and solid phase. In the application of the biphasic theory to hydrogel contact lens, it is as sumed that both the solid and fluid phases are incompressible and the fluid is inviscid. The c onstitutive equation for the bulk material is [62] Tr()2 s sp I I (3-1) where is the strain tensor of the solid matrix (defined by where u is the displacement vector); (T uu )/2 s s are the Lam elastic constants of the solid matrix; and p is the pore fluid pressure. Lam constants are relate d to Youngs modulus and Poisson ratio ( E v ) of the solid matrix, which are used in following sections, by and /[(1)(12)]sEvvv/[2(1)]sEvThe balance of momentum results in the equation (3-2) [Tr()2]ssp I 0Fluid flow is described by Darcys law as skp vv (3-3) where f f ss vvvis the volume-averaged bulk velocity; s v, are the velocity vectors of solid and fluid; fv s f are the volume fraction of the solid and fluid phases in the bulk material; and is the hydraulic permeability (constant k is assumed). Using Eq. (3-3) with the conservation of mass ( ) results in k0 v (Tr()) () kp t (3-4) Equations (3-2) and (3-4) comprise the governing equations for the coupled fluid-solid problem. A up (displacement-pressure) formulation was used in finite element discretization [111]. 76

PAGE 77

3.2.3 Finite Element Modeling Micro-indentation of the lens was modeled as a 2-D, axisymmetric, contact problem with the spherical indenter contacti ng the center of the lens and moving downward at a constant velocity, Fig. 3-3. The contact lens hydrogel wa s modeled as a biphasic isotropic material and the indenter was modeled as a rigid, impermeable body. The hydrogel was assumed to have constant thickness and an equal radius of curv ature on the top and bottom surfaces of the test region. The boundary conditions for the hydrogel were : (1) fixed displacement and no fluid flux at the bottom of the hydrogel (r igid foundation boundary ); (2) free displacement and zero fluid pore pressure at the surface of the hydrogel; (3) im permeable indenter, i.e., zero normal flow flux in the contact region; and (4) frictionless c ontact between the indent er and the hydrogel. A time-varying boundary condition was used to ensure impermeability at points of contact with the indenter. That is, since the contact region increased as the indenter moved downward at a constant velocity, the boundary condition assigned to nodes at the hydrogel surface that were in contact with the indenter had to be reassigned accordingly. Nodal contact points were determined at each time step using a pilot simulation (whi ch initially assumed a zero pressure boundary condition along the contact surface). A MatLab (v ersion 6.5, The MathWorks Inc., Natick, MA) subroutine was developed to generate a FEM mode l at each solution time step as guided by the nodal contact information determined from the pilot simulation. Generated FEM models updated the surface boundary condition, setting normal fluid flux to zero at contact nodes, and imported the solution from the previous time step as the in itial condition. Biphasic equations were solved using the FEM package ADINA (version 8.2.2, ADINA R&D Inc., Watertown, MA), and the solution was transferred to the next ti me step, continuing the simulation. Contact modeling was implemented by defining two potential contact surfaces: the indenter surface and the top surface of the hydrogel, Fig. 3-3. Displacement of the nodes on these 77

PAGE 78

two surfaces was checked and adjusted at each time step such that penetration of a node on one surface into another surface was avoided by im posing a penalty functi on [111]. The microindentation process was modeled as a quasi-static problem. Reac tion force on the indenter was calculated for each time step during simulation. A sparse solver (a direct solution method) was used to solve the discretized fo rm of Eqs. (3-2) and (3-4) [111]. The FEM mesh was generated using nine-node rectangular elements. The final mesh consisted of 1600 elements. To increase the accuracy of the computational results, a finer mesh was adopted in the contact region. Doubling the number of elements resulted in a ne gligible change in pr edicted force response. 3.2.4 Determination of Biphasic Properties The calculated indenter displacement vs. time fr om experiments was used as the input for indenter displacement in the FEM models. The pr edicted reaction force ac ting on the indenter in the vertical direction wa s calculated at each time step by summ ing vertical components of forces acting on all surface elements of the indenter The predicted F-D curve was obtained for a combination of biphasic properties E v and k Given the uncertainty in the material value of the Poisson ratio, a range of values were simulated (0.1, 0.2, 0.3 and 0.4). For each Poisson ratio, 16 20 simulations were carried out; 16 simulations with E varying between 100 to 400 kPa (E = 20 kPa) and 20 simulations varying k between 1.010-17 to 7.510-13 m4/N-s. The choice of these ranges was based on previous studies [103, 105], where Enns measured a Youngs modulus of 255 kPa for this material and Yasuda et al. obtained permeability values between 2.8910-17 to 1.2510-15 m4/N-s for different hydrogels with different EWC from 21% to 64%. Computationally predicted F-D curves were co mpared with experimental F-D data. Since material properties should be independent of i ndentation velocity, the se t of material properties which result in the best fitting F-D curve at the high indenter velocity should also provide the 78

PAGE 79

best fitting results at the slower velocity. F-D curves were simultaneously compared at varying indentation velocities fo r a prescribed set of material properties ( E, v, and k ). The mean square error (MSE) was calculated using the equation, 2 111 ()in m EiCi jj ij/i M SEXXn m (3-5) where m is the total number of F-D curves ( m = 7), is the total number of points to be compared for the ith experimental F-D curve ( n1 = 17 for high velocity and n2 = 32 for low velocity indentation). inEi j X and are the experimental and comput ational reaction force values of the ith experimental F-D curve, respectively. A MSE map for varying E and k was plotted for each v Potential optimal values th at minimize MSE were then identified for each Poisson ratio. Ci jX3.3 Results 3.3.1 Experimental Micro-indentation Results Figure 3-4 shows F-D data obtai ned during micro-indentation of the contact lens for two different indenter velocities: 1.2 m/s ( n = 4) and 20.0 m/s ( n = 5). There was an initial displacement region between 0 to ~12 m, i.e., the toe region, where there was no obvious difference between F-D behaviors between the differe nt indenter velocities. Forces in this region are quite low, up to approximately 1.5 mN. Upon further indentation up to 20 m, two distinct curves were obtained, with the hi gher indenter velocity tests resu lting in a larger force response. Response forces acting on the indenter were as high as 4.8 mN for the 20.0 m/s velocity at maximum displacement (~20% strain). Corresponding reaction forces at 1.2 m/s were ~ 20% lower. 79

PAGE 80

3.3.2 Sensitivity of the Predicted F-D Response Different indenter velocities were simulated while keeping all the other parameters fixed ( E k v and the maximum displacement). As the indent er velocity increased, the response force acting on the indenter also increased, i.e. the FD curve shifted upward, as shown in Fig. 3-5A. Area under the curve (AUC) was calculated to m easure changes in the F-D response. AUC has physical relevance as the work done during micro-i ndentation. Sensitivity of AUC to changes in indenter velocity is shown in Fig. 3-5B. This graph implies that, for a given hydrogel, there is a range of indenter velocities within which the response force increases with indenter velocity. Outside of this range, the F-D response changes lit tle with indenter velocity. For materials with higher permeability, this sensitive range shifts to a range with higher indenter velocity values. These sensitivity results imply that it is necessary to test within the sensitive range of indenter velocities when using micro-indentation to back out hydraulic permeability of a given hydrated material. This can be insured by testing at multiple velocities that result in distinct F-D curves. Micro-indentation of materials with different permeability values was also simulated while keeping all the other parameters fixed ( E v and indenter velocity). F-D response was compared by plotting AUC as a function of permeability, Fig. 3-5C. This gr aph also suggests a sensitive range within which the response force increases as the permeability decreases. This sensitive range of permeability shifts to a range with hi gher values as indenter velocity increases. 3.3.3 Biphasic Parameter Estimates Experimental F-D results were compared with the simulated F-D response over the prescribed range of E v and k A MSE map generated using the experimental F-D response at a single indenter velocity (1.2 m/s) and a fixed Poisson ratio ( v = 0.3) is presented in Fig. 3-6. From this map it is evident that the MS E does not reach a unique minimum, and E and k can not be estimated uniquely. The range of best fitting E and k values was reduced by using the MSE 80

PAGE 81

calculated from multiple indenter velocities (1.2 m/s and 20 m/s). Figure 3-7 shows the corresponding MSE maps for different Poisson ratios. MSE maps did not show a unique value for v with the minimum MSE values being close in value, e.g. 0.0102 for v = 0.4, 0.0098 for v = 0.3, 0.0075 for v = 0.2, and 0.0142 for v = 0.1. So, estimates of k and E were made for each v By using results from a previous study by Yasuda et al. [105], permeability for a high EWC (~ 64%) hydrogel was estimated as approximately 10-15 m4/N-s. By imposing this permeability range, the permeability values estimated using the MSE matrix were narrowed to the closed region with the smaller permeability values. Estimates of E and k are summarized in Table 3-1. Presented ranges correspond to MSE < 0.03. E and k values estimated at varying v were not greatly different The estimated range of k for Etafilcon A was 1.010-15 to 5.010-15 m4/N-s, and the range of estimated E was from 130 to 170 kPa. Simulation curves (with minimal MSE) for the F-D response with the corresponding experimental data are shown in Fig. 3-5. Good fits were achieved with MSE = 0.012 for the indenter velocity of ~20 m/s and MSE = 0.014 for the indenter velocity of ~1.2 m/s. Simulated indentation at diffe rent indenter velocities ( 20, 1.2, and 0.05 m/s) using estimated parameters ( E = 140 kPa, k = 2.510-15 m4/N-s, v = 0.3) is presented in Fig. 3-8. Simulation results show fluid fl ow around the indenter with escap e of pore fluid to the outer surface of the lens. Within the contact region, fl uid pressure gradients developed in the radial direction as expected since both the bottom and t op surface were impermeable. Just outside of the contact region, pressure gradients were sk ewed towards the free drain surface. A maximum pore pressure was found at the botto m center of the contact region. To provide insight into flow dependent behavior, the magnitude of fluid pressure at differe nt indenter velocities were compared. For a slow indenter velocity of 0.05 m/s, the contribution of fluid pressure to the 81

PAGE 82

total stress was quite small (~6%) and load was ma inly borne by the solid. For higher indenter velocities of 1.2 and 20 m/s, fluid pressure co ntributed ~41% and ~51% respectively, to the total stress. In these two cases, both fluid and solid phases support th e indentation load. 3.4 Discussion In this study, the mechanical behavior of a thin contact lens hydrogel under microindentation was investigated. Obse rved rate-dependent F-D behavi or suggests that previously developed micro-indentation anal ysis techniques for elastic/pl astic materials [39] may be inadequate for analysis of soft hydrated materi als, since the resultant bulk modulus values may vary with indenter velocity. Rate-dependent be havior caused by fluid flow can be explained using biphasic analysis. In th is study, a computational bipha sic model was developed which accounts for the varying contact area and the thin, curved geometry of the lens. Indentation rateindependent material properties E and k were determined by fitting model results to experimental measures of F-D response. Sensitivity analyses of F-D behavior showed a range of indenter velocities outside of which, the F-D response changed little with increa ses or decreases in inde nter velocity. A similar result was found when varying the permeability property and holding all other parameters constant. Limits in the mechanical response un der conditions of changing loading conditions and material properties are well known [45] and ex ist because of limitations on the pore fluid response. For slow indentation rate tests or high permeability materials, the fluid within the hydrogel has enough time to redistribute through pores and the load is main ly borne by the solid matrix. For high indentation rate tests or low pe rmeability materials, there is not enough time for pore fluids to redistribute, resulting in signifi cant pressure gradients. Results from sensitivity studies highlight the need to te st in the sensitive range of velocities when determining the hydraulic permeability, otherwise a unique value can not be obtained (Fig. 3-4C). To address this 82

PAGE 83

issue tests at multiple indenter velocities were conducted which result in distinct, nonoverlapping F-D curves. Also, MSE maps fitting to an F-D curve obtained at a single velocity within the sensitive region showed non-unique optimal biphasic properties. This can be explained as a trade-off between properties, e.g., increases in the force response were predicted by increasing E or by decreasing k separately. However by simultaneously fitting F-D curves obtained at multiple indenter velocities, th e common minimum region in MSE mappings was narrowed and these regions provided estimated ranges of E and k No obvious minimum for MSE was found for the wide range of Poisson ratio values input into the biphasic model (0.1 to 0.4). Few publis hed data of Poisson ratios for hydrogel matrices are available with which to narrow the predicted range, though a small Poisson ratio is expected due to the large equilibrium water content. Goldsmith et al. [41] measured a Poisson ratio of 0.0 to 0.307 for the hydrogel polyNVP-MMA (N -vinyl pyrrolidone met hylmethacrylate) with cellulose acetate (CA), which has 50% EWC. However, over the range of Poisson ratio simulated, the estimated ranges of E and k were found to be relatively insensitive to the changes of Poisson ratio, Table 3-1. The estimated Youngs modulus for the solid matrix component of the hydrogel was 130 to 170 kPa. Comparing with bulk modulus values (which measures the combined matrix and pore fluid resistance) provided by the manufacturer [103], the MSE estimated range was lower as expected. The bulk modulus value of 255 kPa was obtained from tensile tests at 850 m/s. Bulk modulus values calculated from biphasic FEM simula tions of the tensile test using the estimated biphasic properties were 193 kPa. This predicted value is comp arable to the experimentally measured value since some difference is expected given the different testing mode and the large difference in testing velocities between the tensile testing and the indentation tests. Estimated 83

PAGE 84

hydraulic permeability range was 110-15 to 510-15 m4/N-s. Yasuda et al. [105] measured the permeability of 21% EWC pHEMA and 64% EWC pGMA. By interpolating their results, a 58% EWC Etafilcon A may be expected to have permeability range of ~0.8-1.010-15 m4/N-s, similar to the range estimated. Our measured hydraulic permeability values were significantly larger than those by Refojo [106] and Monticelli et al. [82]. Differences may, in part, be attributed to different testing procedures. Monticelli et al. [82] used a high pressure gradient to drive flow across the hydrogel membrane, ~35-100 kPa. Permeability has been found to decrease nonlinearly with compression [33, 34, 69], and large pressure gradient may significantly compress the membrane. Differences in measured values may also be due to differences in gel formulation, e.g., crosslinking density, pore size, and percentage of unbound water. Previous studies by Hale et al. [42] and LeRoux et al. [108] also used cu rve-fitting to back out biphasic material properties. Hale et al. [42] fit to an F-D curve obtained at a single velocity and searched for the optimal E v and k of canine articular cartilage within potential ranges. When using this approach, unique values of k or E may not be obtained if property ranges and indentation loading rate are not carefully chosen. Most previous studies including this study, limit analyses to isotropic material properties. When the number of unknown properties increases such as for anisotropic materials, additional di rect experimental measurement is needed. LeRoux et al. [108] have also used a multi-step testing approach for transversely isotropic meniscus. In their study, tensile tests we re used to measure the E and v and then curve-fitting of tensile stress relaxation data was used to find the transversely isotropic permeabilities. Instead of analyzing FD behavior during indentat ion, a number of researchers have used indentation creep tests to extract biphasic properties [ 43, 45]. In this study, analysis was limited on the F-D behavior. However, analysis of creep response provide s a sensitive method for determining biphasic 84

PAGE 85

properties since the effects of E, v and k on the force response are more easily separable. Mow and coworkers [43, 45] have presented an analyt ical solution for indentat ion creep tests using a cylindrical flat-ended porous indenter to which experimental data can be fit. An infinitesimal deformation was assumed in the biphasic formulation, and a maximum 20% strain was considered. Previous studies by Spilker et al. [88] suggest that strain below 25% is reasonably modeled using linear biphasic theo ry. Also, our current biphasic model uses a constant hydraulic permeability assumption. However, local changes of permeability due to compression may be significant and can be m odeled using a strain-dependent relation as presented by Lai and Mow [34]. Implication of the constant hydraulic permeability assumption on the current estimates is that permeability may be underestimated, since fitting was made on data collected from a compressed material. A hi gher permeability is expected at a zero-strain state. FEM boundary conditions assume zero displacement and zero flow flux boundary conditions at the bottom surface of the contact lens. In reality, a thin fluid film may form at the hydrogel-foundation interface, and some relative sliding may occur during indentation. However, no obvious sliding or detachment of the lens wa s observed during micro-i ndentation test, and the no delamination assumption was assumed to be valid. In addition, friction at the indenterhydrogel interface may contribute to shear for ce acting on the indenter. However, measured friction coefficients values for contact with st ainless steel are very small [112]. Even with a perfect adhesion, i.e., no sliding between indenter and sample, Spilker et al. [88] showed only about a 10% increase in force response using a flat-ended solid inde nter compared with a lubricated indenter. Accuracy of the experimental data was also a ffected by several factors. For samples tested in air, the meniscus force due to surface tens ion between the indenter and fluid layer was 85

PAGE 86

estimated to be 0.05 mN (~ 1.0% of maximu m contact force). These small forces were considered to be constant duri ng indentation. However, changing meniscus forces at the initial contact point may have had some effect on cont act point calculation. Submerged configuration tests helped to reduce these effects. The rate of data sampling and envi ronmental noise may also affect the estimated biphasic properties via shifting of the F-D curve. At low sampling rates, an earlier data point, which does not correspond to the physical contac t, may be determined as the contact point, resulting in a smaller E or higher k estimate. Given sensitivity of results to contact point determination, care was taken to minimize error introduced by these effects. Experimental micro-indentation tests were r un at two different inde nter velocities. To check whether these indenter velocities were ade quate to capture fluid fl ow dependent behavior, a posterior analysis was made by using the characteristic time, where a represents the characteristic distance and HA the aggregate modulus ( ) [43, 113, 114]. A small characteristic time corresponds to a fast escape of the fluid from the bottom cente r of the indenter to the outer free drain surface. By using the estimated values for the hydrogel ( E = 130-170 kPa, k = 1-510-15 m4/N-s, v = 0.10.3, and a = 200 m, the maximum contact length), was conservatively ca lculated to be ~ 35 s. Reaching an indentation depth of ~20 m at 1.2 and 20 m/s requires a period of time 2 and 35 times smaller than respectively, indicating that fluid does not have time to redistribute at 20 m/s, and the fluid redistribution time is the same order as the solid deformation time for 1.2 m/s. For a much slower indenter velocity of 0.05 m/s, maximum indentation takes a period of time nearly 11 times larger than So fluid has time to redistribute, and the load is mainly supported by the polymer matrix, Fig. 3-8. This simple analysis using supports our argument that the distinct F-D beha vior exhibited at the indenter velocity of 1.2 m/s is due to being in the *2/()AtaHk 2(1)/[(12)(1)]ss AHEvv v*t*t*t*t 86

PAGE 87

sensitive range for which fluid flow aff ects the F-D response. In practice, such analysis provides an a priori means for estimating indenter veloc ity ranges provided that a range of material is known. *tMicro-indentation in combination with biphas ic analysis may provide a useful way to measure the rate independent bipha sic properties of thin contact lenses. This study demonstrates a method to extract the biphasic properties by fitting to the F-D response during microindentation. Sensitivity analyses using th e biphasic FEM model acts as a guide for understanding of the flow-dependent behavior of hydrated soft materials and improving experimental design. Estimated Youngs modul us and hydraulic permeability for the Etafilcon A contact lens, together with its clinical perfor mance data may provide important parameters for the design of more comfortable c ontact lens. These material properties may also be used in other computational models, e.g., modeling on-eye move ment of contact lenses drug delivery, fluid transport through the lens, and lens buckling. 3.5 Conclusions Biphasic FE simulation of microindentation provides a useful tool to back out biphasic properties via curve-fitting with experimental F-D data. This method is able to test small-sized samples and is able to account for complex micro-indentation conditi ons, such as curved geometry of materials, and impermeabl e indenter. It was also found that There is a sensitive range of indentation veloc ity, within which F-D response is sensitive to the varying indenter velocity. Single F-D curve fitting leads to multiple biph asic properties. Multiple-curve fitting to non-overlapping F-D curves is necessary to narrow down the potential biphasic properties. 87

PAGE 88

Table 3-1. Estimates of biphasic properties at different Poisson ratios. Presented ranges correspond to MSE < 0.03. Poisson ratio Youngs modulus (kPa) Permeability (m4/N-s) 0.4 150-170 1-510-15 0.3 150-170 1-510-15 0.2 130-150 1-510-15 0.1 138-142 2.0-3.010-15 88

PAGE 89

Figure 3-1. Schematic representation of the micr o-indentation experiment on contact lenses. The indenter, located at the end of the stagecontrolled cantilever, was lowered onto the center of the submerged contact lens. Reac tion force was calculated from cantilever flexure which was measured using an optical sensor. Figure 3-2. Cantilever force response dur ing micro-indentation (unsubmerged). 89

PAGE 90

Figure 3-3. Axisymmetric FE mesh of the contact lens and imposed boundary conditions on the top and bottom surfaces. The impermeable spherical indenter (r = 1 mm) which contacts with the hydrogel surface is re presented by the pink arc. Lenses were assumed to have constant thickness (h = 100 m) and radius of curvature (R = 7.68 mm). The indenter surface and the top surface of contact lens were defined as potential contact surfaces. The mesh consisted of ~ 1600 nine-node rectangular elements. Figure 3-4. Fitting between computational and experimental curves at different indenter velocities (1.2 m/s and 20 m/s). The values used are: E=150 kPa, k=2.5x10-15 m4/N-s and v=0.3. 90

PAGE 91

A C B Figure 3-5. Changes in the AUC (area under curve) correspond to changes in F-D behavior A). Sensitivity of F-D curves to B) indenter velocity and C) permeability. Normalized AUC was calculated by dividing AUC with th e asymptotic limit of AUC determined at high loading rate in (B) a nd at low permeability in (C) (E = 270 kPa, v = 0.3). Figure 3-6. Mean-square-error (MSE) map genera ted to compare experimental and simulated FD response for varying E and k. A node on the mesh represents a combination of (E, k) used in the FEM biphasic model. This map (at a fixed v=0.3) compares experimental data at single indenter velocity of 1.2 m/s. Minimum values represent best fit with the experimental F-D da ta. Non-unique optimal values are found. 91

PAGE 92

Figure 3-7. Mean-square-error (MSE) maps generated to compare simulated and experimental FD responses at indenter velocities of 1.2 m/s and 20 m/s. A node on the mesh represents a combination of (E, k) used in the FEM biphasic model. Best-fit parameter values of E and k minimize the MSE at different Poisson ratio, v = 0.1, 0.2, 0.3, and 0.4. 92

PAGE 93

Figure 3-8. Predicted pore pressu re distribution for different inde nter velocities using optimal parameters: E=140 kPa, k=2.5x10-15 m4/N-s, v=0.3, u0=20 m where u0 is the maximum indenter depth. t0 is the ramping time to reach u0. The pink arc is the indenter. For t0 = 1, 16, and 400 s (corresponding to i ndenter velocities of 20, 1.2, and 0.05 m/s, respectively), the magnitude of maximum pore pressure are 50.9, 26.4, and 1.9 kPa, respectively, and the effective stress of the solid are approximately 47.9, 37.7, and 27.1 kPa, respectively. 93

PAGE 94

CHAPTER 4 4. QUANTITATIVE ASSESSMENT OF MACROMOLECULAR CONCENTRATION DURING CED BY USING CONTRAST-EN HANCED MRI: A HYDROGEL PHANTOM STUDY Non-invasive real-time MR imaging provides information on the drug-affected region, and may improve the understanding of drug pharmacokinetics and may aid clinical protocol design. In this chapter, a methodology is developed to quantify the drug concentration distribution during CED using contrast-enhanced MR imaging. This method is able to quantify the spatial concentration profiles in stead of distribution volumes alone. This method utilizes the intrinsic material properties (e.g., T1 and R1) to determine the concentrati on, and reduces the effect of instrumental factors (e.g., magnetic field i nhomogeneity) on concentra tion calculation. As a methodology investigation, Gd-DTPA-labeled albu min was infused into an agarose hydrogel phantom. The MR-derived concentration prof iles were found comparable to both results measured directly using quantitative optical imag ing and results from a computational transport model. Content of this chapter is based on a paper published in the journal of Magnetic Resonance Imaging [115]. This study was collaborated with Garre tt W. Astary (Biomedical Engineering, UF), Hector Sepulveda (Biomedical Engineering, UF), and Dr. Thomas H. Mareci (Biochemistry and Molecular Biology, UF). Garre tt and Hector collected and he lped analyze the MRI data. Dr. Mareci provided technical expert ise in MR imaging techniques. 4.1 Introduction Concentration distribution of therapeutic agents after infusion is significantly related to the CED protocol (e.g., selection of infusion site, infusate concentration, and infusion rate). Concentration distribution also plays an important role in the efficacy of treatment [23, 116, 117]. Real-time monitoring of ag ents transported within nervous tissues will not only provide 94

PAGE 95

information on the drug-affected region, but also improve the understanding of drug pharmacokinetics and aid clinical protocol design Due to its noninvasive nature, experimental contrast-enhanced MRI methods have been deve loped to monitor agents distribution during infusion [6, 9, 47]. For example, Krauze et al [47]used real-time imaging of Gd-labeled liposomes to quantify distribution volumes after in fusion into the primate brain. These previous CED studies have not investig ated the relationship between signal intensity and tissue concentration of Gd-labeled tracers. Also, onl y distribution volumes rather than spatial concentration profiles were quantified. Recently, Kim et al. [50] used a signal-concentration reference table to determine the spatial concentration profile of a drug released from an implant into vitreous of rabbit eyes. The table was obtained by imaging calibration vials with varying concentrations of Gd-DTPA in gels, which mimi c vitreous. Such a method may be difficult to implement in vivo where calibration samples ma y not capture the hete rogeneity of tissues. Accuracy of a reference table may also decrease if the MR excitation or detection field is inhomogeneous. Further examination of the mech anism of signal intensity enhancement by contrast agents may improve and simplify the dete rmination of spatial concentration distribution during direct infusion. More fundamental MR imaging studies [118120] have investigated the mechanism of signal enhancement by contrast agent in tissues or solutions. Early studies [119, 121, 122] have examined the effect of a contrast agent on water relaxation rates, i.e., 1/T1 and 1/T2, in tissue. These studies found that changes in water re laxation rates are proportional to the agent concentration through the contrast agent relaxivities, R1 and R2. Recent studies [120, 123, 124] have investigated the sensitivity of relaxivities to environmental factors, such as magnetic field strength and solvent media. Analysis methods and imaging procedures based on relaxivities 95

PAGE 96

have been investigated to map signal intensity to agent con centration [125, 126]. These studies used fluid media as a solvent, since the concentr ation is uniform and easier to verify than in structured media. Hittmair et al. [126] dete rmined the relative Gd-DTPA concentration in distilled water defined as a relative change of MR signal before and after the addition of a contrast agent. More recently, Morkenborg et al. [125] investig ated the capacity of different pulse sequences to quantify Gd-DTPA concentrations in human plasma. Concentration was determined by numerically solvi ng a nonlinear equation that related the concentration with the ratio of signal intensity before and after doping the plasma with contrast agents. These studies suggest that the absolute spatia l concentration profile of a drug may be quantified during direct infusion using contrast-enhanced MRI. The purpose of this study was to devel op and validate a methodology for contrastenhanced MR imaging to quantify the evolving spatial concentration distribution in a porous media during CED. To the best of our knowledge, MR studies that quantify concentration profiles for CED have not been previously reporte d. We present a data acquisition and analysis method to relate the signal intensity to the infu sate concentration. The method utilizes intrinsic material properties (T1 and R1) and reduces the effect of instrumental factors (e.g., inhomogeneity of MR detection field). As a methodology study, this study used an agarose hydrogel phantom as a tissue substitute. Agaros e hydrogels have been previously used as phantoms for nervous tissue studies [127, 128] a nd is well-characterized. An 11.1 T magnet system was used to obtain preliminary scan s and image the infusion of Gd-DTPA labeled albumin (Gd-albumin), which has a molecular si ze similar to a therapeutic drug, into the hydrogel phantom. Calibration solutions were used to measure the relaxivities of Gd-albumin. Analysis of MR signal dependence on infusate concentration was conducted to optimize the 96

PAGE 97

dynamic range of MR signal. By using T1-weighted scans of the hydrogel before and during infusion, signal enhancement was calculated and radial concentration profiles along the infusion site were derived. As a validation test, MR-der ived concentration profiles were compared with results measured directly us ing quantitative optical imaging and with results from a computational transport model for porous media. 4.2 Theory 4.2.1 Relationship between Concentration and Signal Intensity Signal intensity from conventional spin-echo (SE) MR imaging is expressed as [129] (4-1) 1/ 0(1)TRTTETSSee2/cwhere TR is the time for recovery, TE is the time of echo, and S0 is the maximum signal intensity determined by proton density and instrument factor s, such as the resonance frequency, and the receiving coil geometry. The e ffects of contrast agent on rela xation times are governed by [118, 121, 130] 11011/1/TTR (4-2) 22021/1/ TTR c (4-3) where T10 and T20 are the relaxation times without contrast agent, T1 and T2 are the relaxation times with contrast agent at a concentration c, and R1 and R2 are the longitudinal and transverse relaxivities of the contrast agent on water, respec tively. Substituting Eqs. (4-2) and (4-3) into Eq. (4-1) results in the relation 101 202(1/)(1/) 0()[1]TRTRcTETRcScSee (4-4) which is the enhanced signal af ter addition of a contrast agen t. The signal enhancement is defined as 101 202 10 20(1/)(1/) //()[1] (0)[1]TRTRcTETRc TRTTETScee See (4-5) 97

PAGE 98

where and are the signal intensities at concentrations c and zero, respectively. For T1-weighted imaging, the infusion concentration was selected such that Thus () Sc (0) S2exp()1 TERc 10/ 1111(0)1 [ln (0)()(1)TRTS c0] R TRSSceT (4-6) Equation (4-6) relates the signal intensity to the c ontrast agent concentration for known values of 1 R and (0)S10T4.2.2 Signal in the Presence of Noise A low signal-to-noise ratio ( SNR) may significantly affect the accuracy of estimated concentrations. Zero-mean Gaussian noise wa s assumed and the confounding effect of noise on the measured signal was removed by using [131, 132] 222 SM (4-7) where is the average signal intensity w ithin a region of interest (ROI), S 2 M is the average of the square measured-image-intensity, and is the standard deviation of observed noise. The level of SNR also determines the minimum detect able concentration. The relative signal enhancement ( RSE ) was defined as ()(0) (0) ScS RSE S (4-8) For a given SNR value, we assumed that the enha ncement is not detectable if (0)SRSE where is a factor to increase or d ecrease the stringenc y of criteria (1 was used). That is, if the signal increment, is less than the noise level, ()(0)ScS the concentration is not detectable. The minimum detectable concentration, can be found using minc min()/(0)/(0)1ScSS (4-9) 98

PAGE 99

4.2.3 Porous Media Model of Direct Infusion Infusion into hydrogels can be modeled as an infusion into rigid porous media [59]. Fluid flow in rigid porous media, without source and sink terms, satisfies the continuity equation 0f v (4-10) and Darcys law fkp v (4-11) where is the porosity, i.e., fractional volum e of fluid in the porous media, k is the hydraulic permeability, f v is the fluid velocity in the porous media, and p is the pore pressure. Transport of a non-binding, non-reacting agent, such as albumin, is described by the convection-diffusion equation (f effc cc t vD )0 (4-12) where c is the solute concentr ation in the hydrogel, and Deff is the effective diffusivity of solute in the hydrogel. 4.3 Materials and Methods 4.3.1 Materials A stock solution of 25 mg/mL Gd-albumin (G albumin, BioPAL Inc., Worcester, MA) was used, which has 10-15 Gd-DTPA per albumin molecu le and a molecular weight of 70 kDa. This solution was diluted with deionized water into ~100 L calibration vials at concentrations of 2, 4, 6, 8, 10, 11, 12, 13, 14, 15 mg/mL. Calibra tion vials based on this series were used for T1 and T2 measurements to obtain R1 and R2 values. Hydrogel calibration vi als were not used due to the extremely slow diffusion of album in in the hydrogel. A diluted concentration of 10 mg/mL was used for hydrogel infusion studies In addition, Gd-albumin was mixed with Evans blue (1 mg per 2 mL of Gd-albumin) fo r visualization purposes. 99

PAGE 100

For infusion studies, ~15 mL of 1% (w /w) agarose-based hydrogel (TreviGel 5000, Trevigen, Inc., Gaithersburg, MD) was prepared in 15 mL plastic test tubes. A custom MRcompatible infusion system was designed and bu ilt (shown in Fig. 4-1). A silica cannula (ID = 50 m, OD = 147 m) was inserted ~2 cm into the hydrogel. This cannula was coupled to a non-metallic hydraulic drive which consiste d of modified gas-tight syringes (100 L luer-tip syringe, Hamilton, Reno, NV) and PEEK tubing (U pchurch Scientific, Oak Harbor, WA). The hydraulic drive allowed placement of the syringe pump remote from the high magnetic field. The infusion was delivered at a constant rate of 0.29 L/min over 91 min (total 26 L). 4.3.2 MR Imaging MR imaging procedures were c onducted at room temperature (~25 C) using a Bruker Avance 11.1 T magnet system (Bruker NMR Inst ruments, Billeria, MA) with a custom quadrature birdcage MR coil used for excitation and detection. T1 and T2 values were measured in calibration vials with various Gd-albumin concentrations. For T1 measurements, a SE sequence was used with TR = 250, 500, 1000, 2000, 4000 ms, TE = 15 ms, and 2 averages. For T2 measurements, a SE sequence was used with TE = 15, 30, 45, 60, 75, 100, 125 ms, TR = 2000 ms, and 3 averages. The T1 and T2 values of agarose hydrogel were also measured by using the same method. Multi-slice T1-weighted MR imaging of infused region before and during infusion were performed with TR = 330 ms, TE = 9.4 ms, and 6 averages. For all sequences, the image resolution was 0.229 mm 0.229 mm 1.0 mm per voxel. 4.3.3 MRI Data Processing T1 and T2 values were obtained by fitting mono-exponential curves to signal values of the variable TR and TE experiments using a least square error method. Using the calculated values of T1 and T2, relaxivity values of R1 and R2 were calculated using Eqs. (4-2) and (4-3). T10 and T20 100

PAGE 101

of the hydrogel were also calculated and used together with R1 and R2 values to estimate a range of infusion concentrations which provides distinct signal contrast at different concentrations of contrast agent. Following pre-infusion MR scans to measure S(0), infusion and simultaneous MR scanning were performed to measure S(c) at different times. Spatial concentration profiles were then obtained using Eq. (4-6). 4.3.4 Quantitative Optical Me asurement of Concentration With the same infusion configuration that was used in MRI experiments, Evans-bluebound Gd-albumin concentration was also measur ed directly using optical imaging. After infusing 30 and 91 min, hydrogels were cut into slic es with ~ 1 mm thickness. The slice with the maximum distribution radius at each time point was used for con centration measurement and the slice thickness was also measured. A fluorescent light box was used to backlight the transparent hydrogel slices and images were recorded usi ng a digital SLR camera (Nikon D50) fixed on a tripod. Pixel intensity from each image was c onverted to dye concentration by using Beers law[133], which is log10(I0/I1) = hc, where I0 and I1 are the light intensit y entering and leaving the hydrogel slice, respectively; is the absorption coefficient of the dye; h is the slice thickness, and c is the dye concentration. A calibration solution, w ith a known concentration of Evans-bluebound Gd-albumin (10 mg /mL), was used to measure so that the Evans-blue-bound Gdalbumin concentration in the hydrogel could be qu antified. Figure 4-2 shows the optical imaging method used in this study and typical recorded optical images and calculated concentration maps at two time points. 4.3.5 Simulation of Infusion into Porous Media A finite element model for the infusion in to porous media was implemented using COMSOL (v.3.0, COMSOL, Burlington, MA). We modeled infusion into a spherical isotropic 101

PAGE 102

porous media (radius = 7.5 mm) from a spherical infusion cavity ( radius = 75 m, corresponding to the OD of infusion cannula) at the center. Constant con centration and constant flow rate of 0.29 L/min were applied to the embedded infusion cavity. Zero pore pressure was applied at the outer su rface of the spherical porous media since it is large enough that the pore pressure decayed to approximately zero. Fluid velocity, f v, was first obtained by solving Eqs. (4-10) and (4-11). A hydraulic permeability, k = 1.427 10-12 m4N-1s-1 was used. This value was measured using a custom permeameter system [134], which measured the pressure drop (0.8-7.8 kPa) across a hydrogel membrane for a given flow (5-50L/hr). It should be noted that, for a constant flow rate infusion, f vis not sensitive to changes of k value and only pore pressure magnitude changes. By using the obtained f v, spatial and temporal concentration profiles for Gd-albumin was solved using Eq. (4-12). The transport model was executed using different values of albumin diffusivities, Deff = 1.0 9.0 (10-11m2/s). This range is based on a previous study [135] where Deff = 5.0 and 7.12 (10-11m2/s) at temperatures of 25C and 37C, respectively, were directly measured for bovine serum albumin in 1% (w/w) agarose hydrogel. Also, curve fitting between MRI and simulation results was conducted to find an optimal diffusivity by comparing values of R2, which measures the goodness-of-fit for nonlinear regression. Considering that the signal intensity in an MR image is an average over each voxel, concentrations obtained from porous media simulations were adjusted accordingly by averaging the concentrat ion along the thickness of slice (1.0 mm) in the slice direction. Porosity of the hydr ogel was estimated by the ratio of Vi to Vd. Vd is the distribution volume ca lculated by using the measured distribution diameter after infusion, and Vi is the volume infused. The measured ratio of Vi to Vd was ~0.6. 102

PAGE 103

4.4 Results 4.4.1 Measurement of T10, T20, R1, and R2 for the Hydrogel Relaxation times, T10 = 3.52 0.323 s and T20 = 0.117 0.00141 s, were measured directly in the hydr ogel. Relaxivities, R1 = 22.3 0.557 L/mmol-s and R2 = 42.3 2.53 L/mmols, were measured in water. The R1 and R2 values of the 1% agarose-based hydrogel were assumed to be equivalent to those measured in water because of the high water content ~ 99% (w/w) in the hydrogel. 4.4.2 Infusion Concentration Analysis By using Eq. (4-4) and the measured para meter values, the signal intensity versus concentration relationship for T1-weighted imaging was quantified. As shown in Fig. 4-3, signal intensity was found to increase w ith concentration for low concen tration ranges and decrease at concentrations higher than a threshold value, at which a maximum signal intensity is reached. This threshold value changed for different TR values, e.g., 20 mg/mL for TR = 500 ms, and 29 mg/mL for TR = 330 ms. To map the signal intensity change with chan ging agent concentrati ons, the range of low concentrations is desired due to the nearly linear relationship between signal intensity and concentration. In addition, a larger slope value of dS/dc is desired since it gives a higher dynamic range, i.e., a broader range of signal inte nsity for a given range of concentrations. Longer TR values give higher dynamic ranges but incr ease the imaging tim e. For this study, TR = 330 ms and an infusion concentration of 10 mg/m L were selected to provide a relatively high dynamic range and a short imaging time. 4.4.3 MRI Results Figure 4-4 shows the evoluti on of the MR signal during infusion of Gd-albumin. The location of the silica cannula and infusion site wa s identified as a black dot in the image center 103

PAGE 104

(axial view across the test tube ). MR images show a spherical distribution of Gd-albumin symmetric around the infusion site in the ag arose hydrogel. Contours of MR signal show approximately concentric circles in the infusate-d istributed region (Fig. 44C). The signal profile along the radial direction was obtained by sampli ng in small rectangular ROIs aligned in the radial direction (Fig. 4-4B). Small rectangular RO Is instead of individual pixels were used to calculate the local concentration because the contribution of noise can be reduced by using Eq. (4-7). Figure 4-5 compares MRI-derived Gd-album in concentration profiles with results measured using quantitative optical imaging at two different time points. The MR-derived concentration was obtained by conve rting the signal intensity to Gd-albumin concentration using Eq. (4-6). A maximum MR-derived concentration of 6.4 mg/mL was determined near the infusion site. This value is consistent with the estimated porosity, ~0.6, and the infusate concentration of 10 mg/mL since the dilution effect in porous me dia is described by, average concentration = porosity infusate concentration. The MR-d erived results co rrespond well with concentration profiles measured di rectly using optical imaging (R2 = 0.94 and 0.92 at time = 30 and 91 min, respectively). Figure 4-6 compares Gd-Albumin concentr ation profiles obtained by porous media transport simulations and MRI experiments. MR -derived concentratio n data corresponds well with the simulation profiles at higher values of reported diffusivity, e.g., Deff = 7.12 10-11 m2/s (R2 = 0.91 at time = 30 and 91 min). By fitting MRI and simulation profiles, an optimal diffusivity was found to be Deff = ~6.75 10-11 m2/s (R2 = 0.92 and 0.91 at time = 30 and 91 min, respectively; room temperature). This value is well within the measured range of 5.0 7.12 (10-11 m2/s) for temperatures 25 37C [135]. For the longer time infusion of 91 min, a good 104

PAGE 105

match was found in high concentration regions. MR-d erived concentrations were slightly higher than those predicted by simulation within the lo w concentration region of the advancing front. For the shorter time infusion of 30 min, MR-der ived concentrations we re higher than those predicted by simulation. 4.5 Discussion In this study, we presented a methodology to quantify the concentra tion profiles of Gdalbumin following direct infusion into an agar ose hydrogel. Measurements of relaxation times and relaxivities were first conducted using calibration sample s. Analysis of MR signal dependence on infusate concentration was then conducted to optimize the dynamic range of MR signal intensity. T1-weighted SE imaging was used to obtain the spatial-temporal signal evolution during infusion. Signal intensity was c onverted to Gd-albumin concentration using Eq. (4-6), using the measured 1 R and values. The MR-derived concentration profiles were found comparable to both results measured directly using quantitative optic al imaging and results from a computational transport model. (0)S10TTheoretical infusate concentration analysis s howed that infusate concentration and imaging parameters, such as TR and TE, should be carefully chosen to obtain a broader dynamic range of MR signal and a one-to-one signa l-concentration relationship. Choosing proper agent infusion concentration is important since concentrations that are too low may lead to a poor SNR and may not be detectable, and concentr ations that are too high yield T2 effects that diminish signal intensity and skew data. According to the analysis of minimum detectable concentration [Eq. (49)], Gd-albumin concentrations less than 0.1 mg/mL cannot be separated from noisy signal intensity for this experimental protocol. At high concentrations, the relaxivities will change. But 105

PAGE 106

Tweedle et al. [121] have found that relaxivities were invariant for Gd concentrations from 0 to 2 mmol/L in agar and blood solutions, respectively. The signal-concentration relati on [Eq. (4-6)] is based on T1-weighted imaging, and it assumes that This assumption is reasonable since for the parameters in this study, (R2 42.3 L/mmol-s, TE = 9.4 ms, MW = 70 kDa, and c = 1-6 mg/mL). Also, this term is not sensitive to changes of the R2 value for the range of parameters in this study. The signal-concentration relationship may also be obtai ned by directly using Eq. (44), which necessitate s quantifying the unknown S0 (which may be obtained during T1 or T2 measurements). Hittmair et al. [126] used a modified form of Eq. (4-4) to define an enhancement factor, which required calculating 2exp()1 TERc 2exp()1.01~1.03 TERc 00[(0)]/[(SSSSc)] They used a very low flip angle sequence to obtain S0. However, S0 is environment-dependent, e.g., depending on the receiving coil, and may be different for each MR measurement. Our method uses the relative signal change, and avoids the need of measuring S0. Using the ratio, implies that the concentration calculation is minimally affected by the inhomogeneity of detection field, e.g., surface coils, where the non-uniform field will affect both S(c) and S(0) equally. Also, adiabatic pulses [136] can be used for MR excitati on so that the SE sequence can be used with MR coils that have inhomogeneous fields. In addition, this study used a SE sequence and each scan took ~3 min. This time re solution is acceptable given the relatively slow velocities of albumin transport in the hydrogel (ave rage radial velocity of 0.022 mm/min over the 91 min with a maximum of ~0.060 mm/m in at the beginning of infusion). ()/(0)ScS()/(0)ScS R1 and R2 values measured in water solutions were used as an approximation for the relaxivities of water in agarose hydrogel. Pr evious studies [120, 123, 124, 137] show that R1 values may be media-dependent, and the difference may be determined by the nature of different 106

PAGE 107

media, e.g., the macromolecular content. Stanisz et al. [120] and Rohrer et al. [123] measured different R1 values for Gd-DTPA in water, powdered milk, egg white and plasma. In contrast, Donahue et al. [124] measured similar R1 values in saline, plasma, uncompressed and compressed cartilage. However, the approximation in this study is reasonable given the high water content and low fibrous structure (polysac charide chain network) in the hydrogel. Stanisz et al. [120] have shown that at low m acromolecular content, the variation of R1 in water is negligible. In addition, since Gd -DTPA relaxivity is also aff ected by magnetic field strength, temperature, and the macromolecule to which it is attached, direct comparis on of our results with other studies is difficult. Schm iedl et al. [138] reported a R1 value for Gd-DTPA-Albumin (9 GdDTPA per albumin) to be 110 L/mmol-s measured at 0.25 T, 37 C. Brasch [122] reported a R1 value of 14.4 L/mmol-s per Gd for Albumin-Gd-DTPA at 0.25 T, 25 C. Accurate determination of R1 value is necessary for applying this method to biological tissue. R1 may be measured using phantoms which can properly mimic the biological ti ssue. Alternatively, it may be determined by injecting different concentrations of tracer directly into tissue. The validity of this methodology was first confirmed by comparing th e MRI results with quantitative optical measurement. It should be not ed that the accuracy of optical measurements may be affected by several factors, such as background noise and the homogeneity of the hydrogel slice thickness. Comparison of concen tration profiles obt ained by porous media simulation and MRI experiments also show good agreement. It was found that MR-derived concentrations were slightly higher than those predicted by simulation within the low concentration region of the advancing front. At low concentrations, noise becomes dominant and we cannot accurately separate the low concentr ation from the noisy MRI signal according to our minimum detectable concentration analysis. It was also found th at during short time infusion, 107

PAGE 108

e.g., 30 min, MR-derived concentrations were hi gher than those predicted by simulation. Several factors may contribute to the discrepancy be tween MRI experiments and simulations. (1) A spherical infusion cavity assumption was used in the computational model. In experiments, the infusion cavity may expand preferentially along th e track of the cannula to form a tear-drop shape [28]. In our measurements, an asymmetr ic distribution in the cannula direction was observed at early infusion tim es. However, we did not model this expanding non-spherical source since modeling such a moving boundary problem is technically difficu lt. (2) The tip of the infusion cannula was assumed to be in the middle of an MR slice (along the sl ice or zdirection) within the computational model. This may not ma tch the MR data since exact z-position of the tip within the slice is unknown and as a result, the MRI concentration profiles may be shifted slightly. In addition, by comparing the rati o of infusion volume and distribution volume, a porosity of 0.6 was estimated for the 1% (w/w) ag arose hydrogel. This value is different from those reported by others [139, 140], where volume fraction of water was reported, and the difference may result from the space occupi ed by bounded water which is not accessible to macromolecules. Accuracy of the methodology will also be aff ected by several factors, including the signalnoise ratio and the measurement stability dur ing dynamic contrast-enhanced MRI (DCE-MRI). In this study, we account for noise in our meas urements using the assumption that measured noise has a zero-mean and is Gaussian distri buted. This assumption can be validated by measuring the ratio of the noise magnitude to th e standard deviation of the noise, which should have a value of ~1.913 in an ideal case [141, 142]. We measur ed a value of 1.83 0.07, which is very similar to the ideal case and justifies our assumption. Our DCE-MRI measurements were very stable during infusion. For example, the MR signal from a fixed ROI in the pure hydrogel, 108

PAGE 109

measured at different times during DCE-MRI, were quite consistent with a value of 3.93 0.07. Yet, the noise standard deviation in these measur ements was only 0.02 so the variation is at the level of the noise. In a t ypical hydrogel infusion measurem ent, the SNR was high with a maximum in the infused region of ~19.8 and a mi nimum in the pure hydrogel of ~7.2. Therefore, our measurements were both accurate and stable over the time course of studies. Although a high field magnet system was us ed for MR imaging, the proposed methodology may also be applicable to clinical field strengths. However, the accuracy will be limited by the SNR available at lower field streng ths and will require the use of an optimal infusate concentration. At clinical field strengths data is generally acqui red at a lower resolution than at high field strengths to overcome the decreased SNR. The SNR can be improved by increasing the number of averages, which results in a longer total acquisition time. Typically, this has an undesirable effect in MR tracer experiments since transport phenomena may not be accurately assessed at a lower temporal resolutio n. However for CED into nervous tissue, the convective and diffusive transport may be sl ow enough to allow a decrease in temporal resolution (increased number of averages) us ing a clinical magnet system, which will allow image resolution to be increased. The minimal dete ctable concentration analysis, outlined in this study, can be used to evaluate the sensitivity of the experiment as a function of SNR. The optimal infusate concentration will change with fi eld strength since the optimal concentration is a function of background T1 and T2, as well as relaxivity properties of the contrast agent. As field strength is lowered, de Graaf et al. [143] have shown that T1 decreases and T2 increases in both rat brain white matter and gray matter. Rinck et al. [144] have shown increasing relaxivities, R1 and R2 (decreased relaxation times T1 and T2), of gadolinium-based contrast agents with decreases in field strength. For different contrast agents, Rohrer et al. [123] have also shown 109

PAGE 110

variations of relaxivities with decreases in field strength. Therefore, the infusate concentration should be optimized for the proper ties of the tissue and the contrast agent at a particular magnetic field strength in order to pr ovide the greatest sensitivity to changes in contrast agent concentration to improve the accuracy of the method. Noninvasive visualization and quantification of drug distri bution in tissues using MR imaging are important. Real-time and quantitativ e monitoring may provide physiological insights to drug transport in tissues, help in unders tanding the efficacy of treatment, and aid in optimization of infusion protocols. As a methodol ogy investigation, this study used an agarose hydrogel phantom to mimic a biologi cal tissue and provide a valida tion test which demonstrated that this method can quantify prope rly the spatial concen tration profiles during CED. To the best of our knowledge, MR studies that quantify c oncentration contours for CED have not been previously reported. In future studies, the presented methodology will be applied to direct infusion into biological tissu e, e.g., nervous tissue. 110

PAGE 111

Figure 4-1. Schematic of the infusion system setup. A C B Figure 4-2. Schematic of the optic al imaging method used to meas ure the concentration of Evans blue labeled Gd-Albumin in the hydrogel slice is shown in (A ). Part (B) shows recorded grayscale optical images and pa rt (C) shows the calculated concentration maps. The conversion from the pixel intensity to the dye concentration was based on Beers law [133]. 111

PAGE 112

Figure 4-3. Plot of the theoreti cal relationship between concentra tion and signal intensity used to determine optimal infusate concentration [T1-weighted imaging; Eq. (4-7)]. Parameter values were based on our measurements (see Results). 112

PAGE 113

A 30 min 0 min 4 min 91 min 15 min 60 min B C Figure 4-4. A) Evolution of T1-weighted MR signal during infusion of Gd-Albumin into 1% agarose hydrogel at a rate of 0.29 L/min (transverse image at approximate center of infusion site; the cannula is perpendicular to the transverse plan e; slice thickness = 1 mm). B) The signal profile in radial direction wa s obtained by sampling in the small rectangular regions of interest (ROIs) (~20 pixels in each rectangle). The final signal profile was obtained by sampling in the or thogonal directions symmetric around the infusion site. This figure only shows the samp ling in the positive horizontal direction. The infusion cannula/site was identified by a black dot in the image center. C) Contours of MR signal inte nsity at time = 91 min. 113

PAGE 114

Figure 4-5. Magnetic resonance-derived concen tration of Evans blue labeled Gd-Albumin compared with profiles measured directly using quantitative optical imaging in a 1 mm thick hydrogel slice. R2 = 0.94 and 0.92 at time = 30 and 91 min, respectively. Error bars are the standard deviation (n = 12) of the optical measures. 114

PAGE 115

Figure 4-6. Magnetic resonance-de rived concentration of Gd-Album in compared with predicted porous media transport profiles at time = 30 and 91 min. Concentration was normalized by dividing by the maximum MR-d erived concentration, 6.4 mg/mL. Deff = 7.12 and 5.0 (10-11 m2/s) were experimentally measured at temperatures 37 and 25C, respectively, by Liang et al. [135]. Deff = 6.75 (10-11 m2/s) was the optimal diffusivity obtained in this study. For Deff = 7.12, 6.75, and 5.0 (10-11 m2/s), R2 = 0.91, 0.91, and 0.84 for t = 91 min, and 0.91, 0.92, and 0.92 for t = 30 min, respectively. 115

PAGE 116

CHAPTER 5 5. CONVECTION-ENHANCED DELIVERY IN TO THE RAT SPINAL CORD VIA THE SCIATIC NERVE INFUSION AND REAL-TIME MR IMAGING Convection-enhanced delivery (CED) of drug into the spinal cord th rough peripheral nerve infusion is of potential interest since it may provide a remote delivery site to the spinal cord, require less invasive surgery and re duce the risk of spinal cord inju ry during surgery. It may also allow targeting of specific neurons at nerve root entry and ganglion. In this chapter, we examine the experimental CED parameters and distribut ion characteristics on small animals (rats) by using real-time in vivo MRI. We report a protocol for in vivo CED experiments on the rat sciatic nerve. We also provide a real-time in vivo MRI method to monitor tra cer transport following CED into the peripheral nerve for delivery to the spinal cord. Experimental results provide knowledge of pathways and patterns of drug trans port from the sciatic nerve to the spinal cord. The developed experimental methodologies may be applied to future peripheral nerve and spinal cord CED studies. This study was in collaboration with Garrett W. Astary (Biomedical Engineering, UF) and Dr. Thomas H. Mareci (Biochemistry and Molecular Biology, UF). Garrett helped in collecting the MRI data and designing the CED experiments. Dr. Mareci provided technical expertise and use of the RF coil system and the MR image analysis software, MAS. All MRI data was collected at the Advanced Magnetic Resonance Imaging and Spectroscopy (AMRIS) facility in the McKnight Brain Institute of UF. 5.1 Introduction The efficacy of potentially therapeutic agen ts in clinical use may be limited using traditional delivery techniques, such as subcutan eous, intravenous, and intramuscular injection. Problems associated with systemic administratio ns of drug to central and peripheral nervous systems include systemic toxicity limited transport due to the short half-lif e and poor diffusivity 116

PAGE 117

of macromolecular compounds, and poor epineurial penetration into peripheral nerves. In contrast, direct convection-enhanced delivery (CED) into nervous tissue provides a promising drug delivery method for disease and injury of peripheral and central nervous systems (e.g., peripheral nerve injury, chronic pa in, and spinal cord injury). By introducing bulk flow within the extracellular space, intraneural CED increas es extracellular spread of macromolecules in peripheral and central nervous system tissues and provides localized drug delivery. Direct CED into spinal cord tissues likely involves invasive surgery including a laminectomy to remove protective vertebral bon es. Due to this invasive surgery and the sensitivity of the target area, direct CED into the spinal cord tissue is still not a developed clinical technique. Instead, drug delivery into the spinal cord through peripheral nerve CED is of potential interest since it may provide a remote delivery site to the spinal cord, require less invasive surgery and reduce the risk of spinal co rd injury during surgery. It may also allow targeting of specific neurons at nerve root entry and ga nglion. A similar idea has been used in previous studies which use retr ograde axonal transport for gene delivery to the spinal cord through sciatic nerve and muscle injections [54, 55]. For th is technique, a specific vector is required and a limited number of gene-e xpressing cells may limit delivery. Previous CED studies [7, 32] have investigat ed distribution characteristics after direct CED of tracers into the spinal cords of rats, swin es and non-human primates. Macromolecular infusions into the spinal cord white matter were f ound to spread into the in terstitial space parallel to axonal fibers, reflecting low tran sport resistance in th is direction. Large volumes of the spinal cord were filled with infusate, and the distributions were homogeneous, reflecting clinically significant distribution volumes. In previous studies of CED directly into peripheral nerves [52, 53], similar infusate distribut ion characteristics were observed. Infusate was well contained 117

PAGE 118

within the peripheral nerve and filled a long segmen t of nerve. By taking advantage of this low resistance transport along axonal fibers, CED into a specific region of spinal cord through remote peripheral nerve infusion may be possible. In a li mited number of primates (n=2), Ratiliff et al. [53] conducted CED of Gd-labeled albumin (~ 60-85 L) into the sciatic nerv e, and tracer filling in the spinal cord gray matter was evident in the MR images acquired at the end of infusion. Further studies and characteriza tion of peripheral nerve CED is required in additional animal models. Also, characteristics of remote CED need to be studied including the pathways and dynamics of transport. To capture the transport characteristics of this remote CED technique, real-time in vivo MR imaging is useful to monito r the transport dynamics. Previous studies [9, 54-56] have used contrast-enhanced MRI to visualize the final distribution regions follo wing CED. Only a few have conducted real-time imaging to monitor the transport of MR-visible agents during CED. For example, Krauze et al. [54] used consecutive MR scans to monitor the distribution of Gdlabeled lipsomes during CED into non-human primat e brains. Real-time in vivo MR imaging of the CED transport in the rat peripheral nerve and spinal cord is challenging due to the small dimensions of these features (e.g., ~ 4 mm diamet er for the rat spinal cord and ~1.5 mm diameter for the rat sciatic nerve), and motion-induced artifacts (e.g., rat breathing) in MR images. Particular MR device designs and MR imaging para meters may be evaluated such as the RF coil configuration. Such a real-tim e in vivo MRI methodology to mon itor rat spinal cord infusion may also provide a useful tool for future CED studies in small animals. In this study, we focus on the characterization of spinal cord transport following CED into the peripheral sciatic nerve of MR-visible comp ounds. We present a protocol for in vivo CED into the rat sciatic nerv e. We also report a MR imaging method to monitor tracer infusate 118

PAGE 119

transport in the sciatic nerve a nd spinal cord during CED. This study examines the experimental CED parameters and distributi on characteristics on small animals (rats). The developed experimental methodology may be applied to fu ture peripheral nerve and spinal cord CED studies. Transport of compounds in the peripheral and central nerve system is determined by the anatomy and the underlying microstructure of the nerve system. Figure 5-1 illustrates the anatomical structure in the spinal cord column region. A spinal cord, which consists largely of two types of tissue, gray matter and white matter, is enclosed by three meninges, an innermost pia mater, an arachnoid mater and an outmost dura mater. The arachnoid mater and dura mater are closely adherent to each other. Between the pia mater and arachnoid mater is the subarachnoid space which is filled with cerebrospin al fluid (CSF). An external spinal nerve (i.e., peripheral nerve) enters the C SF by crossing the dura mater and arachnoid mater, and divides into two internal nerves, calle d dorsal and ventral nerve roots. These two nerve roots travel inside the CSF and enter the spinal cord. For simplicity, the region invested by the dura mater is called the spinal column, which encloses the spinal cord, nerve roots and CSF. Connecting to a spinal column are numerous spinal nerves and the connection point is called the spinal nervespinal column junction. Figure 5-2 shows the external anatomy of the sciatic nerve, the spinal column and other connected spinal nerves. The scia tic nerve divides into two major branches, spinal nerves L4 and L5, and enters the spinal column. There is also a thin connection between spinal nerves L4 and L3. It should be noted that spinal nerves L4 and L5 have long internal nerve roots (~4 cm) which travel within the CSF of the spinal column before entering the spinal cord. 119

PAGE 120

5.2 Materials and Methods 5.2.1 Infusion System Setup An MR-compatible, non-compliant infusion system was designed which consists of a gastight syringe (100 L luer-tip syringe, Hamilton, Reno, NV) PEEK tubing (~8 feet), and a silica cannula (OD = 108 m, ID = 41 m) (Fig. 5-3). The infusion system was driven by a syringe pump which was remotely placed from the high ma gnetic field. Two tracer infusates were used in this study, Gd-DTPA and Gd-DTPA-al bumin. Gd-DTPA (2 mg/mL; MW = 573 Da; Omniscan, GE Healthcare, Princeton, NJ) was used to mimic a small molecular compound, and Gd-DTPA-albumin (10 mg/mL; MW = 87 kDa; ~ 35 Gd-DTPA molecules per albumin molecule; purchased from Dr. R. Braschs Laboratory, Univ ersity of California, San Francisco, CA) was used to mimic transport of larger, macromolecu lar compounds. Tracer solutions were mixed with Evans blue dye (MW = 960 Da) for visualiz ation purposes. For Gd-DTPA-albumin, the concentration of Evans blue (1.8 mg Evans bl ue per 5 mL solution of Gd-DTPA-albumin) was low enough that Evans blue was completely bound to the albumin. 5.2.2 Animal Preparation Rat preparation was conducted in accordance to a protocol reviewed and approved by the Animal Care and Use Committee of the University of Florida. Six adult female Sprague-Dawley rats weighting 250270 g were used. Rats were in itially anesthetized using 4% isoflurane, and maintained under anesthesia (2 % isoflurane) during surgery a nd MR imaging. Throughout the course of MR imaging, rat re spiration was monitored using a respiratory sensor pillow, and heated airflow was supplied to keep the rat warm. Two rats were infused with Gd-DTPA and four rats were infused with Gd-DTPA-albumin. The sciatic nerve in the right lower extremity wa s exposed for cannula inse rtion after incising the skin and retracting gluteus muscles (Figs. 5-3A to B). The dissection was extended proximally to 120

PAGE 121

the sciatic notch. Here, the silica cannula was inserted and directed retrogradely into the intrapelvic portion of the sciatic nerve with an insert depth of ~ 1 4 cm (the cannula insertion point on the sciatic nerve was ~ 5 cm distal to the L5 nerve-spinal column junction). Next, the cannula was secured and sealed by tissue adhe sive. The surgical wound was closed using monofilament sutures with a small opening left fo r the cannula to exit. The cannula was left in position for ~60 min before starting the infusion. During this period, hard ware configuration for MR imaging was conducted which include tuning of RF coil circuits and positioning of the rat in the magnet system. An infusion rate between 0.2-0.3 L/min was used. Table 5-1 summarizes the experimental infusion parameters. Follo wing CED infusion and MR imaging procedures, rats were sacrificed immediately and postmortem surgery was conducted to examine the infusate distribution and cannula position. 5.2.3 MR Imaging MR imaging procedures were conducted us ing a Bruker Avance magnet system (11.1 Tesla / 40 cm horizontal bore; Bruker NMR Instruments, Billeria, MA). The imaging apparatus included a custom dual radio frequency (RF) coil system with a large volume coil for excitation and a small surface coil for detection (Fig. 5-3C). For dynamic contrast-enhanced MRI (DCE-MRI), a spin echo sequence and multi-slice T1weighted imaging was used with a fat suppression technique. Altern ative sagittal and transverse scans were used to obtain a complete picture of the infused region. The following imaging parameters were used: for transverse scans, TR = 330 ms, TE = 8 ms, average = 5, FOV = 2 cm 2 cm, image matrix = 96 number of slices = 10; for sagi ttal scans, TR = 330 ms, TE = 8 ms, average = 4, FOV = 2 cm 4 cm, image matrix = 96 number of slices = 10. The resulting image resolution was 0.2 mm 0.2 mm 1 mm per voxel. For all scans, an anatomical 121

PAGE 122

region roughly from vertebral levels L6 to L1 was covered. Two pre-infusion MR scans were conducted followed by simultaneous MR scans a nd infusion of Gd-DTPA or Gd-DTPA-albumin. The pre-infusion MR scans were used as references of the background signal so that the infusate distribution region in the MR s cans during CED could be identified. After CED, high-resolution transverse MR scans were also obtaine d (TR = 1000 ms, TE = 10 ms, FOV = 2 cm 2 cm, image matrix = 128 128, average = 3). 5.2.4 MRI Data Processing Raw MRI data was processed using custom software, written in the Interactive Data Language (Research Systems, Inc., Boulder, CO). The distribution volume (Vd) of infusate in the spinal column was estimated using a Matlab subroutine (version 6.5, The MathWorks Inc., Natick, MA). A voxel whose signal intensity was greater than a threshold value was identified as an infusate-distributed voxel. Only the voxels in the spinal column were counted which may include CSF, nerve roots and spinal cord. The total distribution volume was calculated by adding the volumes of these voxels in each transverse MR scan. In addition, the axial length of the distribution region in the sp inal column was estimated from sagittal scans. 5.3 Results Figure 5-4 shows typical infusate transport along the sciatic nerve at the end of the CED experiment on rat #1, where Gd-DTPA was infused a nd a short insert depth ~ 1 cm was used (the tip of the cannula was ~4 cm distal to the L5 nerve-spinal column junction). Gd-DTPA was found to move up the sciatic nerve from the point of infusion, divide into spinal nerves L4 and L5, and enter the spinal column. Gd-DTPA was we ll contained within the spinal nerves and no leakage into the surrounding tissues was noted for the infusion of 70 L. 122

PAGE 123

Figures 5-5 to 5-8 show distin ct distribution patterns in th e spinal column during CED of Gd-DTPA or Gd-DTPA-albumin into the sciatic nerve. It is evident that the infusate enters the CSF. This can be deduced from the pattern of a large infusate-distributed region in the spinal cavity (Figs. 5-5 and 5-7) or an enhanced edge of the spinal cavity (Figs. 5-6 and 5-8) observed in the transverse MR images. It is also evident that the infusate transports along the nerve root since a distinct bright spot in the spinal column is observed in the transverse MR images and the spot did not expand with time (F igs. 5-6 and 5-8). In addition, no significant difference in distribution patterns was observed between infu sates of Gd-DTPA (Fig. 5-5) and Gd-DTPAalbumin (Fig. 5-6 to 5-8). Also, variations in distribution patterns were observed for each CED experiment. Figure 5-5 shows the temporal evolution of Gd-DTPA distri bution during CED for rat #2. The tip of the cannula was in the L4 spinal ne rve through which infusate entered the spinal column. The enhanced signal region is confined to one side of the spinal column and CSF uptake of the infusate is evident since a large distri bution region is observed in transverse scans. However, it is difficult to tell if infusate stayed in the nerve roots even though bright spots can be identified. Figure 5-6 shows the temporal evolution of Gd-DTPA-albumin distribution during CED for rat #3. The tip of the cannula was in the L4 spinal nerve through which the majority of infusate entered the spinal column. In this case, with increasing infusion volume, some infusate entered the spinal column through the L5 spinal nerve. Gd-DTPA-albumin distribution in both the CSF and nerve root were evident. A bright s pot was first observed at early infusion times in transverse scans and the spot did not expand with time. A tail-like expansion of the spot was observed later during the course of infusion and this expansion is believed to be from infusate 123

PAGE 124

uptake in the CSF instead of infusate leakage from a nerve root. It is also observed that GdDTPA-albumin was contained within a distant nerve root identified by a single bright dot existing in the transverse MR images at different positions. Figures 5-7 and 5-8 show dist ribution patterns for CED experiments on rats #4 to 6. Unlike the distribution pattern in the e xperiment on rat #5 (Fig. 5-8) where a bright edge around the spinal column was observed corresponding to CSF deposition, the distribution pattern in rat #4 (Fig. 5-7) shows that infusate spread into a larg e region within the spinal column and it is likely that the majority of infusate went into the CSF. The experimental results of rat #5 (Fig. 5-8) also shows that infusate leakage into the surrounding tissue occurred since signal enhancement at the outside boundary of verteb ral bone was observed. Figure 5-9 shows a graph of infused Gd-DTP A-albumin volume (Vi) versus distribution volume (Vd) in the spinal cavity. A typical curve includes an initial plateau region followed by a rapid increase in slope. Considerable variation in slope was obs erved between animals. The flat section likely corresponds to the time lag during which the infusate did not yet enter the spinal column, and the duration of this section depe nds on the position of the cannula tip. Following this time lag, the relation between Vd and Vi is approximately linear with an average value of ~ 2.0. However, this ratio has a lower value, ~0.5, in the experiment on rat #3. This may be due to the fact that a shorter insert depth was used in this experiment and more infusate transported through a second nerve, i.e., L5 ne rve (Fig. 5-6), than in other e xperiments where the majority of infusate followed one nerve. 5.4 Discussions In this study, we present an experimental method for CED into rat sciatic nerves for transport to the spinal cord. We also present a real-time in vivo MRI method to monitor the agent transport during CED. Experimental results show that by using DCE-MRI, the remote sciatic 124

PAGE 125

nerve CED can be well characterized. While consid erable variation in distribution patterns were observed, infusate was found to tr ansport towards the sp inal cord and be well contained within the epineurium until entry into the spinal colu mn. Upon entering the spinal column infusate deposition was observed to enter the CSF and nerve roots adjacent to the spinal cord. It is expected that the epineurium, which su rrounds the spinal nerve, and the pia mater, which encloses the nerve root, are poorly pene trable for Gd-DTPA and Gd-DTPA-albumin. This poor penetration may be confirmed by the fact that the infusate is well co ntained in the sciatic nerve (Fig. 5-4) and in the nerve root (Fig. 5-6). This result is c onsistent with a previous study of direct CED into the peripheral nerve by Lonser et al.[52], where infusa te Gd-labeled albumin was well contained in the peripheral nerve, and with a previous study of direct CED into spinal cord by Wood et al.[32], where it was observed that infusate 14C-albumin was well contained in the dorsal white matter column and did not cross the pia mater. CSF uptake of both infusates was clearly observe d in this study. The potential for leakage due to the meninges microstructure at the juncti on between the spinal nerve and nerve roots, where the epineurium developed into three laye rs, pia mater, arachnoid mater and dura mater (Fig. 5-1), is not clear. Given the poor pene tration of infusate across the pia mater and epineurium, this junction point may provide a ch annel through which the in fusate in the spinal nerve enters the CSF. Further study of microstructu re in this junction regi on may be necessary to determine if this is a transport-limiting factor. Infusate was found to enter the spinal column However, no infusate was measured in the gray matter or white matter of the spinal cord. This may be due to seve ral factors. The most important one may be the distributaries for infusate flow and the long travel path from the sciatic nerve to the nerve root entry zone, where the ne rve roots enter the gray matter and white matter. 125

PAGE 126

For example, a typical nerve root length from th e L5 nerve-spinal column junction to the nerve root entry zone is ~ 4 cm. This long travel di stance may require a long infusion time (several hours). But infusate can also outflow into other regions such as the CSF and nerve branches of the sciatic nerve, e.g., the thin connection between L4 and L3, which was observed in the postmortem analysis. Infusion into such a system with multiple sinks may increase the difficulty of transport into the spinal cord tissue. In addi tion, the majority of the infusate may enter the CSF during CED. The fluid resistance of the CSF system (total CSF volume in adult rats is 300 to 400 L [145]) may increase and hinder transport into the spinal cord. Other factors may include infusate leakage to th e surrounding tissue around the tip. This leakage was identified in some experiments (Fig. 5-7) and during post-mort em surgery in which Evans blue dye can be observed in the surrounding tissues, e.g., muscle s near the infusion site Though the epineurium is poorly permeable to the infusate, infusate may penetrate the memb rane since around the cannula tip a high infusion pressure is expected to build, which is the driving force for bulk flow, but this high internal pressure may also change the structure of the epineurium, e.g., increase the pore size. Further study to quantify the infusate leakage around the cannula tip may be needed. Although infusate enters both the CSF and nerve root, it was observed that the percentage of infusate which enters the CSF varied between different experiments on rats #3 to 6 (Figs. 5-6 to 5-8), which have similar experiment parame ters, including the same infusate and similar insertion depths. The radial position of the cannula tip within the cross-section of the sciatic nerve may play an important role in determining th is percentage. If the tip is close to the center in the cross-section plane, more infusate may stay in the nerve since there is likely a much higher transport resistance in the transverse direction than in the axial direct ion [32, 52] which may 126

PAGE 127

confine the infusate in the central region of th e peripheral nerve. However, further study is needed to quantify the effect of the tip radial pos ition on the infusate CSF uptake of infusate. A dual RF coil system was used in the MR imaging. The large volume coil can achieve a higher RF field uniformity but lower sensitivity, and vice versa for the small surface coil. RF field uniformity is essential in the excitation process while sensitivity is important for the reception process [146]. A system combining these two characteristics can achieve a good sensitivity and a relatively large FOV. This is particularly useful fo r rat spinal cord and peripheral nerve imaging due to their small dimens ions. In addition, we used dynamic contrastenhanced MRI to monitor the infusate (tracer) tran sport. It should be noted that to detect the infusate distribution, infusate concentration sh ould be properly selected A tracer concentration that is too high may weaken th e MR signal and distort the dist ribution pattern. Concentrations that are too low may be undetectab le due to the presence of noi se in the MR signal [115]. The infusate concentration used in this study properl y indicated the presence of the infusate in the peripheral nerve and the spinal column. However, further studies may be needed to optimize the infusate concentration in the spinal column and quantitatively determine the concentration distribution [115]. Although delivery of compound agents via remote sciatic nerve CED into gray matter and white matter of spinal cord is technically challe nging, we have demonstrated that the developed methodology provides a way to deliver drug into the spinal column region and nerve roots adjacent to the spinal cord. Our results indicate that drug delivery into the ganglion and nerve root is feasible through peripheral nerve CED. The long distances for infusate tran sport in sciatic nerve roots were an obstacle. Drug delivery into the spinal cord tissue may be possible for CED into other peripheral nerves that have shorter nerv e roots. In addition, we demonstrated that real127

PAGE 128

time in vivo MRI provides a useful tool to mon itor transport dynamics during CED. Such a realtime in vivo MRI methodology may also provide a useful tool for future CED studies on small animals. Future studies may implement remote CE D into the spinal cord via other spinal nerves, which have shorter nerve roots. Future studies may also investigate the anatomical structure at the spinal nerve-spinal column junction region, infusate leak age around the cannula tip and the effect of the tip radial position on the infusa te CSF uptake of infusate. These studies would provide greater understanding of th e transport dynamics from the peripheral nerve to the spinal cord and would provide useful data for potentia l clinical application of this remote CED technique. 128

PAGE 129

Table 5-1. Experimental parameters for CED into rat sciatic nerves Experiment Infusate Location of cannula tip Insert depth (cm) Infusion rate (L/min) Volume infused (L) Axial length of infusate penetration in the spinal column (mm) Rat #1 Gd-DTPA Sciatic nerve ~ 1 0.3 70 ~ 4 Rat #2 Gd-DTPA L4 nerve ~ 4 0.25 21 ~ 16 Rat #3 Gd-DTPAalbumin L4 nerve ~ 3.5 0.3 47 ~ 22 Rat #4 Gd-DTPAalbumin L5 nerve ~ 4 0.3 32 ~ 25 Rat #5 Gd-DTPAalbumin L5 nerve ~ 4 0.3 24 ~ 15 Rat #6 Gd-DTPAalbumin L5 nerve ~ 4 0.3 12 ~ 14 129

PAGE 130

Figure 5-1. Schematic of the anatomy of the spin al column region (GW: gray matter; WM: white matter; CSF: cerebrospinal fluid). Arrows indicate the potential transport pathways for peripheral nerve CED: (Path 1) into the CSF; (Path 2) along the nerve roots. Background picture was ad apted from [147]. Figure 5-2. External anatomy of the sciatic nerv e, the spinal cord and other connected spinal nerves. The single sciatic nerve divides into two major branches, spinal nerves L4 and L5, which enter the spinal column at verteb ral levels L4 and L5, respectively. A thin connection (small arrow) between spinal nerves L4 and L3 is also present. 130

PAGE 131

Figure 5-3. A) Schematic of the sciatic nerve in fusion. Point 1 is the insertion point where the cannula was introduced into the sciatic nerv e. Point 2 is the location of the cannula tip. Cannula length from points 1 to 2 is defined as the inse rt depth. B) The isolated rat sciatic nerve. C) The dual RF coil system with an anesthetized rat used to collect MRI data. 131

PAGE 132

Figure 5-4. A) A typical transver se MR image of the rat spinal cord. The overlap on the left and right sides is a folding artif act (VB: vertebral bone; SC: spinal cord; SN: spinal nerve). B) The distribution of Gd-DTPA with in the sciatic nerve (arrows) at the end of direct nerve infusion in rat #1 (cannula inse rt depth ~1 cm). From left to right and top to bottom, MR images correspond to th e rat vertebral level from L6 to L4 (MR slice thickness 1 mm). Gd-DTPA followed bot h L4 and L5 spinal nerves into the spinal column. 132

PAGE 133

Figure 5-5. Dynamic contrast-enhanced MRI scan s during infusion of Gd-DTPA into the right sciatic nerve of rat #2 (cannula insert dept h ~4 cm; slice thickn ess 1 mm). The left column is sagittal scans covering vertebral levels from L3 to L5. Columns A, B and C are transverse scans (caudal view) at positions A, B and C, respectively. Arrow 1 indicates the infusate distribution originat ing from the L4 spinal nerve. Arrow 2 indicates the infusate present in the C SF since the infusate spread over a large transverse region in the spinal column. (V: ventral; D: dorsal; Cr: cranial; Cd: caudal). 133

PAGE 134

Figure 5-6. Dynamic contrast-e nhanced MRI scans during infusi on of Gd-DTPA-albumin into the right sciatic nerve of rat #3 (cannula insert depth ~4 cm; slice thickness 1 mm). The left column is sagittal scans covering vertebral levels from L2 to L5. Columns A, B and C are transverse scans (caudal view) at positions A, B and C, respectively. Arrow 1 indicates the infusate distribution or iginating from L4 spinal nerve. Arrow 2 indicates the infusate distribution originat ing from L5 spinal nerve when increasing the infusion volume. Arrows 3 and 5 indica tes the infusate present in the CSF. Arrows 4 and 6 indicate the infusate present in the nerve root (V: ventral; D: dorsal; Cr: cranial; Cd: caudal). 134

PAGE 135

Figure 5-7. Evolving Gd-DTPA-al bumin distribution pattern duri ng CED into the right sciatic nerve of rat #4 (cannula insert depth ~ 4 cm ; slice thickness 1 mm ). The left column is sagittal scans. Columns A, B and C are transverse scans (cauda l view) at positions A, B and C, respectively. Arrow 1 indicates the infusate distribu tion originating from L5 spinal nerve. Arrows 2 and 3 also indi cate the infusate transport in L4 spinal nerve. Arrow 4 indicates the infusate pres ent in the CSF since the infusate spreads over a large transverse region in the spinal column (V: ventral; D: dorsal; Cr: cranial; Cd: caudal). 135

PAGE 136

Figure 5-8. Evolving Gd-DTPA-al bumin distribution pattern duri ng CED into the right sciatic nerve in rat #5 (Top) and rat #6 (Bottom) (cannula insert depth ~4 cm; slice thickness 1 mm). The left column is sagittal scans. Columns A, B and C are transverse scans (caudal view) at positions A, B and C, respectively. Arrows 1 and 5 indicate the infusate distribution originating from the L5 spinal nerve. Arrows 2 and 4 indicate the infusate leakage into the surrounding tissue. Arrows 3 and 6 indicates the infusate distribution in the CSF (V: ventral; D: dorsal; Cr: cranial; Cd: caudal). 136

PAGE 137

Figure 5-9. Graph of infused Gd-DTPA-albumin volume versus estimated tissue distribution volume in the spinal cavity (rats #3 to 6). 137

PAGE 138

CHAPTER 6 6. SUMMARY AND FUTURE WORK In this dissertation, fundamental engineering tools and methodologies for CED with the intension to improve CED techni ques and its application have been presented. Specifically, research studies (1) developed a 3-D biphasic finite element model for CED; (2) considered the effect of tissue properties and infusion para meters on solute transport during CED by using simulation analysis; (3) designe d a methodology for biomaterial ch aracterization using FE-based micro-indentation; (4) develope d a methodology for real-time MR imaging and quantification of concentration distribution during CED; and (5) developed a rat CED mode l to investigate the feasibility and transport characteristics of in vivo CED into rat spinal cord via the sciatic nerve infusion, and developed a met hodology of real-time in vivo MRI to monitor tracer transport during rat spinal cord CED. Results provided an infrastructure for futu re CED studies, which include computational models for CED predictions, characterization methods for nervous tissues, and MR imaging methods to monitor CED and quantify the concentration distribution. Figure 6-1 provides a functional relationship between co mputational modeling, MR imaging methods, experimental rat CED models, and tissue characteri zation methods. A computational FE model for direct nervous tissue infusion was presented in Chapter 2. Fund amental real-time in vivo MRI methods and an experimental rat CED model were presented in Chapter 5. A quantitative MRI method for CED was presented in Chapter 4. In addition, a tissue characterization method was presented in Chapter 3. The developed methods may be further integrated in future studies. By using the developed small animal CED model and the real -time quantitative MRI methods, experimental MRI data, such as anatomical boundaries and in vivo tissue transport anisotropy, could be obtained as well as experimental concentration distributions. These MRI data together with 138

PAGE 139

tissue properties and CED parameters, such as infu sion site and infusion rate, could be input into the 3-D FE computational models. Results from the computational models would provide predictions of interstitial flow and transport so that the CED protocol could be optimized. Comparison between experimental concentration distributions and computational predictions may provide important in vivo information, such as in vivo drug diffusivity and drug permeability within tissues. In addition, tissu e characterization methods such as microindentation could be used to pr ovide more accurate values of ne rvous tissue properties and thus improve the accuracy of computational predictions. Future work may expand the application of the developed methodologies, extend the current methodologies, or supplement CED studies on small animals. Detailed future works for each chapter are discussed below. In Chapter 2, a biphasic FE model has been de veloped which integrated tissue swelling and solute transport. An isotropic porous media with a regular geomet ry was used for the benchmark study. Future studies may incorporate realistic anatomical boundaries and transport anisotropy in nerve tissues such as white matter. Further de velopment of the FE model may also include constant flow rat infusion or the backflow dur ing infusion, which have not been addressed. In Chapter 3, an FE-based micro-indentat ion method was developed for biomaterial or tissue characterization. Measured Youngs modulus, E, and hydraulic permeability, k, could be used in biphasic FE models. The method was vali dated on a hydrogel contact lens. Future studies may apply this method to nervous tissues, such as brain tissue slices. Furt her development of the method may consider characteri zation of anisotropic material s, which have additional unknown properties. 139

PAGE 140

In Chapter 4, a quantitative MRI method for CED was developed. The method was validated on a hydrogel agarose phantom. Future studies may apply this method to nervous tissue CED, which requires the measurement of basic MR properties for nervous tissues such as relaxation times, T1 and T2, and relaxivities, R1 and R2. This will require further in vivo MR tracer studies. In Chapter 5, a rat CED model for periphera l nerve infusion was developed as well as a real-time in vivo MRI method to m onitor the tracer transport. CED into the rat sciatic nerve was conducted for delivery into the spinal cord. Futu re studies may use remote CED into the spinal cord via other spinal nerves, which have shorter nerve roots. Future studies may also investigate the anatomical structure at the spinal nerve-sp inal column junction region, infusate leakage around the cannula tip and the effect of the tip radial position on the infusate CSF uptake of infusate. These studies would pr ovide greater understa nding of the transport dynamics from the peripheral nerve to the spinal cord and would provide useful data for potential clinical application of this remote CED technique. These developed CED and imaging methods may also be applied to other CNS regions such as the rat brain. As a whole, fundamental methods were de veloped which may be used to predict biotransport during infusion, aid in CED protocol design, and mon itor transport in nervous tissue during CED. Drug transport during CED may ultimat ely be predicted or measured with more confidence with further developm ent of the computational, imaging, and measurement methods presented in this dissertation. 140

PAGE 141

Figure 6-1. Framework for CED prediction and optimization. 141

PAGE 142

APPENDIX A A. ANALYTICAL SOLUTION OF INFUSION INTO A SPHERICAL POROUS MEDIA Analytical infusion solutions for constant pressure and constant hydraulic permeability were used for our biphasic FE validation anal ysis. We solved for tissue displacement for pressure-controlled infusion using the solutions for pore pressure and fluid velocity determined by Basser [24] 0(,)[1erf()] 4 pa ra prt r t (A-1) 2 0 21( ) (,){[erfc()exp()](1)} 4 4f f fpa rarraa vrtk rt tt t (A-2) where 0 p is the constant infusion pressure, a is the radius of the infusion cavity (constant), and AkH The Laplace transform of th e radial displacement, was given by [24] (,)urt 3 0 21() (,)[ ()(1)] 4 z pa Bs uzs Csze z (zr s ) (A-3) where B(s) and C(s) are unknown functions and can be determined by using the zero contact stress boundary condition, at E 0, ra and the solution of p(r,t), respectively. These were calculated to be 2 00()11A Apapa Bs Hs H s and 01 () s a Apa Cse Hs (A-4) Substituting Eq. (A-4) into Eq. (A-3) results in 32 0000 0 22 222 2 3311111111 11 (,) 4ra ra s s AAAApapapa pa pa urs e e rsHrsHrHrsHr ss (A-5) Taking the inverse Laplace transform provides the solution for displacement (Fig. 2-2) 2() 2 22 0 4 214 4 (,)[ erf()erfc() ] 42 44ra A t Apa Ha trararatra urt at e Hr tt 2 (A-6) 142

PAGE 143

LIST OF REFERENCES [1] Pardridge, W.M., 2005, "Molecular biology of the blood-brain barrier," Mol. Biotechnol., 30(1), pp. 57-69. [2] Langer, R., 1990, "New Methods of Drug Delivery," Science, 249(4976), pp. 1527-1533. [3] Ratner, B.D., Hoffman, A.S., Schoen, F.J., and Lemons, J.E., 2004, Biomaterials Science: An Introduction to Materials in Medicine, Elsevier Academic Press. [4] Jain, R.K., 1989, "Delivery of Novel Ther apeutic Agents in Tumors Physiological Barriers and Strategies," J. Natl. Cancer Inst., 81(8), pp. 570-576. [5] Morrison, P.F., Laske, D.W., Bobo, H., Ol dfield, E.H., Dedrick, R.L., 1994, "High-flow microinfusion: tissue penetration a nd pharmacodynamics," Am. J. Physiol., 266(1), pp. R292-R305. [6] Bobo, R.H., Laske, D.W., Akbasak, A., Morri son, P.F., Dedrick, R.L., and Oldfield, E.H., 1994, "Convection-enhanced delivery of macrom olecules in the brain," Proc. Natl. Acad. Sci., 91(6), pp. 2076-2080. [7] Lonser, R.R., Gogate, N., Morrison, P.F., Wood, J.D., and Oldfield, E.H., 1998, "Direct convective delivery of macromolecules to the spinal cord," J. Neurosurg., 89(4), pp. 616622. [8] Lieberman, D.M., Laske, D.W., Morrison, P.F., Bankiewicz, K.S., and Oldfield, E.H., 1995, "Convection-enhanced distribution of large molecules in gray matter during interstitial drug infusion," J. Neurosurg., 82(6), pp. 1021-1029. [9] Lonser, R.R., Walbridge, S., Garmestani, K., Butman, J.A., Walters, H.A., Vortmeyer, A.O., Morrison, P.F., Brechbiel, M.W., and Oldfield, E.H., 2002, "Successful and safe perfusion of the primate brainstem: in vivo magnetic resonance imaging of macromolecular distribution duri ng infusion," J. Neurosurg., 97(4), pp. 905-913. [10] Hamilton, J.F., Morrison, P.F., Chen, M.Y., Harvey-White, J., Pernaute, R.S., Phillips, H., Oldfield, E., and Bankiewicz, K.S., 2001, "Heparin coinfusion during convectionenhanced delivery (CED) increases the distri bution of the glial-de rived neurotrophic factor (GDNF) ligand family in rat striatum and enhances the pharmacological activity of neurturin," Exp. Neurol., 168(1), pp. 155-161. [11] Gash, D.M., Zhang, Z.M., Ai, Y., Grondin, R., Coffey, R., and Gerhardt, G.A., 2005, "Trophic factor distribution pred icts functional recovery in Parkinsonian monkeys," Ann. Neurol., 58(2), pp. 224-233. [12] Saito, R., Bringas, J.R., Panner, A., Tamas, M., Pieper, R.O., Berger, M.S., and Bankiewicz, K.S., 2004, "Convection-enhanced de livery of tumor necrosis factor-related apoptosis-inducing ligand with systemic administration of temozolomide prolongs 143

PAGE 144

survival in an intracranial glioblas toma xenograft model," Cancer Res., 64(19), pp. 68586862. [13] Khan, A., Jallo, G.I., Liu, Y.J., Carson, B.S., and Guarnieri, M., 2005, "Infusion rates and drug distribution in brain tumor mo dels in rats," J. Neurosurg., 102(1), pp. 53-58. [14] Lonser, R.R., Walbridge, S., Murray, G.J ., Aizenberg, M.R., Vortmeyer, A.O., Aerts, J., Brady, R.O., and Oldfield, E.H., 2005, "Conv ection perfusion of glucocerebrosidase for neuronopathic Gaucher's disease," Ann. Neurol., 57(4), pp. 542-548. [15] Lin, L.F.H., Zhang, T.J., Collins, F., a nd Armes, L.G., 1994, "Pur ification and Initial Characterization of Rat B49 Glial-Cell Li ne-Derived Neurotrophic Factor," J. Neurochem., 63(2), pp. 758-768. [16] Airaksinen, M.S. and Saarma, M., 2002, "The GDNF family: Signalling, biological functions and therapeutic value," Nature Reviews Neuroscience, 3(5), pp. 383-394. [17] Weaver, M. and Laske, D.W ., 2003, "Transferrin receptor ligand-targeted toxin conjugate (Tf-CRM107) for therapy of maligna nt gliomas," J. Neuro-Oncol., 65(1), pp. 3-13. [18] Weber, F., Asher, A., Bucholz, R., Berger, M., Prados, M., Chang, S., Bruce, J., Hall, W., Rainov, N.G., Westphal, M., Warnick, R.E., Ra nd, R.W., Floeth, F., Rommel, F., Pan, H., Hingorani, V.N., and Puri, R.K., 2003, "Safet y, tolerability, and tumor response of IL4Pseudomonas exotoxin (NBI-3001) in patients with recurrent malignant glioma," J. Neuro-Oncol., 64(1), pp. 125-137. [19] Laske, D.W., Youle, R.J., and Oldfield E.H., 1997, "Tumor regression with regional distribution of the targeted toxin TF-CRM107 in patients with mali gnant brain tumors," Nature Medicine, 3(12), pp. 1362-1368. [20] Lidar, Z., Mardor, Y., Jonas, T., Pfeffer, R., Faibel, M., Nass, D., Hadani, M., and Ram, Z., 2004, "Convection-enhanced delivery of p aclitaxel for the treatment of recurrent malignant glioma: a Phase I/II clinical study," J. Neurosurg., 100(3), pp. 472-479. [21] Mardor, Y., Roth, Y., Lidar, Z., Jonas, T., Pfeffer, R., Maier, S.E., Faibel, M., Nass, D., Hadani, M., Orenstein, A., Cohen, J.S., and Ram, Z., 2001, "Monitoring response to convection-enhanced taxol delivery in brai n tumor patients using diffusion-weighted magnetic resonance imaging," Cancer Res., 61(13), pp. 4971-4973. [22] Voges, J., Reszka, R., Gossmann, A., Dittm ar, C., Richter, R., Garlip, G., Kracht, L., Coenen, H.H., Sturm, V., Wienhard, K., He iss, W.D., and Jacobs, A.H., 2003, "Imagingguided convection-enhanced delivery and gene therapy of glioblastoma," Ann. Neurol., 54(4), pp. 479-487. [23] Patel, S.J., Shapiro, W.R., Laske, D.W., Jensen, R.L., Asher, A.L., Wessels, B.W., Carpenter, S.P., and Shan, J.S., 2005, "Safet y and feasibility of convection-enhanced delivery of cotara for the treatment of malignant glioma: Initial experience in 51 patients," Neurosurgery, 56(6), pp. 1243-1252. 144

PAGE 145

[24] Basser, P.J., 1992, "Interstitial pressure, volume, and flow during infusion into brain tissue," Microvasc. Res., 44(2), pp. 143-165. [25] Barry, S.I. and Aldis, G.K., 1992, "Flowinduced deformation from pressurized cavities in absorbing porous tissues," Bull. Math. Biol., 54(6), pp. 977-997. [26] Smith, J.H. and Humphrey, J.A.C., 2007, "I nterstitial transport and transvascular fluid exchange during infusion into brain and tumor tissue," Microvasc. Res., 73(1), pp. 58-73. [27] Chen, Z.-J., Broaddus, W.C., Viswanathan, R.R., Raghavan, R., and Gillies, G.T., 2002, "Intraparenchymal drug delivery via positivepressure infusion: experimental and modeling studies of poroelastic ity in brain phantom gels," IEEE Trans. Biomed. Eng., 49(2), pp. 85-96. [28] Morrison, P.F., Chen, M.Y., Chadwick, R. S., Lonser, R.R., and Oldfield, E.H., 1999, "Focal delivery during direct infusion to brain: role of flow rate, catheter diameter, and tissue mechanics," Am. J. Physiol., 277(4, Part 2), pp. R1218-R1229. [29] Netti, P.A., Travascio, F., and Jain, R. K., 2003, "Coupled Macromolecular Transport and Gel Mechanics: Poroviscoelastic Approach," AIChE J., 49(6), pp. 1580-1596. [30] Neeves, K.B., Lo, C.T., Foley, C.P ., Saltzman, W.M., and Olbricht, W.L., 2006, "Fabrication and characteriza tion of microfluidic probes fo r convection enhanced drug delivery," J. Control Release, 111(3), pp. 252-262. [31] Chen, M.Y., Lonser, R.R., Morrison, P.F ., Governale, L.S., and Oldfield, E.H., 1999, "Variables affecting convection-enhanced delivery to the striatum: a systematic examination of rate of infusion, cannula size, infusate concentration, and tissue-cannula sealing time," J. Neurosurg., 90(2), pp. 315-320. [32] Wood, J.D., Lonser, R.R ., Gogate, N., Morrison, P.F., and Oldfield, E.H., 1999, "Convective delivery of macromolecules into th e naive and traumatized spinal cords of rats," J. Neurosurg., 90(1), pp. 115-120. [33] Gu, W.Y., Yao, H., Huang, C.Y., a nd Cheung, H.S., 2003, "New insight into deformation-dependent hydraulic permeability of gels and cartilage, and dynamic behavior of agarose gels in c onfined compression," J. Biomech., 36(4), pp. 593-598. [34] Lai, W.M. and Mow, V.C., 1980, "Draginduced compression of articular cartilage during a permeation experiment," Biorheology, 17(1-2), pp. 111-123. [35] Zhang, X.Y., Luck, J., Dewhirst, M.W ., and Yuan, F., 2000, "Interstitial hydraulic conductivity in a fibrosarcoma ," Am. J. Physiol.-Heart and Circulatory Physiology, 279(6), pp. H2726-H2734. [36] Ebenstein, D.M. and Pruitt, L.A., 2004b, "N anoindentation of soft hydrated materials for application to vascular tissues," J. Biomed. Mater. Res. A, 69A(2), pp. 222-232. 145

PAGE 146

[37] Ebenstein, D.M., Kuo, A., Rodrigo, J.J ., Reddi, A.H., Ries, M., and Pruitt, L., 2004a, "A nanoindentation technique for functional evaluati on of cartilage repair tissue," J. Mater. Res., 19(1), pp. 273-281. [38] Yang, F.Q. and Fei, P.X., 2004, "Mic roindentation of ground silicon wafers," Semiconductor Science and Technology, 19(9), pp. 1165-1168. [39] Oliver, W.C. and Pharr, G.M., 2004, "Meas urement of hardness and elastic modulus by instrumented indentation: Advances in unde rstanding and refinements to methodology," J. Mater. Res., 19(1), pp. 3-20. [40] Gilbert, J.L., Cumber, J., and Butterfield A., 2002, "Surface micromechanics of ultrahigh molecular weight polyethylene: Microindentation testing, crosslinking, and material behavior," J. Biomed. Mater. Res., 61(2), pp. 270-281. [41] Goldsmith, A.A.J. and Clift, S.E., 1998, "Investigation into the biphasic properties of a hydrogel for use in a cushion form replacement joint," J. Biomech. Eng.-Trans. ASME, 120(3), pp. 362-369. [42] Hale, J.E., Rudert, M.J., and Brown, T. D., 1993, "Indentation Assessment of Biphasic Mechanical Property Deficits in Size-Dep endent Osteochondral De fect Repair," J. Biomech., 26(11), pp. 1319-1325. [43] Mow, V.C., Gibbs, M.C., Lai, W.M., Zhu, W.B., and Athanasiou, K.A., 1989, "Biphasic indentation of articular cartila ge--II. A numerical algorithm and an experimental study," J. Biomech., 22(8-9), pp. 853-861. [44] Elkin, B.S., Azeloglu, E.U., Costa, K.D., and Morrison, B., 2007, "Mechanical heterogeneity of the rat hippocampus measured by atomic force microscope indentation," J. Neurotrauma, 24(5), pp. 812-822. [45] Mak, A.F., Lai, W.M., and Mow, V.C., 1987, "Biphasic indentation of articular cartilage-I. Theoretical analysis," J. Biomech., 20(7), pp. 703-714. [46] Miller, K., Chinzei, K., Orssengo, G., and Bednarz, P., 2000, "Mechanical properties of brain tissue in-vivo: experiment a nd computer simulation," J. Biomech., 33(11), pp. 1369-1376. [47] Krauze, M.T., Forsayeth, J., Park, J.W ., and Bankiewicz, K.S., 2006, "Real-time imaging and quantification of brain delivery of liposomes," Pharm. Res., 23(11), pp. 2493-2504. [48] Mardor, Y., Rahav, O., Zauberman, Y., Lidar, Z., Ocherashvilli, A., Daniels, D., Roth, Y., Maier, S.E., Orenstein, A., and Ram, Z ., 2005, "Convection-enhanced drug delivery: Increased efficacy and magnetic resonance image monitoring," Cancer Res., 65(15), pp. 6858-6863. [49] Nguyen, T.T., Pannu, Y.S., Sung, C., De drick, R.L., Walbridge, S., Brechbiel, M.W., Garmestani, K., Beitzel, M., Yordanov, A. T., and Oldfield, E.H., 2003, "Convective 146

PAGE 147

distribution of macromolecules in the primat e brain demonstrated using computerized tomography and magnetic resonance imaging," J. Neurosurg., 98(3), pp. 584-590. [50] Kim, H., Lizak, M.J., Tansey, G., Csaky, K.G., Robinson, M.R., Yuan, P., Wang, N.S., and Lutz, R.J., 2005, "Study of ocular transpor t of drugs released from an intravitreal implant using magnetic resonance imaging," Ann. Biomed. Eng., 33(2), pp. 150-164. [51] Krauze, M.T., McKnight, T.R., Yamashita Y., Bringas, J., Noble, C.O., Saito, R., Geletneky, K., Forsayeth, J., Berger, M.S., J ackson, P., Park, J.W., and Bankiewicz, K.S., 2005, "Real-time visualization and characte rization of liposomal delivery into the monkey brain by magnetic resonance imag ing," Brain Research Protocols, 16(1-3), pp. 20-26. [52] Lonser, R.R., Weil, R.J., Morrison, P.F ., Governale, L.S., and Oldfield, E.H., 1998, "Direct convective delivery of macromolecule s to peripheral nerves," J. Neurosurg., 89(4), pp. 610-615. [53] Ratliff, J.K. and Oldfield, E.H., 2001, "Convection-enhanced delivery in intact and lesioned peripheral nerve," J. Neurosurg., 95(6), pp. 1001-1011. [54] Boulis, N.M., Turner, D.E., Dice, J.A., Bhatia, V., and Feldman, E.L., 1999, "Characterization of adenovira l gene expression in spinal cord after remote vector delivery," Neurosurgery, 45(1), pp. 131-137. [55] Nakajima, H., Uchida, K., Kobayashi, S ., Kokubo, Y., Yayama, T., Sato, R., and Baba, H., 2005, "Targeted retrograde gene delivery in to the injured cervical spinal cord using recombinant adenovirus vect or," Neuroscience Letters, 385(1), pp. 30-35. [56] Abbott, N.J., Chugani, D.C., Zaharchuk, G ., Rosen, B.R., and Lo, E.H., 1999, "Delivery of imaging agents into brain," Adv. Drug Deliver. Rev., 37(1-3), pp. 253-277. [57] Francis, K., van Beek, J., Canova, C ., Neal, J.W., and Gasque, P., 2003, "Innate immunity and brain inflammati on: the key role of complement," Exp. Rev. Mol. Med., 5, [58] Nicholson, C., 2001, "Diffusion and rela ted transport mechanisms in brain tissue," Reports on Progress in Physics, 64(7), pp. 815-884. [59] Chen, X.M. and Sarntinoranont, M., 2007, "B iphasic Finite Element Model of Solute Transport for Direct Infusion into Nervous Tissue," Ann. Biomed. Eng., 35(12), pp. 2145-2158. [60] Kenyon, D.E., 1979, "A mathematical mode l of water flux through aortic tissue," Bull. Math. Biol., 41(1), pp. 79-90. [61] Oomens, C.W.J., Van Campen, D.H., and Grootenboer, H.J., 1987, "A mixture approach to the mechanics of skin," J. Biomech., 20, pp. 877-885. 147

PAGE 148

[62] Mow, V.C., Kuei, S.C., Lai, W.M., and Armstrong, C.G., 1980, "Biphasic Creep and Stress-Relaxation of Articular-Cartilage in Compression Theory and Experiments," J. Biomech. Eng.-Trans. ASME, 102(1), pp. 73-84. [63] Simon, B.R., Kaufman, M.V., Liu, J ., and Baldwin, A.L., 1998, "Porohyperelastictransport-swelling theory, material properties and finite element models for large arteries," International Jour nal of Solids and Structures, 35(34-35), pp. 5021-5031. [64] Mauck, R.L., Hung, C.T., and Ateshian G.A., 2003, "Modeling of neutral solute transport in a dynamically loaded porous pe rmeable gel: implications for articular cartilage biosynthesis and tissue engin eering," J. Biomech. Eng.-Trans. ASME, 125(5), pp. 602-614. [65] Ferguson, S.J., Ito, K., and Nolte, L.P., 2004, "Fluid flow and convective transport of solutes within the intervertebral disk," J. Biomech., 37, pp. 213-221. [66] Sengers, B.G., Oomens, C.W.J., and Baaijens F.P.T., 2004, "An integrated finite-element approach to mechanics, transport and bios ynthesis in tissue engineering," J. Biomech. Eng.-Trans. ASME, 126(1), pp. 82-91. [67] Yao, H. and Gu, W.Y., 2004, "Physical si gnals and solute trans port in cartilage under dynamic unconfined compression: Finite el ement analysis," Ann. Biomed. Eng., 32(3), pp. 380-390. [68] Johnson, E.M. and Deen, W.M., 1996, "Hydr aulic permeability of agarose gels," AIChE J., 42(5), pp. 1220-1224. [69] Holmes, M.H. and Mow, V.C., 1990, "The nonlinear characteristics of soft gels and hydrated connective ti ssues in ultrafiltr ation," J. Biomech., 23(11), pp. 1145-1156. [70] Mow, V.C., Holmes, M.H., and Lai, W .M., 1984, "Fluid transport and mechanical properties of articular cartil age: a review," J. Biomech., 17(5), pp. 377-394. [71] Kenyon, D.E., 1976, "Transient filtration in a porous elasti c cylinder," ASME Trans. J. Appl. Mech., 43(4), pp. 594-598. [72] Parker, K.H., Mehta, R.V., and Caro, C. G., 1987, "Steady flow in porous, elastically deformable materials," ASME Trans. J. Appl. Mech., 54(4), pp. 794-800. [73] Bradbury, M., 1979, The Concept of a Blood-Brain Barrier, Wiley, New York. [74] Pardridge, W.M., 1995, "Transport of small molecules throu gh the blood-brain barrier: biology and methodology," A dv. Drug Deliver. Rev., 15, pp. 5-36. [75] Johnston, S.T. and Deen, W.M., 1999, "Hinde red convection of proteins in agarose gels," J. Membrane Sci., 153(2), pp. 271-279. 148

PAGE 149

[76] Sugihara-Seki, M. and Fu, B.M.M., 2005, "Blood flow and permeability in microvessels," Fluid Dynamics Research, 37(1-2), pp. 82-132. [77] Hou, J.S., Holmes, M.H., Lai, W.M., a nd Mow, V.C., 1989, "Boundary conditions at the cartilage-synovial fluid interf ace for joint lubrication and th eoretical verifications," J. Biomech. Eng.-Trans. ASME, 111(1), pp. 78-87. [78] Prabhu, S.S., Broaddus, W.C., Gillies, G. T., Loudon, W.G., Chen, Z.J., and Smith, B., 1998, "Distribution of macromolecular dyes in brain using positive pressure infusion: A model for direct controlled delivery of therapeutic agents," Surg. Neurol., 50(4), pp. 367375. [79] Metz, H., McElhaney, J., and Ommaya, A. K., 1970, "A comparison of the elasticity of live, dead, and fixed brain tissue," J. Biomech., 3(4), pp. 453-458. [80] Walsh, E.K. and Schettini, A., 1976, "Elastic Behavior of Brain-Tissue Invivo," Am. J. Physiol., 230(4), pp. 1058-1062. [81] Miller, K. and Chinzei, K., 1997, "Constitutive modelling of brain tissue: Experiment and theory," J. Biomech., 30(11-12), pp. 1115-1121. [82] Monticelli, M.V., Chauhan, A., and Radke C.J., 2005, "The effect of water hydraulic permeability on the settling of a soft contac t lens on the eye," Current Eye Research, 30(5), pp. 329-336. [83] Reulen, H.J., Graham, R., Spatz, M., and Kl atzo, I., 1977, "Role of pressure gradients and bulk flow in dynamics of vasogeni c brain edema," J. Neurosurg., 46, pp. 22-35. [84] Kaczmarek, M., Subramaniam, R.P., and Neff, S.R., 1997, "The hydromechanics of hydrocephalus: steady-state solutions for cy lindrical geometry," Bull. Math. Biol., 59(2), pp. 295-323. [85] Nicholson, C. and Sykova, E., 1998, "Extra cellular space structur e revealed by diffusion analysis," Trends in Neurosciences, 21(5), pp. 207-215. [86] Dykstra, K.H., Hsiao, J.K., Morrison, P.F., Bungay, P.M., Mefford, I.N., Scully, M.M., and Dedrick, R.L., 1992, "Quantitative Examination of Tissue Concentration Profiles Associated with Microdialysis, Journal of Neurochemistry, 58(3), pp. 931-940. [87] Tao, L. and Nicholson, C., 1996, "Diffusion of albumins in rat cortical slices and relevance to volume transmission," Neurosci., 75(3), pp. 839-847. [88] Spilker, R.L., Suh, J.K., and Mow, V.C ., 1992, "A Finite-Element Analysis of the Indentation Stress-Relaxation Response of Linear Biphasic Articular-Cartilage," J. Biomech. Eng.-Trans. ASME, 114(2), pp. 191-201. 149

PAGE 150

[89] Gurbanova, A.A., Aker, R., Berkman, K., Onat, F.Y., van Rijn, C.M., and van Luijtelaar, G., 2006, "Effect of systemic and intracortical administration of phenytoin in two genetic models of absence epilepsy," Br. J. Pharmacol., 148(8), pp. 1076-1082. [90] Sarntinoranont, M., Banerjee, R.K., L onser, R.R., and Morrison, P.F., 2003, "A computational model of direct interstitial infusion of macromolecules into the spinal cord," Ann. Biomed. Eng., 31(4), pp. 448-461. [91] McGuire, S., Zaharoff, D., and Yuan F., 2006, "Nonlinear de pendence of hydraulic conductivity on tissue deformati on during intratumoral infusion," Ann. Biomed. Eng., 34(7), pp. 1173-1181. [92] Sarntinoranont, M., Chen, X.M., Zhao, J. B., and Mareci, T.H., 2006, "Computational model of interstitial transport in the spinal cord using diffusion te nsor imaging," Ann. Biomed. Eng., 34(8), pp. 1304-1321. [93] Chen, X.M., Dunn, A.C., Sawyer, W.G ., and Sarntinoranont, M., 2007, "A biphasic model for micro-indentation of a hydrogel-based contact lens," J. Biomech. Eng.-Trans. ASME, 129(2), pp. 156-163. [94] Nicolson, P.C. and Vogt, J., 2001, "Sof t contact lens polymers: an evolution," Biomaterials, 22, pp. 3273-3283. [95] Yang, W.-H., Smolen, V.F., and Peppas, N.A., 1981, "Oxygen permeability coefficients of polymers for hard and soft contact le ns applications," J. Membrane Sci., 9, pp. 53-67. [96] Jones, L.W. and Jones, D.A., 2000, Common contact lens co mplications : their recognition and management, Butterworth-Heinemann. [97] Holden, B.A., Stephenson, A., Stretton, S., Sankaridurg, P.R., O'Hare, N., Jalbert, I., and Sweeney, D.F., 2001, "Superior epithelial arcuat e lesions with soft contact lens wear," Optometry and Vision Science, 78(1), pp. 9-12. [98] Kikkawa, Y., Soft Contact Lens Kinetics, in Contact Lens Practice, Ruben, M. and Guillon, M., Editors. 1994, Chapman & Hall. p. 113-121. [99] Hinkley, J.A., Morgret, L.D., and Gehr ke, S.H., 2004, "Tensile properties of two responsive hydrogels," Polymer, 45(26), pp. 8837-8843. [100] Drury, J.L., Dennis, R.G., and Mooney, D. J., 2004, "The tensile properties of alginate hydrogels," Biomaterials, 25(16), pp. 3187-3199. [101] Stammen, J.A., Williams, S., Ku, D. N., and Guldberg, R.E., 2001, "Mechanical properties of a novel PVA hydrogel in shear and unconfined compression," Biomaterials, 22(8), pp. 799-806. 150

PAGE 151

[102] Marra, S.P., Ramesh, K.T., and Douglas, A.S., 2001, "Mechanical characterization of active poly(vinyl alcohol)-poly(acrylic ac id) gel," Mat. Sci. Eng. C-Bio. S., 14(1-2), pp. 25-34. [103] Enns, J.B., 1996, "Dynamic mechanical pr operties of hydrogels," Proc. of the 1996 54th Annual Technical Conference, 3, pp. 2852-2856. [104] Anseth, K.S., Bowman, C.N., and BrannonPeppas, L., 1996, "Mechanical properties of hydrogels and their experimental determination," Biomaterials, 17(17), pp. 1647-1657. [105] Yasuda, H., Lamaze, C.E., and Pete rlin, A., 1971, "Diffusive and Hydraulic Permeabilities of Water in Water-Swollen Polymer Membranes," J. Polym. Sci. A2, 9(6), pp. 1117-1131. [106] Refojo, M.F., 1965, "Permeation of Water through Some Hydrogels ," J. Appl. Polym. Sci., 9(10), pp. 3417-3426. [107] Chiarelli, P., Basser, P.J., Derossi, D., and Goldstein, S., 1992, "The dynamics of a hydrogel strip," Biorheology, 29, pp. 383-398. [108] LeRoux, M.A. and Setton, L.A., 2002, "E xperimental and biphasic FEM determinations of the material properties and hydraulic perm eability of the meniscus in tension," J. Biomech. Eng.-Trans. ASME, 124(3), pp. 315-321. [109] Hoffman, A.S., 2002, "Hydrogels for biomedi cal applications," Adv. Drug Deliver. Rev., 54(1), pp. 3-12. [110] Manetti, C., Casciani, L., and Pescos olido, N., 2002, "Diffusive contribution to permeation of hydrogel contact lenses: theore tical model and experimental evaluation by nuclear magnetic resonance techniques," Polymer, 43(1), pp. 87-92. [111] ADINA R&D Inc., 2004, ADINA theory and modeling guide. [112] Rennie, A.C., Dickrell, P.L., and Sawyer, W.G., 2005, "Friction Coefficient of Soft Contact Lenses: Measurements and Modeling," Tribology Letters, 18, pp. 499-504. [113] Bae, W.C., Lewis, C.W., Levenston, M.E ., and Sah, R.L., 2006, "Indentation testing of human articular cartilage: Effects of probe tip geometry and indentation depth on intratissue strain," J. Biomech., 39(6), pp. 1039-1047. [114] Holmes, M.H., Lai, W.M., and Mow, V.C., 198 5, "Singular perturbati on analysis of the nonlinear, flow-dependent compre ssive stress relaxation behavior of articular cartilage," J. Biomech. Eng.-Trans. ASME, 107(3), pp. 206-218. [115] Chen, X.M., Astary, G.W., Sepulveda, H ., Mareci, T.H., and Sarntinoranont, M., 2008, "Quantitative assessment of m acromolecular concentration duri ng direct infusion into an agarose hydrogel phantom using contrast-e nhanced MRI," Magn. Reson. Imaging, In press, 151

PAGE 152

[116] Kunwar, S., Prados, M.D., Chang, S.M., Berger, M.S., Lang, F.F., Piepmeier, J.M., Sampson, J.H., Ram, Z., Gutin, P.H., Gibbons, R.D., Aldape, K.D., Croteau, D.J., Sherman, J.W., and Puri, R.K., 2007, "Dir ect intracerebral delivery of cintredekin besudotox (IL13-PE38QQR) in recurrent maligna nt glioma: A report by the Cintredekin Besudotox Intraparenchymal Study Group," J. Clin. Oncol., 25(7), pp. 837-844. [117] Sampson, J.H., Brady, M.L., Petry, N.A ., Croteau, D., Friedman, A.H., Friedman, H.S., Wong, T., Bigner, D.D., Pastan, I., Puri, R.K., and Pedain, C., 2007, "Intracerebral infusate distribution by convection-enhanced delivery in humans with malignant gliomas: Descriptive effects of target anatomy and catheter positioning," Neurosurgery, 60(2), pp. 89-98. [118] Roberts, T.P.L., 1997, "Physiologic meas urements by contrast-enhanced MR imaging: Expectations and limitations," J. Magn. Reson. Imaging, 7(1), pp. 82-90. [119] Strich, G., Hagan, P.L., Gerber, K.H., and Slutsky, R.A., 1985, "Tissue Distribution and Magnetic-Resonance Spin-Lattice Relaxation Effects of Gadolinium-Dtpa," Radiology, 154(3), pp. 723-726. [120] Stanisz, G.J. and Henkelman, R .M., 2000, "Gd-DTPA relaxivity depends on macromolecular content," Magn. Reson. Med., 44(5), pp. 665-667. [121] Tweedle, M.F., Wedeking, P., Telser, J., So tak, C.H., Chang, C.A., Kumar, K., Wan, X., and Eaton, S.M., 1991, "Dependence of MR Si gnal Intensity on Gd Tissue Concentration over a Broad Dose Range," Magn. Reson. Med., 22(2), pp. 191-194. [122] Brasch, R.C., 1991, "Rationale and Applic ations for Macromolecular Gd-Based Contrast Agents," Magn. Reson. Med., 22(2), pp. 282-287. [123] Rohrer, M., Bauer, H., Mintorovitch, J., Requardt, M., and Weinmann, H.J., 2005, "Comparison of magnetic prope rties of MRI contrast medi a solutions at different magnetic field strengths," Investigative Radiology, 40(11), pp. 715-724. [124] Donahue, K.M., Burstein, D., Manning, W.J ., and Gray, M.L., 1994, "Studies of Gd-Dtpa Relaxivity and Proton-Exchange Rate s in Tissue," Magn. Reson. Med., 32(1), pp. 66-76. [125] Morkenborg, J., Pedersen, M., Jensen, F.T., Stodkilde-Jorgensen, H., Djurhuus, J.C., and Frokiaer, J., 2003, "Quantitativ e assessment of Gd-DTPA contrast agent from signal enhancement: an in-vitro study," Magn. Reson. Imaging, 21(6), pp. 637-643. [126] Hittmair, K., Gomiscek, G., Langenberger, K., Recht, M., Imhof, H., and Kramer, J., 1994, "Method For the Quantitative Assessment of Contrast Agent Uptake in Dynamic Contrast-Enhanced MRI," Magn. Reson. Med., 31(5), pp. 567-571. [127] Johnson, E.M., Berk, D.A., Jain, R. K., and Deen, W.M., 1995, "Diffusion and Partitioning of Proteins in Charged Agarose Gels," Biophys., 68(4), pp. 1561-1568. 152

PAGE 153

[128] Nicholson, C. and Tao, L., 1993, "H indered Diffusion of Hi gh-Molecular-Weight Compounds in Brain Extracellular Microenvironment Measured with Integrative Optical Imaging," Biophys., 65(6), pp. 2277-2290. [129] Haacke, E.M., Brown, R.W., Thompson, M.R., and Venkatesan, R., 1999, Magnetic Resonance Imaging: Physical Pr inciples and Sequence Design, John Wiley & Sons, Inc, New York. [130] Okuhata, Y., 1999, "Delivery of diagnostic agents for magne tic resonance imaging," Adv. Drug Deliver. Rev., 37(1-3), pp. 121-137. [131] Miller, A.J. and Joseph, P.M., 1993, "The use of power images to perform quantitative analysis on low SNR MR images," Magn. Reson. Imaging, 11, pp. 1051-1056. [132] McGibney, G. and Smith, M.R., 1993, "An unbiased signal-to-noise ratio measure for magnetic resonance images," Med. Phys., 20, pp. 1077-1078. [133] Smith, B.C., 2002, Quantitative Spectroscopy: Theory and Practice, Academic Press. 200. [134] Pishko, G.L., Lee, S.J., Wanakule, P., and Sarntinoranont, M., 2007, "Hydraulic permeability of a hydrogel-based contact lens membrane for low flow rates," J. Appl. Polym. Sci., 104(6), pp. 3730-3735. [135] Liang, S.M., Xu, J., Weng, L.H., Dai, H.J., Zhang, X.L., and Zhang, L.N., 2006, "Protein diffusion in agarose hydrogel in situ measur ed by improved refractive index method," J. Controlled Release, 115(2), pp. 189-196. [136] Tannus, A. and Garwood, M., 1997, "Adiabatic pulses," NMR Biomed., 10(8), pp. 423434. [137] Pintaske, J., Martirosian, P., Graf, H., Erb, G., Lodemann, K.P., Claussen, C.D., and Schick, F., 2006, "Relaxivity of gadopent etate dimeglumine (Magnevist), gadobutrol (Giadovist), and gadobenate dimeglumine (Mu ltiHance) in human blood plasma at 0.2, 1.5, and 3 Tesla," Investigative Radiology, 41(3), pp. 213-221. [138] Schmiedl, U., Ogan, M., Paajanen, H., Maro tti, M., Crooks, L.E., Brito, A.C., and Brasch, R.C., 1987, "Albumin Labeled with Gd-Dtpa as an Intravascular, Blood Pool Enhancing Agent for Mr Imaging Biodistributi on and Imaging Studies," Radiology, 162(1), pp. 205-210. [139] Gu, W.Y., Yao, H., Vega, A.L., and Flagler, D., 2004, "Diffusivity of ions in agarose gels and intervertebral disc: Effect of porosity," Ann. Biomed. Eng., 32(12), pp. 1710-1717. [140] Pluen, A., Netti, P.A., Jain, R.K., and Berk, D.A., 1999, "Diffusion of macromolecules in agarose gels: Comparison of linear a nd globular configur ations," Biophys., 77(1), pp. 542-552. 153

PAGE 154

[141] Koay, C.G. and Basser, P.J., 2006, "Ana lytically exact correc tion scheme for signal extraction from noisy ma gnitude MR signals," J. Magn. Reson., 179(2), pp. 317-322. [142] Henkelman, R.M., 1985, "Measurement of Signa l Intensities in the Presence of Noise in Mr Images," Med. Phys., 12(2), pp. 232-233. [143] de Graaf, R.A., Brown, P.B., McIntyre, S ., Nixon, T.W., Behar, K.L., and Rothman, D.L., 2006, "High magnetic field wate r and metabolite proton T-1 and T-2 relaxation in rat brain in vivo," Magn. Reson. Med., 56(2), pp. 386-394. [144] Rinck, P.A. and Muller, R.N., 1999, "Fie ld strength and dose dependence of contrast enhancement by gadolinium-based MR contrast agents," Eur. Radiol., 9(5), pp. 998-1004. [145] van den Berg, M.P., Romeijn, S.G., Verhoe f, J.C., and Merkus, F.W., 2002, "Serial cerebrospinal fluid sampling in a rat model to study drug uptake from the nasal cavity," J. Neurosci. Meth., 116, pp. 99-107. [146] Boskamp, E.B., 1985, "Improved Surface Coil Imaging in Mr Decoupling of the Excitation and Receiver Coils," Radiology, 157(2), pp. 449-452. [147] Williams, P.L. and Warwick, R., eds, 1980, Gray's anatomy, 36 ed. Longman Group Ltd. 154

PAGE 155

BIOGRAPHICAL SKETCH Xiaoming Chen was born in Jinhua, Zhejiang, P. R. China. His early education was received at Jinhua First High School. He earned his Bachelor of Engineering degree from Tsinghua University in Beijing, P. R. China, in 1998. He received his Master of Engineering degree from National University of Singapore, Singapore, in 2003 and his Master of Science degree from the University of Florida, Gaines ville, in 2005. He earned his Doctor of Philosophy degree in mechanical engi neering from the University of Florida, Gainesville, in August 2008. His research interests include computational m echanics and magnetic resonance imaging with an emphasis on biotransport and drug de livery in the nervous tissue. 155